Abstract
Background
Immunosuppressant therapeutic drug monitoring (TDM) usually requires outpatient travel to hospitals or phlebotomy sites for venous blood collection; however Mitra® Microsampling Device (MSD) sampling could allow self-collection and shipping of samples to a laboratory for analysis. This study examined the feasibility of using volumetric microsampling by MSD for TDM of tacrolimus (TaC) and cyclosporin A (CsA) in transplant patients, along with their feedback on the process.
Methods
MSD was used to collect TaC and CsA from venous (VB) or capillary (CB) blood. The MSDs were rehydrated, extracted, and analyzed using on-line solid phase extraction coupled to tandem mass spectrometry (SPE-MS/MS). We report an abbreviated method validation of the MSD including: accuracy, precision, linearity, carry-over, and stability using residual venous whole blood (VB) samples. Subsequent clinical validation compared serially collected MSD + CB against VB (200 µL) from transplant patients.
Results
Accuracy comparing VB vs. MSD+VB showed high clinical concordance (TaC = 89% and CsA = 98%). Inter- and intra-precision was ≤11.5 %CV for TaC and CsA. Samples were stable for up to 7 days at room temperature with an average difference of <10%. Clinical validation with MSD+CB correlated well with VB for CsA (slope = 0.95, r2 = 0.88, n = 47) and TaC (slope = 0.98, r2 = 0.82, n = 49). CB vs. VB gave concordance of 94% for CsA and 79% for TaC. A satisfaction survey showed 82% of patients preferred having the capillary collection option.
Conclusion
Transplant patients favored having the ability to collect capillary samples at home for TaC/CsA monitoring. Our results demonstrate good concordance between MSD+CB and VB for TaC and CsA TDM, but additional studies are warranted.