scholarly journals Cyclic Fatigue Resistance of Three Single-use NiTi Instruments after Immersion in Sodium Hypochlorite

2014 ◽  
Vol 3 (2) ◽  
pp. 67-72 ◽  
Author(s):  
Alberto Dagna ◽  
Riccardo Beltrami ◽  
Marco Colombo ◽  
Marco Chiesa ◽  
Stefano Bianchi ◽  
...  

ABSTRACT The aim of this in vitro study was to compare the cyclic fatigue resistance of three single-use nickel-titanium (NiTi) instruments (One Shape, Reciproc and Wave One) after immersion in sodium hypochlorite (NaOCl) solution over several time periods. A total of 135 One Shape, Reciproc R25 and Wave One Primary were tested. The 45 files of the same brand were randomly assigned to three groups (n = 15) and submitted to the following immersion protocol: 1 minute in 5% NaOCl at 37°C, 5 minutes in 5% NaOCl at 37°C, no immersion (as control). Cyclic fatigue test was performed rotating/reciprocating all instruments in a stainless steel curved artificial canal (60° angle of curvature and 5 mm radius). The number of cycles to fracture (NCF) was determined by measuring the time to fracture. The data were compared for differences by using 2-way analysis of variance (p < 0.05). In general, resistance to cyclic fatigue was not significantly affected by immersion in NaOCl. Reciproc R2 showed the highest cyclic fatigue resistance in all groups compared to One Shape and Wave One. Immersion in NaOCl for 1 or 5 minutes did not reduce the cyclic fatigue resistance of NiTi single-use instruments significantly. Reciproc were more resistant, but the new rotary single-use One Shape instruments showed mechanical resistance similar to NiTi files developed for reciprocating motion. How to cite this article Dagna A, Beltrami R, Colombo M, Chiesa M, Bianchi S, Poggio C. Cyclic Fatigue Resistance of Three Single-use NiTi Instruments after Immersion in Sodium Hypochlorite. Int J Experiment Dent Sci 2014;3(2):67-72.

2020 ◽  
Vol 14 (2) ◽  
pp. 111-115 ◽  
Author(s):  
Gülsah Uslu ◽  
Mustafa Gündoğar ◽  
Taha Özyürek ◽  
Gianluca Plotino

Background. This study was conducted to compare the cyclic fatigue resistance of VDW.ROTATE,TruNatomy Prime, HyFlex CM, and 2Shape nickel-titanium (NiTi) rotary instruments in doublecurvedcanals in a simulated clinical environment. Methods. Eighty NiTi files were used for the fatigue testing in stainless steel canals compatible withinstrument sizes until fracture occurred (n=20): VDW.ROTATE (04./25#), TruNatomy Prime (04./26#),HyFlex CM (04./25#) and 2Shape TS04./25#( 1). For each instrument, the number of cycles to fracture(NCF) was calculated, and the fractured fragment length (FL) was measured. To verify that the fileswere fractured due to cyclic fatigue, the fractured surfaces of the files were evaluated under a scanningelectron microscope. Data were statistically analyzed using the Kruskal–Wallis and Student’s t-tests atthe %95 confidence level. Results. The failure of the files due to cyclic fatigue was first seen in the apical curvature before thecoronal curvature (P<0.05). The highest fatigue resistance was observed in VDW.ROTATE and HyFlexCM files in both curvatures (P<0.05). There were no significant differences in the fatigue resistancebetween the HyFlex CM and VDW.ROTATE files or between the 2Shape and the TruNatomy files(P>0.05). There was no difference in the fractured lengths of the files between the apical and coronalcurvatures (P>0.05). Conclusion. In artificial S-shaped root canals, VDW.ROTATE and HyFlex CM files exhibited higherfatigue resistance compared to 2Shape and TruNatomy files.


2013 ◽  
Vol 24 (2) ◽  
pp. 117-120 ◽  
Author(s):  
Oscar Faciola Pessoa ◽  
Juliana Melo da Silva ◽  
Giulio Gavini

The aim of this study was to assess cyclic fatigue resistance in rotary nickel-titanium instruments after simulated clinical use in curved root canals (40-degree, 5-mm radius curve). Thirty-six RaCe rotary NiTi files, size #5, taper 0.04, were divided into 3 groups: Groups A, B and C with one, three and five cycles of use, respectively. Time to failure was recorded with a stopwatch in seconds and subsequently converted to number of cycles to fracture. The data were analyzed statistically by ANOVA and Tukey's test (α=0.05). Five sets of clinically used files (group C) reached significantly lower cycle-numbers before fracture (mean=197.5 cycles) when compared with one set of clinically used files (mean=309.2) and three sets (mean=287.5). Results showed that the number of simulated clinical uses of RaCe instruments for shaping curved canals affects adversely the fatigue resistance of these instruments after five uses.


2018 ◽  
Vol 12 (4) ◽  
pp. 283-287 ◽  
Author(s):  
Gülşah Uslu ◽  
Taha Özyürek ◽  
Mustafa Gündoğar ◽  
Koray Yılmaz

Background. The aim of this study was to compare the cyclic fatigue resistance of 2Shape, Twisted File (TF) and EndoSequence Xpress (ESX) nickel-titanium rotary files at intracanal temperature (35°C). Methods. Twenty 2Shape TS1 (25/.04), 20 TF (25/.04) and 20 ESX (25/.04) files were tested for cyclic fatigue at intracanal temperature (35°C). All the instruments were rotated in artificial canals which were made of stainless steel with an inner diameter of 1.5 mm, 60° angle of curvature and a radius curvature of 5 mm until fracture occurred; the time to fracture was recorded in seconds using a digital chronometer and the number of cycles to fracture (NCF) for each file was calculated. Kruskal-Wallis test with Bonferroni correction was performed to statistically analyze data using SPSS 21.0. Statistical significance was set at P<0.05. Results. NCF values revealed that the 2Shape had significantly the highest cyclic fatigue resistance, followed by TF and ESX at intracanal temperature (P<0.05). The difference was significant between the TF and ESX groups (P<0.05). There was no significant difference among the 2Shape, TF and ESX files with respect to the lengths of the fractured file fragments (P>.05). Conclusion. Within the limitations of present study, it was concluded that the cyclic fatigue resistance of 2Shape files at the intracanal temperature is higher than that of TF and ESX files.


Author(s):  
C Ruiz-Sanchez ◽  
V Faus-Matoses ◽  
T Alegre-Domingo ◽  
I Faus-Matoses ◽  
VJ Faus-Llacer

2020 ◽  
Vol 2020 ◽  
pp. 1-5
Author(s):  
Hind F. Abuhulaibah ◽  
Ammar AbuMostafa

Background/purpose. To assess the resistance to cyclic fatigue (CF) displayed by two nickel-titanium (NiTi) files, One Curve (OC) and ProTaper Gold (PTG), after their immersion in 2.5% sodium hypochlorite (NaOCl) solution at body temperature. Materials and Methods. Forty-five files from each brand were randomly assigned to three groups (n = 15) and subjected to the following: no immersion (control), 1-minute immersion, and 5-minute immersion in 2.5% NaOCl at 37°C. CF for all the files was tested within a well-lubricated stainless-steel artificial canal in a water bath at 37°C simulating body temperature. The procedure was video recorded, and the number of cycles to fracture (NCF) was calculated by multiplying the time taken to fracture, with the number of rotations per second. The data were analyzed for normality, and thereafter, one-way ANOVA with multiple Bonferroni was used as the post hoc test. Results. The resistance to CF of OC files was significantly higher than PTG files in all groups. In groups immersed in NaOCl for one minute, OC files showed significant drop in the CF resistance; PTG files showed a drop as well but not significantly. Both files demonstrated a significant drop in resistance to CF after immersion in NaOCl for five minutes compared to the control group. Conclusion. OC files were significantly more resistant to CF compared to PTG in all groups. Immersion in 2.5% NaOCl for 5 minutes significantly reduced the resistance to CF for all the files.


2021 ◽  
Vol 2021 ◽  
pp. 1-11
Author(s):  
Suhad Jabbar Hamed Al-Nasrawi ◽  
Zuha Ayad Jaber ◽  
Nibrass Talib Al-Quraine ◽  
Abtesam Imhemed Aljdaimi ◽  
Sattar Jabbar Abdul-Zahra Al-Hmedat ◽  
...  

Peracetic acid (PAA) is widely used as a sterilizing/disinfecting agent, and, in endodontics, it has been introduced as a promising irrigant in root canal treatment. It has been used at different concentrations to achieve various functions. However, endodontic instruments in contact with PAA of a certain concentration may affect their fatigue resistance. Therefore, the aim of this study was to investigate the impact of PAA on the cyclic fatigue resistance of three commercial heat-treated nickel-titanium (NiTi) rotary files. Three types of heat-treated NiTi rotary files were selected: One Curve (OC), ProTaper Gold (PTG), and Wave One Gold (WOG). Each type was divided into three subgroups (n = 6 for each file type): (1) untreated instruments; (2) files immersed in 0.002% PAA; and (3) files immersed in 0.35% PAA. The performance of each file type was tested in a simulated canal. The number of cycles to fracture (NCF) was determined to assess cyclic fatigue resistance of the files. Independent sample t-test was applied to compare each treated file within a subgroup with its respective control group, and one-way ANOVA was used for comparison among the main groups. All types of tested files revealed a significant decline in the cyclic fatigue resistance after exposure to 0.002% PAA except the PTG ( P = 0.209 ). After exposure of the files to a higher concentration (0.35% PAA), a dramatic reduction was demonstrated by all the groups. Before and after exposure of the files to PAA, PTG displayed the highest cyclic fatigue resistance, followed by the WOG, while the OC showed the lowest resistance. Exposure of heated-treated NiTi files to PAA in a relatively high or low concentration adversely affects the cyclic fatigue resistance. The PTG files demonstrated the best performance among the tested types and can be disinfected with 0.002% PAA for clinical purpose.


2021 ◽  
Vol 2021 ◽  
pp. 1-6
Author(s):  
Pooja D. Khandagale ◽  
Prashant P. Shetty ◽  
Saleem D. Makandar ◽  
Pradeep A. Bapna ◽  
Mohmed Isaqali Karobari ◽  
...  

Introduction. The main aims of root canal instrumentation are to provide an environment that will lead to healing and to provide a root canal shape that is comfortable to clean and seal. When working with rotary endodontic instruments, the most significant concerns are that the instrument might fracture in the root canal, thus affecting the treatment outcome. Hence, it is of immense importance to know which file systems have more cyclic fatigue resistance. Methodology. This study evaluated the effect of the curved segment length of the artificial canal (the arch), and the number of cycles necessary in fracture of Hyflex EDM, Twisted files, and ProTaper Gold were recorded. Sixty NiTi rotary instruments of 25 mm length (Hyflex EDM (20), Twisted files (20), and ProTaper Gold (20)) were tested in a metal block with simulated canal having 90° angle of curvature. The study was performed with a specific radius and degree of curvature, i.e., 8 mm radius and 90⁰ angle of curvature, and data obtained were subsequently subjected to statistical evaluation using one-way analysis of variance and Tukey’s post hoc test. Result. The Hyflex EDM (774.29) exhibited the maximum cyclic fatigue resistance compared to Twisted files (654.875) and ProTaper Gold (375.575). A statistically significant difference was observed between the tested groups. Conclusion. The Hyflex EDM files showed the highest cyclic fatigue resistance, followed by Twisted files and ProTaper Gold files.


2020 ◽  
Author(s):  
Amr M. Elnaghy ◽  
Shaymaa E Elsaka ◽  
Ayman Mandorah

Abstract Background: The purpose of this study was to compare the cyclic fatigue resistance of newly developed TruNatomy instruments (TRN) in single and double (S-shaped) curvature canals with HyFlex CM (HCM), Vortex Blue (VB) and RaCe (RC) instruments. Methods: Size 20/.04 taper and size 25/0.04 of HCM, VB and RC were used. For TRN instruments, size 20/.04 taper (small) and size 26/.04 taper (prime) were used. The instruments were tested in artificial canals with double curvature (coronal curve; 60° curvature, 5 mm radius and apical curve; 70° curvature and 2 mm radius) and single curvature (60° curvature, 5 mm radius). The number of cycles to failure (NCF) was recorded. Data were statistically analyzed by Kruskal-Wallis and Dunn’s multiple comparison tests. Weibull analysis was performed on NCF data. Statistical significant was set at p < 0.05. Results: TRN and HCM revealed higher NCF compared with the other instruments for both tested sizes in single and double curvature canals ( p < 0.05). TRN and HCM showed no statistically significant difference in the NCF ( p > 0.05). The probability of survival was higher for HCM and TRN instruments than VB and RC instruments. Conclusions: HCM and TRN instruments were more resistant to cyclic fatigue than VB and RC instruments in single and double curvature canals. HCM and TRN instruments were anticipated to survive with higher number of cycles than the other tested instruments. RC instrument had the lowest fatigue resistance than the other instruments.