A simple, precise, specific, and accurate reversed phase high performance liquid chromatographic (RP-HPLC) method was developed and validated for determination of vinpocetine in pure and pharmaceutical dosage forms. The different analytical performance parameters such as linearity, accuracy, specificity, precision, and sensitivity (limit of detection and limit of quantitation) were determined according to International Conference on Harmonization ICH Q2 (R1) guidelines. RP-HPLC was conducted on Zorbax C18 (150 mm length × 4.6 mm ID, 5 μm) column. The mobile phase was consisting of buffer (containing 1.54% w/v ammonium acetate solution) and acetonitrile in the ratio (40 : 60, v/v), and the flow rate was maintained at 1.0 mLmin−1. Vinpocetine was monitored using Agilent 1200 series equipped with photo diode array detector (λ = 280 nm). Linearity was observed in concentration range of 160–240 μgmL−1, and correlation coefficient was found excellent (R2 = 0.999). All the system suitability parameters were found within the range. The proposed method is rapid, cost-effective and can be used as a quality-control tool for routine quantitative analysis of vinpocetine in pure and pharmaceutical dosage forms.