scholarly journals Comparison of anterior chamber depth measurements by 3-dimensional optical coherence tomography, partial coherence interferometry biometry, Scheimpflug rotating camera imaging, and ultrasound biomicroscopy

2012 ◽  
Vol 38 (7) ◽  
pp. 1207-1213 ◽  
Author(s):  
Shunsuke Nakakura ◽  
Etsuko Mori ◽  
Nozomi Nagatomi ◽  
Hitoshi Tabuchi ◽  
Yoshiaki Kiuchi
2020 ◽  
Author(s):  
Tommy Chan ◽  
Marco Yu ◽  
Vivian Chiu ◽  
Gilda Lai ◽  
Christopher Leung ◽  
...  

Abstract Purpose To evaluate the repeatability and agreement of cornea and biometry measurements obtained with two swept-source optical coherence tomography (SSOCT) and a partial coherence interferometry-based device.METHODS This is a cross-sectional study. Forty-eight eyes of 48 patients had three consecutive measurements for ANTERION (Heidelberg Engineering, Germany), CASIAII (Tomey, Japan) and IOLMaster500 (Carl Zeiss Meditec, USA) on the same visit. Mean keratometry (Km), central cornea thickness (CCT), anterior chamber depth (ACD) and axial length (AL) were recorded. Corneal astigmatic measurements were converted into vector components – J0 and J45. Intra-device repeatability and agreements of measurements amongst the devices were evaluated using repeatability coefficients (RCs) and Bland-Altman plots, respectively. RESULTS All devices demonstrated comparable repeatability for Km (p≥0.138). ANTERION had the lowest RC for J0 amongst the devices (p≤0.039). Systematic difference was found for the Km and J0 obtained with IOLMaster500 compared to either SSOCTs (p≤0.010). The ACD and AL measured by IOLMaster500 showed a higher RC compared with either SSOCTs (p<0.002). Systematic difference was found in CCT and ACD between the two SSOCTs (p<0.001), and in AL between ANTERION and IOLMaster500 (p<0.001), with a mean difference of 1.6 µm, 0.022mm and 0.021 mm, respectively. CONCLUSIONS Both SSOCTs demonstrated smaller test-retest variability for measuring ACD and AL compared with IOLMaster500. There were significant disagreement in keratometry and AL measurements between the SSOCTs and PCI-based device; their measurements should not be considered as interchangeable but the disagreement could be refined by readjustment of intraocular lens constant in clinical practice.


2019 ◽  
Vol 47 (12) ◽  
pp. 5950-5956 ◽  
Author(s):  
Yiyong Qian ◽  
Lin Liu ◽  
Yuehui Shi ◽  
Minli Wang ◽  
Min Li ◽  
...  

Objective To investigate the clinical indications of ultrasound biomicroscopy (UBM) and anterior segment optical coherence tomography (AS-OCT) in eyes with inflammatory glaucoma. Methods Sixteen patients (16 eyes) with inflammatory glaucoma were analysed retrospectively. All of the subjects underwent UBM and AS-OCT scans, measuring the following parameters: central corneal thickness (CCT), central anterior chamber depth (ACD), angle-opening distance (AOD500), and trabecular-iris angle (TIA500) in four quadrants. Results Both the UBM and AS-OCT generated detailed, high-resolution images of the anterior segments of the eyes with inflammatory glaucoma. Moreover, the UBM and AS-OCT exhibited statistically similar measurement results for all of the indices. A Bland–Altman analysis showed a high level of agreement between the two imaging techniques. Additionally, the UBM was able to visualize the ciliary body and pars plana, while the cells and the flare in the anterior chamber appeared more pronounced in the UBM images. Conclusions Both the UBM and AS-OCT are useful tools for anterior segment imaging and taking measurements needed for evaluating inflammatory glaucoma. The AS-OCT has the advantages of being noninvasive and fast, whereas the UBM is presently unparalleled in the visualization of the ciliary body and pars plana.


2019 ◽  
Author(s):  
Ziyan Yu ◽  
Ting Huang ◽  
Qiao Yu ◽  
Jun Kong ◽  
Bo Qu

Abstract Purpose To compare anterior chamber depth (ACD) and central corneal thickness (CCT) measurements by anterior segment optical coherence tomography (AS-OCT), Lenstar and ultrasound biomicroscopy (UBM). Methods A retrospective cross section study. A total of 83 eyes from 54 patients were included. Central corneal thickness (CCT) and anterior chamber depth (ACD) values obtained from three devices (AS-OCT, Lenstar and UBM) were recorded. All patients were collected from Oct. 2015 to Aug. 2016 in the Fourth Affiliated Hospital of China Medical University. Differences of these three devices (ACD and CCT) were statistically analyzed by randomized blocks analysis. Pearson correlative analysis and Bland-Altman analysis were used to compare the correlation and agreement. Results There’s no significant difference in measuring ACD by Lenstar and UBM (P>0.05), but there were significant differences in other two groups (AS-OCT and UBM, AS-OCT and Lenstar) (P<0.05). There were significant differences in the mean CCT measured by the three devices (P<0.05). There were high correlation in measuring ACD (all P<0.001, r1=0.928,r2=0.982,r3=0.932, respectively), and CCT(all P<0.001,r1=0.957,r2=0.949,r3=0.928, respectively). Conclusion AS-OCT, Lenstar and UBM yielded comparable results and are all can be used interchangeably for anterior segment measurements.


2021 ◽  
Vol 11 (1) ◽  
Author(s):  
Tommy C. Y. Chan ◽  
Marco C. Y. Yu ◽  
Vivian Chiu ◽  
Gilda Lai ◽  
Christopher K. S. Leung ◽  
...  

AbstractTo evaluate the repeatability and agreement of corneal and biometry measurements obtained with two swept-source optical coherence tomography (SSOCT) and a partial coherence interferometry-based device. This is a cross-sectional study. Forty-eight eyes of 48 patients had three consecutive measurements for ANTERION (Heidelberg Engineering, Germany), CASIAII (Tomey, Japan) and IOLMaster500 (Carl Zeiss Meditec, USA) on the same visit. Mean keratometry (Km), central corneal thickness (CCT), anterior chamber depth (ACD) and axial length (AL) were recorded. Corneal astigmatic measurements were converted into vector components—J0 and J45. Intra-device repeatability and agreements of measurements amongst the devices were evaluated using repeatability coefficients (RCs) and Bland–Altman plots, respectively. All devices demonstrated comparable repeatability for Km (p ≥ 0.138). ANTERION had the lowest RC for J0 amongst the devices (p ≤ 0.039). Systematic difference was found for the Km and J0 obtained with IOLMaster500 compared to either SSOCTs (p ≤ 0.010). The ACD and AL measured by IOLMaster500 showed a higher RC compared with either SSOCTs (p < 0.002). Systematic difference was found in CCT and ACD between the two SSOCTs (p < 0.001), and in AL between ANTERION and IOLMaster500 (p < 0.001), with a mean difference of 1.6 µm, 0.022 mm and 0.021 mm, respectively. Both SSOCTs demonstrated smaller test–retest variability for measuring ACD and AL compared with IOLMaster500. There were significant disagreement in keratometry and AL measurements between the SSOCTs and PCI-based device; their measurements should not be considered as interchangeable.


2018 ◽  
Vol 1 ◽  
pp. 3
Author(s):  
Joshua S Agranat ◽  
Yoshihiro Yonekawa

Iris pigment epithelial (IPE) cysts are a subset of iris cysts that arise from the IPE. They are spontaneously erupting epithelial-lined cavities that are found in various anatomic locations of the iris, including the iris pupillary margin, midzone, periphery, and free floating in the vitreous or anterior chamber. We report the case of an asymptomatic 13-year-old boy with an incidental finding of a dislodged anterior chamber cyst diagnosed on routine examination. Modern multimodal image analysis of the cyst including anterior segment optical coherence tomography and ultrasound biomicroscopy (UBM) was utilized to characterize the microstructural anatomy of the lesion. The patient was managed conservatively without complications. Cysts of the IPE typically do not affect vision or ocular health and can be monitored and observed after ascertaining no associated malignancy. Initial diagnostic investigation can include UBM and anterior segment optical coherence tomography. Intervention should be reserved only for cases where the cyst growth leads to obstruction of the visual axis and/or other secondary complications.


Author(s):  
Alireza Khodabande ◽  
Massood Mohammadi ◽  
Hamid Riazi-Esfahani ◽  
Shahab Karami ◽  
Massood Mirghorbani ◽  
...  

Abstract Background To evaluate changes in anterior segment morphology on anterior segment optical coherence tomography (AS-OCT) following pars plana vitrectomy (PPV) without tamponade. Methods Patients who underwent PPV without tamponade for epiretinal membrane were evaluated. Eligible patients underwent intraocular pressure (IOP) measurement and AS-OCT preoperatively as well as 1 month and 6 months post-operatively. Anterior chamber width (ACW), anterior chamber depth (ACD), trabecular iris angle (TIA), angle opening distance at 500 and 750 µm (AOD), and trabecular iris space area at 500 and 750 µm (TISA) at four quadrants were recorded and analyzed. Additionally, the mean values of TIA (MTIA), AOD (MAOD), and TISA (MTISA) for each eye (mean of four quadrants) were analyzed. Results 23 patients completed the study. The mean age of participants was 56.4 ± 3.6 years of age and 13/23 (56%) were female. Mean IOP of patients was 18.1 ± 1.1, 18.3 ± 1.1, and 18.1 ± 1.2 preoperatively,1 month post-operatively, and 6 months post-operatively, respectively. (p = 0.83). No difference was detected post-operatively in measurements of ACW, ACD, MTIA, MAOD500, MAOD750, MTISA500, and MTISA750. Conclusion Pars plana vitrectomy without tamponade was not associated with changes in anterior chamber morphology.