microbeam radiation therapy
Recently Published Documents


TOTAL DOCUMENTS

180
(FIVE YEARS 36)

H-INDEX

32
(FIVE YEARS 4)

2022 ◽  
Vol 29 (1) ◽  
Author(s):  
Jeremy Davis ◽  
Andrew Dipuglia ◽  
Matthew Cameron ◽  
Jason Paino ◽  
Ashley Cullen ◽  
...  

Successful transition of synchrotron-based microbeam radiation therapy (MRT) from pre-clinical animal studies to human trials is dependent upon ensuring that there are sufficient and adequate measures in place for quality assurance purposes. Transmission detectors provide researchers and clinicians with a real-time quality assurance and beam-monitoring instrument to ensure safe and accurate dose delivery. In this work, the effect of transmission detectors of different thicknesses (10 and 375 µm) upon the photon energy spectra and dose deposition of spatially fractionated synchrotron radiation is quantified experimentally and by means of a dedicated Geant4 simulation study. The simulation and experimental results confirm that the presence of the 375 µm thick transmission detector results in an approximately 1–6% decrease in broad-beam and microbeam peak dose. The capability to account for the reduction in dose and change to the peak-to-valley dose ratio justifies the use of transmission detectors as thick as 375 µm in MRT provided that treatment planning systems are able to account for their presence. The simulation and experimental results confirm that the presence of the 10 µm thick transmission detector shows a negligible impact (<0.5%) on the photon energy spectra, dose delivery and microbeam structure for both broad-beam and microbeam cases. Whilst the use of 375 µm thick detectors would certainly be appropriate, based upon the idea of best practice the authors recommend that 10 µm thick transmission detectors of this sort be utilized as a real-time quality assurance and beam-monitoring tool during MRT.


2021 ◽  
Author(s):  
Jean Albert Laissue ◽  
Sébastien Barré ◽  
Stefan Bartzsch ◽  
Hans Blattmann ◽  
Audrey M. Bouchet ◽  
...  

Microbeam radiation therapy, an alternative radiosurgical treatment under preclinical investigation, aims to safely treat muzzle tumors in pet animals. This will require data on the largely unknown radiation toxicity of microbeam arrays for bones and teeth. To this end, the muzzle of six young adult New Zealand rabbits was irradiated by a lateral array of microplanar beamlets with peak entrance doses of 200, 330 or 500 Gy. The muzzles were examined 431 days postirradiation by computed microtomographic imaging (micro-CT) ex vivo, and extensive histopathology. The boundaries of the radiation field were identified histologically by microbeam tracks in cartilage and other tissues. There was no radionecrosis of facial bones in any rabbit. Conversely, normal incisor teeth exposed to peak entrance doses of 330 Gy or 500 Gy developed marked caries-like damage, whereas the incisors of the two rabbits exposed to 200 Gy remained unscathed. A single, unidirectional array of microbeams with a peak entrance dose ≤200 Gy (valley dose14 Gy) did not damage normal bone, teeth and soft tissues of the muzzle of normal rabbits longer than one year after irradiation. Because of that, Microbeam radiation therapy of muzzle tumors in pet animals is unlikely to cause sizeable damage to normal teeth, bone and soft tissues, if a single array as used here delivers a limited entrance dose of 200 Gy and a valley dose of ≤14 Gy.


2021 ◽  
Vol 11 (19) ◽  
pp. 9101
Author(s):  
Jason Paino ◽  
Micah Barnes ◽  
Elette Engels ◽  
Jeremy Davis ◽  
Susanna Guatelli ◽  
...  

Synchrotron microbeam radiation therapy is a promising pre-clinical radiation treatment modality; however, it comes with many technical challenges. This study describes the image guidance protocol used for Australia’s first long-term pre-clinical MRT treatment of rats bearing 9L gliosarcoma tumours. The protocol utilises existing infrastructure available at the Australian Synchrotron and the adjoining Monash Biomedical Imaging facility. The protocol is designed and optimised to treat small animals utilising high-resolution clinical CT for patient specific tumour identification, coupled with conventional radiography, using the recently developed SyncMRT program for image guidance. Dosimetry performed in small animal phantoms shows patient dose is comparable to standard clinical doses, with a CT associated dose of less than 1.39cGy and a planar radiograh dose of less than 0.03cGy. Experimental validation of alignment accuracy with radiographic film demonstrates end to end accuracy of less than ±0.34mm in anatomical phantoms. Histological analysis of tumour-bearing rats treated with microbeam radiation therapy verifies that tumours are targeted well within applied treatment margins. To date, this technique has been used to treat 35 tumour-bearing rats.


2021 ◽  
Vol 28 (5) ◽  
Author(s):  
Jessie A. Posar ◽  
Matthew Large ◽  
Saree Alnaghy ◽  
Jason R. Paino ◽  
Duncan J. Butler ◽  
...  

Spatially fractionated ultra-high-dose-rate beams used during microbeam radiation therapy (MRT) have been shown to increase the differential response between normal and tumour tissue. Quality assurance of MRT requires a dosimeter that possesses tissue equivalence, high radiation tolerance and spatial resolution. This is currently an unsolved challenge. This work explored the use of a 500 nm thick organic semiconductor for MRT dosimetry on the Imaging and Medical Beamline at the Australian Synchrotron. Three beam filters were used to irradiate the device with peak energies of 48, 76 and 88 keV with respective dose rates of 3668, 500 and 209 Gy s−1. The response of the device stabilized to 30% efficiency after an irradiation dose of 30 kGy, with a 0.5% variation at doses of 35 kGy and higher. The calibration factor after pre-irradiation was determined to be 1.02 ± 0.005 µGy per count across all three X-ray energy spectra, demonstrating the unique advantage of using tissue-equivalent materials for dosimetry. The percentage depth dose curve was within ±5% of the PTW microDiamond detector. The broad beam was fractionated into 50 microbeams (50 µm FHWM and 400 µm centre-to-centre distance). For each beam filter, the FWHMs of all 50 microbeams were measured to be 51 ± 1.4, 53 ± 1.4 and 69 ± 1.9 µm, for the highest to lowest dose rate, respectively. The variation in response suggested the photodetector possessed dose-rate dependence. However, its ability to reconstruct the microbeam profile was affected by the presence of additional dose peaks adjacent to the one generated by the X-ray microbeam. Geant4 simulations proved that the additional peaks were due to optical photons generated in the barrier film coupled to the sensitive volume. The simulations also confirmed that the amplitude of the additional peak in comparison with the microbeam decreased for spectra with lower peak energies, as observed in the experimental data. The material packaging can be optimized during fabrication by solution processing onto a flexible substrate with a non-fluorescent barrier film. With these improvements, organic photodetectors show promising prospects as a cost-effective high spatial resolution tissue-equivalent flexible dosimeter for synchrotron radiation fields.


2021 ◽  
Vol 129 (24) ◽  
pp. 244902
Author(s):  
Jeremy A. Davis ◽  
Elette Engels ◽  
Marco Petasecca ◽  
Jason Paino ◽  
Moeava Tehei ◽  
...  

Cancers ◽  
2021 ◽  
Vol 13 (5) ◽  
pp. 936
Author(s):  
Laura Eling ◽  
Audrey Bouchet ◽  
Alexandre Ocadiz ◽  
Jean-François Adam ◽  
Sarvenaz Kershmiri ◽  
...  

Delivery of high-radiation doses to brain tumors via multiple arrays of synchrotron X-ray microbeams permits huge therapeutic advantages. Brain tumor (9LGS)-bearing and normal rats were irradiated using a conventional, homogeneous Broad Beam (BB), or Microbeam Radiation Therapy (MRT), then studied by behavioral tests, MRI, and histopathology. A valley dose of 10 Gy deposited between microbeams, delivered by a single port, improved tumor control and median survival time of tumor-bearing rats better than a BB isodose. An increased number of ports and an accumulated valley dose maintained at 10 Gy delayed tumor growth and improved survival. Histopathologically, cell death, vascular damage, and inflammatory response increased in tumors. At identical valley isodose, each additional MRT port extended survival, resulting in an exponential correlation between port numbers and animal lifespan (r2 = 0.9928). A 10 Gy valley dose, in MRT mode, delivered through 5 ports, achieved the same survival as a 25 Gy BB irradiation because of tumor dose hot spots created by intersecting microbeams. Conversely, normal tissue damage remained minimal in all the single converging extratumoral arrays. Multiport MRT reached exceptional ~2.5-fold biological equivalent tumor doses. The unique normal tissue sparing and therapeutic index are eminent prerequisites for clinical translation.


2021 ◽  
Vol 109 (2) ◽  
pp. 626-636
Author(s):  
Franziska Treibel ◽  
Mai Nguyen ◽  
Mabroor Ahmed ◽  
Annique Dombrowsky ◽  
Jan J. Wilkens ◽  
...  

2020 ◽  
Vol 152 ◽  
pp. S874
Author(s):  
M. Petasecca ◽  
M. Duncan ◽  
M. Donzelli ◽  
P. Pellicioli ◽  
E. Brauer-Krisch ◽  
...  

Sign in / Sign up

Export Citation Format

Share Document