scholarly journals Cyclic fatigue resistance of reduced-taper nickel-titanium (NiTi)instruments in doubled-curved (S-shaped) canals at body temperature

2020 ◽  
Vol 14 (2) ◽  
pp. 111-115 ◽  
Author(s):  
Gülsah Uslu ◽  
Mustafa Gündoğar ◽  
Taha Özyürek ◽  
Gianluca Plotino

Background. This study was conducted to compare the cyclic fatigue resistance of VDW.ROTATE,TruNatomy Prime, HyFlex CM, and 2Shape nickel-titanium (NiTi) rotary instruments in doublecurvedcanals in a simulated clinical environment. Methods. Eighty NiTi files were used for the fatigue testing in stainless steel canals compatible withinstrument sizes until fracture occurred (n=20): VDW.ROTATE (04./25#), TruNatomy Prime (04./26#),HyFlex CM (04./25#) and 2Shape TS04./25#( 1). For each instrument, the number of cycles to fracture(NCF) was calculated, and the fractured fragment length (FL) was measured. To verify that the fileswere fractured due to cyclic fatigue, the fractured surfaces of the files were evaluated under a scanningelectron microscope. Data were statistically analyzed using the Kruskal–Wallis and Student’s t-tests atthe %95 confidence level. Results. The failure of the files due to cyclic fatigue was first seen in the apical curvature before thecoronal curvature (P<0.05). The highest fatigue resistance was observed in VDW.ROTATE and HyFlexCM files in both curvatures (P<0.05). There were no significant differences in the fatigue resistancebetween the HyFlex CM and VDW.ROTATE files or between the 2Shape and the TruNatomy files(P>0.05). There was no difference in the fractured lengths of the files between the apical and coronalcurvatures (P>0.05). Conclusion. In artificial S-shaped root canals, VDW.ROTATE and HyFlex CM files exhibited higherfatigue resistance compared to 2Shape and TruNatomy files.

2013 ◽  
Vol 24 (2) ◽  
pp. 117-120 ◽  
Author(s):  
Oscar Faciola Pessoa ◽  
Juliana Melo da Silva ◽  
Giulio Gavini

The aim of this study was to assess cyclic fatigue resistance in rotary nickel-titanium instruments after simulated clinical use in curved root canals (40-degree, 5-mm radius curve). Thirty-six RaCe rotary NiTi files, size #5, taper 0.04, were divided into 3 groups: Groups A, B and C with one, three and five cycles of use, respectively. Time to failure was recorded with a stopwatch in seconds and subsequently converted to number of cycles to fracture. The data were analyzed statistically by ANOVA and Tukey's test (α=0.05). Five sets of clinically used files (group C) reached significantly lower cycle-numbers before fracture (mean=197.5 cycles) when compared with one set of clinically used files (mean=309.2) and three sets (mean=287.5). Results showed that the number of simulated clinical uses of RaCe instruments for shaping curved canals affects adversely the fatigue resistance of these instruments after five uses.


2017 ◽  
Vol 31 (2) ◽  
pp. 78-82
Author(s):  
Massimo Amato ◽  
Giuseppe Pantaleo ◽  
Dina Abdellatif ◽  
Andrea Blasi ◽  
Roberto Lo Giudice ◽  
...  

Aim: The aim of present study was to compare cyclic fatigue resistance of three modern Ni-Ti instruments used with continuous rotation. Materials and methods: For this study 3 groups of rotating instruments with continuous rotation (HyFlex EDM, Twisted File Adaptive, Revo S SU) have been used, each group consisted of 20 files. The various groups were subjected to cyclic fatigue testing through an artificial metal device. A statistical analysis with Kruskal-Wallis test and Mann-Whitney test was performed. Results: There were statistically significant differences between the three groups. The HyFlex EDM instruments have a fracture resistance slightly higher than the Twisted file and far higher than Revo S SU. Conclusions: Modern Ni-Ti alloys increase resistance of the rotating instruments to cyclic fatigue.


Metals ◽  
2020 ◽  
Vol 10 (10) ◽  
pp. 1359
Author(s):  
Ajinkya M. Pawar ◽  
Unmesh Khanvilkar ◽  
Shivani Mallishery ◽  
Kashmira Sawant ◽  
Ricardo Machado ◽  
...  

The cyclic fatigue, torsional resistance, and angular deflection of a new Fire-Wire rotary file (CricENDO) were compared. A total of 20 files of each type were tested. Cyclic fatigue testing was performed for each group (n = 10) by measuring the number of cycles to fracture (NCF) in an artificial stainless-steel canal (60° angle of curvature, with a 6-mm radius) for each group. The torque and angle of rotation at the failure of each group (n = 10) were measured according to ISO 3630-1. The fractured surfaces were examined using scanning electron microscopy. Statistical analysis was carried out utilizing Student’s t-test at a significance level set at 5%. The Fire-Wire CricENDO rotary files were associated with a significantly higher number of cycles to fracture and time to failure (in seconds) compared to the M-Wire Protaper Next (p < 0.05). A significantly higher angular deflection to fracture was observed for CricENDO compared to Protaper Next (p < 0.05). The new Fire-Wire CricENDO rotary files exhibit higher cyclic fatigue resistance and angle of rotation to fracture than M-Wire Protaper Next. Without warning, file fracture may occur as a result of cyclic fatigue, torsional stress, or a combined effect of both. CricENDO rotary files may be an effective alternative in curved root canals as they exhibited elevated cyclic and torsional resistance. It will be helpful in eliminating one of the reasons for file fracture during the root canal treatment.


2020 ◽  
Vol 10 (2) ◽  
pp. 55-59
Author(s):  
Mehmet Adıgüzel ◽  
Faruk Öztekin

Aim: To evaluate the fracture resistance to cyclic fatigue of One Curve (Micro-Mega, Besancon, Cedex, France), One Shape (Micro-Mega), 2Shape (Micro-Mega) and EdgeFile X3 (EdgeEndo, Albuquerque, NM) files in simulated root canals. Methodology: One hundred and sixty nickel-titanium endodontic files were tested in simulated canals with single and s-shaped (double) curvatures. The One Curve (25/.06), One Shape (25/.06), 2Shape (25/.06) and EdgeFile X3 (25/.06) files were assessed (n=20 for each test). All the files were rotated up to fracture occurred. The number of cycles to fracture (NCF) was determined and also, the length of the fractured file part was measured. For statistical analysis, one-way analysis of variance and Tukey HSD tests were used (p < 0.05).   Results: NCF for all instrument groups were significantly higher in the single curvature canal compared to the s-shaped curvature canal (p < 0.05). The EdgeFile X3 had the highest fatigue resistance in both curvature groups. The NCF of the One Curve were higher than that of the One Shape and 2Shape files (p < 0.05). In the s-shaped curvature canal, the files inclined to fracture more often in the apical curvature compared to the coronal curvature. There was no statistically considerable difference in the length of the fractured parts among the four experiment groups in either the single or s-shaped canal curvatures (p > 0.05). Conclusion: Within the limitations of this study, fracture resistance to cyclic fatigue was the highest for the EdgeFile X3 group followed by the One Curve group and lowest for the 2Shape and One Shape group.   How to cite this article: Adıgüzel M, Öztekin F. Comparison of the resistance to cyclic fatigue of One Curve, One Shape, 2Shape and EdgeFile X3 Files in simulated single and S-shaped (double) curvatures. Int Dent Res 2020;10(2):55-59. https://doi.org/10.5577/intdentres.2020.vol10.no2.5   Linguistic Revision: The English in this manuscript has been checked by at least two professional editors, both native speakers of English.


Materials ◽  
2020 ◽  
Vol 13 (22) ◽  
pp. 5246
Author(s):  
Sangmi Ahn ◽  
Jung-Hong Ha ◽  
Sang Won Kwak ◽  
Hyeon-Cheol Kim

Nickel-titanium (NiTi) endodontic rotary instruments are used extensively in root canal procedures by both general dentists and specialists. However, their vulnerability to fracture is the major reason for clinicians’ concern regarding their use. The objective of this study was to investigate the potential effects of spring machining of the file shaft on the fatigue and torsional resistances of NiTi rotary instruments. Three types of NiTi rotary systems with (S) and without (NS) spring machining were used in this study (n = 15 each): a spring file (SPR; #25/.06, SPR-S, SPR-NS), a ProTaper Next X2 (PTN; #25/variable taper, PTN-S, PTN-NS), and a ProTaper Gold F2 (PTG; #25/variable taper, PTG-S, PTG-NS). Spring machining was adjusted on the 6 mm of each file system’s shaft via a laser cutting process. The number of rotation cycles until fracture (i.e., cyclic fatigue resistance), ultimate torsional strength, the distortion angle, and the toughness of each subgroup were estimated with specially designed devices. The results were analyzed using a paired t-test at a significance level of 95%. NiTi rotary instruments with spring machining exhibited a higher cyclic fatigue resistance than instruments without spring machining. The groups with spring machining exhibited a higher toughness and larger distortion angle than the groups without it (p < 0.05). In conclusion, spring machining on the shank of NiTi instruments may provide a stress-bearing area and attenuate the torsional and cyclic fatigue of NiTi rotary instruments.


2014 ◽  
Vol 3 (2) ◽  
pp. 67-72 ◽  
Author(s):  
Alberto Dagna ◽  
Riccardo Beltrami ◽  
Marco Colombo ◽  
Marco Chiesa ◽  
Stefano Bianchi ◽  
...  

ABSTRACT The aim of this in vitro study was to compare the cyclic fatigue resistance of three single-use nickel-titanium (NiTi) instruments (One Shape, Reciproc and Wave One) after immersion in sodium hypochlorite (NaOCl) solution over several time periods. A total of 135 One Shape, Reciproc R25 and Wave One Primary were tested. The 45 files of the same brand were randomly assigned to three groups (n = 15) and submitted to the following immersion protocol: 1 minute in 5% NaOCl at 37°C, 5 minutes in 5% NaOCl at 37°C, no immersion (as control). Cyclic fatigue test was performed rotating/reciprocating all instruments in a stainless steel curved artificial canal (60° angle of curvature and 5 mm radius). The number of cycles to fracture (NCF) was determined by measuring the time to fracture. The data were compared for differences by using 2-way analysis of variance (p < 0.05). In general, resistance to cyclic fatigue was not significantly affected by immersion in NaOCl. Reciproc R2 showed the highest cyclic fatigue resistance in all groups compared to One Shape and Wave One. Immersion in NaOCl for 1 or 5 minutes did not reduce the cyclic fatigue resistance of NiTi single-use instruments significantly. Reciproc were more resistant, but the new rotary single-use One Shape instruments showed mechanical resistance similar to NiTi files developed for reciprocating motion. How to cite this article Dagna A, Beltrami R, Colombo M, Chiesa M, Bianchi S, Poggio C. Cyclic Fatigue Resistance of Three Single-use NiTi Instruments after Immersion in Sodium Hypochlorite. Int J Experiment Dent Sci 2014;3(2):67-72.


Materials ◽  
2020 ◽  
Vol 13 (24) ◽  
pp. 5823
Author(s):  
Saulius Drukteinis ◽  
Vytaute Peciuliene ◽  
Ruta Bendinskaite ◽  
Vilma Brukiene ◽  
Rasmute Maneliene ◽  
...  

The better understanding of the clinically important behavioral features of new instrument systems has an important significance for the clinical endodontics. This study aimed to investigate the shaping and centering ability as well as cyclic fatigue resistance of HyFlex CM (CM), HyFlex EDM (EDM) and EdgeFile (EF) thermally treated nickel–titanium (NiTi) endodontic instrument systems. Sixty curved root canals of the mesial roots of mandibular molars were randomly assigned into three groups (n = 20) and shaped using CM, EDM and EF files up to the size 40 and taper 04 of the instruments. µCT scanning of the specimens before and after preparation was performed and the morphometric 2D and 3D parameters were evaluated in the apical, middle and coronal thirds of root canals. In each group, 40.04 instruments (n = 20) were subjected to the cyclic fatigue resistance test in artificial root canals at 37 °C temperature until fractures occurred, and the number of cycles to failure (NCF) was calculated. The fractographic analysis was performed using a scanning electron microscope, evaluating topographic features and surface profiles of the separated instruments. The one-way analysis of variance with post hoc Tuckey’s test was used for statistical analysis of the data; the significance level was set at 5%. All systems prepared the comparable percentage of root canal surface with the similar magnitude of canal transportation in all root thirds (p > 0.05), but demonstrated significantly different resistance to cyclic fatigue (p < 0.05). The most resistant to fracture was EF, followed by EDM and CM. The length of the fractured fragments was not significantly different between the groups, and fractographic analysis by SEM detected the typical topographic features of separated thermally treated NiTi instrument surfaces.


2018 ◽  
Vol 12 (4) ◽  
pp. 283-287 ◽  
Author(s):  
Gülşah Uslu ◽  
Taha Özyürek ◽  
Mustafa Gündoğar ◽  
Koray Yılmaz

Background. The aim of this study was to compare the cyclic fatigue resistance of 2Shape, Twisted File (TF) and EndoSequence Xpress (ESX) nickel-titanium rotary files at intracanal temperature (35°C). Methods. Twenty 2Shape TS1 (25/.04), 20 TF (25/.04) and 20 ESX (25/.04) files were tested for cyclic fatigue at intracanal temperature (35°C). All the instruments were rotated in artificial canals which were made of stainless steel with an inner diameter of 1.5 mm, 60° angle of curvature and a radius curvature of 5 mm until fracture occurred; the time to fracture was recorded in seconds using a digital chronometer and the number of cycles to fracture (NCF) for each file was calculated. Kruskal-Wallis test with Bonferroni correction was performed to statistically analyze data using SPSS 21.0. Statistical significance was set at P<0.05. Results. NCF values revealed that the 2Shape had significantly the highest cyclic fatigue resistance, followed by TF and ESX at intracanal temperature (P<0.05). The difference was significant between the TF and ESX groups (P<0.05). There was no significant difference among the 2Shape, TF and ESX files with respect to the lengths of the fractured file fragments (P>.05). Conclusion. Within the limitations of present study, it was concluded that the cyclic fatigue resistance of 2Shape files at the intracanal temperature is higher than that of TF and ESX files.


Author(s):  
Jader Camilo Pinto ◽  
Fernanda Ferrari Esteves Torres ◽  
Airton Oliveira Santos-Junior ◽  
Marco Antonio Hungaro Duarte ◽  
Juliane Maria Guerreiro-Tanomaru ◽  
...  

Abstract Objective The aim of this study was to investigate the effect of additional apical preparation using the ProDesign Logic (PDL) 50/.01 rotary heat-treated nickel–titanium (NiTi) file with a larger diameter and minimal taper for retreatment of curved root canals. Materials and Methods Mesial curved root canals of 12 mandibular molars were prepared using PDL 25/.06 and filled using the continuous wave of condensation technique and AH Plus sealer. After retreatment using ProDesign S (PDS) 25/.08, PDL 25/.06 and PDL 35/05, a complementary procedure was performed with PDL 50/.01. Microcomputed tomography (micro-CT) scanning was performed before and after retreatment procedures. The cyclic fatigue resistance of unused PDS 25/.08, PDL 25/.06, PDL 35/.05 and PDL 50/.01 instruments (n = 12) was evaluated in a stainless-steel device. Statistical Analysis Data on the volumes of the root canals and the remaining filling materials were submitted to the paired t-test. Cyclic fatigue resistance data was submitted to one-way ANOVA and Tukey’s tests (α = 0.05). Results Use of PDL 50/.01 decreased the remaining filling materials in the apical third (p < 0.05). The root canal volume was similar in the cervical and middle thirds before and after preparation using PDL 50/.01 (p > 0.05). PDL 50/.01 presented the highest cyclic fatigue resistance (p < 0.05). Conclusions Use of the PDL 50/.01 instrument as an additional apical preparation for retreatment of curved root canals improved filling material removal in the apical third, while maintaining the dentin in the cervical and middle thirds. In addition, PDL 50/.01 presented high-flexural resistance.