scholarly journals A Validated RP-HPLC Method for the Determination of Diltiazem in Raw Material and Pharmaceutical Dosage Form

Author(s):  
B M S Kumar ◽  
B. Rajkamal ◽  
B. Chandramowli

The objective of this work is to develop and validate a reverse phase high performance liquid chromatography (RP-HPLC) method for the quantitative analysis of Diltiazem in bulk and pharmaceutical dosage form. Chromatographic analyses were performed on RP C-18 column with a mobile phase consisting of 0.01M ammonium acetate in water, methanol and acetonitrile in the ratio 700:240:60 at a flow rate of 1 mL/min. The Diltiazem was detected and quantitated using a photodiode array detector at a wavelength of 295 nm with a retention time of 11.57 min. The detector response was linear in the concentration of 20-60μg/ml, the respective linear regression equation being Y=3000181x+356238.2. The limit of detection and limit of quantification were 0.5μg/ml and 0.15μg/ml respectively. The assay of Diltiazem in bulk was found to be 99.85%. From the recovery studies it was found that about 101% on average of Diltiazem was recovered which indicates high accuracy of the method. The method was validated by determining its accuracy, precision and system suitability. The method fulfilled the requirements for reliability and feasibility for application to the quantitative analysis of Diltiazem in bulk and pharmaceutical dosage form.

2008 ◽  
Vol 5 (s2) ◽  
pp. 1098-1102 ◽  
Author(s):  
N. Appala Raju ◽  
J. Venkateswara Rao ◽  
K. Vanitha Prakash ◽  
K. Mukkanti ◽  
K. Srinivasu

A simple, precise, rapid and accurate reverse phase HPLC method developed for the estimation of levetiracetam in tablet dosage form. A Sun Fire C18, 250 x 4.6 mm, 5 μm partical size, with mobile phase consisting of acetonitrile and 0.03 M potassium dihydrogen phosphate (pH adjusted to 3.0 with orthophosphoric acid) in the ratio of 15:85 v/v was used. The flow rate was 1 mL /min and the effluents were monitored at 210 nm. The retention time was 5.53 min. The detector response was linear in the concentration of 20-240 μg/mL. The respective linear regression equation being Y= 22119.684 x 6829.3428. The limit of detection and limit of quantification was 0.16 and 0.5 μg/mL respectively. The percentage assay of levetiracetam was 99.87%. The method was validated by determining its accuracy, precision and system suitability. The results of the study showed that the proposed RP-HPLC method is simple, rapid, precise and accurate, which is useful for the routine determination of levetiracetam in bulk drug and in its pharmaceutical dosage form.


Author(s):  
MADHURIMA BASAK ◽  
Santhosh Reddy Gouru ◽  
Animesh Bera ◽  
Krishna veni Nagappan

Objective: The present study aims at developing an accurate precise, rapid and sensitive Reverse Phase High-Performance Liquid Chromatography (RP-HPLC) method for assessing Empagliflozin in bulk drug and in the pharmaceutical dosage form. Methods: The proposed method employs a Reverse Phase Shim Pack C18 column (250 mm × 4.6 mm id; 5 µm) using a mobile phase comprising of acetonitrile and water in the ratio of 60:40 v/v flushed at a flow rate of 1 ml/min. The eluents were monitored at 223 nm. Results: Empagliflozin was eluted at a retention time of 5.417 min and established a co-relation co-efficient (R2>0.999) over a concentration ranging from 0.0495-100µg/ml. Percentage recovery was obtained between 98-102% which indicated that the method is accurate. The Limit of Detection (LOD) and Limit of Quantitation (LOQ) were found at 0.0125µg/ml and 0.0495µg/ml, respectively. Conclusion: An RP-HPLC method which was relatively simple, accurate, rapid and precise was developed and its validation was performed for the quantitative analysis of empagliflozin in bulk and tablet dosage form (10 and 25 mg) in accordance to International Conference of Harmonization (ICH) Q2 (R1) guidelines. The proposed method may aid in routinely analyzing empagliflozin in pharmaceuticals.


Author(s):  
G.M. Kadam ◽  
A.L. Puyad ◽  
T.M. Kalyankar

A new, economical, simple, accurate, and precise RP-HPLC method was developed for simultaneous assay and content uniformity determination of Sacubitril and Valsartan in bulk and pharmaceutical dosage form. The separation of Sacubitril and Valsartan was achieved within 6 minutes on Phenomenex Luna C18 250 mm x 4.6mm and 5µm Particle Size, column using Acetonitrile: Methanol: Water (30:55:15% v/v/v) as the mobile phase. Detection was carried out at 250 nm wavelength. The retention time of Sacubitril and Valsartan was found to be 2.361 and 3.304 min, respectively. The validation of the developed method was performed in terms of specificity, accuracy, precision, linearity, the limit of detection, the limit of quantification as mentioned in International Conference on Harmonization (ICH) guidelines. The method showed adequate sensitivity concerning linearity, accuracy, and precision over the range 12-36 μg/ml and 13-39 μg/ml for Sacubitril and Valsartan, respectively. The percentage recoveries obtained for Sacubitril and Valsartan were found to be in the range of 98.00 – 102.00 %. The proposed method is suitable for use in quality-control laboratories for quantitative analysis.


Author(s):  
Heena Ar Shaikh ◽  
Vandana Jain

Objective: A simple, accurate, precise, robust reverse phase high performance liquid chromatography (RP-HPLC) method was developed for the estimation of telmisartan and nebivolol hydrochloride (HCl) simultaneously in its combined dosage form.Methods: The compounds were well resolved in an isocratic method using the mobile phase composition of acetonitrile: Buffer (potassium dihydrogen orthophosphate pH adjusted 3.1 with orthophosphoric acid) in a ratio of 40:60 v/v at a flow rate of 1.2 ml/min using C18 Shim-pack (150 mm × 4.6 mm, 5 μ) column. The detection was carried out at 280 nm.Results: The retention time of telmisartan and nebivolol HCl was 4.8 min and 6.5 min, respectively. The developed method was validated by evaluating various validation parameters such as linearity, precision, accuracy, robustness, specificity, limit of detection, and limit of quantification according to the international council for harmonization guidelines. The standard calibration curve was obtained in the concentration range of 24–56 μg/ml for telmisartan and 3–7 μg/ml for nebivolol HCl. The overall average % recovery was found out to be 100.35 for telmisartan and 98.84 for nebivolol HCl.Conclusion: Statistical analysis of the data showed that the method is reproducible and selective for the estimation of telmisartan and nebivolol HCl. The proposed method could be used for analysis of telmisartan and nebivolol HCl in their dosage form.


2017 ◽  
Vol 9 (4) ◽  
pp. 76
Author(s):  
Panchumarthy Ravisankar ◽  
Shaheem Sulthana ◽  
Inturi Mary Thanuja ◽  
A. Dihitha Chowdary ◽  
J. Vyshnavi

Objective: The objective of the current study was to develop and validate a novel RP-HPLC method for determination of bamifylline hydrochloride in pharmaceutical dosage form.Methods: Chromatographic separation was conducted on Agilent technologies-1260 series with the G1311C quaternary pump, eclipse XDB C18 column (4.6 mm i.d. X 250 mm, 5 µm particle sizes) and equipped with photodiode array detector G1315D. Mobile phase consisted of methanol and acetonitrile were mixed in the ratio of 90:10 v/v, was used at a flow rate of 1 ml/min and detection wavelength was set at 263 nm.Results: The retention time for bamifylline hydrochloride was found to be 2.913 min. The calibration was linear (r2= 0.9996) in the concentration range of 2-10 µg/ml. The limit of detection and the limit of quantitation were found to be 0.4825 μg/ml and 1.4621 µg/ml respectively. Recovery of bamifylline hydrochloride in tablet formulation was observed in the range of 99.6-99.8 %. Percentage assay of bamifylline hydrochloride (Bamifix) was found to be 99.4 % w/w.Conclusion: Thus the novel proposed method for bamifylline hydrochloride was found to be feasible for the estimation of bamifylline hydrochloride in bulk as well as a pharmaceutical dosage form. 


Author(s):  
Sayyed Nazifa Sabir Ali ◽  
Lajporiya Mobina ◽  
Manjra Mehfuza ◽  
Patel Seema ◽  
Aejaz Ahmed ◽  
...  

Aims: To develop and validate a new, simple, rapid, precise, and accurate An Eco-friendly RP-HPLC and UV-Method Development and Validation for an estimation of Favipiravir in Bulk and pharmaceutical dosage form followed by Forced Degradation Studies. Study Design: This was employed for UV-visible (200-400 nm and 400-800 nm respectively) and RP-HPLC method development using C 18 inertsil column and optimization of variables for Favipiravir estimation in bulk and formulations. Place and Duration of the Study: The present work was carried out at Ali-allana College of Pharmacy, Akkalkuwa between the duration of November-2020 to February-2021. Methodology: UV-Spectroscopic method was developed for the estimation of Favipiravir in the bulk and pharmaceutical dosage form. The solvent selected for the Favipiravir UV analysis was water, the solution in a range of 2-10µg/ml was scanned in the UV region from 200-400 nm and the λmax value was determined. The RP-HPLC method was developed on inertsil ODS-3V C18 150 mm x 4.6mm x 5μ column using buffer pH 3.5: acetonitrile [90:10] as mobile phase at flow rate 1.0 ml/min and PDA detection at 358 nm. Results: The maximum absorbance was observed at 358 nm. The wavelength 358 nm was selected for further analysis of Favipiravir. The calibration curve was determined using drug concentrations ranging from 2-10 µg/ml. The % recovery for accuracy was 100.50-100.76%. The method was to be precise with a % RSD value 0.51-1.37 and 0.77-1.78 for intraday and Interday respectively. The limit of detection (LOD) and limit of quantification (LOQ) was found to be 0.0723 &0.219 µg/ml respectively by UV method. The RP-HPLC method was shown to be linear in the 50-250 μg/ml concentration range. The limit of detection (LOD) and limit of quantification (LOQ) was found to be 2.186 & 6.626 μg/ml respectively. The method was to be precise with a % RSD value 0.25-1.53 and 0.86-1.68 for intraday and inter-day respectively. Conclusion: Here we conclude that the developed UV and RP-HPLC methods are precise, accurate, sensitive, and reproducible for the quantitative estimation of Favipiravir bulk and its formulation. The developed method can be used by the pharmaceutical industries for the routine analysis of Favipiravir, in particular by UV and RP-HPLC. The main features of the proposed method are economic and eco-friendly with less retention time around 5.0 min.


2010 ◽  
Vol 7 (3) ◽  
pp. 807-812 ◽  
Author(s):  
Vanita Somasekhar ◽  
D. Gowri Sankar

A reverse phase HPLC method is described for the determination of esmolol hydrochloride in bulk and injections. Chromatography was carried on a C18column using a mixture of acetonitrile, 0.05 M sodium acetate buffer and glacial acetic acid (35:65:3 v/v/v) as the mobile phase at a flow rate of 1 mL/min with detection at 275 nm. The retention time of the drug was 4.76 min. The detector response was linear in the concentration of 1-50 μg/mL. The limit of detection and limit of quantification was 0.614 and 1.86 μg/mL respectively. The method was validated by determining its sensitivity, linearity, accuracy and precision. The proposed method is simple, economical, fast, accurate and precise and hence can be applied for routine quality control of esmolol hydrochloride in bulk and injections.


INDIAN DRUGS ◽  
2014 ◽  
Vol 51 (12) ◽  
pp. 32-36
Author(s):  
T. Vishalakhi ◽  
◽  
S. K Kumar ◽  
K Sujana ◽  
P Rani

A simple validated RP HPLC method for the estimation of rizatriptan benzoate in pharmaceutical dosage form and bulk was developed for routine analysis. This method was developed by selecting Agilent TC C18 (250 x 4.6 mm, 5 μ) column as stationary phase and acrylonibrile:water (45:55), pH adjusted to 3, as mobile phase. Flow rate of mobile phase was maintained at 4: 1 mL/min at ambient temperature throughout the experiment. Quantification was achieved with ultraviolet (DAD) detection at 220 nm. The retention time obtained for rizatriptan was 2.8 min. The detector response was linear in the concentration range of 2-25μg/mL. This method was validated and shown to be specific, sensitive, precise, linear, accurate, rugged and robust. Hence, this method can be applied for routine quality control of rizatriptan benzoate in dosage forms as well as in bulk drug.


2017 ◽  
Vol 9 (5) ◽  
pp. 33 ◽  
Author(s):  
Mitali V. Verma ◽  
Chirag J. Patel ◽  
M. M. Patel

Objective: To develop precise, accurate and reproducible stability assay method by RP-HPLC for estimation of dapagliflozin in API and pharmaceutical dosage form.Methods: The adequate separation was carried using agilent C18 (4.6 ml (millimeter)*150,5 µm (micromiter), mixture of acetonitrile: di-potassium hydrogen phosphate with pH-6.5 adjusted with OPA (40:60 %v/v) as a mobile phase with the flow rate of 1 ml/min (milliliter/minute) and the effluent was monitored at 222 nm (nanometer) using photo diode array detector. The retention time of dapagliflozin API and dapagliflozin tablet were 3.160 min (minute) and 3.067 min (minute) respectively.Results: Linearity for dapagliflozin was found in the range of 50-150µg/ml (microgram/milliliter) (R2 = 0.99) respectively. The accuracy of the present method was evaluated at 50 %, 100% and 150%. The % recoveries of dapagliflozin API and tablet were found to be in the range of 99.00–99.99 % and 98.50–99.99 % respectively. Precision studies were carried out and the relative standard deviation values were less than two. The method was found to be robust.Conclusion: The proposed method was found to be specific, accurate, precise and robust can be used for estimation of dapagliflozin in API and Pharmaceutical dosage form.


INDIAN DRUGS ◽  
2021 ◽  
Vol 58 (01) ◽  
pp. 28-34
Author(s):  
Yarasani Prashanthi ◽  
◽  
Faheem Ahmed ◽  
Tentu Nageswara Rao ◽  
Botsa Parvatamma ◽  
...  

A novel approach was used for develop and validate a rapid, accurate, and an isocratic RP-HPLC method with PDA detector for the estimation of ixazomib drug in pharmaceutical dosage forms. Ixazomib was seperated using Agilent 4.6*150 mm, 5μm analytical column, a Waters HPLC system (USA) and a mobile phase consisting of water and acetonitrile in the ratio of 40:60 V/V. The flow rate was set to 0.7 mL/min with 10µL injection volume. The column was maintained at ambient temperature, detector was set at wavelength of 274 nm. The retention time of ixazomib was found to be 2.17 min. The system suitability parameters for ixazomib such as theoretical plates and tailing factor were found to be 4146. Linearity was established for ixazomib such as theoretical plates and telling factor were found to be 4146. Linearilty was established for ixazomib in the range of 50-250 µg/ml concentration levels with correlation coefficients (r2) of 0.999. The intra-and inter-day precision % RSD values were found to be 0.47 and 0.31, respectively. The limit of detection (LOD) and limit of quantification (LOQ) were found to be 2.03 and 6.17 µg/mL respectively. The method was validated for all of the above parameters according to the International Conference on Harmonization (ICH) guidelines. This method can be used for estimation and analysis of ixazomib drug in active pharmaceutical ingredients and pharmaceuticals.