scholarly journals Simultaneous quantification of six main active constituents in Chinese Angelica by high-performance liquid chromatography with photodiode array detector

2013 ◽  
Vol 9 (34) ◽  
pp. 114 ◽  
Author(s):  
Wang Si-Wang ◽  
Tu Hong-Hai ◽  
Cao Wei ◽  
Li Ying
Author(s):  
Kanan G Gamit ◽  
Niraj Y Vyas ◽  
Nishit D Patel ◽  
Manan A Raval

Objective: A study was aimed to estimate guggulsterone-Z (GZ) in Gokshuradi Guggulu (GG).Methods: An analytical method was developed and validated using Waters Alliance high-performance liquid chromatography system (Empower software), equipped with photodiode array detector. Separation was achieved using Phenomenex, C-18 (250 mm×4.6 mm, 5 μ) column. Mobile phase consisted of acetonitrile:water (70:30,v/v). Flow rate was set to 1 ml/min and detection was performed at 251 nm.Results and Discussion: Validation parameters such as linearity, precision, accuracy, limit of detection, limit of quantification, and robustness were performed. Amount of GZ was estimated using linearity equation.Conclusion: GG was found to contain 0.815±0.03 g% w/w GZ. Validated method may be used as one of the parameters to standardize the formulation.


Sign in / Sign up

Export Citation Format

Share Document