scholarly journals Validation of the wrist blood pressure measuring device Omron RS6 (HEM-6221-E) among obese Sudanese patients according to the European Society of Hypertension International Protocol Revision 2010

F1000Research ◽  
2020 ◽  
Vol 9 ◽  
pp. 1284
Author(s):  
Elrazi A. Ali ◽  
Saeed M. Omar ◽  
Yassin Ibrahim ◽  
Osama Al-Wutayd ◽  
Ishag Adam

Background: Electronic devices for measuring blood pressure (BP) need to go through independent clinical validation as recommended by different authorities, both in general and specific populations. The aim of this study was to assess the validity of the Omron RS6 (HEM-6221-E) wrist oscillometric devices in obese Sudanese patients. Methods: Of 90 obese individuals invited for recruitment, 33 were included in the study, and had their BP at the level of the wrist measured using Omron RS6 and standard mercury sphygmomanometer. Two observations were made and the mean was taken. BP differences between the two methods for the 33 participants were classified into three categories (≤5, ≤10, and ≤15 mmHg), according to the European Society of Hypertension-International Protocol revision 2010 (ESH-IP2) criteria. This was then used to assess the validity of the tested Omron RS6 device. Results: Participants had a mean age of 56.97 years (standard deviation (SD), 8.75; range, 36-79). Average systolic blood pressure (SBP) was 146.21 mmHg (SD, 23.07; range, 107-182), and average diastolic blood pressure (DBP) was 93.82 mmHg (SD, 16.06; range, 67-128). There was a good agreement between the two observations using the OMRON RS6 and the standard sphygmomanometer: −4 to + 3 mmHg for SBP and −4 to +4 mmHg for DBP, with the mean difference of 1.73±1.11 mmHg for SBP and 1.49±1.02 mmHg for DBP. Conclusion: Thus, the Omron RS6 (HEM-6221-E) is a valid and suitable measure of BP according to ESH-IP2.

Author(s):  
Yasmine Winda Kusumaningtyas ◽  
Tri Bowo Indrato ◽  
M. Prastawa Assalim T.P ◽  
Bedjo Utomo

Sphygmomanometer is a human blood pressure measuring device which has several types such as mercury, digital, and aneroid. At this time, mercury Sphygmomanometer is not allowed to be used anymore considering the side effects of mercury which can harm the skin and even respiration. This study helps to reduce the use of mercury type Sphygmomanometer. The researchers make arduino TFT LCD display based Sphyhmomanometer so that the use of mercury can be reduced and make it easier for users to process measurement. The researchers used the oscillmetry method in making the device to calculate the systolic and diastolic pressure. The result of blood pressure measurement has a systolic error rate of 0,08% and a diastolic error rate 0,09%.


Sign in / Sign up

Export Citation Format

Share Document