scholarly journals Glucose Meter Performance Criteria for Tight Glycemic Control Estimated by Simulation Modeling

2010 ◽  
Vol 56 (7) ◽  
pp. 1091-1097 ◽  
Author(s):  
Brad S Karon ◽  
James C Boyd ◽  
George G Klee

Abstract Background: Glucose meter analytical performance criteria required for safe and effective management of patients on tight glycemic control (TGC) are not currently defined. We used simulation modeling to relate glucose meter performance characteristics to insulin dosing errors during TGC. Methods: We used 29 920 glucose values from patients on TGC at 1 institution to represent the expected distribution of glucose values during TGC, and we used 2 different simulation models to relate glucose meter analytical performance to insulin dosing error using these 29 920 initial glucose values and assuming 10%, 15%, or 20% total allowable error (TEa) criteria. Results: One-category insulin dosing errors were common under all error conditions. Two-category insulin dosing errors occurred more frequently when either 20% or 15% TEa was assumed compared with 10% total error. Dosing errors of 3 or more categories, those most likely to result in hypoglycemia and thus patient harm, occurred infrequently under all error conditions with the exception of 20% TEa. Conclusions: Glucose meter technologies that operate within a 15% total allowable error tolerance are unlikely to produce large (≥3-category) insulin dosing errors during TGC. Increasing performance to 10% TEa should reduce the frequency of 2-category insulin dosing errors, although additional studies are necessary to determine the clinical impact of such errors during TGC. Current criteria that allow 20% total allowable error in glucose meters may not be optimal for patient management during TGC.

2012 ◽  
Vol 58 (12) ◽  
pp. 1703-1710 ◽  
Author(s):  
Yeo-Min Yun ◽  
Julianne Cook Botelho ◽  
Donald W Chandler ◽  
Alex Katayev ◽  
William L Roberts ◽  
...  

BACKGROUND Testosterone measurements that are accurate, reliable, and comparable across methodologies are crucial to improving public health. Current US Food and Drug Administration–cleared testosterone assays have important limitations. We sought to develop assay performance requirements on the basis of biological variation that allow physiologic changes to be distinguished from assay analytical errors. METHODS From literature review, the technical advisory subcommittee of the Partnership for the Accurate Testing of Hormones compiled a database of articles regarding analytical and biological variability of testosterone. These data, mostly from direct immunoassay-based methodologies, were used to specify analytical performance goals derived from within- and between-person variability of testosterone. RESULTS The allowable limits of desirable imprecision and bias on the basis of currently available biological variation data were 5.3% and 6.4%, respectively. The total error goal was 16.7%. From recent College of American Pathologists proficiency survey data, most currently available testosterone assays missed these analytical performance goals by wide margins. Data from the recently established CDC Hormone Standardization program showed that although the overall mean bias of selected certified assays was within 6.4%, individual sample measurements could show large variability in terms of precision, bias, and total error. CONCLUSIONS Because accurate measurement of testosterone across a wide range of concentrations [approximately 2–2000 ng/dL (0.069–69.4 nmol/L)] is important, we recommend using available data on biological variation to calculate performance criteria across the full range of expected values. Additional studies should be conducted to obtain biological variation data on testosterone from women and children, and revisions should be made to the analytical goals for these patient populations.


Author(s):  
Snježana Kos ◽  
Arie van Meerkerk ◽  
Joke van der Linden ◽  
Theo Stiphout ◽  
Remi Wulkan

AbstractPoint-of-care (POC) glucose devices are widely used for insulin-dosage decision-making although such an application is not always permitted. In this study, we have evaluated a new generation of POC glucose device, the HemoCueThis study was performed according to the CLSI/STARD criteria. The 201DMRT was compared to the laboratory hexokinase glucose method (Siemens Dimension VistaThe 201DMRT showed a good agreement with the laboratory reference method. This was examined using Deming regression analysis, percentage Bland-Altman plot and a modified Clarke-error grid. The total analytical error at the clinically critical glucose concentrations of 5.6, 7.0 and 11.1 mmol/L (101, 126 and 200 mg/dL) was 6.4%, 4.3% and 3.0%, respectively. The total error among the different POC devices and among different cuvette lot numbers was <6.5%. Glucose measurements on the 201DMRT were not affected by changes in partial pressure of oxygen, whereas changes in hematocrit had influence on the results (3.4% for every 0.10 L/L change in hematocrit).The 201DMRT device can be used for glycemic control based on analytical results presented. However, the clinical applicability for tight glycemic control must be confirmed in a clinical study.