scholarly journals Harmonization of Measurement Results of the Alcohol Biomarker Carbohydrate-Deficient Transferrin by Use of the Toolbox of Technical Procedures of the International Consortium for Harmonization of Clinical Laboratory Results

2014 ◽  
Vol 60 (7) ◽  
pp. 945-953 ◽  
Author(s):  
Cas Weykamp ◽  
Jos Wielders ◽  
Anders Helander ◽  
Raymond F Anton ◽  
Vincenza Bianchi ◽  
...  

Abstract BACKGROUND The need for equivalent results of routine measurement procedures for the alcohol biomarker carbohydrate-deficient transferrin (CDT) has been recognized by the IFCC. This article describes a project to harmonize CDT as conducted by an IFCC working group initiated for this purpose. METHODS We used procedures for achieving harmonization as developed by the Consortium for Harmonization of Clinical Laboratory Results to assess the suitability of a candidate reference measurement procedure (cRMP), candidate reference materials (cRMs), and the success of efforts to achieve harmonization. RESULTS CDT measurement procedures in routine use showed good reproducibility (CV 1.1%–2.8%) and linearity (r > 0.990) with variable slopes (0.766–1.065) and intercepts (−0.34 to 0.92) compared to the cRMP. Heterogeneity after simulated harmonization was 4.7%. cRMs of frozen human native sera demonstrated commutability and 3-year stability for routine measurement procedures. The cRMP provided reproducible value assignment to cRMs with an expanded uncertainty (k = 2) of 0.03% at the 1.2% CDT level and 0.06% at the 4.4% CDT level. Harmonization efforts reduced the intermeasurement CV from 8.8% to 3.4%, allowed 99% recovery of the values assigned with the cRMP, and demonstrated 99% of results within the desirable allowable total error. Harmonization was less successful in samples with low CDT and high trisialotransferrin concentrations. CONCLUSIONS Harmonization of CDT is possible with frozen human native sera as cRMs with values assigned by use of the cRMP. We propose the cRMP as a candidate international conventional reference measurement procedure and cRMs as candidate international calibrators.

Author(s):  
Daisuke Manita ◽  
Hiroshi Yoshida ◽  
Isao Koyama ◽  
Masakazu Nakamura ◽  
Yuji Hirowatari

Abstract Background A new lipoprotein testing method based on anion-exchange HPLC (AEX-HPLC) was recently established. We verified the accuracy of LDL-C levels, a primary therapeutic target for the prevention of cardiovascular disease (CVD), measured by AEX-HPLC comparing with LDL-C levels measured by beta quantification-reference measurement procedure (BQ-RMP), homogenous assays, and calculation methods. Methods We compared LDL-C levels measured by AEX-HPLC (adLDL-Ch: LDL-Ch and IDL-Ch) and BQ-RMP using blood samples from 52 volunteers. AdLDL-Ch levels were also compared with those measurements by homogeneous assays and calculation methods (Friedewald equation, Martin equation, and Sampson equation) using blood samples from 411 participants with dyslipidemia and/or type 2 diabetes. Results The precision and accuracy of adLDL-Ch were verified by BQ-RMP. The mean percentage bias [bias (%)] for LDL-C was 1.2%, and the correlation was y = 0.990x + 3.361 (r = 0.990). These results met the acceptable range of accuracy prescribed by the National Cholesterol Education Program. Additionally, adLDL-Ch levels were correlated with LDL-C levels measured by the 2 homogeneous assays (r > 0.967) and the calculation methods (r > 0.939), in serum samples from patients with hypertriglyceridemia. Conclusions AEX-HPLC is a reliable method for measuring LDL-C levels for CVD risk in daily clinical laboratory analyses.


2008 ◽  
Vol 132 (5) ◽  
pp. 838-846
Author(s):  
W. Greg Miller ◽  
Gary L. Myers ◽  
Edward R. Ashwood ◽  
Anthony A. Killeen ◽  
Edward Wang ◽  
...  

Abstract Context.—Harmonization and standardization of results among different clinical laboratories is necessary for clinical practice guidelines to be established. Objective.—To evaluate the state of the art in measuring 10 routine chemistry analytes. Design.—A specimen prepared as off-the-clot pooled sera and 4 conventionally prepared specimens were sent to participants in the College of American Pathologists Chemistry Survey. Analyte concentrations were assigned by reference measurement procedures. Participants.—Approximately 6000 clinical laboratories. Results.—For glucose, iron, potassium, and uric acid, more than 87.5% of peer groups meet the desirable bias goals based on biologic variability criteria. The remaining 6 analytes had less than 52% of peer groups that met the desirable bias criteria. Conclusions.—Routine measurement procedures for some analytes had acceptable traceability to reference systems. Conventionally prepared proficiency testing specimens were not adequately commutable with a fresh frozen specimen to be used to evaluate trueness of methods compared with a reference measurement procedure.


2011 ◽  
Vol 57 (4) ◽  
pp. 614-622 ◽  
Author(s):  
Selvin H Edwards ◽  
Mary M Kimberly ◽  
Susan D Pyatt ◽  
Shelton L Stribling ◽  
Kara D Dobbin ◽  
...  

BACKGROUND Our purpose was to establish a mass spectrometry reference measurement procedure (RMP) for cholesterol to use in the CDC's standardization programs. We explored a gas chromatography–isotope dilution mass spectrometry (GC-IDMS) procedure using a multilevel standard calibration curve to quantify samples with varying cholesterol concentrations. METHODS We calibrated the mass spectrometry instrument by isotope dilution with a pure primary standard reference material and an isotopically enriched cholesterol analog as the internal standard (IS). We diluted the serum samples with Tris-HCl buffer (pH 7.4, 0.05 mol/L, 0.25% Triton X-100) before analysis. We used 17 serum pools, 10 native samples, and 2 standard reference materials (SRMs). We compared the GC-IDMS measurements with the CDC's modified Abell–Levy–Brodie–Kendall (AK) RMP measurements and assessed method accuracy by analyzing 2 SRMs. We evaluated the procedure for lack of interference by analyzing serum spiked with a mixture of 7 sterols. RESULTS The mean percent bias between the AK and the GC-IDMS RMP was 1.6% for all samples examined. The mean percent bias from NIST's RMP was 0.5% for the SRMs. The total %CVs for SRM 1951b levels I and II were 0.61 and 0.73%, respectively. We found that none of the sterols investigated interfered with the cholesterol measurement. CONCLUSIONS The low imprecision, linear response, lack of interferences, and acceptable bias vs the NIST primary RMP qualifies this procedure as an RMP for determining serum cholesterol. The CDC will adopt and implement this GC-IDMS procedure for cholesterol standardization.


2018 ◽  
Vol 64 (9) ◽  
pp. 1296-1307 ◽  
Author(s):  
Alexandra S Whale ◽  
Gerwyn M Jones ◽  
Jernej Pavšič ◽  
Tanja Dreo ◽  
Nicholas Redshaw ◽  
...  

Abstract BACKGROUND Genetic testing of tumor tissue and circulating cell-free DNA for somatic variants guides patient treatment of many cancers. Such measurements will be fundamental in the future support of precision medicine. However, there are currently no primary reference measurement procedures available for nucleic acid quantification that would support translation of tests for circulating tumor DNA into routine use. METHODS We assessed the accuracy of digital PCR (dPCR) for copy number quantification of a frequently occurring single-nucleotide variant in colorectal cancer (KRAS c.35G>A, p.Gly12Asp, from hereon termed G12D) by evaluating potential sources of uncertainty that influence dPCR measurement. RESULTS Concentration values for samples of KRAS G12D and wild-type plasmid templates varied by <1.2-fold when measured using 5 different assays with varying detection chemistry (hydrolysis, scorpion probes, and intercalating dyes) and <1.3-fold with 4 commercial dPCR platforms. Measurement trueness of a selected dPCR assay and platform was validated by comparison with an orthogonal method (inductively coupled plasma mass spectrometry). The candidate dPCR reference measurement procedure showed linear quantification over a wide range of copies per reaction and high repeatability and interlaboratory reproducibility (CV, 2%–8% and 5%–10%, respectively). CONCLUSIONS This work validates dPCR as an SI-traceable reference measurement procedure based on enumeration and demonstrates how it can be applied for assignment of copy number concentration and fractional abundance values to DNA reference materials in an aqueous solution. High-accuracy measurements using dPCR will support the implementation and traceable standardization of molecular diagnostic procedures needed for advancements in precision medicine.


2017 ◽  
Vol 63 (10) ◽  
pp. 1642-1652 ◽  
Author(s):  
Linde A C De Grande ◽  
Katleen Van Uytfanghe ◽  
Dries Reynders ◽  
Barnali Das ◽  
James D Faix ◽  
...  

Abstract BACKGROUND The IFCC Committee for Standardization of Thyroid Function Tests intended to standardize free thyroxine (FT4) immunoassays. We developed a Système International d'Unités traceable conventional reference measurement procedure (RMP) based on equilibrium dialysis and mass spectrometry. We describe here the latest studies intended to recalibrate against the RMP and supply a proof of concept, which should allow continued standardization efforts. METHODS We used the RMP to target the standardization and reference interval (RI) panels, which were also measured by 13 manufacturers. We validated the suitability of the recalibrated results to meet specifications for bias (3.3%) and total error (8.0%) determined from biological variation. However, because these specifications were stringent, we expanded them to 10% and 13%, respectively. The results for the RI panel were reported as if the assays were recalibrated. We estimated all but 1 RI using parametric statistical procedures and hypothesized that the RI determined by the RMP was suitable for use by the recalibrated assays. RESULTS Twelve of 13 recalibrated assays had a bias, meeting the 10% specification with 95% confidence; for 7 assays, this applied even for the 3.3% specification. Only 1 assay met the 13% total error specification. Recalibration reduced the CV of the assay means for the standardization panel from 13% to 5%. The proof-of-concept study confirmed our hypothesis regarding the RI but within constraints. CONCLUSIONS Recalibration to the RMP significantly reduced the FT4 immunoassays' bias, so that the RI determined by the RMP was suitable for common use within a margin of 12.5%.


Sign in / Sign up

Export Citation Format

Share Document