Analytical Method Validation and Determination of Iron and Phosphorus in Vegetable Oil by Inductively Coupled Plasma–Mass Spectrometry with Microwave Assisted Digestion

2018 ◽  
Vol 51 (11) ◽  
pp. 1774-1788 ◽  
Author(s):  
Rizwan Shariff ◽  
Ayyappan A. Aachary ◽  
Lawrence H. Pacquette ◽  
Ashutosh K. Mittal ◽  
Rajesh Girdhar
2015 ◽  
Vol 98 (1) ◽  
pp. 160-164 ◽  
Author(s):  
Nancy Morris Hepp

Abstract Specifications in the Code of Federal Regulationslimit the levels of As, Pb, and Hg in most certifiable color additives, and some have additional specification limits for Cr and Mn. A new ICP/MS method wasdeveloped and validated for the quantitative determination of these elements in various certifiable color additives. One dissolution and two microwave-assisted digestion sample preparation procedures were optimized to address initial low Hg and enhanced As recoveries. Results using the three sample preparation procedures were generally equivalent for all of the elements determined. LOQ values were 0.3 mg/kg for As, 0.7 mg/kg for Cr, 0.4 mg/kg for Pb, 0.7 mg/kg forMn, and 0.1 mg/kg for Hg.