DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR ESTIMATION OF EDOXABAN TOSYLATE IN TABLET DOSAGE FORMS

INDIAN DRUGS ◽  
2020 ◽  
Vol 57 (07) ◽  
pp. 47-51
Author(s):  
B. Anupama ◽  
P. Tejaswi ◽  
KNV. Chenchu Lakshmi ◽  
A. Vishwanadh

A rapid, simple and precise reversed phase high performance liquid chromatography (RP-HPLC) method was developed for the estimation of edoxabantosylate in tablets. The quantification was carried out using a Phenomenex C-18 column (250×4.6 mm i.d., 5µm particle size) in isocratic mode with mobile phase comprising of ammonium acetate buffer andacetonitrile in the ratio of 50:50 (V/V) at a flow rate 1 mL/min. The eluent was monitored at 240 nm. The retention time of the drug was 3.486 min. The calibration curve was linear in the concentration range of 5-25 µg/mL and per cent recovery ranged from 98.25-101.6.The developed method was validated as per ICH guidelines and the results obtained were satisfactory.The method can be applied for routine quality control analysis of edoxabantosylate in tablet dosage forms.

Author(s):  
Sireesha D ◽  
Sai Lakshmi E ◽  
Sravya E ◽  
Vasudha Bakshi

A new simple, rapid, specific, accurate, precise and novel Reverse Phase High Performance Liquid Chromatography (RP-HPLC) method has been developed for the estimation of Sitagliptin Phosphate in the pharmaceutical dosage form. The chromatographic separation for Sitagliptin was achieved with mobile phase containing methanol, Thermoscientific C18 column, (250x4.6 particle size of 5μ) at room temperature and UV detection at 248 nm. The compounds were eluted in the isocratic mode at a flow rate of 1ml/min. The retention time of Sitagliptin was 1.91min. The above method was validated in terms of linearity, accuracy, precision, LOD and LOQ in accordance with ICH guidelines.


INDIAN DRUGS ◽  
2018 ◽  
Vol 55 (04) ◽  
pp. 77-80
Author(s):  
M Vijaya Lakshmi ◽  
◽  
K Hima Bindu ◽  
M. Pravallika ◽  
B. N. Nalluri

A simple and precise RP-HPLC method has been developed and validated for the estimation of sarpogrelate hydrochloride, an anti-platelet drug in bulk and pharmaceutical dosage forms. Sarpogrelate is an antagonist at 5HT2A and 5HT2B receptors which blocks serotonin induced platelet aggregation and has application in the treatment of diseases including diabetes mellitus, Raynaud’s disease, angina pectoris and atherosclerosis. Chromatography was carried out on a Phenomenex C18 (250 x 4.6mm, 5μm) column with a mobile phase of 10mM ammonium acetate and acetonitrile (45:55% v/v). The flow rate was 1.2mL/min. The detection wavelength was carried out at 220nm. The retention time is 3.356 minutes for sarpogrelate hydrochloride. The linearity was found in the range of 10-50 μg/ml (R = 0.999) and % RSD is less than 2%. The mean recoveries obtained for sarpogrelate hydrochloride were in the range of 98.73-100.67%. The method is validated as per ICH guidelines and can be applied for the estimation of percentage purity in Sarpogrelate hydrochloride for quality control analysis in bulk and its dosage forms.


Author(s):  
Bijithra Cholaraja ◽  
Shanmugasundaram P ◽  
Ragan G ◽  
Sankar Ask ◽  
Sumithra M

ABSTRACTObjective: To development and validation of a reversed-phase high-performance liquid chromatography (RP-HPLC) for the determination of modafinilin bulk and pharmaceutical dosage forms.Methods: A simple, precise, rapid, and accurate RP-HPLC method was developed for the estimation of modafinil in bulk and pharmaceutical dosageforms. Xterra RP 18 (250 mm × 4.6 mm, 5 µ particle size) with a mobile phase consisting of methanol:water 70:30 V/V was used. The flow rate1.0 ml/min and the effluents were monitored at 260 nm. The retention time and recovery time was 12 minutes. The detector response was linear inthe concentration of 10-50 µg/ml. The respective linear regression equation being Y=452.1x+65237. The limit of detection and limit of quantificationwere 4.547 and 1.377 mcg, respectively. The method was validated by determining its accuracy, precision, and system suitability.Result: The objective of the present work is to develop simple, precise, and reliable HPLC method for the analysis of modafinil in bulk andpharmaceutical dosage forms. This is achieved using the most commonly employed Xterra RP 18 (250 mm × 4.6 mm, 5 μ particle size) columndetection at 260 nm. The present method was validated according to ICH guidelines.Conclusion: In this study, a simple, fast and reliable HPLC method was developed and validated for the determination of modafinil in pharmaceuticalformulations.Keywords: Modafinil, Reversed-phase high-performance liquid chromatography, Estimation, ICH guidelines, Tablets. 


2019 ◽  
Vol 6 (1) ◽  
Author(s):  
Ananda Thangadurai Subramaniam ◽  
Devi Velmurugan ◽  
Sambathkumar Ramanathan ◽  
Kamalakannan Dhanabalan ◽  
Jambulingam Munusamy ◽  
...  

Recent study was conducted to develop and validate analytical methods for estimation of Risedronate sodium in pure and pharmaceutical dosage form using UV Spectroscopy and               RP- HPLC method. The first method (Method A) based on the UV Spectroscopy using 0.1M Hcl as diluent lambda max was found at 261 nm. Linearity existed perceived in the concentration between 10-50 μg/ml (r 2 = 0.999) for the method. The method was validated pertaining to linearity, precision and accuracy studies, LOD and LOQ consistent with ICH guidelines. The second method (Method B), based on determination of Risedronate sodium tablet dosage form by RP-HPLC method.  Chromatography separation was carried out on a C18 (150X4.6 mm x5 µ) SS Column using Methanol: Ammonium formate (85:15) as the mobile phase at a flow rate of   1.0 ml/min. The chromatographic analysis was carried out in the reflectance and absorbance mode at 254 nm and retention time of the drug was found to be 1.11 ml/min for standard and tablet. Linear responses of the drug were in the concentration range of 200-1000 µg/ml. The accuracy of the method was assessed by standard dilution method and found to be 98% to              102% .The results of the analysis were validated statistically prism software. The method established was found to be simple, precise, linear, accurate and sensitive. The developed method can be used for routine quality control analysis of Risedronate sodium in pure and pharmaceutical dosage form.


2020 ◽  
Vol 10 (4) ◽  
pp. 97-101
Author(s):  
Mithun Rudrapal ◽  
Nazim Hussain

A simple, precise and accurate RP-HPLC method was developed and validated for the estimation of gemifloxacin in the tablet dosage form. The separation was achieved on a reversed-phase C-18 column (250 x 4.6 mm i.d., 5 µm) using a mobile phase consisting of acetonitrile/acetate buffer of pH 4.5 (70:30 v/v) at a flow rate of 1.0 ml/min and a detection wavelength of 244 nm. The separation was carried out on an isocratic mode at room temperature. The method was validated as per ICH guidelines for linearity, accuracy, precision, robustness, LOD, LOQ and specificity. The developed method showed good linearity over the concentration range of 50-150 µg/ml (r2=0.995). The average percentage recovery was 99.77%. The LOD and LOQ were 12.678 µg/ml and 14.261 µg/ml, respectively. Based upon validation studies, the developed method can be successfully applied for the routine analysis of gemifloxacin in bulk drugs as well as pharmaceutical dosage forms. Keywords: Gemifloxacin, Tablet dosage form, RP-HPLC, Validation, ICH guidelines


2020 ◽  
Vol 10 (6) ◽  
pp. 6610-6618

A Simple, selective, accurate, precise, linear, and stability-indicating RP-HPLC method was developed and validated for the estimation of Cinacalcet hydrochloride in bulk and tablet dosage forms. Chromatographic separation was achieved on X-Terra Symmetry C18 (4.6 x 150mm; 5 m) with mobile phase containing Phosphate buffer: Acetonitrile (40:60 v/v) pH adjusted to 3.0 ±0.05 with diluted ortho-phosphoric acid. The flow rate was maintained at 0.9 mL/min. The eluent was monitored at 282 nm. Moreover, the retention time of Cinacalcet was 2.8 minutes. The method was validated for linearity, accuracy, precision, and robustness as per ICH guidelines. The developed method was found linear between 25-150 μg/ml, and the linear regression coefficient was 0.999. The % RSD values are less than 2 % indicating the accuracy and precision of the method. The percentage of recovery was obtained from 98-102%. The system suitability parameters were found to be within the limit. Forced degradation studies were conducted under various conditions. The proposed method is simple, rapid, precise, and accurate. It can be used for the quantitation of Cinacalcet hydrochloride in bulk and commercial pharmaceutical dosage forms.


INDIAN DRUGS ◽  
2015 ◽  
Vol 52 (01) ◽  
pp. 26-32
Author(s):  
A Singh ◽  
◽  
C. L Singh ◽  
S. Kumar ◽  
M. Kumar

A sensitive and accurate reversed– phase high performance liquid chromatography (RP-HPLC) method with UV absorbance detection at 289 nm was developed and validated for the determination and quantification of besifloxacin (BSF) in rat plasma. Ofloxacin was used as an internal standard (IS). The sample was prepared by liquid extraction of BSF from plasma, using methanol and acetonitrile (70:30). The chromatographic separation was achieved with octadecylsilane (ODS-3), Hypersil® C18 column (250 mm×6mm×5μm). The chromatographic runtime was less than 5 minutes where the retention time of internal standard and the drug was 2.15 min and 3.30 min respectively. A standard curve with a regression coefficient (r2) 0.999 was obtained in the range of 0.025-20 μg/mL. The method was validated with respect to linearity, range, precision, accuracy and robustness according to ICH guidelines. The method was found to be accurate and robust with a runtime of less than 5 minutes. Hence, the present method was rapid and economical to use for clinical studies as well as to analyze the drug in different plasma samples.


Author(s):  
Deepthi R ◽  
Gowri Sankar D

Objective: The present study aimed to develop a stability-indicating reverse-phase high performance-liquid chromatography (RP-HPLC) method for the estimation of Sofosbuvir, Velpatasvir, and Voxilaprevir in tablet dosage form and validated in accordance with ICH guidelines. Methods: The optimized conditions for the developed RP-HPLC method are Agilent C18 (250 mm×4.6mm, 5μ) column maintained at 30ºC with a mobile phase consisting of Buffer(0.1%OPA) and Acetonitrile taken in the ratio 55:45%v/v on isocratic mode at flow rate 1.0ml/min. The sample was detected at 220 nm. Results: The retention time of Sofosbuvir, Velpatasvir, and Voxilaprevir was found to be 2.17, 2.731 and 3.55 min respectively. The developed method was validated for accuracy, precision, specificity, ruggedness, robustness and solution stability.


INDIAN DRUGS ◽  
2012 ◽  
Vol 49 (05) ◽  
pp. 61-64
Author(s):  
A. L Rao ◽  
◽  
R. V Bhaskara

A reverse phase high performance liquid chromatographic method was developed and validated asper ICH guidelines for estimation of trandolapril and verapamil in combined tablet dosage form. Theseparation was obtained using a mobile phase consisting of acetonitrile and phosphate buffer adjustingpH to 3.0 in the ratio of 70:30 v/v and using Waters C18 (250 x 4.6 mm, 5 mcm) column maintained atambient temperature. The flow rate was 1.2 mL min-1 and UV detection was monitored at 215 nm. Theretention time (min) and linearity range (mcg mL-1) for trandolapril and verapamil were (5.12, 2.70) and(20-60, 20-60), respectively. The method validation results are within the acceptance criteria for precision,accuracy and linearity. The proposed method was found to be suitable for routine quality control ofmarketed formulation containing these APIs.


2020 ◽  
Vol 2020 ◽  
pp. 1-9
Author(s):  
Phoebe Esinam Goku ◽  
Emmanuel Orman ◽  
Anna Naa Kwarley Quartey ◽  
Joseph Kwasi Adu ◽  
Reimmel Kwame Adosraku

An accurate and rapid reverse HPLC method has been developed and validated for the simultaneous quantification of lamivudine, nevirapine, and tenofovir disoproxil fumarate. Suitable separation was achieved on Phenomenex Synergi C18 (250 × 4.6 mm, 4 μm) using mobile phase, methanol (50%): ammonium acetate buffer (adjusted to pH 2.80) (40%): acetonitrile (10%) in an isocratic mode. The drugs were detected at 270 nm with a flow rate of 1.0 ml/min, and the retention times were found to be 3.26, 5.42, and 7.55 minutes for lamivudine, nevirapine, and tenofovir disoproxil fumarate, respectively. The developed method was validated per ICH guidelines. Good linearity was obtained within the concentration ranges of 10–59 µg/ml, 7–42 µg/ml, and 15–90 µg/ml with a correlation coefficient of not less than 0.990. The % RSD values for precision (intraday and interday) and accuracy studies were found to be less than 2%. The results obtained from quantitative analysis conform to USP content requirements for marketed tablet dosage forms, RICOVIR-LN, and tenofovir disoproxil fumarate/lamivudine tablets. The method is therefore useful for routine quality control of antiretroviral tablet dosage forms containing tenofovir disoproxil fumarate, lamivudine, and nevirapine.