scholarly journals Method Development and Its Validation for Simultaneous Estimation of Domperidone and Esomeprazole by RP-HPLC in Combination Tablet Dosage Form

Author(s):  
V. Tejasvi Reddy ◽  
S. Hari Hara Kumar ◽  
M Akiful Haque ◽  
Dr. Vasudha Bakshi

A simple and validated reverse phase high performance liquid chromatography (RP-HPLC) method was developed for simultaneous estimation of Esomeprazole and Domperidone. Different HPLC Chromatographic conditions were used to find out the optimum chromatographic condition for best elution of drugs. Good chromatographic separation was achieved on Develosil ODS HG-5 RP C18 (15cmX4.6mm, 5mm Particle size), analytical column using a mixture of Potassium dihydrogen phosphate buffer pH: acetonitrile in the ratio of 45:55 v/v used as mobile phase at the flow rate of 1 ml/min and detector wavelength at 284nm. The validation of the proposed method was carried out for specificity, linearity and range, accuracy, precision, limit of detection and limit of quantification. Linearity range was found to be 0-140 µg/ml and 0-150 µg/ml with correlation coefficients of 0.995 & 0.994 and retention time of 2.59 and 3.85 min for Domperidone and Esomeprazole respectively. The LOD was found to be 0.341 mg/ml and 2.031 mg/ml and LOQ was found to be 1.023 mg/ml and 6.093 mg/ml respectively. The result shows the developed method is yet another suitable method for assay and stability studies which can help in the analysis of Domperidone & Esomeprazole in different formulations.

Author(s):  
V. Tejasvi Reddy ◽  
S. Hari Hara Kumar ◽  
M Akiful Haque ◽  
Vasudha Bakshi

A simple and validated reverse phase high performance liquid chromatography (RP-HPLC) method was developed for simultaneous estimation of Esomeprazole and Domperidone. Different HPLC Chromatographic conditions were used to find out the optimum chromatographic condition for best elution of drugs. Good chromatographic separation was achieved on Develosil ODS HG-5 RP C18 (15cmX4.6mm, 5m Particle size), analytical column using a mixture of Potassium dihydrogen phosphate buffer pH: acetonitrile in the ratio of 45:55 v/v used as mobile phase at the flow rate of 1 ml/min and detector wavelength at 284nm. The validation of the proposed method was carried out for specificity, linearity and range, accuracy, precision, limit of detection and limit of quantification. Linearity range was found to be 0-140 μg/ml and 0-150 μg/ml with correlation coefficients of 0.995 and 0.994 and retention time of 2.59 and 3.85 min for Domperidone and Esomeprazole respectively. The LOD was found to be 0.341 g/ml and 2.031 g/ml and LOQ was found to be 1.023 g/ml and 6.093 g/ml respectively. The result shows the developed method is yet another suitable method for assay and stability studies which can help in the analysis of Domperidone and Esomeprazole in different formulations.


Author(s):  
V. Tejasvi Reddy ◽  
S. Hari Hara Kumar ◽  
M Akiful Haque ◽  
Dr. Vasudha Bakshi

A simple and validated reverse phase high performance liquid chromatography (RP-HPLC) method was developed for simultaneous estimation of Esomeprazole and Domperidone. Different HPLC Chromatographic conditions were used to find out the optimum chromatographic condition for best elution of drugs. Good chromatographic separation was achieved on Develosil ODS HG-5 RP C18 (15cmX4.6mm, 5μm Particle size), analytical column using a mixture of Potassium dihydrogen phosphate buffer pH: acetonitrile in the ratio of 45:55 v/v used as mobile phase at the flow rate of 1 ml/min and detector wavelength at 284nm. The validation of the proposed method was carried out for specificity, linearity and range, accuracy, precision, limit of detection and limit of quantification. Linearity range was found to be 0-140 μgg/ml and 0-150 μgg/ml with correlation coefficients of 0.995 & 0.994 and retention time of 2.59 and 3.85 min for Domperidone and Esomeprazole respectively. The LOD was found to be 0.341 μg/ml and 2.031 μg/ml and LOQ was found to be 1.023 μg/ml and 6.093 μg/ml respectively. The result shows the developed method is yet another suitable method for assay and stability studies which can help in the analysis of Domperidone & Esomeprazole in different formulations.


Author(s):  
ANKITA SHARMA ◽  
INDER KUMAR ◽  
KARAN RANA

Objective: The present study was undertaken to develop and validate an RP-HPLC method for the combination of imiquimod and salicylic acid Methods: The method was carried out on Nucleodur C18 (250 mm × 4.6 mm I.D., 5 ????m) using low-pressure gradient elution mode. The mobile phase was used as 30M potassium dihydrogen phosphate and acetonitrile (45:55) pH 6.5 adjusted using ortho-phosphoric acid. The concentration of solvents was 1-20 µg/ml and the volume of injection was 20 mcl with the flow rate of 1.0 ml/min. The absorption maxima of salicylic acid and imiquimod were found 234 nm and 226 nm, respectively. Results: The method was validated and showed the linearity greater than 0.99% and with precision (RSD%<1). The limit of detection (LOD) and limit of quantification (LOQ) of salicylic acid was found to be 0.09756 µg/ml and 0.2956 µg/ml, respectively, and imiquimod was found to be 0.044031 µg/ml and 0.13334 µg/ml, respectively. Conclusion: The method developed in the present study was found to be sensitive, specific, and can be applied for the simultaneous estimation of imiquimod and salicylic acid.


Author(s):  
Kumar S. Ashutosh ◽  
Debnath Manidipa ◽  
Rao J.V.L.N. Seshagiri ◽  
Sankar D. Gowri

This paper is concern with a reverse phase high performance liquid chromatography (RP-HPLC) bio-analytical method development and validation for Prasugrel in human plasma using photo diode array detector (PDA detector). The HPLC separation was carried out in an isocratic mode on an X-Terra C18 column (4.6 x 150 mm; 5 μm) with a mobile phase consisting of potassium dihydrogen phosphate [pH 3.0] and acetonitrile in the ratio of 30:70 v/v at a flow rate of 1.0 mL/min. The run time was maintained for 5 mins and the detection was monitored at 210 nm. The percentage recovery was found 99.61-100.06 in human plasma. This reveals that the method is quite accurate. The linearity was found 15-40 μg/mL in human plasma. The inter-day and intra-day precision in plasma was found within the limits. The lower limit of quantification (LLOQ) obtained by the proposed method was 0.05 μg/mL. The percentage relative standard deviation (%RSD) obtained for the drug spiked in plasma for stability studies were less than 2 %.Kathmandu University Journal of Science, Engineering and TechnologyVol. 13, No. 1, 2017, Page: 65-75


2019 ◽  
Vol 9 (1-s) ◽  
pp. 265-274
Author(s):  
Bindusar Kalia ◽  
Uttam Singh Baghel

This article refers to simple isocratic reverse-phase high-performance liquid chromatographic method (RP-HPLC) developed for the simultaneous quantification of Escitalopram Oxalate (EST) and Clonazepam (CZP) in active pharmaceutical ingredient and pharmaceuticals. The separation of the two drugs was attained using a C₁₈ column (250mm×4.6mm, 5µ) as a stationary phase. The mobile phase was used as a mixture of methanol; acetonitrile; and 0.05M potassium dihydrogen orthophosphate buffer (pH 4 adjusted by orthophosphoric acid) with an isocratic ratio of 40:20:40 v/v. Detection was made by using PDA detector at 210 nm. Escitalopram Oxalate (RT= 4.428 minutes) and Clonazepam (RT= 6.532 minutes) were separated in a single chromatographic run with resolution of 8.719. The calibration plot indicated good linear relationship with r2 = 0.998 for Escitalopram Oxalate in concentration range of 32 µg/ml - 48 µg/ml and r2 = 0.999 for Clonazepam in concentration range of 16 µg/ml - 24 µg/ml. The retrievals for Escitalopram Oxalate and Clonazepam were found to be 99.75% and 99.00%, respectively. The established analytical method was validated and found acceptable as per ICH guidelines for linearity, precision, accuracy, specificity, limit of detection, limit of quantification, robustness and stability. Escitalopram Oxalate and Clonazepam individually as well as in combination were exposed to different stress conditions like acid, base, thermal, photolytic and oxidation degradation and peaks of a degraded product were well determined from peaks of pure drug. This method is modest, quick and appropriate for routine quality control analysis. Keywords: Reverse Phase – HPLC; Escitalopram Oxalate; Clonazepam; Validation; Degradation study.


2020 ◽  
Vol 10 (1) ◽  
pp. 31-38
Author(s):  
Rahul Suryawanshi ◽  
Siddiqua Shaikh ◽  
Snehal Patil

A new, simple, precise, accurate and reproducible Reverse Phase High Performance Liquid Chromatography (RP-HPLC) method for Simultaneous estimation of bulk and pharmaceutical formulations. Separation of Mirabegron was successfully achieve , C18, 250X4.6mm, 5µm or equivalent in an isocratic mode utilizing methanol water (70:30) at pH 5.0 Adjusted to OPA at a flow rate of 1.0ml/min and eluate was monitored at 243nm, with a retention time of 2.584 minutes for Mirabegron. The method was validated and the response was found to be linear in the drug concentration range of 50µg/ml to150 µg/ml for Mirabegron. The values of the correlation coefficient were found to 0.999for Mirabegron. The Limit of Detection(LOD) and Limit of Quantification (LOQ) for Mirabegron were found to be 0.149 and 0.498 respectively. This method was found to be good percentage recovery were found to be 99 indicates that the proposed method is highly accurate. The specificity of the method shows good correlation between retention times of standard with the sample so, the method specifically determines the analyte in the sample without interference from excipients of tablet dosage forms. The method was extensively validated according to International Council for Harmonisation(ICH) guidelines for Linearity, Accuracy, Precision, Specificity and


Author(s):  
Ramesh Jayaprakash ◽  
Senthil Kumar Natesan

Objective: The present study was aimed to develop a rapid, accurate, linear, sensitive and validate stability-indicating high performance liquid chromatographic [RP-HPLC] method for determination of vildagliptin and metformin in pharmaceutical dosage form.Methods: The chromatographic separation was performed on kromasil-C18 column [4.5 x 250 mm; 5 µm] using a mobile phase consisting of 0.05 mmol potassium dihydrogen phosphate buffer: acetonitrile [80:20 v/v], [pH adjusted to 3.5 using orthophosphoric acid]. The flow rate is 0.9 ml/min and the detection was carried out at 263 nm.Results: The chromatographic condition, the peak retention time of metformin and vildagliptin were found to be 2.215 min and 2.600 min respectively. Stress testing was performed in accordance with an international conference on harmonization [ICH] Q1A R2 guidelines. The method was validated as per ICH Q2 R1 guidelines. The calibration curve was found to be linear in the concentration range of 5-17.5 µg/ml and 50-175 µg/ml for vildagliptin and metformin. The limit of detection and quantification was found to be 0.0182 µg/ml and 0.0553 µg/ml for vildagliptin and 0.4451 µg/ml and 1.3490 µg/ml for metformin respectively.Conclusion: A new sensitive, simple and stability indicating reverse-phase high-performance liquid chromatography [RP-HPLC] method has been developed and validated for the determination of vildagliptin and metformin. The proposed method can be used for routine determination of vildagliptin and metformin.


2020 ◽  
Vol 10 (1) ◽  
pp. 92-96
Author(s):  
Balakrishna Tiwari ◽  
Mrunal K. Shirsat ◽  
Amol Kulkarni

Brinzolamide is inhibitor of carbonic anhydride and is highly specific and non-competitive. The aim of the present study is to develop a simple, precise, accurate, sensitive RP-HPLC method for the determination of bulk drug. The objective of the method validation is to demonstrate whether the method was suited for the intended purpose. The method was validated as per the ICH guidelines. The method was validated for linearity, precision (repeatability, intermediate precision), accuracy, specificity, robustness, ruggedness, limit of detection and limit of quantification. Cosmosil (4.6X250mm, 5 μ) column was used for separation. The selected wavelength for Brinzolamide was 254 nm. The mobile phase consists of Acetonitrile: Potassium dihydrogen phosphate buffer (40:60). Flow rate was delivered at 1.0 mL/min. Appropriate dilutions of standard stock solutions were prepared to get desired concentrations in the range of 100-500 mcg/ml. The equation od standard curve was y = 441.8x + 1132 and R2 = 0.998. The RT obtained was 6.6167 minutes. Keywords: Brinzolamide, UV spectroscopy, RP-HPLC, ICH


INDIAN DRUGS ◽  
2012 ◽  
Vol 49 (08) ◽  
pp. 33-37
Author(s):  
T. Venkatachalam ◽  
◽  
K. G. Lalitha

A chromatographic method has been developed as per ICH norms for the simultaneous estimation of cefixime and ofloxacin from pharmaceutical formulations. The method was carried out on a column -Gemini (250 x. 4.6, 5 mc) with a mobile phase consisting of 0.2 M potassium dihydrogen phosphate buffer (adjusted to pH 7 with 1 % w/w triethylamine), acetonitrile and methanol in 70:20:10 ratio and filtered through 0.45 mc cellulose nitrate filters. The flow rate 1.0 mL/min. Detection was carried out at 288 nm. The retention time of CEF and OFL was 2.16 and 7.86 min respectively. The developed method was validated in terms of accuracy, precision, linearity, limit of detection, limit of quantification and solution stability. The proposed method can be used for the estimation of these drugs in combined dosage forms.


Author(s):  
Bhupender Tomar ◽  
Ankita Sharma ◽  
Inder Kumar ◽  
Sandeep Jain ◽  
Pallavi Ahirrao

A simple, precise, and accurate reverse phase high performance liquid chromatographic method (RP-HPLC) was developed and validated for the estimation of the combination of 5- Fluorouracil (5-FU) and Imiquimod in active pharmaceutical ingredients (APIs). The method was carried out on Phenomenex C18 (250 × 4.6mm I.D., 5𝜇m) using isocratic elution mode. The mobile phase was used as Acetonitrile: 10mM potassium dihydrogen orthophosphate: triethylamine (40:59.9:0.1, v/v, pH 4.5 with orthophosphoric acid) and Water: ACN (50:50 v/v) was used as a diluent. The concentration of solvents was 1-20µg/ml and the volume of injection was 20µl with the flow rate of 1.2ml/min. The retention times for 5-FU and Imiquimod were found to be 1.9±0.5 and 6.6±0.5 min respectively. The absorption maxima of 5FU and Imiquimod were found 267nm and 227nm respectively. The method was validated as per ICH guidelines. All the data were found within the specified limits. The limit of detection (LOD) and limit of quantification (LOQ) of 5- Fluorouracil were found to be 0.015μg/mL and 0.048 μg/mL, respectively, and Imiquimod was found to be 0.078μg/mL and 0.237μg/mL, respectively. The method developed in the present study was found to be sensitive, specific, and precise and can be applied for the simultaneous estimation of 5-FU and Imiquimod.