scholarly journals Hearing preservation in children with electric-acoustic stimulation after cochlear implantation

HNO ◽  
2018 ◽  
Vol 66 (S2) ◽  
pp. 56-62 ◽  
Author(s):  
T. Rader ◽  
A. Bohnert ◽  
C. Matthias ◽  
D. Koutsimpelas ◽  
M-A. Kainz ◽  
...  

Abstract Background Cochlear implantation in patients with functional residual low-frequency hearing is performed according to an established hearing-preserving surgical technique in order to cause minimal trauma of inner ear structures. Due to the increasing number of cochlear implants in children, the preservation of residual hearing is becoming increasingly important in this patient collective. Objectives Short- and mid-term hearing preservation outcome in pediatric patients is investigated. Materials and methods A group of 9 children (12 ears) between 5 and 12 years of age were examined after hearing-assisted cochlear implantation with respect to the pure tone audiometric thresholds. Retrospectively, short-term hearing preservation (up to 3 months after surgery) was examined. In a subgroup of 5 children, mid-term hearing preservation (7.5 to 16 months after surgery) was also analyzed. The mean values of hearing preserved (HL%) and hearing loss (HL) due to electrode insertion were calculated as measured values. Results In the whole group, the mean values of the preoperative PTAlow were 29.8 dB and the short-term postoperative PTAlow 42.6 dB. The mean value of the HL% was 73.6%, corresponding to an HL of 9.4 dB. In the subgroup, the mean PTAlow postoperatively was 46.0 dB in the mid-term and the HL% at 80.7% with a HL of 6.6 dB. Conclusions The results in children are consistent with the results in adults. Electric-acoustic stimulation (EAS) should be used in the treatment of children with existing low-frequency residual hearing, as good residual hearing preservation can also be achieved in children after implantation.

2019 ◽  
Vol 277 (2) ◽  
pp. 367-375 ◽  
Author(s):  
Matti Iso-Mustajärvi ◽  
Sini Sipari ◽  
Heikki Löppönen ◽  
Aarno Dietz

Abstract Purpose To evaluate the insertion results and hearing preservation of a novel slim modiolar electrode (SME) in patients with residual hearing. Methods We retrospectively collected the data from the medical files of 17 patients (18 ears) implanted with a SME. All patients had functional low frequency hearing (PTA (0.125–0.5 kHz) ≤ 80 dB HL). The insertion results were re-examined from the postoperative cone-beam computed tomography scans. Postoperative thresholds were obtained at the time of switch-on of the sound processors (mean 43 days) and at latest follow-up (mean 582 days). The speech recognition in noise was measured with the Finnish matrix sentence test preoperatively and at follow-up. Results The mean insertion depth angle (IDA) was 395°. Neither scala dislocations nor tip fold over were detected. There were no total hearing losses. Functional low-frequency hearing was preserved in 15/18 (83%) ears at switch-on and in 14/17 (82%) ears at follow-up. According to HEARRING classification, 55% (10/18) had complete HP at switch-on and 41% (7/17) still at follow-up. Thirteen patients (14 ears) were initially fitted with electric–acoustic stimulation and seven patients (8 ears) continued to use it after follow-up. Conclusions The preliminary hearing preservation results with the SME were more favorable than reported for other perimodiolar electrodes. The results show that the array may also be feasible for electro-acoustic stimulation; it is beneficial in that it provides adequate cochlear coverage for pure electrical stimulation in the event of postoperative or progressive hearing loss.


2014 ◽  
Vol 2014 ◽  
pp. 1-6 ◽  
Author(s):  
Priscila Carvalho Miranda ◽  
André Luiz Lopes Sampaio ◽  
Rafaela Aquino Fernandes Lopes ◽  
Alessandra Ramos Venosa ◽  
Carlos Augusto Costa Pires de Oliveira

In the past, it was thought that hearing loss patients with residual low-frequency hearing would not be good candidates for cochlear implantation since insertion was expected to induce inner ear trauma. Recent advances in electrode design and surgical techniques have made the preservation of residual low-frequency hearing achievable and desirable. The importance of preserving residual low-frequency hearing cannot be underestimated in light of the added benefit of hearing in noisy atmospheres and in music quality. The concept of electrical and acoustic stimulation involves electrically stimulating the nonfunctional, high-frequency region of the cochlea with a cochlear implant and applying a hearing aid in the low-frequency range. The principle of preserving low-frequency hearing by a “soft surgery” cochlear implantation could also be useful to the population of children who might profit from regenerative hair cell therapy in the future. Main aspects of low-frequency hearing preservation surgery are discussed in this review: its brief history, electrode design, principles and advantages of electric-acoustic stimulation, surgical technique, and further implications of this new treatment possibility for hearing impaired patients.


2017 ◽  
Vol 3 (2) ◽  
pp. 119-122
Author(s):  
Wouter J. van Drunen ◽  
Sarra Kacha Lachheb ◽  
Anatoly Glukhovskoy ◽  
Jens Twiefel ◽  
Marc C. Wurz ◽  
...  

AbstractFor patients suffering from profound hearing loss or deafness still having respectable residual hearing in the low frequency range, the combination of a hearing aid with a cochlear implant results in the best quality of hearing perception (EAS – electric acoustic stimulation). In order to optimize EAS, ongoing research focusses on the integration of these stimuli in a single implant device. Within this study, the performance of piezoelectric actuators, particularly the dual actuator stimulation, in a scaled uncoiled test rig was investigated.


2015 ◽  
Vol 26 (08) ◽  
pp. 732-740 ◽  
Author(s):  
Margaret T. Dillon ◽  
Andrea L. Bucker ◽  
Marcia C. Adunka ◽  
English R. King ◽  
Oliver F. Adunka ◽  
...  

Background: Candidacy criteria for cochlear implantation are expanding to include patients with substantial low-to-mid frequency hearing sensitivity. Postoperative hearing preservation has been achieved in cochlear implant recipients, though with variable outcomes. Previous investigations on postoperative hearing preservation outcomes have evaluated intraoperative procedures. There has been limited review as to whether electric stimulation influences hearing preservation. Purpose: The purpose of this analysis was to evaluate whether charge levels associated with electric stimulation influence postoperative hearing preservation within the first year of listening experience. Research Design: Retrospective analysis of unaided residual hearing and charge levels. Study Sample: Twenty-eight cochlear implant recipients with postoperative residual hearing in the operative ear and at least 12 mo of listening experience with electric-acoustic stimulation (EAS). Data Collection and Analysis: Assessment intervals included initial cochlear implant activation, initial EAS activation, and 3-, 6-, and 12-mo postinitial EAS activation. A masked low-frequency bone-conduction (BC) pure-tone average (PTA) was calculated for all participants at each assessment interval. Charge levels for each electrode were determined using the most comfortable loudness level and pulse width values. Charge levels associated with different regions of the electrode array were compared to the change in the low-frequency BC PTA between two consecutive intervals. Results: Charge levels had little to no association with the postoperative change in low-frequency BC PTA within the first year of listening experience. Conclusions: Electric charge levels do not appear to be reliably related to the subsequent loss of residual low-frequency hearing in the implanted ear within the first year of EAS listening experience.


2020 ◽  
Vol 29 (1) ◽  
pp. 1-5 ◽  
Author(s):  
Margaret T. Dillon ◽  
Emily Buss ◽  
Brendan P. O'Connell ◽  
Meredith A. Rooth ◽  
English R. King ◽  
...  

Purpose The goal of this work was to evaluate the low-frequency hearing preservation of long electrode array cochlear implant (CI) recipients. Method Twenty-five participants presented with an unaided hearing threshold of ≤ 80 dB HL at 125 Hz pre-operatively in the ear to be implanted. Participants were implanted with a long (31.5-mm) electrode array. The unaided hearing threshold at 125 Hz was compared between the preoperative and postoperative intervals (i.e., initial CI activation, and 1, 3, 6, 9, and 12 months after activation). Results Eight participants maintained an unaided hearing threshold of ≤ 80 dB HL at 125 Hz postoperatively. The majority ( n = 5) demonstrated aidable low-frequency hearing at initial activation, whereas 3 other participants experienced an improvement in unaided low-frequency hearing thresholds at subsequent intervals. Conclusions CI recipients can retain residual hearing sensitivity with fully inserted long electrode arrays, and low-frequency hearing thresholds may improve during the postoperative period. Therefore, unaided hearing thresholds obtained within the initial weeks after surgery may not reflect later hearing sensitivity. Routine measurement of postoperative unaided hearing thresholds—even for patients who did not demonstrate aidable hearing thresholds initially after cochlear implantation—will identify CI recipients who may benefit from electric–acoustic stimulation. Supplemental Material https://doi.org/10.23641/asha.11356637


2016 ◽  
Vol 37 (9) ◽  
pp. e353-e359 ◽  
Author(s):  
Silke Helbig ◽  
Youssef Adel ◽  
Tobias Rader ◽  
Timo Stöver ◽  
Uwe Baumann

2021 ◽  
Vol 10 (19) ◽  
pp. 4305
Author(s):  
Farnaz Matin ◽  
Eralp-Niyazi Artukarslan ◽  
Angelika Illg ◽  
Anke Lesinski-Schiedat ◽  
Thomas Lenarz ◽  
...  

This retrospective study aimed to investigate the range of hearing levels in a cochlear implant (CI) elderly population receiving electric-acoustic-stimulation (EAS) or electric-stimulation (ES) alone. The investigation evaluates the degree of hearing preservation (HP) and the speech comprehension resulting from EAS or ES-only to identify audiometric factors that predict adequate EAS and ES use. We analyzed the pure tone audiometry and speech perception in quiet and noise preoperatively and 12-months after activation of 89 elderly adults (age of 65 years old or older), yielding in total 97 CIs. Thirty-two (33.1%) patients were potential EAS candidates preoperatively, of which 18 patients used EAS at the time of first fitting and the other 14 patients continued to use their residual hearing for EAS at 12-months. Post-treatment, patients with EAS system and ES-only users’ with longer electrodes showed better results in monosyllable word scores in quiet than ES-only users with shorter electrodes. A similar trend was revealed for the speech recognition in noise. Patients with an EAS system benefit from maintaining their natural residual hearing. Nevertheless, strict preoperative patient selection is warranted particularly in elderly patients, in whom the hearing thresholds for EAS indication differ slightly from that in younger adults.


2017 ◽  
Vol 2 (6) ◽  
pp. 54-63
Author(s):  
Conor Kelly ◽  
Lina A. J. Reiss

Hearing preservation cochlear implants (CIs) are specifically designed to preserve residual low-frequency acoustic hearing for use together with electrically stimulated high-frequency hearing. This combined electro-acoustic stimulation (EAS) provides a promising treatment option for patients with severe high-frequency hearing loss, but with some residual low-frequency hearing, and has been shown to improve speech perception, especially in background noise, music perception, and sound source localization. Thus, preservation of residual hearing should be a priority in treatment. Although residual low-frequency hearing is successfully preserved to varying degrees in many patients, some patients experience a loss of residual hearing following implantation. A wide range of potential causes of, or contributors to, loss of residual hearing in EAS CI users have been proposed. In this paper, we review the evidence for several of the proposed causes of hearing loss with EAS CI. We conclude that its etiology is likely a multifactorial, heterogeneous phenomenon. Furthermore, we suggest that studies to further elucidate effects of ischemia on lateral wall function and maintenance of endocochlear potential in the context of EAS CI implantation and use are needed.


2018 ◽  
Vol 23 (4) ◽  
pp. 222-228
Author(s):  
Christof Buhl ◽  
Dmitri Artemiev ◽  
Flurin Pfiffner ◽  
Jaap Swanenburg ◽  
Dorothe Veraguth ◽  
...  

Objectives: (1) To assess dynamic postural stability before and after cochlear implantation using a functional gait assessment (FGA). (2) To evaluate the correlation between loss of residual hearing and changes in dynamic postural stability after cochlear implantation. Methods: Candidates for first-sided cochlear implantation were prospectively included. The FGAs and pure-tone audiograms were performed before and 4–6 weeks after cochlear implantation. Results: Twenty-three subjects were included. Forty-eight percent (n = 11) showed FGA performance below the age-referenced norm before surgery. One subject had a clinically relevant decrease of the FGA score after cochlear implantation. No significant difference between the mean pre- and postoperative FGA scores was detectable (p = 0.4). Postoperative hearing loss showed no correlation with a change in FGA score after surgery (r = 0.3, p = 0.3, n = 16). Conclusion: Single-sided cochlear implantation does not adversely affect dynamic postural stability 5 weeks after surgery. Loss of functional residual hearing is not correlated with a decrease in dynamic postural stability.


2013 ◽  
Vol 2013 ◽  
pp. 1-8 ◽  
Author(s):  
Guilherme Machado de Carvalho ◽  
Alexandre C. Guimaraes ◽  
Alexandre S. M. Duarte ◽  
Eder B. Muranaka ◽  
Marcelo N. Soki ◽  
...  

Background. Electric-acoustic stimulation (EAS) is an excellent choice for people with residual hearing in low frequencies but not high frequencies and who derive insufficient benefit from hearing aids. For EAS to be effective, subjects' residual hearing must be preserved during cochlear implant (CI) surgery.Methods. We implanted 6 subjects with a CI. We used a special surgical technique and an electrode designed to be atraumatic. Subjects' rates of residual hearing preservation were measured 3 times postoperatively, lastly after at least a year of implant experience. Subjects' aided speech perception was tested pre- and postoperatively with a sentence test in quiet. Subjects' subjective responses assessed after a year of EAS or CI experience.Results. 4 subjects had total or partial residual hearing preservation; 2 subjects had total residual hearing loss. All subjects' hearing and speech perception benefited from cochlear implantation. CI diminished or eliminated tinnitus in all 4 subjects who had it preoperatively. 5 subjects reported great satisfaction with their new device.Conclusions. When we have more experience with our surgical technique we are confident we will be able to report increased rates of residual hearing preservation. Hopefully, our study will raise the profile of EAS in Brazil and Latin/South America.


Sign in / Sign up

Export Citation Format

Share Document