A biomechanical comparison of steel screws versus PLLA and magnesium screws for the Latarjet procedure
Abstract Introduction The fixation of the coracoid process onto the glenoid is an important step of the Latarjet procedure, and implant-associated complications are a relevant and severe problem. This study compares the fixation strength and failure mode of two biodegradable materials with stainless-steel screws. Methods 24 Fresh-frozen cadaveric scapulae were divided into three groups of equal size and received a coracoid transfer. Cadavers were matched according to their bone mineral density (BMD). In group 1, small-fragment screws made of stainless steel were used. In the second group, magnesium screws were used, and in the third group, screws consisted of polylactic acid (PLLA). A continuously increasing sinusoidal cyclic compression force was applied until failure occurred, which was defined as graft displacement relative to its initial position of more than 5 mm. Results At 5-mm displacement, the axial force values showed a mean of 374 ± 92 N (range 219–479 N) in group 1 (steel). The force values in group 2 (magnesium) had a mean of 299 ± 57 N (range 190–357 N). In group 3 (PLLA), failure occurred at 231 ± 83 N (range 109–355 N). The difference between group 1 (steel) and group 2 (magnesium) was not statistically significant (P = 0.212), while the difference between group 1 (steel) and group 3 (PLLA) was significant (P = 0.005). Conclusion Stainless-Steel screws showed the highest stability. However, all three screw types showed axial force values of more than 200 N. Stainless steel screws and PLLA screws showed screw cut-out as the most common failure mode, while magnesium screws showed screw breakage in the majority of cases. Evidence Controlled laboratory study.