Development and Validation of RP-HPLC Method for Estimation of Teneligliptin and its Impurity in Tablet
The objective of the study is a simple, precise and accurate stability RP-HPLC method has been developed and subsequently validated for the estimation of Teneligliptin and its impurity in tablet formulation. The adequate separation was carried out using Grace Smart C18 column (250mm x 4.6mm, 5?m particle size), mixture of 0.05M Potassium dihydrogen phosphate PH 4.0 and Acetonitrile 80:20 % v/v as a mobile phase with a flow rate of 1 ml/min and the effluent was monitored at 242 nm using PDA detector. The retention time of Teneligliptin, Impurity B and Impurity G were 7.443 min, 6.650 min and 8.473 min respectively. Linearity for Teneligliptin, Impurity B and Impurity G were found in the range of 500-3000 µg/ml (R2 = 0.998), 5-15 µg/ml (R2 = 0.994) and 5-15 µg/ml (R2 = 0.998) respectively. The accuracy of the present method was evaluated at 50%, 100% and 150%. The % recoveries of drug were found to be in range of 99.315 ± 0.283 for Teneligliptin. Precision studies were carried out and the RSD values were less than two. The method was found to be robust. The proposed method was found to be specific, accurate, precise and robust can be used for simultaneous estimation of these drugs in tablet dosage form.