Resolution of a Transmitted Electron Image Formed by A Scanning Electron Microscope

Author(s):  
S. Takashima ◽  
H. Hashimoto ◽  
S. Kimoto

The resolution of a conventional transmission electron microscope (TEM) deteriorates as the specimen thickness increases, because chromatic aberration of the objective lens is caused by the energy loss of electrons). In the case of a scanning electron microscope (SEM), chromatic aberration does not exist as the restrictive factor for the resolution of the transmitted electron image, for the SEM has no imageforming lens. It is not sure, however, that the equal resolution to the probe diameter can be obtained in the case of a thick specimen. To study the relation between the specimen thickness and the resolution of the trans-mitted electron image obtained by the SEM, the following experiment was carried out.

Author(s):  
K. Shibatomi ◽  
T. Yamanoto ◽  
H. Koike

In the observation of a thick specimen by means of a transmission electron microscope, the intensity of electrons passing through the objective lens aperture is greatly reduced. So that the image is almost invisible. In addition to this fact, it have been reported that a chromatic aberration causes the deterioration of the image contrast rather than that of the resolution. The scanning electron microscope is, however, capable of electrically amplifying the signal of the decreasing intensity, and also free from a chromatic aberration so that the deterioration of the image contrast due to the aberration can be prevented. The electrical improvement of the image quality can be carried out by using the fascionating features of the SEM, that is, the amplification of a weak in-put signal forming the image and the descriminating action of the heigh level signal of the background. This paper reports some of the experimental results about the thickness dependence of the observability and quality of the image in the case of the transmission SEM.


1997 ◽  
Vol 3 (S2) ◽  
pp. 1243-1244 ◽  
Author(s):  
Raynald Gauvin ◽  
Steve Yue

The observation of microstructural features smaller than 300 nm is generally performed using Transmission Electron Microscopy (TEM) because conventional Scanning Electron Microscopes (SEM) do not have the resolution to image such small phases. Since the early 1990’s, a new generation of microscopes is now available on the market. These are the Field Emission Gun Scanning Electron Microscope with a virtual secondary electron detector. The field emission gun gives a higher brightness than those obtained using conventional electron filaments allowing enough electrons to be collected to operate the microscope with incident electron energy, E0, below 5 keV with probe diameter smaller than 5 nm. At 1 keV, the electron range is 60 nm in aluminum and 10 nm in iron (computed using the CASINO program). Since the electron beam diameter is smaller than 5 nm at 1 keV, the resolution of these microscopes becomes closer to that of TEM.


Author(s):  
J. R. Sellar

To study chemical reactions in situ and observe biological specimens in the hydrated state, it is necessary to isolate the specimen from the vacuum of the electron microscope by means of a cell. Figure 1 represents such a cell in a conventional transmission electron microscope (CTEM) with apertures or windows A, specimen mounting B, and specimen T. O labels the objective lens, L the cell gas, and S is the width of the layer of gas.Calculations have been carried out involving the contrast available when thick specimens are imaged or an environmental cell is used at 1000 keV. For the limitations considered, including multiple scattering, chromatic aberration and the effects of various geometrical configurations, it appears that, in the absence of noise effects due to loss of intensity, multiple scattering causes the largest disc of confusion.


Author(s):  
K. Shi rota ◽  
A. Yonezawa ◽  
K. Shibatomi ◽  
T. Yanaka

As is well known, it is not so easy to operate a conventional transmission electron microscope for observation of magnetic materials. The reason is that the instrument requires re-alignment of the axis and re-correction of astigmatism after each specimen shift, as the lens field is greatly disturbed by the specimen. With a conventional electron microscope, furthermore, it is impossible to observe magnetic domains, because the specimen is magnetized to single orientation by the lens field. The above mentioned facts are due to the specimen usually being in the lens field. Thus, special techniques or systems are usually required for magnetic material observation (especially magnetic domain observation), for example, the technique to switch off the objective lens current and Lorentz microscopy. But these cannot give high image quality and wide magnification range, and furthermore Lorentz microscopy is very complicated.


Author(s):  
Oliver C. Wells

The low-loss electron (LLE) image in the scanning electron microscope (SEM) is useful for the study of uncoated photoresist and some other poorly conducting specimens because it is less sensitive to specimen charging than is the secondary electron (SE) image. A second advantage can arise from a significant reduction in the width of the “penetration fringe” close to a sharp edge. Although both of these problems can also be solved by operating with a beam energy of about 1 keV, the LLE image has the advantage that it permits the use of a higher beam energy and therefore (for a given SEM) a smaller beam diameter. It is an additional attraction of the LLE image that it can be obtained simultaneously with the SE image, and this gives additional information in many cases. This paper shows the reduction in penetration effects given by the use of the LLE image.


Author(s):  
K. Tsuno ◽  
Y. Harada ◽  
T. Sato

Magnetic domains of ferromagnetic amorphous ribbon have been observed using Bitter powder method. However, the domains of amorphous ribbon are very complicated and the surface of ribbon is not flat, so that clear domain image has not been obtained. It has been desired to observe more clear image in order to analyze the domain structure of this zero magnetocrystalline anisotropy material. So, we tried to observe magnetic domains by means of a back-scattered electron mode of high voltage scanning electron microscope (HVSEM).HVSEM method has several advantages compared with the ordinary methods for observing domains: (1) high contrast (0.9, 1.5 and 5% at 50, 100 and 200 kV) (2) high penetration depth of electrons (0.2, 1.5 and 8 μm at 50, 100 and 200 kV). However, image resolution of previous HVSEM was quite low (maximum magnification was less than 100x), because the objective lens cannot be excited for avoiding the application of magnetic field on the specimen.


Author(s):  
Shaopeng Hu ◽  
Jianhua Wang ◽  
Zhen Li ◽  
Huei Chen ◽  
Fei Cu ◽  
...  

Gastritis from returning bile is a common disease, but the reason for the disease is not clear. As the pathologic ultrastructure research progresses, it has drawn attention to the ultrastructural change of cells in gastric mucosa by clinical workers. We observed gastric mucosa tissues of 15 patients suffering from gastritis with a transmission electron microscope (TEM) and a scanning electron microscope (SEM). It is the first report in China that fungus exists in the lamina propria of gastric mucosa tissue. The result is as follows.The gastric mucosa tissues of 15 patients suffering from gastritis were acquired by stomachoscopy. Both TEM and SEM specimens were prepared by the usual methods. Under the TEM, the epithelial surface became higher and larger. Mitochondria of the cells were swollen and cristae were disrupted. There were vacuoles in the cells. The nucleus showed disorder, heterochromatin became darker, and nucleolae could be observed.


Author(s):  
W. P. Wergin ◽  
S. Roy ◽  
E. F. Erbe ◽  
C. A. Murphy ◽  
C. D. Pooley

Larvae of the nematode, Steinernema carpocapsae Weiser strain All, were cryofixed and freezesubstituted for 3 days in acetone containing 2% osmium tetroxide according to established procedures. Following chemical fixation, the nematodes were brought to room temperature, embedded in Spurr's medium and sectioned for observation with a Hitachi S-4100 field emission scanning electron microscope that was equipped with an Oxford CT 1500 Cryotrans System. Thin sections, about 80 nm thick, similar to those generally used in conventional transmission electron microscope (TEM) studies were mounted on copper grids and stained with uranyl acetate for 30 min and lead citrate for 5 min. Sections about 2 μm thick were also mounted and stained in a similar fashion. The grids were mounted on an Oxford grid holder, inserted into the microscope and onto a cryostage that was operated at ambient temperature. Thick and thin sections of the larvae were evaluated and photographed in the SEM at different accelerating voltages. Figs. 4 and 5 have undergone contrast conversion so that the images would resemble transmitted electron micrographs obtained with a TEM.


Author(s):  
William P. Wergin ◽  
Eric F. Erbe ◽  
Terrence W. Reilly

Although the first commercial scanning electron microscope (SEM) was introduced in 1965, the limited resolution and the lack of preparation techniques initially confined biological observations to relatively low magnification images showing anatomical surface features of samples that withstood the artifacts associated with air drying. As the design of instrumentation improved and the techniques for specimen preparation developed, the SEM allowed biologists to gain additional insights not only on the external features of samples but on the internal structure of tissues as well. By 1985, the resolution of the conventional SEM had reached 3 - 5 nm; however most biological samples still required a conductive coating of 20 - 30 nm that prevented investigators from approaching the level of information that was available with various TEM techniques. Recently, a new SEM design combined a condenser-objective lens system with a field emission electron source.


Author(s):  
Gao Fengming

Transmission electron microscope(TEM) and scanning electron microscope(SEM) were widely used in experimental tumor studies. They are useful for evaluation of cellular transformation in vitro, classification of histological types of tumors and treating effect of tumors. We have obtained some results as follows:1. Studies on the malignant transformation of mammalian cells in vitro. Syrian golden hamster embryo cells(SGHEC) were transformed in vitro by ThO2 and/or ore dust. In a few days after dust added into medium, some dust crystals were phagocytized. Two weeks later, malignant transformation took place. These cells were of different size, nuclear pleomorphism, numerous ribosomes, increasing of microvilli on cell surface with various length and thickness, and blebs and ruffles(Figs. 1,2). Myelomonocytic leukemic transformation of mouse embryo cells(MEC) was induced in vitro by 3H-TdR. Transformed cells were become round from fusiform. The number of mitochondria and endoplasmic reticulum was reduced, ribosomes and nucleoli increased, shape of nuclei irregular, microvilli increased, and blebs and ruffles appeared(Fig. 3).


Sign in / Sign up

Export Citation Format

Share Document