Solid-phase extraction of alkylphosphonic and O-alkyl alkylphosphonic acids followed by HPLC separation using porous graphitic carbon sorbent

2021 ◽  
pp. 462420
Author(s):  
E.N. Goncharova ◽  
M.A. Statkus ◽  
P.N. Nesterenko ◽  
G.I. Tsysin ◽  
YuA. Zolotov
2005 ◽  
Vol 51 ◽  
pp. 23-28 ◽  
Author(s):  
Violeta Ivanova ◽  
Dragica Zendelovska ◽  
Marina Stefova

A simple, rapid and precise HPLC method has been developed for the assay of hydrochlorothiazide in urine. The clean-up of the urine samples was carried out by solid-phase extraction using HLB cartridges. Extraction recovery was 94.00-100.28 %. HPLC separation was performed with isocratic elution on Hypersil BDS C18 column (100 x 4.0 mm I.D., 3 µm particle size) protected with appropriate guard column. The mobile phase was 18 % acetonitrile and 0.025 mol/L solution of KH2PO4, pH 4 at flow rate of 0.3 mL/min. Detection of the substances was performed at 220 nm. The calibration curves were linear in the range of 2-50 µg/mL. The developed method is validated by checking its accuracy, precision and stability. The detection limit is 2 µg/mL hydrochlorothiazide. The method is proved to be convenient for routine analysis of hydrochlorothiazide in urine.