scholarly journals Biomechanical Evaluation of Unilateral Versus Bilateral C1 Lateral Mass-C2 Intralaminar Fixation

2017 ◽  
Vol 7 (3) ◽  
pp. 239-245 ◽  
Author(s):  
Nitin Bhatia ◽  
Asheen Rama ◽  
Brandon Sievers ◽  
Ryan Quigley ◽  
Michelle H. McGarry ◽  
...  

Study Design: Biomechanical, cadaveric study. Objectives: To compare the relative stiffness of unilateral C1 lateral mass-C2 intralaminar fixation to intact specimens and bilateral C1 lateral mass-C2 intralaminar constructs. Methods: The biomechanical integrity of a unilateral C1 lateral mass-C2 intralaminar screw construct was compared to intact specimens and bilateral C1 lateral mass-C2 intralaminar screw constructs. Five human cadaveric specimens were used. Range of motion and stiffness were tested to determine the stiffness of the constructs. Results: Unilateral fixation significantly decreased flexion/extension range of motion compared to intact ( P < .001) but did not significantly affect axial rotation ( P = .3) or bending range of motion ( P = .3). There was a significant decrease in stiffness in extension for both unilateral and bilateral fixation techniques compared to intact ( P = .04 and P = .03, respectively). There was also a significant decrease in stiffness for ipsilateral rotation for the unilateral construct compared to intact ( P = .007) whereas the bilateral construct significantly increased ipsilateral rotation stiffness compared to both intact and unilateral fixation ( P < .001). Conclusion: Bilateral constructs did show improved biomechanical properties compared to the unilateral constructs. However, unilateral C1-C2 fixation using a C1 lateral mass and C2 intralaminar screw-rod construct decreased range of motion and improved stiffness compared to the intact state with the exception of extension and ipsilateral rotation. Hence, a unilateral construct may be acceptable in clinical situations in which bilateral fixation is not possible, but an external orthosis may be necessary to achieve a fusion.

2008 ◽  
Vol 9 (3) ◽  
pp. 296-300 ◽  
Author(s):  
Michael A. Finn ◽  
Daniel R. Fassett ◽  
Todd D. Mccall ◽  
Randy Clark ◽  
Andrew T. Dailey ◽  
...  

Object Stabilization with rigid screw/rod fixation is the treatment of choice for craniocervical disorders requiring operative stabilization. The authors compare the relative immediate stiffness for occipital plate fixation in concordance with transarticular screw fixation (TASF), C-1 lateral mass and C-2 pars screw (C1L-C2P), and C-1 lateral mass and C-2 laminar screw (C1L-C2L) constructs, with and without a cross-link. Methods Ten intact human cadaveric spines (Oc–C4) were prepared and mounted in a 7-axis spine simulator. Each specimen was precycled and then tested in the intact state for flexion/extension, lateral bending, and axial rotation. Motion was tracked using the OptoTRAK 3D tracking system. The specimens were then destabilized and instrumented with an occipital plate and TASF. The spine was tested with and without the addition of a cross-link. The C1L-C2P and C1L-C2L constructs were similarly tested. Results All constructs demonstrated a significant increase in stiffness after instrumentation. The C1L-C2P construct was equivalent to the TASF in all moments. The C1L-C2L was significantly weaker than the C1L-C2P construct in all moments and significantly weaker than the TASF in lateral bending. The addition of a cross-link made no difference in the stiffness of any construct. Conclusions All constructs provide significant immediate stability in the destabilized occipitocervical junction. Although the C1L-C2P construct performed best overall, the TASF was similar, and either one can be recommended. Decreased stiffness of the C1L-C2L construct might affect the success of clinical fusion. This construct should be reserved for cases in which anatomy precludes the use of the other two.


2017 ◽  
Vol 30 (06) ◽  
pp. 430-437 ◽  
Author(s):  
Nadja Bösch ◽  
Martin Hofstetter ◽  
Alexander Bürki ◽  
Beatriz Vidondo ◽  
Fenella Davies ◽  
...  

Abstract Objective To study the biomechanical effect of facetectomy in 10 large breed dogs (>24 kg body weight) on the fourth canine cervical functional spinal unit. Methods Canine cervical spines were freed from all muscles. Spines were mounted on a six-degrees-of-freedom spine testing machine for three-dimensional motion analysis. Data were recorded with an optoelectronic motion analysis system. The range of motion wasdetermined inall threeprimary motionsaswellasrange of motion of coupled motions on the intact specimen, after unilateral and after bilateral facetectomy. Repeated-measures analysis of variance models were used to assess the changes of the biomechanical properties in the three treatment groups considered. Results Facetectomy increased range of motion of primary motions in all directions. Axial rotation was significantly influenced by facetectomy. Coupled motion was not influenced by facetectomy except for lateral bending with coupled motion axial rotation. The coupling factor (coupled motion/primary motion) decreased after facetectomy. Symmetry of motion was influenced by facetectomy in flexion–extension and axial rotation, but not in lateral bending. Clinical Significance Facet joints play a significant role in the stability of the cervical spine and act to maintain spatial integrity. Therefore, cervical spinal treatments requiring a facetectomy should be carefully planned and if an excessive increase in range of motion is expected, complications should be anticipated and reduced via spinal stabilization.


Neurosurgery ◽  
2013 ◽  
Vol 72 (6) ◽  
pp. 1021-1030 ◽  
Author(s):  
Luis Perez-Orribo ◽  
Andrew S. Little ◽  
Richard D. Lefevre ◽  
Phillip R. Reyes ◽  
Anna G.U.S. Newcomb ◽  
...  

Abstract BACKGROUND: Endoscopic endonasal approaches to the craniovertebral junction and clivus, which are increasingly performed for ventral skull base pathology, may require disruption of the occipitocondylar joint. OBJECTIVE: To study the biomechanical implications at the craniovertebral junction of progressive unilateral condylectomy as would be performed through an endonasal exposure. METHODS: Seven upper cervical human cadaveric specimens (C0-C2) underwent nondestructive biomechanical flexibility testing during flexion-extension, axial rotation, and lateral bending at C0-C1 and C1-C2. Each specimen was tested intact, after an inferior one-third clivectomy, and after stepwise unilateral condylectomy with an anterior approach. Angular range of motion (ROM), lax zone, and stiff zone were determined and compared with the intact state. RESULTS: At C0-C1, mobility during flexion-extension and axial rotation increased significantly with progressive condylectomy. ROM increased from 14.3 ± 2.7° to 20.4 ± 5.2° during flexion and from 6.7 ± 3.5° to 10.8 ± 3.0° during right axial rotation after 75% condyle resection (P &lt; .01). At C1-C2, condylectomy had less effect, with ROM increasing from 10.7 ± 2.0° to 11.7 ± 2.0° during flexion, 36.9 ± 4.8° to 37.1 ± 5.1° during right axial rotation, and 4.3 ± 1.9° to 4.8 ± 3.3° during right lateral bending (P = NS). Because of marked instability, the 100% condylectomy condition was untestable. Changes in ROM were a result of changes more in the lax zone than in the stiff zone. CONCLUSION: Lower-third clivectomy and unilateral anterior condylectomy as would be performed in an endonasal approach cause progressive hypermobility at the craniovertebral junction. On the basis of biomechanical criteria, craniocervical fusion is indicated for patients who undergo &gt; 75% anterior condylectomy.


2008 ◽  
Vol 8 (2) ◽  
pp. 143-152 ◽  
Author(s):  
Nicholas C. Bambakidis ◽  
Iman Feiz-Erfan ◽  
Eric M. Horn ◽  
L. Fernando Gonzalez ◽  
Seungwon Baek ◽  
...  

Object The stability provided by 3 occipitoatlantal fixation techniques (occiput [Oc]–C1 transarticular screws, occipital keel screws rigidly interconnected with C-1 lateral mass screws, and suboccipital/sublaminar wired contoured rod) were compared. Methods Seven human cadaveric specimens received transarticular screws and 7 received occipital keel–C1 lateral mass screws. All specimens later underwent contoured rod fixation. All conditions were studied with and without placement of a structural graft wired between the skull base and C-1 lamina. Specimens were loaded quasistatically using pure moments to induce flexion, extension, lateral bending, and axial rotation while recording segmental motion optoelectronically. Flexibility was measured immediately postoperatively and after 10,000 cycles of fatigue. Results Application of Oc–C1 transarticular screws, with a wired graft, reduced the mean range of motion (ROM) to 3% of normal. Occipital keel–C1 lateral mass screws (also with graft) offered less stability than transarticular screws during extension and lateral bending (p < 0.02), reducing ROM to 17% of normal. The wired contoured rod reduced motion to 31% of normal, providing significantly less stability than either screw fixation technique. Fatigue increased motion in constructs fitted with transarticular screws, keel screws/lateral mass screw constructs, and contoured wired rods, by means of 19, 5, and 26%, respectively. In all constructs, adding a structural graft significantly improved stability, but the extent depended on the loading direction. Conclusions Assuming the presence of mild C1–2 instability, Oc–C1 transarticular screws and occipital keel–C1 lateral mass screws are approximately equivalent in performance for occipitoatlantal stabilization in promoting fusion. A posteriorly wired contoured rod is less likely to provide a good fusion environment because of less stabilizing potential and a greater likelihood of loosening with fatigue.


2019 ◽  
Vol 9 (7) ◽  
pp. 697-707 ◽  
Author(s):  
Paul M. Arnold ◽  
Ivan Cheng ◽  
Jonathan A. Harris ◽  
Mir M. Hussain ◽  
Chengmin Zhang ◽  
...  

Study Design: In vitro cadaveric biomechanical study. Objective: To compare the biomechanics of integrated anchor and blade versus traditional screw fixation techniques for interbody fusion. Methods: Fifteen cadaveric cervical spines were divided into 3 equal groups (n = 5). Each spine was tested: intact, after discectomy (simulating an injury model), interbody spacer alone (S), integrated interbody spacer (iSA), and integrated spacer with lateral mass screw and rod fixation (LMS+iS). Each treatment group included integrated spacers with either screw, anchor, or blade integrated spacers. Constructs were tested in flexion-extension (FE), lateral bending (LB), and axial rotation (AR) under pure moments (±1.5 N m). Results: Across all 3 planes, the following range of motion trend was observed: Injured > Intact > S > iSA > LMS+iS. In FE and LB, integrated anchor and blade significantly decreased motion compared with intact and injured conditions, before and after supplemental posterior fixation ( P < .05). Comparing tested devices revealed biomechanical equivalence between screw, anchor, and blade fixation methods in all loading modes ( P > .05). Conclusion: All integrated interbody devices reduced intact and injured motion; lateral mass screws and rods further stabilized the single motion segment. Comparing screw, anchor, or bladed integrated anterior cervical discectomy and fusion spacers revealed no significant differences.


2006 ◽  
Vol 4 (2) ◽  
pp. 160-164 ◽  
Author(s):  
Jeremy C. Wang ◽  
David Spenciner ◽  
James C. Robinson

Object The authors studied the biomechanical properties of a novel spinous process stabilization plate (CD HORIZON SPIRE Spinal System) and present the results in comparison with those of other posterior fixation methods. Methods Ten functional cadaveric lumbar segments were subjected to nondestructive quasistatic loading forces in 10 different conditions: intact, destabilized (discectomy), fitted with spinous process plate (SPP) alone, with anterior-column support (ACS) alone, ACS with SPP, ACS with posterior translaminar facet screw (PTFS) fixation, ACS with unilateral pedicle screw and rod (UPSR) fixation, ACS with bilateral pedicle screw and rod (BPSR) fixation, UPSR alone, or BPSR alone. Stiffness and range of motion (ROM) data were compared using a repeated-measures, one-way analysis of variance. The construct with greatest mean limitation of flexion–extension ROM was ACS/SPP at 4.14° whereas it was 5.75° for ACS/UPSR fixation, 5.03° for ACS/BPSR fixation, and 10.13° for the intact spine. The SPIRE plate alone also provided greater flexion and extension stiffness, with less ROM than other posterior stabilization options. Fixation with BPSR with or without ACS resulted in the stiffest construct in lateral bending and axial rotation. The SPP and UPSR fixation groups were equivalent in resisting lateral bending and axial rotation forces with or without ACS. Conclusions The SPIRE plate effectively stabilized the spine, and the test results compare favorably with other fixation techniques that are more time consuming to perform and have greater inherent risks.


Author(s):  
Timothy L Lasswell ◽  
John B Medley ◽  
Jack P Callaghan ◽  
Duane S Cronin ◽  
Colin D McKinnon ◽  
...  

The aim of this experimental study was to assess the biomechanical performance of a novel C1 posterior arch (C1PA) clamp compared with C1 lateral mass (C1LM) screws in constructs used to treat atlantoaxial instability. These constructs had either C2 pedicle (C2P) screws or C2 translaminar (C2TL) screws. Eight fresh-frozen human cadaveric ligamentous spine specimens (C0-C3) were tested under six conditions: the intact state, the destabilized state after a simulated odontoid fracture, and when instrumented with four constructs (C1LM-C2P, C1LM-C2TL, C1PA-C2P, C1PA-C2TL). Each specimen was tested in a spinal loading simulator that separately applied axial rotation, flexion-extension and lateral bending. In each test condition, displacement controlled angular motion was applied in both directions at a speed of 2 deg/s until a resulting moment of 1.5 Nm was achieved. The measured ranges of motion (ROM) of the C1-C2 segments were compared for each test condition using nonparametric Friedman tests. The destabilized state had significantly more C1-C2 motion ( p < 0.05) than the intact state in all cases, and all constructs greatly reduced this motion. C2 pedicle screw constructs that used the C1PA clamp had significantly less C1-C2 motion ( p < 0.05) than those with C1LM screws in flexion-extension as well as axial rotation and no statistically significant difference was detected in lateral bending. C2 translaminar screw constructs that used the C1PA clamp had significantly less C1-C2 motion ( p < 0.05) than those with C1LM screws in flexion-extension and no statistically significant difference was detected in axial rotation or in lateral bending. Data from the current study suggested that constructs using the novel C1PA clamp would provide as good, or improved, biomechanical stability to the C1-C2 segment compared with constructs using C1LM screws.


2010 ◽  
Vol 13 (2) ◽  
pp. 253-259 ◽  
Author(s):  
Fatih Ersay Deniz ◽  
Leonardo B. C. Brasiliense ◽  
Bruno C. R. Lazaro ◽  
Phillip M. Reyes ◽  
Anna G. U. Sawa ◽  
...  

Object The authors investigated the biomechanical properties of transpedicular discectomy in the thoracic spine and compared the effects on spinal stability of a partial and total facetectomy. Methods Human thoracic specimens were tested while intact, after a transpedicular discectomy with partial facetectomy, and after an additional total facetectomy was incorporated. Nonconstraining pure moments were applied under load control (maximum 7.5 Nm) to induce flexion, extension, lateral bending, and axial rotation while spinal motion was measured at T8–9 optoelectronically. The range of motion (ROM) and lax zone were determined in each specimen and compared among conditions. Results Transpedicular discectomy with and without a total facetectomy significantly increased the ROM and lax zone in all directions of loading compared with the intact spine (p < 0.008). The segmental increase in ROM observed with the transpedicular discectomy was 25%. The additional total facetectomy created an insignificant 3% further increase in ROM compared with medial facetectomy (p > 0.2). Conclusions Transpedicular discectomy can be performed in the thoracic spine with a modest decrease in stability expected. Because the biomechanical behavior of a total facetectomy is equivalent to that of a medial facetectomy, the additional facet removal may be incorporated without further biomechanical consequences.


Author(s):  
Héctor E Jaramillo S

The annulus fibrosus has substantial variations in its geometrical properties (among individuals and between levels), and plays an important role in the biomechanics of the spine. Few works have studied the influence of the geometrical properties including annulus area, anterior / posterior disc height, and over the range of motion, but in general these properties have not been reported in the finite element models. This paper presents a probabilistic finite element analyses (Abaqus 6.14.2) intended to assess the effects of the average disc height ( hp) and the area ( A) of the annulus fibrosus on the biomechanics of the lumbar spine. The annulus model was loaded under flexion, extension, lateral bending, and axial rotation and analyzed for different combinations of hpand A in order to obtain their effects over the range of motion. A set of 50 combinations of hp(mean = 18.1 mm, SD = 3.5 mm) and A (mean = 49.8%, SD = 4.6%) were determined randomly according to a normal distribution. A Yeoh energy function was used for the matrix and an exponential function for the fibers. The range of motion was more sensitive to hpthan to A. With regard to the range of motion the segment was more sensitive in the following order: flexion, axial rotation, extension, and lateral bending. An increase of the hpproduces an increase of the range of motion, but this decreases when A increases. Comparing the range of motion with the experimental data, on average, 56.0% and 73.0% of the total of data were within the experimental range for the L4–L5 and L5–S1 segments, respectively. Further, an analytic equation was derived to obtain the range of motion as a function of the hpand A. This equation can be used to calibrate a finite element model of the spine segment, and also to understand the influence of each geometrical parameter on the range of motion.


2019 ◽  
Vol 10 (8) ◽  
pp. 973-981
Author(s):  
Raymond J. Hah ◽  
Ram Alluri ◽  
Paul A. Anderson

Study Design: Biomechanics study. Objectives: To evaluate the biomechanical advantage of interfacet allograft spacers in an unstable single-level and 2-level anterior cervical discectomy and fusion (ACDF) pseudoarthrosis model. Methods: Nine single-level and 8 two-level ACDF constructs were tested. Range of motion in flexion-extension (FE), lateral bending (LB), and axial rotation (AR) at 1.5 N m were collected in 4 testing configurations: (1) intact spine, (2) ACDF with interbody graft and plate/screw, (3) ACDF with interbody graft and plate/loosened screws (loose condition), and (4) ACDF with interbody graft and plate/loosened screws supplemented with interfacet allograft spacers (rescue condition). Results: All fixation configurations resulted in statistically significant decreases in range of motion in all bending planes compared with the intact spine ( P < .05). One Level. Performing ACDF with interbody graft and plate on the intact spine reduced FE, LB, and AR 60.0%, 64.9%, and 72.9%, respectively. Loosening the ACDF screws decreased these reductions to 40.9%, 44.6%, and 52.1%. The addition of interfacet allograft spacers to the loose condition increased these reductions to 74.0%, 84.1%, and 82.1%. Two Level. Performing ACDF with interbody graft and plate on the intact spine reduced FE, LB, and AR 72.0%, 71.1%, and 71.2%, respectively. Loosening the ACDF screws decreased these reductions to 55.4%, 55.3%, and 51.3%. The addition of interfacet allograft spacers to the loose condition significantly increased these reductions to 82.6%, 91.2%, and 89.3% ( P < .05). Conclusions: Supplementation of a loose ACDF construct (pseudarthrosis model) with interfacet allograft spacers significantly increases stability and has potential applications in treating cervical pseudarthrosis.


Sign in / Sign up