Analytical evaluation of the new Beckman Coulter Access high sensitivity cardiac troponin I immunoassay

Author(s):  
Giuseppe Lippi ◽  
Anna Ferrari ◽  
Giorgio Gandini ◽  
Matteo Gelati ◽  
Claudia Lo Cascio ◽  
...  

AbstractBackground:This study was aimed to evaluate the analytical performance of the novel chemiluminescent and fully-automated Beckman Coulter Access hsTnI high-sensitivity immunoassay for measurement of cardiac troponin I (cTnI).Methods:The study, using lithium heparin samples, included assessment of limit of blank (LOB), limit of detection (LOD), functional sensitivity, linearity, imprecision (within run, between-run and total), calculation of 99th percentile upper reference limit (URL) in 175 healthy blood donors (mean age, 36±12 years; 47% women) and comparison with two other commercial cTnI immunoassays.Results:The LOB, LOD and functional sensitivity of Access hsTnI were 0.14, 0.34 and 1.35 ng/L, respectively. The within-run, between-run and total imprecision was 2.2%–2.9%, 4.6%–5.4%, and 5.4%–6.1%, respectively. The linearity was excellent in the range of cTnI values between 0.95 and 4195 ng/L (r=1.00). The 99th percentile URL was 15.8 ng/L. Measurable cTnI values were found in 173/175 healthy subjects (98.9%). Good agreement of cTnI values was found with AccuTnI+3 (r=0.97; mean bias, −9.3%), whereas less satisfactory agreement was found with Siemens Dimension Vista cTnI (r=0.95; mean bias, −55%).Conclusions:The results of our evaluation of the Beckman Coulter Access hsTnI indicate that the analytical performance of this fully-automated immunoassay is excellent.

2021 ◽  
pp. emermed-2020-210812
Author(s):  
Rob Meek ◽  
Louise Cullen ◽  
Zhong Xian Lu ◽  
Arthur Nasis ◽  
Lisa Kuhn ◽  
...  

BackgroundHigh-sensitivity cardiac troponin I (hs-cTnI) assays promise high diagnostic accuracy for myocardial infarction (MI). In an ED where conventional cTnI was in use, we evaluated an assessment pathway using the new Access hsTnI assay.MethodsThis retrospective analysis recruited ED patients with suspected MI between June and September 2019. All patients received routine care with a conventional cTnI assay (AccuTnI +3: limit of detection (LoD) 10 ng/L, 99th centile upper reference limit (URL) 40 ng/L, abnormal elevation cut-point 80 ng/L). Arrival, then 90-minute or 360-minute cTnI levels for low and non-low risk patients, respectively (ED Assessment of Chest pain score) guided diagnosis and disposition which was at treating physician discretion. The same patients had arrival and 90-minute or 180-minute samples drawn for hs-cTnI levels (Access hsTnI: LoD 2 ng/L, 99th centile URL 10 ng/L (females) and 20 ng/L (males); abnormal elevation above the URL and delta >30%). Treating physicians were blinded to the hs-cTnI results. Using the hs-cTnI values, investigators retrospectively assigned likely diagnosis, disposition and likelihood of a 30-day major adverse cardiac event (MACE). Admission was recommended for significantly rising hs-cTnI elevations. The primary objective was to demonstrate an acceptable unexpected 30-day post-discharge MACE rate of <1%. cTnI elevation rates, diagnostic outcomes and ED disposition were also compared between pathways.ResultsFor the 935 patients, unexpected 30-day post-discharge MACE rates were 0/935 (0%, 95% CI 0% to 0.4%) with the conventional or novel pathway. For the high-sensitivity and conventional assays, respectively, abnormal elevation rates were 29% (95% CI 26% to 32%) and 19% (95% CI 17% to 22%), for MI were 9% (95% CI 8% to 11%) and 8% (95% CI 6% to 10%), and for hospital admission were 42% (95% CI 39% to 45%) and 43% (95% CI 40% to 47%).ConclusionThe novel pathway using the Access hsTnI assay has an acceptably low 30-day MACE rate.


2016 ◽  
Vol 462 ◽  
pp. 193-200 ◽  
Author(s):  
Magdalena Krintus ◽  
Marek Kozinski ◽  
Tomasz Fabiszak ◽  
Magdalena Kuligowska-Prusinska ◽  
Ewa Laskowska ◽  
...  

Author(s):  
Silvia Masotti ◽  
Veronica Musetti ◽  
Concetta Prontera ◽  
Simona Storti ◽  
Claudio Passino ◽  
...  

Abstract Bakground The aim of this study was to evaluate both analytical characteristics and clinical results of a new chemiluminescent method for the measurement of cardiac troponin I (cTnI), named VITROS ® High Sensitivity Troponin I Assay, using the VITROS® 3600 automated platform. The results found with this new method were compared to those observed with hs-cTnI ARCHITECT and ECLIA hs-cTnT ELECSYS methods. Methods For evaluation of analytical performance and comparison of clinical results, plasma samples (lithium-heparin), were collected from apparently healthy subjects and patients with cardiovascular diseases. Results The hs-cTnI VITROS method showed values for limit of blank (LoB 0.33 ng/L), limit of detection (LoD, 0.91 ng/L), limit of quantifications at 20% (LoQ 20% CV, 1.82 ng/L), and 10% (LoQ 10% CV, 4,74 ng/L), which are comparable to those previously reported for other hs-cTnI methods. Moreover, the clinical results of the hs-cTnI VITROS method were found to be closely correlated to those of hs-cTnI ARCHITECT (R = 0,9883, N = 198) and ECLIA hs-cTnT Elecsys (R = 0,9704, N = 293) methods. Conclusions The hs-cTnI VITROS method shows analytical performance comparable to other cTnI and cTnT assay. The results of this study confirm that there are significant systematic differences among hs-cTnI methods. Further multicenter studies using larger reference populations are needed in order to obtain a better estimation, especially of the 99° percentile URL values categorized for sex and age of hs-cTnI and hs-cTnT methods.


Author(s):  
Peter A. Kavsak ◽  
Tara Edge ◽  
Chantele Roy ◽  
Paul Malinowski ◽  
Karen Bamford ◽  
...  

AbstractObjectivesTo analytically evaluate Ortho Clinical Diagnostics VITROS high-sensitivity cardiac troponin I (hs-cTnI) assay in specific matrices with comparison to other hs-cTn assays.MethodsThe limit of detection (LoD), imprecision, interference and stability testing for both serum and lithium heparin (Li-Hep) plasma for the VITROS hs-cTnI assay was determined. We performed Passing-Bablok regression analyses between sample types for the VITROS hs-cTnI assay and compared them to the Abbott ARCHITECT, Beckman Access and the Siemens ADVIA Centaur hs-cTnI assays. We also performed Receiver-operating characteristic curve analyses with the area under the curve (AUC) determined in an emergency department (ED)-study population (n=131) for myocardial infarction (MI).ResultsThe VITROS hs-cTnI LoD was 0.73 ng/L (serum) and 1.4 ng/L (Li-Hep). Stability up to five freeze-thaws was observed for the Ortho hs-cTnI assay, with the analyte stability at room temperature in serum superior to Li-Hep with gross hemolysis also affecting Li-Hep plasma hs-cTnI results. Comparison of Li-Hep to serum concentrations (n=202), yielded proportionally lower concentrations in plasma with the VITROS hs-cTnI assay (slope=0.85; 95% confidence interval [CI]:0.83–0.88). In serum, the VITROS hs-cTnI concentrations were proportionally lower compared to other hs-cTnI assays, with similar slopes observed between assays in samples frozen <−70 °C for 17 years (ED-study) or in 2020. In the ED-study, the VITROS hs-cTnI assay had an AUC of 0.974 (95%CI:0.929–0.994) for MI, similar to the AUCs of other hs-cTn assays.ConclusionsLack of standardization of hs-cTnI assays across manufacturers is evident. The VITROS hs-cTnI assay yields lower concentrations compared to other hs-cTnI assays. Important differences exist between Li-Hep plasma and serum, with evidence of stability and excellent clinical performance comparable to other hs-cTn assays.


Author(s):  
Pascale Beyne ◽  
Erik Bouvier ◽  
Patrick Werner ◽  
Pierre Bourgoin ◽  
Damien Logeart ◽  
...  

AbstractThe aim of this study was to define the use of a new cardiac troponin I (cTnI) assay for emergency patients with chest pain and no specific electrocardiographic changes consistent with the presence of ischemia. Patients (n=106) admitted in Emergency/Cardiology Departments for chest pain and suspicion of acute coronary syndrome (ACS) were randomized into two diagnosis groups (ACS or non-ACS) by two independent cardiologists. cTnI measurements were performed at admission, and 6 hours and 12 hours later with a new generation assay (Access AccuTnI, Beckman Coulter). Using an upper reference limit of 0.04 μg/l, 27 patients had a cTnI elevation not related to the final diagnosis of ischemia; the positive predictive value (PPV) was 67% with specificity 48%. The decisional value was re-defined and set at 0.16 μg/l, a concentration corresponding to the 99th percentile of the non-ACS patient group. Precision (coefficient of variation) was 8% at this level, PPV 97% and specificity 98%. This new decisional value is now used in our institution and could be included in standard care guidelines to improve the management of patients presenting chest pain in emergency departments.


1997 ◽  
Vol 43 (11) ◽  
pp. 2047-2051 ◽  
Author(s):  
Fred S Apple ◽  
Alireza Falahati ◽  
Pamela R Paulsen ◽  
Elizabeth A Miller ◽  
Scott W Sharkey

Abstract This study compared the diagnostic accuracy of the measurement of serum cardiac troponin I (cTnI) with creatine kinase (CK) MB mass in patients with minor myocardial injury whose measured total CK activity did not exceed twice the upper reference limit (300 U/L for men; 200 U/L for women). Forty-eight consecutive patients presenting with chest pain and with in-hospital documentation of myocardial injury were enrolled. Electrocardiogram, echocardiogram, and serial serum CK-MB mass, cTnI, and total CK were measured over 36 h after admission. Peak total CK activity was within normal limits in 28 patients (58%). The mean (±SD) peak CK-MB mass and cTnI concentrations were: 16.4 (11.8) μg/L and 132 (13.0) μg/L; respectively. The peak biochemical marker index (defined as CK-MB or cTnI divided by its respective upper reference limit) was significantly (P &lt;0.05) higher for cTnI than for CK-MB from 7 to 36 h. The clinical sensitivity for detection of myocardial injury for cTnI was 100% [95% confidence interval (CI): 87.2% to 100%], compared with 81.8% (CI: 67.3% to 91.8%) for CK-MB. Thus, cTnI was more sensitive than CK-MB mass for detection of myocardial injury in patients with small increases of total CK.


2021 ◽  
Vol 10 (1) ◽  
Author(s):  
Victor J. van den Berg ◽  
Rohit M. Oemrawsingh ◽  
Victor A. W. M. Umans ◽  
Isabella Kardys ◽  
Folkert W. Asselbergs ◽  
...  

Background Detailed insights in temporal evolution of high‐sensitivity cardiac troponin following acute coronary syndrome (ACS) are currently missing. We aimed to describe and compare the post‐ACS kinetics of high‐sensitivity cardiac troponin I (hs‐cTnI) and high‐sensitivity cardiac troponin T (hs‐cTnT), and to determine their intra‐ and interindividual variation in clinically stable patients. Methods and Results We determined hs‐cTnI (Abbott) and hs‐cTnT (Roche) in 1507 repeated blood samples, derived from 191 patients with ACS (median, 8/patient) who remained free from adverse cardiac events during 1‐year follow‐up. Post‐ACS kinetics were studied by linear mixed‐effect models. Using the samples collected in the 6‐ to 12‐month post‐ACS time frame, patients were then considered to have chronic coronary syndrome. We determined (differences between) the average hs‐cTnI and average hs‐cTnT concentration, and the intra‐ and interindividual variation for both biomarkers. Compared with hs‐cTnT, hs‐cTnI peaked higher (median 3506 ng/L versus 494 ng/L; P <0.001) and was quicker below the biomarker‐specific upper reference limit (16 versus 19 days; P <0.001). In the post–6‐month samples, hs‐cTnI and hs‐cTnT showed modest correlation ( r spearman =0.60), whereas the average hs‐cTnT concentration was 5 times more likely to be above the upper reference limit than hs‐cTnI. The intraindividual variations of hs‐cTnI and hs‐cTnT were 14.0% and 18.1%, while the interindividual variations were 94.1% and 75.9%. Conclusions Hs‐cTnI peaked higher after ACS and was quicker below the upper reference limit. In the post–6‐month samples, hs‐cTnI and hs‐cTnT were clearly not interchangeable, and average hs‐cTnT concentrations were much more often above the upper reference limit than hs‐cTnI. For both markers, the within‐patient variation fell largely below beween‐patient variation. Registration URL: https://www.trialregister.nl ; unique identifiers: NTR1698 and NTR1106.


2020 ◽  
Vol 10 (1) ◽  
Author(s):  
Jungchan Park ◽  
Cheol Won Hyeon ◽  
Seung-Hwa Lee ◽  
Sangmin Maria Lee ◽  
Junghyun Yeo ◽  
...  

Abstract Preoperative high-sensitivity cardiac troponin (hs-cTn) above the 99th-percentile upper reference limit (URL) is associated with mortality after noncardiac surgery. This study aimed to evaluate whether preoperative hs-cTn concentrations above the lowest limit of detection (LOD) but below the 99th-percentile URL can predict mortality after noncardiac surgery.From January 2010 to April 2019, a total of 12,415 noncardiac surgical patients with preoperative hs-cTn I below the 99th-percentile URL were enrolled. The patients were divided into two groups according to preoperative hs-cTn I concentration: (1) [hs-cTn] below the LOD (6 ng/L), and (2) mildly elevated [hs-cTn] but below the 99th-percentile URL (40 ng/L). The primary outcome was 30-day mortality. Of the 12,415 patients enrolled, 7958 (64.1%) were in the LOD group whereas 4457 (35.9%) were in the mild elevation group. The incidence of 30-day mortality was significantly greater in the mild elevation group (2.1% vs. 4.0% hazard ratio [HR] 1.73; 95% confidence interval [CI] 1.39–2.16; p < 0.001) in the multivariate analyses. The propensity score matched analyses also produced a similar result (2.6% vs. 4.2% HR 1.61; 95% CI 1.26–2.07; p < 0.001). The threshold at which the risk of mortality increased corresponded to a preoperative hs-cTn I ≥ 12 ng/L. Patients with preoperative hs-cTn I above the LOD and below the 99th-percentile URL had greater 30-day mortality after noncardiac surgery.


2018 ◽  
Vol 56 (3) ◽  
pp. 492-501 ◽  
Author(s):  
Silvia Masotti ◽  
Concetta Prontera ◽  
Veronica Musetti ◽  
Simona Storti ◽  
Rudina Ndreu ◽  
...  

AbstractBackground:The study aim was to evaluate and compare the analytical performance of the new chemiluminescent immunoassay for cardiac troponin I (cTnI), called Access hs-TnI using DxI platform, with those of Access AccuTnI+3 method, and high-sensitivity (hs) cTnI method for ARCHITECT platform.Methods:The limits of blank (LoB), detection (LoD) and quantitation (LoQ) at 10% and 20% CV were evaluated according to international standardized protocols. For the evaluation of analytical performance and comparison of cTnI results, both heparinized plasma samples, collected from healthy subjects and patients with cardiac diseases, and quality control samples distributed in external quality assessment programs were used.Results:LoB, LoD and LoQ at 20% and 10% CV values of the Access hs-cTnI method were 0.6, 1.3, 2.1 and 5.3 ng/L, respectively. Access hs-cTnI method showed analytical performance significantly better than that of Access AccuTnI+3 method and similar results to those of hs ARCHITECT cTnI method. Moreover, the cTnI concentrations measured with Access hs-cTnI method showed close linear regressions with both Access AccuTnI+3 and ARCHITECT hs-cTnI methods, although there were systematic differences between these methods. There was no difference between cTnI values measured by Access hs-cTnI in heparinized plasma and serum samples, whereas there was a significant difference between cTnI values, respectively measured in EDTA and heparin plasma samples.Conclusions:Access hs-cTnI has analytical sensitivity parameters significantly improved compared to Access AccuTnI+3 method and is similar to those of the high-sensitivity method using ARCHITECT platform.