Assessing matrix, interferences and comparability between the Abbott Diagnostics and the Beckman Coulter high-sensitivity cardiac troponin I assays

2018 ◽  
Vol 56 (7) ◽  
pp. 1176-1181 ◽  
Author(s):  
Peter A. Kavsak ◽  
Paul Malinowski ◽  
Chantele Roy ◽  
Lorna Clark ◽  
Shana Lamers

Abstract Background: Analytical evaluation of high-sensitivity cardiac troponin (hs-cTn) assays, with particular attention to imprecision, interferences and matrix effects, at normal cTn concentrations, is of utmost importance as many different clinical algorithms use concentration cutoffs <10 ng/L for decision-making. The objective for the present analytical study was to compare the new Beckman Coulter hs-cTnI assay (Access hsTnI) to Abbott’s hs-cTnI assay in different matrices and for different interferences, with a focus on concentrations <10 ng/L. Methods: The limit of blank (LoB) and the limit of detection (LoD) were determined in different matrices for the Beckman hs-cTnI assay. Passing-Bablok regression and difference plots were determined for 200 matched lithium heparin and EDTA plasma samples for the Beckman assay and 200 lithium heparin samples for the Abbott assay. Both EDTA and heparin plasma samples were also evaluated for stability under refrigerated conditions, for endogenous alkaline phosphatase interference and for hemolysis and icterus. Results: The Beckman hs-cTnI assay LoB was 0.5 ng/L with the following range of LoDs=0.8–1.2 ng/L, with EDTA plasma yielding lower concentrations as compared to lithium heparin plasma (mean difference=−14.9%; 95% CI=−16.9 to 12.9). Below 10 ng/L, lithium heparin cTnI results from the Beckman assay were on average 1.1 ng/L (95% CI=0.7 to 1.5) higher than the Abbott results, with no difference between the methods when using EDTA plasma (mean difference =−0.1 ng/L; 95% CI=−0.3 to 0.2). Low cTnI concentrations were less effected by interferences in EDTA plasma. Conclusions: The Access hsTnI method can reliably detect normal cTnI concentrations with both lithium heparin and EDTA plasma being suitable matrices.

2019 ◽  
Vol 57 (5) ◽  
pp. 745-751 ◽  
Author(s):  
Peter A. Kavsak ◽  
Chantele Roy ◽  
Paul Malinowski ◽  
Lorna Clark ◽  
Shana Lamers ◽  
...  

Abstract Background Manufacturers of high-sensitivity cardiac troponin (hs-cTn) assays have restricted use of what sample types or matrices are acceptable to use for measurement. Our goal was to evaluate the comparability of the Siemens ADVIA Centaur hs-cTnI assay across different matrices and under different storage conditions. Methods Three different QC-plasma matrices were evaluated for imprecision <10 ng/L. Passing-Bablok regression and difference plots were determined for cTnI concentrations spanning the reference interval (limit of quantification to male 99th-percentile: 2.5 ng/L to <60 ng/L) between serum and lithium heparin plasma, lithium heparin and EDTA plasma and between the Siemens and Abbott hs-cTnI assays. Stability at room temperature (RT) and 2–8 °C was also assessed across the three matrices. Results Over 16-weeks the SDs were ≤1.0 ng/L for QCs ranging from 5.0 to 8.3 ng/L. Across the reference interval there was excellent agreement between lithium heparin plasma and serum for the Siemens hs-cTnI assay (slope=0.98/intercept=–0.1), however, cTnI concentrations were proportionally lower in EDTA as compared to lithium heparin plasma (slope=0.90, 95% CI: 0.88–0.92). In lithium heparin plasma the Siemens hs-cTnI concentrations were higher than the Abbott hs-cTnI concentrations (slope=1.26/intercept=–0.2). Stability of cTnI in lithium heparin plasma as compared in serum and EDTA plasma appeared more labile, with decreases ≥20% in concentrations evident as early as 1-day in storage at RT. Conclusions There is excellent agreement in concentrations between lithium heparin plasma and serum with the Siemens ADVIA Centaur hs-cTnI assay; however, cTnI concentrations in EDTA plasma are lower. Reference intervals and clinical studies in EDTA plasma for the Centaur hs-cTnI assay are required before clinical use.


2018 ◽  
Vol 56 (3) ◽  
pp. 492-501 ◽  
Author(s):  
Silvia Masotti ◽  
Concetta Prontera ◽  
Veronica Musetti ◽  
Simona Storti ◽  
Rudina Ndreu ◽  
...  

AbstractBackground:The study aim was to evaluate and compare the analytical performance of the new chemiluminescent immunoassay for cardiac troponin I (cTnI), called Access hs-TnI using DxI platform, with those of Access AccuTnI+3 method, and high-sensitivity (hs) cTnI method for ARCHITECT platform.Methods:The limits of blank (LoB), detection (LoD) and quantitation (LoQ) at 10% and 20% CV were evaluated according to international standardized protocols. For the evaluation of analytical performance and comparison of cTnI results, both heparinized plasma samples, collected from healthy subjects and patients with cardiac diseases, and quality control samples distributed in external quality assessment programs were used.Results:LoB, LoD and LoQ at 20% and 10% CV values of the Access hs-cTnI method were 0.6, 1.3, 2.1 and 5.3 ng/L, respectively. Access hs-cTnI method showed analytical performance significantly better than that of Access AccuTnI+3 method and similar results to those of hs ARCHITECT cTnI method. Moreover, the cTnI concentrations measured with Access hs-cTnI method showed close linear regressions with both Access AccuTnI+3 and ARCHITECT hs-cTnI methods, although there were systematic differences between these methods. There was no difference between cTnI values measured by Access hs-cTnI in heparinized plasma and serum samples, whereas there was a significant difference between cTnI values, respectively measured in EDTA and heparin plasma samples.Conclusions:Access hs-cTnI has analytical sensitivity parameters significantly improved compared to Access AccuTnI+3 method and is similar to those of the high-sensitivity method using ARCHITECT platform.


2016 ◽  
Vol 62 (8) ◽  
pp. 1115-1119 ◽  
Author(s):  
Sara A Love ◽  
Yader Sandoval ◽  
Stephen W Smith ◽  
Jennifer Nicholson ◽  
Jing Cao ◽  
...  

Abstract INTRODUCTION We compared the incidence of undetectable [below the limit of detection (LoD)], measurable (LoD to 99th percentile), and increased cardiac troponin I (cTnI) concentrations above the 99th percentile between Abbott high-sensitivity cTnI (hs-cTnI) and contemporary cTnI assays in a US emergency department population. METHODS Patients (n = 2100) presenting to the emergency department who had serial cTnI (0, 3, 6, 9 h) measurements ordered on clinical indication were enrolled. Contemporary cTnI [Abbott Architect used clinically; 99th percentile: 0.030 μg/L (30 ng/L)] and hs-cTnI [Abbott investigational; sex-specific 99th percentiles: female (F) 16 ng/L, male (M) 34 ng/L] assays simultaneously measured fresh EDTA plasma. RESULTS The hs-cTnI assay measured fewer undetectable cTnI concentrations compared to the contemporary cTnI assay across baseline (F: 31% vs 47%, M: 22% vs 40%) and serial (F: 21% vs 46%; M: 19% vs 54%) measurements. Conversely, the proportion of measurable cTnI concentrations was higher using hs-cTnI compared to contemporary cTnI assay across both baseline (F: 46% vs 31%; M: 60% vs 33%) and serial (F: 48% vs 28%; M: 83% vs 40%) measurements. The overall proportion of patients with increased cTnI concentrations above the 99th percentile was not significantly different between the contemporary (31%) and hs-cTnI (26%) assays (P = 0.09). CONCLUSIONS In patients presenting to the emergency department, the use of the Abbott hs-cTnI assay provides clinicians with more numeric cTnI concentrations. This occurs via a shift from results below the LoD to those between the LoD and the 99th percentile and does not increase in the number of cTnI concentrations above the 99th percentile.


2009 ◽  
Vol 55 (3) ◽  
pp. 573-577 ◽  
Author(s):  
Peter A Kavsak ◽  
Andrew R MacRae ◽  
Marie-Jeanne Yerna ◽  
Allan S Jaffe

Abstract Background: Improvements in cardiac troponin (cTn) assays have increased the rapidity with which clinicians can identify patients with changing cTn concentrations (rise or fall) indicative of acute myocardial injury. The aim of the present study was to characterize a new, high-sensitivity cTnI (hs-cTnI) assay and examine whether increased sensitivity can result in still earlier detection of evolving injury. Methods: We determined the limit of detection, precision profiles, and preliminary estimates of the 99th percentile for the Beckman Coulter hs-cTnI assay in 125 healthy individuals (age &lt;55 years, 54% male). We compared AccuTnI® and hs-cTnI to assess whether change criteria for early concentration changes (i.e., ≥3SD for low concentrations and 20% difference for concentrations &gt;0.10 μg/L) were exceeded in the first 2 specimens (median time between specimens, 1 h; 25th–75th percentile, 1–3 h) from subjects with symptoms suggestive of cardiac ischemia (n = 290). Results: The limit of detection for the hs-cTnI assay was 2.06 ng/L, and the 20% CV and 10% CV concentrations were 2.95 and 8.66 ng/L, respectively. The preliminary 99th percentile estimates in lithium heparin, serum, and EDTA plasma were 9.20, 8.00, and 8.60 ng/L, respectively. In 108 patients with myocardial injury based on the peak AccuTnI concentration, applying the change criteria on the 2 earliest specimens identified 81% (95% CI 73%–88%) of patients using the hs-cTnI assay compared to 62% (53%–71%) using the AccuTnI assay (P &lt; 0.001). Conclusions: Although more extensive validation studies are required, this Beckman Coulter hs-cTnI assay appears to detect patients with evolving myocardial injury earlier.


Author(s):  
Peter A. Kavsak ◽  
Tara Edge ◽  
Chantele Roy ◽  
Paul Malinowski ◽  
Karen Bamford ◽  
...  

AbstractObjectivesTo analytically evaluate Ortho Clinical Diagnostics VITROS high-sensitivity cardiac troponin I (hs-cTnI) assay in specific matrices with comparison to other hs-cTn assays.MethodsThe limit of detection (LoD), imprecision, interference and stability testing for both serum and lithium heparin (Li-Hep) plasma for the VITROS hs-cTnI assay was determined. We performed Passing-Bablok regression analyses between sample types for the VITROS hs-cTnI assay and compared them to the Abbott ARCHITECT, Beckman Access and the Siemens ADVIA Centaur hs-cTnI assays. We also performed Receiver-operating characteristic curve analyses with the area under the curve (AUC) determined in an emergency department (ED)-study population (n=131) for myocardial infarction (MI).ResultsThe VITROS hs-cTnI LoD was 0.73 ng/L (serum) and 1.4 ng/L (Li-Hep). Stability up to five freeze-thaws was observed for the Ortho hs-cTnI assay, with the analyte stability at room temperature in serum superior to Li-Hep with gross hemolysis also affecting Li-Hep plasma hs-cTnI results. Comparison of Li-Hep to serum concentrations (n=202), yielded proportionally lower concentrations in plasma with the VITROS hs-cTnI assay (slope=0.85; 95% confidence interval [CI]:0.83–0.88). In serum, the VITROS hs-cTnI concentrations were proportionally lower compared to other hs-cTnI assays, with similar slopes observed between assays in samples frozen <−70 °C for 17 years (ED-study) or in 2020. In the ED-study, the VITROS hs-cTnI assay had an AUC of 0.974 (95%CI:0.929–0.994) for MI, similar to the AUCs of other hs-cTn assays.ConclusionsLack of standardization of hs-cTnI assays across manufacturers is evident. The VITROS hs-cTnI assay yields lower concentrations compared to other hs-cTnI assays. Important differences exist between Li-Hep plasma and serum, with evidence of stability and excellent clinical performance comparable to other hs-cTn assays.


Author(s):  
Giuseppe Lippi ◽  
Anna Ferrari ◽  
Giorgio Gandini ◽  
Matteo Gelati ◽  
Claudia Lo Cascio ◽  
...  

AbstractBackground:This study was aimed to evaluate the analytical performance of the novel chemiluminescent and fully-automated Beckman Coulter Access hsTnI high-sensitivity immunoassay for measurement of cardiac troponin I (cTnI).Methods:The study, using lithium heparin samples, included assessment of limit of blank (LOB), limit of detection (LOD), functional sensitivity, linearity, imprecision (within run, between-run and total), calculation of 99th percentile upper reference limit (URL) in 175 healthy blood donors (mean age, 36±12 years; 47% women) and comparison with two other commercial cTnI immunoassays.Results:The LOB, LOD and functional sensitivity of Access hsTnI were 0.14, 0.34 and 1.35 ng/L, respectively. The within-run, between-run and total imprecision was 2.2%–2.9%, 4.6%–5.4%, and 5.4%–6.1%, respectively. The linearity was excellent in the range of cTnI values between 0.95 and 4195 ng/L (r=1.00). The 99th percentile URL was 15.8 ng/L. Measurable cTnI values were found in 173/175 healthy subjects (98.9%). Good agreement of cTnI values was found with AccuTnI+3 (r=0.97; mean bias, −9.3%), whereas less satisfactory agreement was found with Siemens Dimension Vista cTnI (r=0.95; mean bias, −55%).Conclusions:The results of our evaluation of the Beckman Coulter Access hsTnI indicate that the analytical performance of this fully-automated immunoassay is excellent.


2021 ◽  
pp. emermed-2020-210812
Author(s):  
Rob Meek ◽  
Louise Cullen ◽  
Zhong Xian Lu ◽  
Arthur Nasis ◽  
Lisa Kuhn ◽  
...  

BackgroundHigh-sensitivity cardiac troponin I (hs-cTnI) assays promise high diagnostic accuracy for myocardial infarction (MI). In an ED where conventional cTnI was in use, we evaluated an assessment pathway using the new Access hsTnI assay.MethodsThis retrospective analysis recruited ED patients with suspected MI between June and September 2019. All patients received routine care with a conventional cTnI assay (AccuTnI +3: limit of detection (LoD) 10 ng/L, 99th centile upper reference limit (URL) 40 ng/L, abnormal elevation cut-point 80 ng/L). Arrival, then 90-minute or 360-minute cTnI levels for low and non-low risk patients, respectively (ED Assessment of Chest pain score) guided diagnosis and disposition which was at treating physician discretion. The same patients had arrival and 90-minute or 180-minute samples drawn for hs-cTnI levels (Access hsTnI: LoD 2 ng/L, 99th centile URL 10 ng/L (females) and 20 ng/L (males); abnormal elevation above the URL and delta >30%). Treating physicians were blinded to the hs-cTnI results. Using the hs-cTnI values, investigators retrospectively assigned likely diagnosis, disposition and likelihood of a 30-day major adverse cardiac event (MACE). Admission was recommended for significantly rising hs-cTnI elevations. The primary objective was to demonstrate an acceptable unexpected 30-day post-discharge MACE rate of <1%. cTnI elevation rates, diagnostic outcomes and ED disposition were also compared between pathways.ResultsFor the 935 patients, unexpected 30-day post-discharge MACE rates were 0/935 (0%, 95% CI 0% to 0.4%) with the conventional or novel pathway. For the high-sensitivity and conventional assays, respectively, abnormal elevation rates were 29% (95% CI 26% to 32%) and 19% (95% CI 17% to 22%), for MI were 9% (95% CI 8% to 11%) and 8% (95% CI 6% to 10%), and for hospital admission were 42% (95% CI 39% to 45%) and 43% (95% CI 40% to 47%).ConclusionThe novel pathway using the Access hsTnI assay has an acceptably low 30-day MACE rate.


2018 ◽  
Vol 64 (7) ◽  
pp. 1044-1053 ◽  
Author(s):  
Martin P Than ◽  
Sally J Aldous ◽  
Richard W Troughton ◽  
Christopher J Pemberton ◽  
A Mark Richards ◽  
...  

Abstract BACKGROUND Increased cardiac troponin I or T detected by high-sensitivity assays (hs-cTnI or hs-cTnT) confers an increased risk of adverse prognosis. We determined whether patients presenting with putatively normal, detectable cTn concentrations [&gt; limit of detection and &lt; upper reference limit (URL)] have increased risk of major adverse cardiovascular events (MACE) or all-cause mortality. METHODS A prospective 5-year follow-up of patients recruited in the emergency department with possible acute coronary syndrome (ACS) and cTn concentrations measured with hs-cTnI (Abbott) and hs-cTnT (Roche) assays. Cox regression models were generated with adjustment for covariates in those without MACE on presentation. Hazard ratios (HRs) for hs-cTn were calculated relative to the HRs at the median concentration. RESULTS Of 1113 patients, 836 were without presentation MACE. Of these, 138 incurred a MACE and 169 died during a median 5.8-year follow-up. HRs for MACE at the URLs were 2.3 (95% CI, 1.7–3.2) for hs-cTnI and 1.8 (95% CI, 1.3–2.4) for hs-cTnT. Corresponding HRs for mortality were 1.7 (95% CI, 1.2–2.2) for hs-cTnI and 2.3 (95 % CI, 1.7–3.1) for hs-cTnT. The HR for MACE increased with increasing hs-cTn concentration similarly for both assays, but the HR for mortality increased at approximately twice the rate for hs-cTnT than hs-cTnI. Patients with hs-cTnI ≥10 ng/L or hs-cTnT ≥16 ng/L had the same percentage of MACE at 5-year follow-up (33%) as patients with presentation MACE. CONCLUSIONS Many patients with ACS ruled out and putatively normal but detectable hs-cTnI concentrations are at similar long-term risk as those with MACE. hs-cTnT concentrations are more strongly associated with 5-year mortality than hs-cTnI.


Author(s):  
Fred S. Apple ◽  
Karen Schulz ◽  
Christian W. Schmidt ◽  
Trees S. Y. van Domburg ◽  
Judith M. Fonville ◽  
...  

Abstract Objectives High sensitivity (hs) cardiac troponin (cTn) assays are defined per the IFCC Committee on Clinical Application of Cardiac Biomarker (C-CB) by the ability to measure ≥ 50% of concentrations greater than the limit of detection (LoD) with an impression of ≤10% at sex-specific 99th percentiles. Our study determined the sex-specific 99th percentile upper reference limits for males and females utilizing heparinized plasma from AACC universal sample bank for the Siemens point of care (POC) Atellica® VTLi hs-cTnI immunoassay. Methods Apparently healthy subjects, included overall 693, males 363, and females 330, following exclusionary surrogate biomarker use of hemoglobin A1c, NT-proBNP, and eGFR, along with statin medication. hs-cTnI was measured in a central laboratory, on multiple POC Atellica® VTLi immunoassay analyzers. The LoD was 1.24 ng/L and limit of quantitation (CV 20%) was 6.7 ng/L. 99th percentile URLs were determined by the nonparametric (NP) method. Results Histograms of the hs-cTnI concentrations (ng/L) for males and females were used to visualize the distributions and concentrations in men and women and differed significantly (pre- and post-exclusion, both p <0.001). 99th percentile URLs were: overall 23 ng/L (90% CI 20–32 ng/L); male 27 ng/L (CI 21–37 ng/L); female 18 ng/L (CI 9–78 ng/L). The percentages of subjects having a measurable concentration ≥ the LoD were: overall 83.7%, male 87.3%, female 79.7%. Conclusions Our findings show the novel POC Atellica® VTLi hs-cTnI assay meets the designation of a ‘high-sensitivity’ assay using heparinized plasma.


2012 ◽  
Vol 58 (11) ◽  
pp. 1574-1581 ◽  
Author(s):  
Fred S Apple ◽  
Ranka Ler ◽  
MaryAnn M Murakami

BACKGROUND Between-assay comparability of 99th percentiles for cardiac troponin concentrations has not been assessed systematically in a single population for a large number of assays. METHODS We determined 99th percentiles for 19 cardiac troponin assays in heparin plasma samples from a population of 272 and 252 presumably healthy males and females, respectively. The assays evaluated included 1 cardiac troponin T (cTnT) assay from Roche and 18 cTnI assays from Abbott, Alere, Beckman, bioMerieux, Instrumentation Laboratory, Ortho-Clinical Diagnostics, Singulex, Siemens, and Roche. Five of these assays were categorized as high-sensitivity, 9 as sensitive-contemporary, and 5 as point-of-care (POC) assays. RESULTS For high-sensitivity cTnI (hs-cTnI) assays 99th percentiles varied from 23 to 58 ng/L. At least 80% of individuals had measurable hs-cTnI, whereas only 25% had measurable high-sensitivity cTnT. All high-sensitivity cardic troponin assays had 99th percentiles that were 1.2–2.4-fold higher in males than females. For the 9 sensitive-contemporary cTnI assays, 99th percentiles varied from 12 to 392 ng/L, and only the Beckman assay gave measurable concentrations in a substantial portion of the population (35% vs ≤6% for the others). Seven of these 9 assays had 1.3–5.0-fold higher 99th percentiles for males than females. For 5 cTnI POC assays, 99th percentiles varied from &lt;10 to 40 ng/L. The Instrumentation Laboratory assay gave measurable results in 27.8% of study participants vs ≤6% for the others. Correlations were generally poor among assays. CONCLUSIONS Among cardiac troponin assays 99th percentile concentrations appear to differ. High-sensitivity assays provide measurable cardiac troponin results in a substantially greater fraction of presumably healthy individuals.