Innovative high-performance liquid chromatography method development for the screening of 19 antimalarial drugs based on a generic approach, using design of experiments, independent component analysis and design space

2011 ◽  
Vol 1218 (31) ◽  
pp. 5205-5215 ◽  
Author(s):  
B. Debrus ◽  
P. Lebrun ◽  
J. Mbinze Kindenge ◽  
F. Lecomte ◽  
A. Ceccato ◽  
...  
Author(s):  
Sufiyan Ahmad ◽  
Mohammed Rageeb Mohammed Usman

  Objective: A simple rapid, accurate, precise, and reproducible validated reversed-phase high performance liquid chromatography method was developed for the determination of emtricitabine (EMB) and tenofovir (TEN) in bulk and tablet dosage forms.Methods: The quantification was carried out using symmetry Premsil C18 (250 mm×4.6 mm, 5 μm) Younglin (S.K.) gradient way using mobile phase comprising of methanol:water (70:30 v/v) pH 3 and a detection wavelength of 273 nm, and injection volume of 20 μL, with a flow rate of 1 ml/minutes.Results: In the developed method, the retention time of EMB and TEN were found to be 3.1667 minutes and 7.5000 minutes. The developed method was validated according to the International Conference on Harmonization (ICH) guidelines.Conclusion: The linearity, precision, range, robustness was within the limits as specified by the ICH guidelines. Hence, the method was found to be simple, accurate, precise, economic, and reproducible. Hence, it is worthwhile that the proposed methods can be successfully utilized for the routine quality control analysis EMB and TEN in bulk drug as well as in formulations.