scholarly journals Fracture strength analysis of titanium insert-reinforced zirconia abutments according to the axial height of the titanium insert with an internal connection

PLoS ONE ◽  
2021 ◽  
Vol 16 (4) ◽  
pp. e0249208
Author(s):  
Seung-rye Song ◽  
Kyeong-Mee Park ◽  
Bock-Young Jung

This study aimed to analyze fracture strength in vitro by varying the axial height of the titanium insert and the labial height of the zirconia abutment in an internal connection implant to identify the titanium insert axial height with optimal mechanical stability. Sixty implants with an internal connection system were used. Two-piece zirconia abutments were used with the titanium inserts. Combinations of different titanium insert axial heights (mm) and zirconia abutment labial heights (mm) constituted five groups: Gr1 (1–3), Gr2 (3–3), Gr3 (3–5), Gr4 (5–3), and Gr5 (5–5). After thermocycling, a fracture load test was performed with a universal testing machine. The initial deformation load and the fracture load were measured and analyzed. The fractured surface and cross-section of the specimens were examined by scanning electron microscopy (SEM). The groups of titanium inserts with axial heights of 3 mm and 5 mm showed significantly greater initial deformation load and fracture load than the group with an axial height of 1 mm (p < 0.05), but there was no significant difference between the two groups with axial heights of 3 mm and 5 mm. The labial height of the zirconia abutment had no significant influence on the initial deformation load and fracture load. In some specimens in Gr4 and Gr5, cracking or bending of the titanium insert and abutment screw was observed on SEM. The axial height of the titanium insert should be designed to not be less than 3 mm to increase the fracture strength and promote the long-term stability of implants.

2018 ◽  
Vol 43 (2) ◽  
pp. 162-169 ◽  
Author(s):  
ME Hshad ◽  
EE Dalkılıç ◽  
GC Ozturk ◽  
I Dogruer ◽  
F Koray

SUMMARY Objective: The purpose of this study was to determine the fracture strength of endodontically treated mandibular premolar teeth restored with composites and different reinforcement techniques. Methods and Materials: Forty-eight freshly extracted human mandibular premolar teeth were randomly divided into four groups: group IN, group CR, group FRC, and group PRF. Group IN consisted of teeth with intact crowns and served as the control group. In the other three groups, endodontic treatment was performed and standard mesio-occluso-distal (MOD) cavities were prepared. Then cavities were restored with hybrid resin composite only, flowable composite and hybrid resin composite, and Ribbond, flowable composite and hybrid resin composite in groups CR, FRC and PRF, respectively. All of the teeth were subjected to fracture by means of a universal testing machine, and compressive force was applied with a modified stainless-steel ball at a crosshead speed at 0.5 mm/min. Results: The highest values were observed in group IN, while the lowest values were determined in group CR. There was not any statistically significant difference between group CR and group FCR (p&gt;0.05). When groups CR, FCR, and PRF were compared, group PRF showed significantly better fracture strength than did groups CR and FCR (p&lt;0.05). It was determined that there was not any significant difference between group IN and group PRF (p&gt;0.05). Conclusions: Polyethylene ribbon fiber considerably increases the fracture strength of mandibular premolar teeth with MOD cavities restored with composite.


2007 ◽  
Vol 18 (2) ◽  
pp. 134-138 ◽  
Author(s):  
José Mondelli ◽  
Fábio Sene ◽  
Renata Pereira Ramos ◽  
Ana Raquel Benetti

This study evaluated, in vitro, the loss of tooth substance after cavity preparation for direct and indirect restorations and its relationship with fracture strength of the prepared teeth. Sixty sound human maxillary first premolars were assigned to 6 groups (n=10). MOD direct composite cavities (Groups I, II and III) and indirect inlay cavities (Groups IV, V and VI) were prepared maintaining standardized dimensions: 2-mm deep pulpal floors, 1.5-mm wide gingival walls and 2-mm high axial walls. Buccolingual width of the occlusal box was established at 1/4 (Groups I and IV), 1/3 (Groups II and V) or 1/2 (Groups III and VI) of the intercuspal distance. Teeth were weighed (digital balance accurate to 0.001 g) before and after preparation to record tooth substance mass lost during cavity preparation. The prepared teeth were submitted to occlusal loading to determine their fracture strength using a universal testing machine at a crosshead speed of 0.5 mm/min. Data were analyzed by two-way ANOVA and Tukey test (alpha= 0.05). 1/4-inlay cavities had higher percent mean mass loss (9.71%) than composite resin cavities with the same width (7.07%). 1/3-inlay preparations also produced higher percent mean mass loss (13.91%) than composite resin preparations with the same width (10.02%). 1/2-inlay cavities had 21.34% of mass loss versus 16.19% for the 1/2-composite resin cavities. Fracture strength means (in kgf) were: GI = 187.65; GII = 143.62; GIII = 74.10; GIV = 164.22; GV = 101.92; GVI = 50.35. Statistically significant difference (p<0.05) were observed between Groups I and IV, II and V, III and VI. Higher tooth structure loss and lower fracture strength were recorded after preparation of inlay cavities, regardless of the width of the occlusal box, compared to the direct composite resin cavities.


2021 ◽  
Vol 45 (3) ◽  
pp. 171-176
Author(s):  
Brent Lin ◽  
Amit Khatri ◽  
Michael Hong

The purpose of this study was to determine and compare the shear force (N) required to fracture or dislodge an all-ceramic zirconia-based crown using different luting cement with those of polycarbonate crown and strip crown for the primary anterior teeth in vitro. Study design: Four groups of esthetic restoration for primary anterior teeth were tested for fracture strength: 1) Fifteen all-ceramic zirconia-based crowns cemented with glass ionomer cement, 2) Fifteen all-ceramic zirconia-based crowns bonded with a self-adhesive resin cement, 3) Fifteen polycarbonate crowns cemented with a polymer reinforced zinc-oxide eugenol and 4) Fifteen resin strip crowns. All restorations were placed and cemented on reproductions of dies in an independent laboratory at Delhi, India. All samples underwent loading until fracture or dislodgement with the Universal Testing Machine. The force in Newton (N) required to produce failure was recorded for each sample and the type of failures was also noted and characterized. One-way analysis of variance (ANOVA) test and the Tukey and Scheffe’s post hoc comparisons were used for statistical analyses. Results: In this invitro study, results were measured in Newtons (N). Group 1 (410.9±79.5 N) and Group 2 (420.5±57.8 N) had higher fracture strength than Group 3 (330.3±85.6 N) and Group 4 (268.4±28.2 N). These differences were statistically significant at P≤.05 among the sample groups. No significant difference was found between groups 1 and 2 (P = 0.984) nor between groups 3 and 4 (P =0.104). Among type of failures, majority of restoration fractures for zirconia-based crowns and resin strip crowns were due to cohesive failures and polycarbonate crowns had predominantly mixed failures. Conclusions: Under the limitations of this in vitro study, it could be concluded that all-ceramic zirconia-based crowns attained the highest fracture strength among all restorative samples tested regardless of the type of luting agent employed (P&lt;.01). Cohesive failures were commonly observed in the zirconia crowns and resin strip crowns, whereas polycarbonate crowns revealed predominately mixed failures.


Author(s):  
Marwah Ismael Abdulazeez ◽  
Manhal A. Majeed

Abstract Objective The aim of this study was to evaluate the influence of different marginal designs (deep chamfer, vertical, and modified vertical with reverse shoulder) on the fracture strength and failure modes of monolithic zirconia crowns. Materials and Methods Thirty sound human maxillary first premolar teeth with comparable size were used in this study. The teeth were divided randomly into three groups according to the preparation design (n = 10): (1) group A: teeth prepared with a deep chamfer finish line; (2) group B: teeth prepared with vertical preparation; and (3) group C: teeth prepared with modified vertical preparation, where a reverse shoulder of 1 mm was placed on the buccal surface at the junction of middle and occlusal thirds. All samples were scanned by using an intraoral scanner (CEREC Omnicam, Sirona, Germany), and then the crowns were designed by using Sirona InLab 20.0 software and milled with a 5-axis machine. Each crown was then cemented on its respective tooth with self-adhesive resin cement by using a custom-made cementation device. A single load to failure test was used to assess the fracture load of each crown by using a computerized universal testing machine that automatically recorded the fracture load of each sample in Newton (N). Statistical Analysis The data were analyzed statistically by using one-way analysis of variance test and Bonferroni test at a level of significance of 0.05. Results The highest mean of fracture load was recorded by chamfer (2,969.8 N), which followed by modified vertical (2,899.3 N) and the lowest mean of fracture load was recorded by vertical (2,717.9 N). One-way ANOVA test revealed a significant difference among the three groups. Bonferroni test showed a significant difference between group A and group B, while a nonsignificant difference was revealed between group C with group A and group B. Conclusion Within the limitations of this in vitro study, the mean values of fracture strength of monolithic zirconia crowns of all groups were higher than the maximum occlusal forces in the premolar region. The modification of the vertical preparation with a reverse shoulder placed at the buccal surface improved the fracture strength up to the point that it was statistically nonsignificant with the chamfer group.


2012 ◽  
Vol 13 (2) ◽  
pp. 210-215
Author(s):  
Firas Al Quran ◽  
Reem Haj-Ali

ABSTRACT Purpose The purpose of this study was to investigate the fracture loads and mode of failure of all-ceramic crowns fabricated using Top-Ceram and compare it with all-ceramic crowns fabricated from well-established systems: IPS-Empress II, In-Ceram. Materials and methods Thirty all-ceramic crowns were fabricated; 10 IPS-Empress II, 10 In-Ceram alumina and 10 Top-Ceram. Instron testing machine was used to measure the loads required to introduce fracture of each crown. Results Mean fracture load for In-Ceram alumina [941.8 (± 221.66) N] was significantly (p < 0.05) higher than those of Top-Ceram and IPS-Empress II. There was no statistically significant difference between Top-Ceram and IPS-Empress II mean fracture loads; 696.20 (+222.20) and 534 (+110.84) N respectively. Core fracture pattern was highest seen in Top- Ceram specimens. How to cite this article Quran AF, Haj-Ali R. Fracture Strength of Three All-Ceramic Systems: Top-Ceram Compared with IPSEmpress and In-Ceram. J Contemp Dent Pract 2012;13(2): 210-215.


2021 ◽  
Vol 15 (1) ◽  
pp. 664-671
Author(s):  
Noor Nawafleh ◽  
Shareen Elshiyab ◽  
Andreas Öchsner ◽  
Roy George

Objective: This study evaluated the performance of zirconia and lithium disilicate crowns supported by implants or cemented to epoxy resin dies. Methods: Eigthy zirconia and lithium disilicate crowns each were prepared and assigned in four groups according to the crown material and supporting structure combinations (implant-supported zirconia, die-supported zirconia, implant-supported lithium disilicate, and die-supported lithium disilicate). Ten crowns in each group acted as control while the rest (n=10) underwent thermocycling and fatigue with 100 N loading force for 1.5 million cycles. Specimens were then loaded to fracture in a universal testing machine. Data were analysed using one-way ANOVA and Tukey multiple comparison test with a 95% level of significance. Results: No implants or crown failure occurred during fatigue. The mean fracture load values (control, fatigued) in newton were as follows: (4054, 3344) for implant-supported zirconia, (3783, 3477) for die-supported zirconia, (2506, 2207) for implant-supported lithium disilicate, and (2159, 1806) for die-supported lithium disilicate. Comparing the control with the corresponding fatigued subgroup showed a significantly higher fracture load mean of the control group in all cases. Zirconia showed a significantly higher fracture load mean than lithium disilicate (P=0.001, P<0.001). However, comparing crowns made from the same material according to the supporting structure showed no significant difference (P=0.923, P=0.337). Conclusion: Zirconia and lithium disilicate posterior crowns have adequate fatigue and fracture resistance required for posterior crowns. However, when heavy fatigue forces are expected, zirconia material is preferable over lithium disilicate. Zirconia and lithium disilicate implant-supported crowns cemented to hybrid abutments should have satisfactory clinical performance.


2007 ◽  
Vol 8 (1) ◽  
pp. 44-51 ◽  
Author(s):  
Asude Yilmaz ◽  
Seyfettin Baydaş

Abstract Aim The aim of this study was to evaluate the fracture resistance of various provisional crown materials using an in vitro model test system. Methods and Materials In the present study polycarbonate crowns, prefabricated by the manufacturer (3M Polycarbonate Crown), and the temporary crowns, fabricated in the dental laboratory environment, were fabricated using bis-acryl composite (Protemp II), autopolymerizing PMMA resin (BISICO Temp S), and heatpolymerized PMMA resin (Major C&B-V Dentine). All temporary crowns were stored in distilled water for 24 hours at room temperature prior to testing. The crowns were seated on metal dies, fabricated from Cr-Co alloy (AZ Dental, Konstanz, Germany), and then tested using the indenter of a Hounsfield testing machine (Hounsfield Tensometer, Hounsfield Test Equipment, Raydon, England). The tip of the indenter was located at a position one-third of the way down the inciso-palatine surface at 135°. The data were statistically analyzed for differences using one-way analysis of variance (ANOVA) and the Tukey HSD test (P<.05). Additionally, the types of failure obtained from the fracture load test were examined using 10X magnification with a stereo microscope. Results The results of the present study indicated polycarbonate crowns were significantly different from the BISICO Temp S, Protemp II, and Major C&B-V Dentine (P<.05) groups. Conclusion This in vitro study shows polycarbonate crowns may be preferable to the other types of temporary crowns used in this study. Citation Yilmaz A, Baydaş S. Fracture Resistance of Various Temporary Crown Materials. J Contemp Dent Pract 2007 January;(8)1:044-051.


F1000Research ◽  
2019 ◽  
Vol 8 ◽  
pp. 1491 ◽  
Author(s):  
Bushra Mohammed ◽  
Jylan EL-Guindy

Background: Cerasmart hybrid material offers specific advantages such as less fragility and more flexibility than glass ceramics. This material also has the option of readily modifying or repairing the surface and favorable stress-absorbing characteristics. In our study, Cerasmart hybrid and lithium disilicate ceramic laminate veneers with two different preparation designs were compared with regards to their fracture resistance. Methods: A total of 52 of comparable human central maxillary incisors were used. Group A (n=26) was made up of Cerasmart hybrid ceramic laminate veneers were fabricated from Cerasmart blocks, while Group B (n=26) was made up of lithium disilicate ceramic laminate veneers were made of IPS e.max pressable ingots. Each group was subdivided in two equal subgroups according to preparation designs. Subgroup I comprised Featheredge preparation design and subgroup II: Wraparound preparation design. All samples were subjected to thermocycling between 5°C and 55°C in a water bath for a total of 1750 cycle with 10 seconds dwell time at each bath. The fracture load strength test was performed using a universal testing machine. Results: There was no statistically significant difference between all groups. E.max wraparound group recorded the highest fracture resistance mean value (422.1 N) followed by Cerasmart wraparound group (317.23 N), then e.max featheredge group (289.6 N), and finally Cerasmart featheredge group (259.3 N) had the lowest value as analyzed by one-way ANOVA. Conclusions: The Cerasmart hybrid material could be considered as a valid alternative to the widely used IPS e.max material. The fracture resistance of laminate veneers is not influenced by different type of preparation designs.


2015 ◽  
Vol 03 (02) ◽  
pp. 080-084
Author(s):  
Vijay Singh ◽  
Poonam Bogra ◽  
Saurabh Gupta ◽  
Navneet Kukreja ◽  
Neha Gupta

AbstractFracture resistance of endodontically treated teeth restored with post. Aims: This study aims to compare the fracture resistance of endodontically treated teeth restored with resin fiber and stainless steel post. Commercially available prefabricated resin fiber post(Dentsply Maillefer Easy Post), prefabricated stainless steel post(Coltene/Whaledent Parapost) were used. Methods and Material: Forty five maxillary central incisors were obturated and divided into 3 groups: Control Group (Group I) without any post (n = 15), Resin Fiber Post Group (Group II) (n = 15) and Stainless Steel Post Group (Group III) (n = 15). In all Groups except control group, post space was prepared; a post was cemented, and a core build-up was provided. All the specimens were subjected to compressive force under a universal testing machine until fracture. Statistical analysis used: The results were analyzed using the variable analysis test (ANOVA). Results: One-way analysis of variance revealed significant difference among test groups. The control group demonstrated highest fracture resistance (925.2183 N), followed by the resin fiber post group (486.7265 N) and stainless steel post group (423.539N). Conclusions: Teeth restored with resin fiber post showed higher fracture resistance values than prefabricated stainless steel post.


2016 ◽  
Vol 6 (1) ◽  
pp. 19-23
Author(s):  
Amol Mhatre ◽  
VK Ravindranath ◽  
Sachin Doshi ◽  
Girish Karandikar ◽  
PS Vivek

ABSTRACT Aim The aim of this in vitro study was to investigate the efficiency of the new generation of elastomeric ligatures with innovative designs (SlideTM and AlastiKTM Easy-to-Tie) in reducing frictional resistance (FR) during sliding mechanics as compared with conventional ligatures. Materials and Methods Sixty ligature samples divided into four groups were used for the study. Group A: QuiK-StiK™ (3M Unitek, Monrovia, CA, USA), Group B: AlastiK™ Easy-to-Tie (3M Unitek, Monrovia, CA, USA), Group C: Slide™ (Leone, Firenze, Italy), and Group D: SS ligatures 0.010” (Libral Traders, New Delhi, India). Universal Testing Machine, Instron was used for measuring FR at the bracket-wire interface. Results There was statistically significant difference in FR among all the four groups of ligatures tested (p < 0.001). Slide ligatures produced the least amount of FR followed by SS ligatures, Easy-to-Tie, and QuiK-StiK in the increasing order of the FR values registered. Conclusion SlideTM ligatures may represent a valid alternative to passive self-ligating brackets when minimal amount of friction is desired. Angulation introduced into the elastomeric ligatures reduces the friction in comparison to conventional elastomeric ligatures. How to cite this article Vivek PS, Ravindranath VK, Karandikar G, Doshi S, Mhatre A, Sonawane M. Frictional Characteristics of the Newer Low-friction Elastomeric Ligatures. J Contemp Dent 2016;6(1):19-23.