Transmission electron microscopy of rapidly-frozen, hydrated specimens (cryo-TEM) is a powerful way of examining labile microstructures. This technique avoids some artifacts associated with conventional preparative methods. Use of a controlled environment vitrification system (CEVS) for specimen preparation reduces the risk of unwanted sample changes due to evaporation, and permits the examination of specimens vitrified from a defined temperature. Studies of dynamic processes with time resolution on the order of seconds, in which the process was initiated by changes in sample pH, have been conducted. We now report the development of an optical method for increasing specimen temperature immediately before vitrification. Using our method, processes that are regulated by temperature can be initiated in less than 500 msec on the specimen grid. The ensuing events can then be captured by plunge-freezing within an additional 200 msec.Dimyristoylphosphatidylcholine (DMPC) liposomes, produced by extrusion, were used as test specimens. DMPC undergoes a gel/liquid crystalline transition at 24°C, inducing a change in liposome morphology from polyhedral to spherical. Five-μl aliquots of DMPC dispersions were placed on holey-carbon-filmed copper grids mounted in the CEVS environmental chamber, and maintained at 6-8°C and 80% relative humidity. Immediately before the temperature jump most of the sample was blotted away with filter paper, leaving a thin specimen film on the grid. Upon pressing the trigger, an electronic control circuit generated this timed sequence of events. First, a solenoid-activated shutter was opened to heat the specimen by exposing it for a variable time to the focused beam of a 75W Xenon arc lamp. Simultaneously, a solenoid-activated cryogen shutter in the bottom of the CEVS was opened. Next, the lamp shutter was closed after the desired heating interval. Finally, a solenoid-activated cable release was used to trigger a spring-loaded plunger in the CEVS, propelling the sample into a reservoir of liquid ethane. Vitrified samples were subsequently transferred to a Zeiss EM902 TEM, operated in zero-loss brightfield mode, for examination at −163°C.