atomic force microscope cantilever
Recently Published Documents


TOTAL DOCUMENTS

108
(FIVE YEARS 12)

H-INDEX

27
(FIVE YEARS 1)

2020 ◽  
Vol 31 (18) ◽  
pp. 185702
Author(s):  
Steven J Eppell ◽  
David Friedenberg ◽  
Oliver Payton ◽  
Loren Picco ◽  
Fredy R Zypman

2019 ◽  
Vol 116 (50) ◽  
pp. 25008-25012 ◽  
Author(s):  
Dan Daniel ◽  
Chee Leng Lay ◽  
Anqi Sng ◽  
Coryl Jing Jun Lee ◽  
Darren Chi Jin Neo ◽  
...  

There is a huge interest in developing superrepellent surfaces for antifouling and heat-transfer applications. To characterize the wetting properties of such surfaces, the most common approach is to place a millimetric-sized droplet and measure its contact angles. The adhesion and friction forces can then be inferred indirectly using Furmidge’s relation. While easy to implement, contact angle measurements are semiquantitative and cannot resolve wetting variations on a surface. Here, we attach a micrometric-sized droplet to an atomic force microscope cantilever to directly measure adhesion and friction forces with nanonewton force resolutions. We spatially map the micrometer-scale wetting properties of superhydrophobic surfaces and observe the time-resolved pinning–depinning dynamics as the droplet detaches from or moves across the surface.