scholarly journals Early Benchmarking Total Hip Arthroplasty Implants Using Data from the Michigan Arthroplasty Registry Collaborative Quality Initiative (MARCQI)

2021 ◽  
Vol Volume 13 ◽  
pp. 215-228
Author(s):  
Heather A Chubb ◽  
Eric R Cornish ◽  
Brian R Hallstrom ◽  
Richard E Hughes
2020 ◽  
pp. 112070002096964
Author(s):  
Kirill Gromov ◽  
Nanna H Sillesen ◽  
Thomas Kallemose ◽  
Henrik Husted ◽  
Henrik Malchau ◽  
...  

Background: Introduction of new implants should be monitored closely to capture any signs of compromising patient safety. Vitamin E infused highly-crosslinked polyethylene liners (VEPE) offer the potential for reduced wear. Highwall liners have been hypothesised to result in increased wear and potential liner fractures. The aim of this study was to determine the 3–7-year follow-up of highwall VEPE for primary total hip arthroplasty (THA), focusing on liner-related complications. Methods: We included 1221 consecutive THA operations from July 2010 to May 2014 with minimum follow-up of 3 (3.1–6.8) years Data collected included demographics, implant data, complications, reoperations, and deaths. Data were cross-referenced with the Danish Hip Arthroplasty Registry in order to ensure validity and completeness. Acetabular shell position was measured using Martell Hip Analysis Suite in a subgroup of 931 THAs. Results: Cumulative stem revision and shell revision at 3-year follow-up was 3.4% and 0.4% respectively. There were no revisions due to liner failure. Reason for revision included 11 dislocations, 15 soft-tissue revisions for infection, 44 stem revisions of which 34 were periprosthetic fractures and 13 shell revisions of which 6 were combined shell and stem revisions. Conclusion: Early follow-up of VEPE liners for primary THA have not shown any revisions associated with liner failure. Continued monitoring of new materials are necessary to capture any signs of compromised patient safety.


2018 ◽  
Vol 43 (8) ◽  
pp. 1815-1821 ◽  
Author(s):  
Francesco Castagnini ◽  
Barbara Bordini ◽  
Susanna Stea ◽  
Pierina Paola Calderoni ◽  
Claudio Masetti ◽  
...  

Author(s):  
Francesco Castagnini ◽  
Barbara Bordini ◽  
Monica Cosentino ◽  
Cristina Ancarani ◽  
Federica Mariotti ◽  
...  

Abstract Introduction Recurrent dislocations are still the most frequent reason for revision in total hip arthroplasty (THA). The impact of bearing surfaces on dislocations is still controversial. We hypothesized that: (1) bearing surfaces influence the revisions due to dislocations; (2) ceramic-on-ceramic reduced the revisions for dislocations in adjusted models; (3) Delta-on-Delta bearings reduced the revisions for dislocations in comparison to surfaces with cross-linked polyethylene. Materials and methods The regional arthroplasty registry was enquired about bearing surfaces and revisions for dislocations and instability. Unadjusted and adjusted rates were provided, including sex, age (<65 years or ≥65 years), head diameter (≤28 mm or >28 mm; <36 mm or ≥36 mm) as variables. 44,065 THAs were included. Results The rate of revisions for dislocations was significantly lower in ceramic-on-ceramic and metal-on-metal bearings (unadjusted rates). After adjusting for age, sex, and head size (36 and 28 mm), hard-on-hard bearings were protective (p < 0.05): ceramic-on-ceramic had a lower risk of revisions due to dislocation than ceramic-on-polyethylene (HR 1.6, 95% CI 1.2–2.2 p = 0.0009). The rate of revisions for dislocation was similar in bearings with cross-linked polyethylene and Delta-on-Delta articulations, in unadjusted and adjusted models. Conclusion Bearings with conventional polyethylene were more predisposed to dislocations. Currently adopted bearings exerted no significant influence on revisions due to dislocations. These findings could be primarily related to wear, but due to the time distribution, soft tissue envelopes and surface tension may also play a role. Pre-clinical biomechanical evaluations and prospective matched cohort studies are required to draw definitive conclusions.


2011 ◽  
Vol 32 (3) ◽  
pp. 296-297 ◽  
Author(s):  
LJ Worth ◽  
AL Bull ◽  
MJ Richards

The risk of surgical site infection (SSI) is greater after revision hip arthroplasty than after primary procedures. While this is accepted as a clinical phenomenon, standardized surveillance strategies for healthcare-associated infections, including SSIs, do not currently take this into consideration. Most notably, the National Nosocomial Infections Surveillance (NNIS) risk index for stratification does not differentiate between primary and revision surgeries. Using data from a single US center, Leekha et al. recently demonstrated that the risk for SSI was almost twice as high after revision total hip arthroplasty when compared to primary total hip arthroplasty and that risk was even greater when deep incisional or organ/space infections were analyzed. The objective of this study was to compare SSI rates following primary and revision hip arthroplasty in a much larger Australian population to determine whether differences are accounted for by current risk indexing.


2015 ◽  
Vol 2015 ◽  
pp. 1-4
Author(s):  
Zuned Hakim ◽  
Claire Rutherford ◽  
Elizabeth Mckiernan ◽  
Tony Helm

Obesity is becoming a critical problem in the developed world and is associated with an increased incidence of osteoarthritis of the hip. The Oxford Hip Score was used to determine if Body Mass Index (BMI) is an independent factor in determining patient outcome following primary total hip arthroplasty (THA). Using data from 353 operations we found that patients with BMI ≥ 30 had an absolute score that was lower preoperatively and postoperatively compared to those with a BMI < 30. There was no difference in pre- and postoperative point score change within each group; Kendall’s rank correlation was 0.00047 (95% CI, −0.073 to 0.074 (p=0.99)) and demonstrated no trend. There was no statistically significant difference in change between those with BMI ≥ 30 and < 30 (p=0.65). We suggest that those with a higher BMI be considered for THA as they can expect the same degree of improvement as those with a lower BMI. Given the on-going increase in obesity these findings could be significant for the future of THA.


2022 ◽  
pp. oemed-2021-107970
Author(s):  
Elena Zaballa ◽  
Georgia Ntani ◽  
E Clare Harris ◽  
Anne Lübbeke ◽  
Nigel K Arden ◽  
...  

ObjectivesTo investigate the rates of return to work and workability among working-age people following total hip arthroplasty (THA).MethodsParticipants from the Geneva Arthroplasty Registry and the Clinical Outcomes for Arthroplasty Study aged 18–64 years when they had primary THA and with at least 5 years’ follow-up were mailed a questionnaire 2017–2019. Information was collected about preoperative and post-THA employment along with exposure to physically demanding activities at work or in leisure. Patterns of change of job were explored. Survival analyses using Cox proportional hazard models were created to explore risk factors for having to stop work because of difficulties with the replaced hip.ResultsIn total, 825 returned a questionnaire (response 58%), 392 (48%) men, mean age 58 years, median follow-up 7.5 years post-THA. The majority (93%) of those who worked preoperatively returned to work, mostly in the same sector but higher rates of non-return (36%–41%) were seen among process, plant and machine operatives and workers in elementary occupations. 7% reported subsequently leaving work because of their replaced hip and the risk of this was strongly associated with: standing >4 hours/day (HR 3.81, 95% CI 1.62 to 8.96); kneeling/squatting (HR 3.32, 95% CI 1.46 to 7.55) and/or carrying/lifting ≥10 kg (HR 5.43, 95% CI 2.29 to 12.88).ConclusionsIt may be more difficult to return to some (particularly physically demanding) jobs post-THA than others. Rehabilitation may need to be targeted to these types of workers or it may be that redeployment or job change counselling are required.


2018 ◽  
Vol 188 (3) ◽  
pp. 873-877 ◽  
Author(s):  
Gerard A. Sheridan ◽  
Raymond M. Kelly ◽  
Suzanne M. McDonnell ◽  
Fionnuala Walsh ◽  
John M. O’Byrne ◽  
...  

2021 ◽  
Vol 10 (6) ◽  
pp. 1193
Author(s):  
Lisa Renner ◽  
Carsten Perka ◽  
Oliver Melsheimer ◽  
Alexander Grimberg ◽  
Volkmar Jansson ◽  
...  

Periprosthetic joint infection (PJI) is one of the most common complications in total hip arthroplasty (THA). The influence of bearing material on the risk of PJI remains unclear to date. This registry-based matched study investigates the role of bearing partners in primary cementless THA. Primary cementless THAs recorded in the German Arthroplasty Registry since 2012 with either a ceramic-on-ceramic (CoC) or ceramic-on-polyethylene (CoP) bearings were included in the analysis. Using propensity score matching (PSM) for age, sex, obesity, diabetes mellitus, Elixhauser comorbidity index, year of surgery and head size, we compared the risk for revision for PJI for CoC and CoP. Within the 115,538 THAs (87.1% CoP; 12.9% CoC), 977 revisions were performed due to PJI. There was a significantly higher risk for revision for PJI for CoP compared with CoC over the whole study period (p < 0.01) after 2:1 matching (CoP:CoC) with a hazard ratio of 1.41 (95% confidence interval (CI), 1.09 to 1.80) After 3 years, the risk for revision for PJI was 0.7% (CI 0.5–0.9%) for CoC and 0.9% (CI 0.8–1.1%) for CoP. The risk for revision for all other reasons except PJI did not significantly differ between the two groups over the whole study period (p = 0.4). Cementless THAs with CoC bearings were less likely to be revised because of infection in mid-term follow-up. In the future, registry-embedded studies focusing on long-term follow-up, including clinical data, as well as basic science studies, may give a deeper insight into the influence of the bearing partners.


Sign in / Sign up

Export Citation Format

Share Document