Equipment-independent reference values for dopamine transporter imaging with 123I-FP-CIT
Summary Aim: Reliable reference values are helpful to interpret and compare the results of dopamine transporter imaging with SPECT. Since semi-quantitative reference values cannot be easily transferred between imaging equipments, this study aimed to establish equipment independent normal values for the true striatal binding of 123I-FP-CIT. Patients, methods: Specific striatal FP-CIT binding of 6 healthy volunteers and 26 patients with essential tremor were used to generate a reference range by applying an equipment specific resolution dependent factor to compensate for recovery effects. This factor has been determined previously by a series of standardized phantom measurements of an anthropomorphic basal ganglia phantom. Herewith, the resulting DAT binding values represent the expected true specific binding in the striatum. Results: On average, true specific striatal binding was 5.83 ± 0.96 in healthy controls, 5.25 ± 0.67 in patients with essential tremor and 5.36 ± 0.75 in the entire study cohort. Conclusion: These preliminary results may serve as a basis for the generation of a generally accepted equipment independent reference range for dopamine transporter imaging with 123I-FP-CIT. By a simple phantom measurement that can be accomplished within one day factors related to specific imaging equipment and processing can be corrected for, resulting in specific binding values which may enable a more standardized interpretation of dopamine transporter scans.