Electrophoretic patterns post daratumumab

Author(s):  
Joanna Sheldon ◽  
Rachel D Wheeler ◽  
Ray Powles

Background Daratumumab (Darzalex) is a human IgG1 kappa monoclonal antibody targeting CD38 that has been recently approved for the treatment of refractory multiple myeloma. As it is a monoclonal protein, it can be detected on routine serum protein electrophoresis and by immunofixation. Methods Serum samples from four patients were analysed by serum protein electrophoresis immediately pre- and post-treatment with daratumumab. Results For all four patients, daratumumab was visible on serum protein electrophoresis as an additional small band (approximately 1 g/L) in the slow gamma region. Conclusion Diagnostic laboratories should be aware that daratumumab can be detected on routine serum protein electrophoresis of myeloma patients and should liaise closely with clinicians to ensure the presence of daratumumab is not misinterpreted as development of a new monoclonal protein.

Blood ◽  
2010 ◽  
Vol 116 (21) ◽  
pp. 5025-5025
Author(s):  
Gregory D. Bianchi ◽  
Peter M Voorhees ◽  
Shirley A. Hainsworth ◽  
Herbert C. Whinna ◽  
John F. Chapman ◽  
...  

Abstract Abstract 5025 Background and objectives: Serum protein electrophoresis (SPE) is widely used for the detection and surveillance of patients with plasma cell dyscrasias. This study was initiated following the detection of a new IgG heavy-chain immunoglobulin in a 51-year old female patient with a known history of IgDk multiple myeloma. The laboratory finding was unusual because of the rarity of an apparent IgG heavy-chain gammopathy in combination with an IgDk as well as a low quantitative serum IgG of 255 mg/dL (reference interval 600–1700). Interference was suspected and after contacting the treating physician it was established that the patient was enrolled in a phase II clinical trial with the monoclonal antibody Situximab (Centocor, Inc., Malvern, PA, USA) in combination with dexamethasone. Siltuximab is a high affinity antibody to human interleukin-6 (IL-6) currently used in clinical trials for the treatment of a number of malignancies, including multiple myeloma. The objectives of this study were to determine if monoclonal immunoglobulin therapies interfere with SPE testing and commonly used immunoassays. Methods: Siltuximab was obtained from the manufacturer and tested as an interferent using two different SPE platforms: Capillarys II (Sebia, USA) and Hydrasys (Sebia, USA). Siltuximab was analyzed alone or spiked into normal reference sera at concentrations between 50–600 mcg/mL to reflect the reported range of therapeutic circulating drug concentrations. We also examined additional monoclonal therapies at their reported mean peak serum drug concentrations. The therapies tested represented both chimeric human-mouse immunoglobulins (Rituximab, Siltuximab, Infliximab, Cetuximab) and humanized antibodies (Trastuzumab, Bevacizumab, Adalimumab). We also tested the effect of Situximab on a variety of automated serum immunoassays including hCG, iPTH, troponin T (Roche Diagnostics), cortisol, Hs-CRP, PSA, RF, and TSH (Ortho Clinical Diagnostics). Results: Siltuximab was evident as an IgGk monoclonal protein at a threshold of 100 mcg/mL and detected by all methods tested including capillary electrophoresis, immunosubtraction, agarose gel electrophoresis, and immunofixation (IFE). We also tested other widely prescribed monoclonal therapies at their reported mean peak serum drug concentrations: Rituximab (Rituxan), Trastuzumab (Herceptin), Bevacizumab (Avastin), Infliximab (Remicade), Cetuximab (Erbitux), and Adalimumab (Humira). Each drug was readily detected using IFE as an IgGk monoclonal protein (Fig 1A). Review of immunofixation results from 13 additional patients receiving Siltuximab revealed a discrete IgGk monoclonal protein corresponding to the migration of the drug in 11/13 patients (4 examples are shown in Fig 1B). In the other two cases, endogenous IgGk monoclonal immunoglobulins overlapped with the electrophoretic patterns of Siltuximab making them indistinguishable. We therefore tested two different approaches to cope with interference from monoclonal therapies: 1) adsorption of drug with specific antisera and 2) patient testing after allowing time for drug clearance. Pre-incubation of Siltuximab with anti-drug antibodies shifted the drug electrophoretic pattern such that it could potentially be differentiated from endogenous monoclonal proteins. We also observed that Siltuximab was undetectable by IFE in patients 3–4 months after cessation of therapy. This time period corresponds to approximately 5 half-lives (median half-life of siltuximab is 17.8 days). We observed no effect (<5% bias) from monoclonal therapies in commonly used immunoassays (hCG, PTH, troponin T, cortisol, Hs-CRP, PSA, RF, TSH) or quantitative immunoglobulin tests (IgG, IgA, IgM). Conclusions: Interference of monoclonal antibody therapies may be of clinical significance in patients being investigated or monitored for plasma cell dyscrasias. Individuals receiving these therapies may be subjected to unnecessary follow-up testing and patients with an IgGk gammopathy may be misconstrued as having recurrent or resistant disease. This study highlights that the possibility of false positive testing on IFE needs to be considered given the increasing use of these agents in clinical practice. Pre-incubation with anti-drug antibodies or testing after clearance of the therapeutic immunoglobulins can help clarify concerns of monoclonal therapy interference with IFE testing. Disclosures: Voorhees: Millennium Pharmaceuticals: Speakers Bureau; Celgene: Speakers Bureau.


Author(s):  
Rachel D Wheeler ◽  
Micsha V Costa ◽  
Asante Crichlow ◽  
Fenella Willis ◽  
Yasmin Reyal ◽  
...  

Multiple myeloma is a haematological cancer caused by malignant plasma cells in the bone marrow that can result in organ dysfunction and death. Recent novel treatments have contributed to improved survival rates, including monoclonal antibody therapies that target the CD38 protein on the surface of plasma cells. Anti-CD38 therapies are IgG kappa monoclonal antibodies that are given in doses high enough for the drug to be visible on serum protein electrophoresis as a small paraprotein. We present a case where isatuximab, the most recent anti-CD38 monoclonal antibody to be approved for treatment of myeloma, obscured the patient’s paraprotein on gel immunofixation, so that complete remission could not be demonstrated. This was resolved using the isatuximab Hydrashift assay. The interference on gel immunofixation was unexpected because isatuximab migrated in a position distinct from the patient’s paraprotein on capillary zone electrophoresis. We demonstrate the surprising finding that isatuximab migrates in a different position on gel electrophoresis compared to capillary zone electrophoresis. It is vital that laboratories are aware of the possible interference on electrophoresis from anti-CD38 monoclonal antibody therapies, and are able to recognise these drugs on protein electrophoresis. The difference in isatuximab’s electrophoretic mobility on capillary and gel protein electrophoresis makes this particularly challenging. Laboratories should have a strategy for alternative analyses in the event that the drugs interfere with assessment of the patient’s paraprotein.


2018 ◽  
Vol 51 ◽  
pp. 61-65 ◽  
Author(s):  
Mina Salamatmanesh ◽  
Christopher R. McCudden ◽  
Arleigh McCurdy ◽  
Ronald A. Booth

Blood ◽  
2009 ◽  
Vol 114 (22) ◽  
pp. 1803-1803
Author(s):  
Melissa Snyder ◽  
Angela Dispenzieri ◽  
S.Vincent Rajkumar ◽  
Robert Kyle ◽  
Joanne Benson ◽  
...  

Abstract Abstract 1803 Poster Board I-829 Background Plasma cell proliferative disorders are monitored by a variety of methods. Serum protein electrophoresis (SPEP) and/or urine PEP M-spike quantitation are commonly assessed in patients with monoclonal gammopathy of undetermined significance (MGUS), smoldering multiple myeloma (SMM), and multiple myeloma (MM) to determine disease progression, response, or relapse. Serum immunoglobin (Ig) concentrations can be quantitated when the M-spike is large or if the migration is obscured within the SPEP beta fraction. Serum FLC quantitation provides a rapid indicator of response, will detect the rare occurrence of FLC escape, and will allow disease monitoring in the absence of a measurable serum or urine M-spike. The International Myeloma Working Group (IMWG) has recommended that the serum and urine M-spike should be used to monitor disease, and that FLC quantitation should be used only if there is no measurable disease by electrophoresis and if the monoclonal FLC concentration is greater than 10 mg/dL in the context of an abnormal FLC K/L ratio. We have studied serial samples in clinically stable patients in order to assess the total variability (analytic and biologic) of these assays and to confirm the IMWG recent recommendations. Methods Serial data from stable MGUS patients (n=35) were identified by the availability of all 3 serum test results (M-spike, Ig, FLC) in at least 4 serial samples that were obtained 9 months to 5 years apart and whose serum M-spikes varied by less than 25%. For MM (n=60) and SMM (n=48) the samples were within 9-15 months and serum M-spikes varied by less than 0.5 g/dL. Among the 60 MM, 48 SMM, and 35 MGUS patients, there were 23, 41, and 18 patients with measurable disease by serum M-spike (i.e. M-spike >1 g/dL); 19, 10, and 10 patients with an evaluable FLC (i.e. monoclonal FLC > 10 mg/dL and an abnormal FLC ratio); and 5, 5, and 1 patient with an evaluable urine (i.e. M-spike > 200mg/24 hr). The FLC data was analyzed as the involved FLC concentration (iFLC), the difference between the involved and uninvolved FLC concentration (dFLC), and the FLC K/L ratio (rFLC). The coefficients of variability (CV) were determined for each methodology in each patient sample set, and the average CVs were determined. Igs were quantitated by immunonephelometry using a Siemens BNII and Siemens reagent sets; kappa and lambda FLC were quantitated using a Siemens BNII and Freelite reagent sets from The Binding Site; M-spikes were quantitated using Helena SPIFE SPE and reagent sets. Results The CVs for the Ig quantitation, SPEP M-spike, FLC quantitation, and urine M-spike in each of the patient groups are listed in the table: Our laboratory's interassay analytic CV for replicate samples are 4-5% for Ig quantitation, 6-8% for SPEP M-spikes, 6-7% for FLC quantitation, and 5-7% for urine M-spikes. The analytic CVs of the methods are similar, but the total (analytic + biologic) CVs are very different. The samples have been selected by restricting the variability of serum M-spike values; when we apply the same criteria to the IgG quantitation, the IgG total CV comes closer to the serum M-spike CVs. The remaining differences, however, may be due to biologic variability contributed by polyclonal Ig. The total CV for iFLC is similar to the urine M-spike CV and suggests a previously unrecognized large biologic CV for serum FLC. The iFLC and dFLC CVs were comparable but were smaller than the rFLC CV. Conclusion The variability of the serum and urine M-spike, Ig, and FLC measurements confirm the IMWG recommendations for patient monitoring. If a patient has a measurable M-spike >1 g/dL, then the serum M-spike should be followed. If there is no measurable disease, then the iFLC can be monitored, provided that the rFLC is abnormal and the iFLC concentration is >10 mg/dL. Although the number of patients with evaluable urine M-spikes in this study is small, larger studies may confirm the utility of serum FLC compared to urine M-spike for monitoring patients with monoclonal gammopathies. Disclosures No relevant conflicts of interest to declare.


2020 ◽  
Vol 30 (2) ◽  
pp. 265-271
Author(s):  
Anamarija Rade ◽  
Anamarija Đuras ◽  
Irena Kocijan ◽  
Patricija Banković Radovanović ◽  
Ana Turčić

Introduction: Serum samples of haemodialysed patients collected through vascular access devices, e.g. central venous catheter (CVC) can contain residual heparin, which can cause incomplete clotting and consequently fibrinogen interference in serum protein electrophoresis (SPE). We hypothesized that this problem may be overcome by addition of thrombin and aimed to find a simple thrombin-based method for fibrinogen interference removal. Materials and methods: Blood samples of 51 haemodialysed patients with CVC were drawn through catheter into Clot Activator Tube (CAT) and Rapid Serum Tube Thrombin (RST) vacutainers (Becton Dickinson, New Jersey, USA) following the routine hospital protocols and analysed with gel-electrophoresis (Sebia, Lisses, France). Samples were redrawn in the CAT tubes and re-analysed after being treated with thrombin using two methods: transferring CAT serum into RST vacutainer and treatment of CAT serum with fibrinogen reagent (Multifibren U, Siemens, Marburg, Germany). Results: Direct blood collection in RST proved to be slightly more efficient than CAT in removing the interfering band in beta fraction (CAT removed 6/51 and RST removed 12/51, P = 0.031). Transferring CAT serum into the RST vacutainer proved to be more efficient for subsequent removal of interfering band from CAT serum than the addition of fibrinogen reagent (39/45 vs. 0/45 samples with efficiently removed interfering band, P &lt; 0.001). Conclusion: Fibrinogen interference caused by incomplete clotting because of residual heparin can be overcome by addition of thrombin. Transferring CAT serum into the RST vacutainer was the most efficient method.


2019 ◽  
Vol 152 (Supplement_1) ◽  
pp. S146-S147
Author(s):  
Roula Katerji ◽  
Tamera Paczos ◽  
Li Liu

Abstract Objectives Serum protein electrophoresis (SPE) and immunofixation electrophoresis (IFE) are commonly used to screen and monitor monoclonal gammopathies. Currently, there are no consensus guidelines on optimal testing frequency leading to overutilization. Here we examined the testing frequencies of SPE and IFE in our institution to provide an evidence-based perspective on efficient test utilization. Methods We retrospectively reviewed all SPE and IFE tests performed in 2018. Ordering patterns and testing frequencies were analyzed. In cases with more than monthly repeats, electronic medical records were reviewed to follow the result changes over time. Results There were 10,054 SPE and 4,248 IFE orders in 2018. The 4,248 IFE cases represented 2,439 patients, among whom 104 patients (4.3%) had IFE repeated more frequently than every 2 months and 50 patients (2%) more frequently than monthly. The 10,054 SPE cases represented 5,472 patients; 127 patients (2.3%) had SPE performed more than 12 times. Rare cases (0.1%) had SPE and IFE repeated every 1 to 2 weeks. Most IFE tests were ordered together with SPE (89% of all IFE orders), among which 28% of cases had normal SPE findings. Among the cases with more than monthly SPE and IFE tests, IFE results showed meaningful interpretation changes in a minimum period of 2 to 3 months; 35% cases had no IFE interpretation changes throughout the year. In contrast, during active treatment period of multiple myeloma, SPE detected paraprotein level change weekly. Conclusion IFE is overutilized and monthly monitoring does not add value even during active treatment of multiple myeloma. Our results support the need for the development of testing frequency guidelines to avoid overutilization and provide more cost-effective patient care.


2019 ◽  
Vol 3 (5) ◽  
pp. 857-863 ◽  
Author(s):  
Katie L Thoren ◽  
Matthew J Pianko ◽  
Youssef Maakaroun ◽  
C Ola Landgren ◽  
Lakshmi V Ramanathan

Abstract Background Daratumumab, a monoclonal antibody used to treat relapsed or refractory multiple myeloma, can interfere with protein electrophoresis and immunofixation assays. False-positive immunofixation results due to daratumumab can cause uncertainty regarding the status of a patient's disease and lead to potential misclassification of their response to therapy. The Hydrashift 2/4 Daratumumab assay (Sebia) was recently cleared by the Food and Drug Administration for resolving daratumumab interference on immunofixation. Here, we evaluate the performance of the Hydrashift assay in multiple myeloma patients receiving treatment with daratumumab-based regimens. Methods Waste serum samples from multiple myeloma patients (n = 40) receiving daratumumab were analyzed by standard immunofixation and the Hydrashift assay. Results from these tests were compared and were evaluated along with pretreatment serum protein electrophoresis and immunofixation results, if available. Results The Hydrashift assay shifted the migration of daratumumab in patient samples. In 27 cases, the patient's M protein was distinguishable from daratumumab by standard immunofixation. In these cases, the Hydrashift assay confirmed that the IgGκ band was daratumumab and helped identify the presence of treatment-related oligoclonal bands. There were 11 instances in which the patient's IgGκ M protein comigrated with daratumumab. In all 11 cases, the Hydrashift assay confirmed the presence of residual M protein. Finally, in 2 patients whose pretreatment immunofixation results were not available, the Hydrashift assay confirmed that the IgGκ band visible on immunofixation was due to daratumumab alone. Conclusions The Hydrashift 2/4 Daratumumab assay is a useful tool to clarify the source of an IgGκ band on immunofixation and allow a patient's M protein to be viewed without interference.


Blood ◽  
2004 ◽  
Vol 104 (11) ◽  
pp. 4856-4856
Author(s):  
Arthur R. Bradwell ◽  
Jean Garbincius ◽  
Earle W. Holmes

Abstract Serum free light chain measurements have been shown to be useful in the diagnosis and monitoring of patients with monoclonal gammopathies. The present study was undertaken to evaluate the effect of adding the measurement of serum free light chain kappa to lambda ratios to the serum protein electrophoresis evaluation that we typically use as an initial screen for the detection of monoclonal proteins. We retrospectively tested 347 consecutive samples from individuals who had no previous history of plasma cell dyscrasia and had not previously had a serum or urine electrophoresis or immunofixation electrophoresis test at our institution. The quantitative serum protein electrophoresis test that was ordered was performed using Hydragel Beta 1- Beta 2 gels and Hydrasis instrument (Sebia, Inc., Norcross, GA). The protein content of the electrophoresis zones were quantitated by scanning densitometry and the electrophoresis pattern of each sample was qualitatively examined for abnormal bands and suspicious findings by a single, experienced observer. Serum free light chain concentrations and the serum free light chain kappa to lambda ratios were determined using the Freelite Human Kappa and Lambda Kits (The Binding Site Ltd, Birmingham, UK) and the Immage analyzer (Beckman Coulter Inc., Brea, CA). The serum free light chain kappa to lambda ratios were outside the reference interval (0.25 to1.65) in 23 of the samples. Ten abnormal ratios were observed among a group of 57 samples that had either positive or suspicious qualitative evaluations for the presence of a restriction or that demonstrated hypo-gammaglobulinemia. Both abnormalities led to recommendations for follow-up testing, which confirmed the presence of a monoclonal protein in 21 of the samples. Six abnormal ratios were observed among a group of 159 specimens that had quantitative abnormalities in albumin or one or more of globulin fractions (hypo-gammaglobulinemia excepted) and normal qualitative evaluations. Seven abnormal ratios were observed among a group of 131 samples that had normal quantitative results and normal qualitative evaluations. Follow-up testing is not usually recommended for serum protein electrophoresis results like those in the latter two groups. We found that the addition of the serum free light chain kappa to lambda ratio to the serum protein electrophoresis test increased the number of abnormal screens that would have required further clinical and/or laboratory evaluation by 23%(i.e. from 57 to 70). Given the high specificity of the serum free light chain kappa to lambda ratio for monoclonal light chains, the additional 13 abnormal samples identified by this test are expected to have a high likelihood of harboring a monoclonal protein that would have otherwise eluded detection. Pending a definitive prospective study, we estimate that the addition of a serum free light chain kappa to lambda ratio to the serum protein electrophoresis screen would increase the rate of detection of serum monoclonal proteins by as much as 1.6-fold.


Sign in / Sign up

Export Citation Format

Share Document