scholarly journals Comparison of Image Quality and Spatial Resolution Between 18F, 68Ga and 64Cu Phantom Measurements Using a Digital Biograph Vision PET/CT

Author(s):  
Anja Braune ◽  
Liane Oehme ◽  
Robert Freudenberg ◽  
Frank Hofheinz ◽  
Jörg van den Hoff ◽  
...  

Abstract Background: The PET nuclide and reconstruction method can have a considerable influence on spatial resolution and image quality of PET/CT scans, which can, for example, influence the diagnosis in oncology. The individual impact of the positron energy of 18F, 68Ga and 64Cu on spatial resolution and image quality of PET/CT scans acquired using a clinical, digital scanner was compared. Furthermore, the impact of different reconstruction parameters on image quality and spatial resolution was evaluated for 18F-FDG PET/CT scans acquired with a scanner of the newest generation. Methods: PET/CT scans of a Jaszczak phantom and a NEMA PET body phantom, filled with 18F-FDG, 68Ga-HCl and 64Cu-HCl, respectively, were performed on a Siemens Biograph Vision. Images were assessed using spatial resolution and image quality (Recovery Coefficients (RC), coefficient of variation within the background, Contrast Recovery Coefficient (CRC), Contrast-Noise-Ratio (CNR), and relative count error in lung insert). In a subsequent analysis, the scan of the NEMA PET body phantom filled with 18F-FDG was reconstructed applying different parameters (with/without the application of Point Spread Function (PSF), Time of Flight (ToF) or post-filtering; matrix size). Spatial resolution and quantitative image quality were compared between reconstructions. Results: We found that image quality was comparable between 18F-FDG and 64Cu-HCl PET/CT measurements featuring similar maximal endpoint energy. In comparison, RC, CRC and CNR were worse in 68Ga-HCl data, despite similar count rates. Spatial resolution was up to 18 % worse in 68Ga-HCl compared to 18F-FDG images. Post-filtering of 18F-FDG acquisitions changed image quality the most and reduced spatial resolution by 52 % if a Gaussian filter with 5 mm FWHM was applied. ToF measurements especially improved the recovery of the smallest lesion (RCmean = 1.07 compared to 0.65 without ToF) and improved spatial resolution by 29 %.Conclusions: The positron energy of PET nuclides influences spatial resolution and image quality of digital PET/CT scans. Image quality of 68Ga-HCl PET/CT images was worse compared to 18F-FDG and 64Cu-HCl, respectively, despite similar count rates. Reconstruction parameters have a high impact on image quality and spatial resolution and should be considered when comparing images of different scanners or centers.

2021 ◽  
pp. 20201356
Author(s):  
Feng-Jiao Yang ◽  
Shu-Yue Ai ◽  
Runze Wu ◽  
Yang Lv ◽  
Hui-Fang Xie ◽  
...  

Objectives: To investigate the impact of total variation regularized expectation maximization (TVREM) reconstruction on the image quality of 68Ga-PSMA-11 PET/CT using phantom and patient data. Methods: Images of a phantom with small hot sphere inserts and 20 prostate cancer patients were acquired with a digital PET/CT using list-mode and reconstructed with ordered subset expectation maximization (OSEM) and TVREM with seven penalisation factors between 0.01 and 0.42 for 2 and 3 minutes-per-bed (m/b) acquisition. The contrast recovery (CR) and background variability (BV) of the phantom, image noise of the liver, and SUVmax of the lesions were measured. Qualitative image quality was scored by two radiologists using a 5-point scale (1-poor, 5-excellent). Results: The performance of CR, BV, and image noise, and the gain of SUVmax was higher for TVREM 2 m/b groups with the penalization of 0.07 to 0.28 compared to OSEM 3 m/b group (all p < 0.05). The image noise of OSEM 3 m/b group was equivalent to TVREM 2 and 3 m/b groups with a penalization of 0.14 and 0.07, while lesions’ SUVmax increased 15 and 20%. The highest qualitative score was attained at the penalization of 0.21 (3.30 ± 0.66) for TVREM 2 m/b groups and the penalization 0.14 (3.80 ± 0.41) for 3 m/b group that equal to or greater than OSEM 3 m/b group (2.90 ± 0.45, p = 0.2 and p < 0.001). Conclusions: TVREM improves lesion contrast and reduces image noise, which allows shorter acquisition with preserved image quality for PSMA PET/CT. Advances in knowledge: TVREM reconstruction with optimized penalization factors can generate higher quality PSMA-PET images for prostate cancer diagnosis.


2015 ◽  
Vol 2015 ◽  
pp. 1-11 ◽  
Author(s):  
Dustin R. Osborne ◽  
Derek W. Austin

Multi-modality imaging provides coregistered PET-CT and SPECT-CT images; however such multi-modality workflows usually consist of sequential scans from the individual imaging components for each modality. This typical workflow may result in long scan times limiting throughput of the imaging system. Conversely, acquiring multi-modality data simultaneously may improve correlation and registration of images, improve temporal alignment of the acquired data, increase imaging throughput, and benefit the scanned subject by minimizing time under anesthetic. In this work, we demonstrate the feasibility and procedure for modifying a commercially available preclinical SPECT-CT platform to enable simultaneous SPECT-CT acquisition. We also evaluate the performance of simultaneous SPECT-CT tomographic imaging with this modified system. Performance was accessed using a 57Co source and image quality was evaluated with Tc99m phantoms in a series of simultaneous SPECT-CT scans.


2018 ◽  
Vol 4 (Supplement 2) ◽  
pp. 68s-68s
Author(s):  
B. Ali ◽  
A. Afshan ◽  
M.B. Kakakhel

Background: In PET imaging both quantitative and qualitative interpretations are used. Qualitative and quantitative interpretations depend upon PET/CT image quality that along with many biologic factors strongly depends upon image reconstruction parameters. Aim: The objective of this experimental work was to study the impact of one of the key reconstruction parameter, i.e., number of reconstruction iteration, on standardized uptake value and image quality of PET/CT scan. Methods: Images of NEMA IEC Image Quality Phantom were acquired in list mode for 10 mins on Discovery STE PET/CT scanner, using tumor to background ratio of 4:1 and 18F-FDG as radiotracer. List mode data were further transformed into data sets of varying acquisition time (0.5, 1.0, 1.5, 2.0, 2.5, 3.0, 3.5, 4.0, 4.5, 5.0 mins) per bed position. Transformed data set of 5.0 mins were used to study the impact of varying number of iterations (2, 3, 4, 5, 7, 10, 15, 20) using OSEM approach of iterative reconstruction. Standardized uptake value (SUV) and underestimation in SUV were calculated as quantitative measures, while hot lesion contrast, cold lesion contrast and background variability were calculated as qualitative measures. Results: Standardized uptake value, hot and cold lesion contrast, image spatial resolution and background variability showed increasing trend with increase in reconstruction iterations. Maximum increase of 20.25%, 16.33%, 9.79% and 6.88% was observed in SUV for 10 mm, 13 mm, 17 mm and 22 mm lesions as number of iteration change from 2 to 3. Smallest and the largest diameter lesions showed maximum underestimations of 54.67% and 8.20% at 2 iterations respectively. Percentage hot lesion contrast showed rapid increase as the number of iteration change from 2 to 7 and increased slowly afterward. Background variability range from 4.4% to 6.4%, 4.1%–5.7%, 3.6%–4.6%, 3%–3.8%, 2.7%–3.2%, 2.4%–2.7% for 10.0 mm, 13.0 mm, 17.0 mm, 22.0 mm, 28.0 mm and 37.0 mm sphere respectively. Conclusion: Optimized reconstruction parameters for routine clinical studies 3 iterations with image matrix size of 128 × 128 with filters FWHM of 6 mm and for high resolution studies 3 iterations image matrix size of 256×256 with filters FWHM of 6 mm.


2021 ◽  
Author(s):  
Sebastijan Rep ◽  
Petra Tomše ◽  
Luka Jensterle ◽  
Katja Zaletel ◽  
Luka Ležaič

Abstract Background PET/CT imaging is widely used in oncology and provides both metabolic and anatomic information. Because of the relatively poor spatial resolution of PET/CT imaging technique the detection of small lesions is limited. The low spatial resolution introduces the partial-volume effect (PVE) which negatively affects images both visually and quantitatively. The aim of our research was to investigate the effect of 4 mm and 2 mm voxel size on image quality and on detection of small spheres. MethodsWe used the NEMA body phantom with six fillable spheres. The spheres and background were filled with a solution of 18F-FDG, in ratio spheres vs background 2:1, 3:1, 4:1 and 8:1 In all images reconstructed with 2 mm and 4 mm voxel size the contrast recovery coefficient (CRC), contrast to noise ratio (CNR) in standardized uptake value (SUV) were evaluated.ResultsFor phantom spheres ≤ 13 mm, we found significant higher CRC, SUV and CNR using small-voxel reconstructions. CRC and SUV did not differ for large spheres (≥ 17 mm) using 2 mm and 4 mm voxel size. On the other hand, CNR for large spheres (≥ 17 mm) was significantly decreased in 2 mm voxel size images compared to the 4 mm.ConclusionAccording to our results, the reconstruction with 2 mm voxel size can improve precise lesion localization, image contrast, and image quality.


2015 ◽  
Vol 49 (3) ◽  
pp. 227-233 ◽  
Author(s):  
Alen Suljic ◽  
Petra Tomse ◽  
Luka Jensterle ◽  
Damijan Skrk

Abstract Background. The aim of the study was to explore the influence of various time-of-flight (TOF) and non-TOF reconstruction algorithms on positron emission tomography/computer tomography (PET/CT) image quality. Materials and methods. Measurements were performed with a triple line source phantom, consisting of capillaries with internal diameter of ~ 1 mm and standard Jaszczak phantom. Each of the data sets was reconstructed using analytical filtered back projection (FBP) algorithm, iterative ordered subsets expectation maximization (OSEM) algorithm (4 iterations, 24 subsets) and iterative True-X algorithm incorporating a specific point spread function (PSF) correction (4 iterations, 21 subsets). Baseline OSEM (2 iterations, 8 subsets) was included for comparison. Procedures were undertaken following the National Electrical Manufacturers Association (NEMA) NU-2-2001 protocol. Results. Measurement of spatial resolution in full width at half maximum (FWHM) was 5.2 mm, 4.5 mm and 2.9 mm for FBP, OSEM and True-X; and 5.1 mm, 4.5 mm and 2.9 mm for FBP+TOF, OSEM+TOF and True-X+TOF respectively. Assessment of reconstructed Jaszczak images at different concentration ratios showed that incorporation of TOF information improves cold contrast, while hot contrast only slightly, however the most prominent improvement could be seen in background variability - noise reduction. Conclusions. On the basis of the results of investigation we concluded, that incorporation of TOF information in reconstruction algorithm mostly affects reduction of the background variability (levels of noise in the image), while the improvement of spatial resolution due to incorporation of TOF information is negligible. Comparison of traditional and modern reconstruction algorithms showed that analytical FBP yields comparable results in some parameter measurements, such as cold contrast and relative count error. Iterative methods show highest levels of hot contrast, when TOF and PSF corrections were applied simultaneously.


2021 ◽  
Vol 15 ◽  
pp. 174830262110080
Author(s):  
Changjun Zha* ◽  
Qian Zhang* ◽  
Huimin Duan

Traditional single-pixel imaging systems are aimed mainly at relatively static or slowly changing targets. When there is relative motion between the imaging system and the target, sizable deviations between the measurement values and the real values can occur and result in poor image quality of the reconstructed target. To solve this problem, a novel dynamic compressive imaging system is proposed. In this system, a single-column digital micro-mirror device is used to modulate the target image, and the compressive measurement values are obtained for each column of the image. Based on analysis of the measurement values, a new recovery model of dynamic compressive imaging is given. Differing from traditional reconstruction results, the measurement values of any column of vectors in the target image can be used to reconstruct the vectors of two adjacent columns at the same time. Contingent upon characteristics of the results, a method of image quality enhancement based on an overlapping average algorithm is proposed. Simulation experiments and analysis show that the proposed dynamic compressive imaging can effectively reconstruct the target image; and that when the moving speed of the system changes within a certain range, the system reconstructs a better original image. The system overcomes the impact of dynamically changing speeds, and affords significantly better performance than traditional compressive imaging.


2015 ◽  
Vol 19 (4) ◽  
pp. 791-813 ◽  
Author(s):  
Zilia Iskoujina ◽  
Joanne Roberts

Purpose – This paper aims to add to the understanding of knowledge sharing in online communities through an investigation of the relationship between individual participant’s motivations and management in open source software (OSS) communities. Drawing on a review of literature concerning knowledge sharing in organisations, the factors that motivate participants to share their knowledge in OSS communities, and the management of such communities, it is hypothesised that the quality of management influences the extent to which the motivations of members actually result in knowledge sharing. Design/methodology/approach – To test the hypothesis, quantitative data were collected through an online questionnaire survey of OSS web developers with the aim of gathering respondents’ opinions concerning knowledge sharing, motivations to share knowledge and satisfaction with the management of OSS projects. Factor analysis, descriptive analysis, correlation analysis and regression analysis were used to explore the survey data. Findings – The analysis of the data reveals that the individual participant’s satisfaction with the management of an OSS project is an important factor influencing the extent of their personal contribution to a community. Originality/value – Little attention has been devoted to understanding the impact of management in OSS communities. Focused on OSS developers specialising in web development, the findings of this paper offer an important original contribution to understanding the connections between individual members’ satisfaction with management and their motivations to contribute to an OSS project. The findings reveal that motivations to share knowledge in online communities are influenced by the quality of management. Consequently, the findings suggest that appropriate management can enhance knowledge sharing in OSS projects and online communities, and organisations more generally.


Author(s):  
Balakrishnan A ◽  
Dr. KVM Varambally

The impact of Globalization creates a competitive horizon with new markets, new products, new mindsets, new competencies, and new ways of thinking about business. As an outcome of this, the market places become more complex, uncertain, competitive, and transformational. In this era, the major challenge faced by any organization is to maintain a peaceful as well positive culture and climate inside the organization. The human resource development climate of an organization plays a significant role in ensuring the competency, motivation, and development of its employees. Here the term climate denotes the quality of the internal environment. It influences morale and the attitudes of the individual toward his work and his environment. Culture refers to the deep structure of organizations, which is rooted in the values, beliefs, and assumptions held by organizational members. The study was aimed to analyze the Human Resource Development culture and climate of chemical-based public sector enterprises in Kerala. The results indicated that the Human Resource Development climate in an organization is average and the perception of employees regarding Human Resource Development culture and climate does not differs significantly based on age, job approval status, and qualification. But it differs significantly based on gender and experiences.


2015 ◽  
Vol 15 (7) ◽  
pp. 88-98
Author(s):  
J. Dezert ◽  
A. Tchamova ◽  
P. Konstantinova

Abstract The main purpose of this paper is to apply and to test the performance of a new method, based on belief functions, proposed by Dezert et al. in order to evaluate the quality of the individual association pairings provided in the optimal data association solution for improving the performances of multisensor-multitarget tracking systems. The advantages of its implementation in an illustrative realistic surveillance context, when some of the association decisions are unreliable and doubtful and lead to potentially critical mistake, are discussed. A comparison with the results obtained on the base of Generalized Data Association is made.


Sign in / Sign up

Export Citation Format

Share Document