strehl ratio
Recently Published Documents


TOTAL DOCUMENTS

131
(FIVE YEARS 22)

H-INDEX

17
(FIVE YEARS 2)

Photonics ◽  
2021 ◽  
Vol 8 (12) ◽  
pp. 559
Author(s):  
Francesco D’Oria ◽  
Ali Nowrouzi ◽  
Jorge L. Alio del Barrio ◽  
Francesco Versaci ◽  
Jorge L. Alio

(1) Background: This study aimed to evaluate and compare the clinical optical image quality following implantation with different premium IOLs by the analysis of the point spread function (PSF) Strehl ratio using a Pyramidal WaveFront-based sensor (PWS) aberrometer at two different pupil sizes. (2) Methods: This study included 96 eyes of 70 patients implanted with: (1) 19 AcrySof SA60AT (control group); (2) 24 LENTIS Mplus LS-313 MF30; (3) 33 LENTIS Mplus LS-313 MF15; and (4) 20 Precizon Presbyopic. Main outcome measures were PSF Strehl ratio, PSF Strehl ratio excluding second-order aberrations (PSFw2), total root-mean-square (RMS), and low- and high-order aberrations’ RMS measured by PWS aberrometer. Results: SA60AT had the highest significant PSFw2 Strehl ratio at both 3- and 4-mm pupil size (0.41 ± 0.11 and 0.28 ± 0.07) followed by LENTIS Mplus 15 (group C, 0.35 ± 0.1 and 0.21 ± 0.06) and a near tie between LENTIS MPLUS 30 (group B, 0.27 ± 0.08 and 0.18 ± 0.06) and Precizon Presbyopic (group D, 0.27 ± 0.07 and 0.17 ± 0.04). MPlus MF15 was found to be significantly better than MPlus MF30 at both 3.00 mm (p < 0.0001) and 4.00 mm (p = 0.002). (4) Conclusions: The PSFw2 represents a new tool to objectively evaluate the far distance retinal image quality of multifocal IOLs, and the far distance clinical image quality parameters measured by PWS aberrometer differed significantly according to the technology of the implanted lens.


2021 ◽  
Vol 21 (10) ◽  
pp. 245
Author(s):  
Shi-Dong Shen ◽  
Xiang-Qun Cui ◽  
Yong Zhang

Abstract In this paper, co-phasing errors of a segmented primary mirror tiled by hexagonal segments are successfully calculated for the 12-meter Large aperture Optical/infrared Telescope (LOT). Co-phasing errors including out-of-plane errors are simulated separately and comprehensively based on several software simulation platforms. PAOLA simulation results show that the Strehl Ratio (SR) of LOT is larger than 0.8 when the RMS value of tip-tilt obeying a normal distribution is less than 0.018 arcsec, and the SR of LOT is larger than 0.8 when the RMS value of piston obeying a normal distribution is less than 40 nm. Besides, simulation results of Zemax show that the SR of LOT is larger than 0.8 when the RMS value of tip-tilt obeying a normal distribution is less than 0.02 arcsec, and the SR of LOT is larger than 0.8 when the RMS value of piston obeying a normal distribution is less than 40 nm. These simulation results successfully lay a solid foundation for LOT (especially the segmented primary mirror with active optics).


2021 ◽  
Vol 8 (1) ◽  
Author(s):  
Jorge L. Alio ◽  
Francesco D’Oria ◽  
Francesca Toto ◽  
Joan Balgos ◽  
Antonio Palazon ◽  
...  

Abstract Background To study and compare the clinical optical image quality following implantation with different premium IOLs by analysing the point spread function (PSF) Strehl ratio using a pyramidal wavefront sensor (PWS)-based aberrometer. Methods This study included 194 eyes implanted with: (a) 19 AcrySof SA60AT (control group); (b) 19 Miniwell; (c) 24 LENTIS Mplus LS-313 MF30; d) 33 LENTIS Mplus LS-313 MF15; (e) 17 AkkoLens Lumina; (f) 31 AT LISA Tri 839MP; (g) 20 Precizon Presbyopic; (h) 20 AcrySof IQ PanOptix; (i) 11 Tecnis Eyhance. Main outcome measures were PSF Strehl ratio, PSF Strehl ratio excluding second-order aberrations (PSFw2), total root mean square (RMS), low-order aberration (LOA) and high-order aberration (HOA) RMS measured by PWS aberrometer. Results AT LISA Tri had the highest PSFw2 Strehl ratio at both 3.0- and 4.0-mm pupil size (0.52 ± 0.14 and 0.31 ± 0.10; P < 0.05), followed by SA60AT (0.41 ± 0.11 and 0.28 ± 0.07) and PanOptix (0.4 ± 0.07 and 0.26 ± 0.04). AT LISA Tri was found to provide a significantly better retinal image quality than PanOptix at both 3.0 mm (P < 0.0001) and 4.0 mm (P = 0.004). Mplus MF15 was found to be significantly better than Mplus MF30 at both 3.0 mm (P < 0.0001) and 4.0 mm (P = 0.002). Total RMS, LOA RMS, HOA RMS, PSF Strehl ratio and PSFw2 varied significantly between the studied groups (P < 0.001). Conclusions Far distance clinical image quality parameters measured by PWS aberrometer differed significantly according to the technology of the implanted lens. AT LISA Tri, SA60AT and PanOptix showed the highest values of far distance retinal image quality, while the lowest PSFw2 Strehl ratios were displayed by Miniwell, Mplus MF30 and Precizon Presbyopic.


Photonics ◽  
2021 ◽  
Vol 8 (9) ◽  
pp. 387
Author(s):  
Luka Černe ◽  
Jaka Petelin ◽  
Rok Petkovšek

In this manuscript, an implementation of a tunable nonlinear phase compensation method is demonstrated on a typical femtosecond hybrid laser consisting of a fiber pre-amplifier and an additional solid-state amplifier. This enables one to achieve constant laser pulse parameters over a wide range of pulse repetition rates in such a laser. As the gain in the solid-state amplifier is inversely proportional to the input power, the shortfall in the solid-state gain at higher repetition rates must be compensated for with fiber pre-amplifier to ensure constant pulse energy. This increases the accumulated nonlinear phase and consequently alters the laser pulse parameters such as pulse duration and Strehl ratio. To overcome this issue, the nonlinear phase must be compensated for, and what is more it should be compensated for to a different extent at different pulse repetition rates. This is achieved with a tunable CFBG, used also as a pulse stretcher. Using this concept, we demonstrate that constant laser pulse parameters such as pulse energy, pulse duration and Strehl ratio can be achieved in a hybrid laser regardless of the pulse repetition rate.


2021 ◽  
Vol Volume 14 ◽  
pp. 1541-1546
Author(s):  
Xiaoyan Bian ◽  
Yuxia Guo ◽  
Shixuan Guo ◽  
Song Zhang ◽  
Guisen Zhang ◽  
...  

2021 ◽  
Author(s):  
Tan Long ◽  
Xin Gu ◽  
Wei Wei ◽  
Ting Ma ◽  
Rui Wang

Abstract Background: To investigate the distribution of the center of the intraocular lens (IOL) after phacoemulsification, and to assess the correlation between the center of IOL and preoperative angle kappa, angle alpha, and objective internal visual quality, respectively, in cataract patients with monofocal and bifocal IOLs implantation. Methods: Prospective cross-section cases series. One hundred and thirty-seven eyes of 107 patients who underwent phacoemulsification were included. Preoperative angle kappa and alpha, postoperative internal ocular aberrations, internal objective visual quality, and the center of IOL relative to the visual axis (CIV) was evaluated using iTrace system. Independent sample t-tests and Pearson correlations were performed.Results: Locations of CIV were scattered in all directions centered on corneal light reflection for both C-Loop designed IOL and plate-haptic designed IOL. No correlations were found between CIV and preoperative angle kappa and alpha in both magnitude and orientation. No correlations were found between CIV and postoperative internal ocular aberrations (astigmatism, coma, and trefoil). In the bifocal IOLs group, the CIV was negatively correlated to the internal Strehl ratio at 3mm; however, it was not correlated to the Strehl ratio at 5mm. The magnitude of CIV was positively correlated to the length of the optic axis.Conclusions: CIV was not predictable according to angle kappa and alpha before cataract surgery. CIV was not related to internal ocular aberration, but large CIV may lead to light scattering due to steps between diffractive rings in patients with small pupil sizes. The magnitude of CIV may be greater in cataract patients with longer optic axis.Trial registration: retrospectively registered.


2020 ◽  
Vol 27 (5) ◽  
pp. 1131-1140 ◽  
Author(s):  
Grant Cutler ◽  
Daniele Cocco ◽  
Elaine DiMasi ◽  
Simon Morton ◽  
Manuel Sanchez del Rio ◽  
...  

This paper presents a novel cantilevered liquid-nitrogen-cooled silicon mirror design for the first optic in a new soft X-ray beamline that is being developed as part of the Advanced Light Source Upgrade (ALS-U) (Lawrence Berkeley National Laboratory, USA). The beamline is optimized for photon energies between 400 and 1400 eV with full polarization control. Calculations indicate that, without correction, this design will achieve a Strehl ratio greater than 0.85 for the entire energy and polarization ranges of the beamline. With a correction achieved by moving the focus 7.5 mm upstream, the minimum Strehl ratio is 0.99. This design is currently the baseline plan for all new ALS-U insertion device beamlines.


2020 ◽  
Vol 27 (5) ◽  
pp. 1121-1130
Author(s):  
Frank Seiboth ◽  
Dennis Brückner ◽  
Maik Kahnt ◽  
Mikhail Lyubomirskiy ◽  
Felix Wittwer ◽  
...  

Modern subtractive and additive manufacturing techniques present new avenues for X-ray optics with complex shapes and patterns. Refractive phase plates acting as glasses for X-ray optics have been fabricated, and spherical aberration in refractive X-ray lenses made from beryllium has been successfully corrected. A diamond phase plate made by femtosecond laser ablation was found to improve the Strehl ratio of a lens stack with a numerical aperture (NA) of 0.88 × 10−3 at 8.2 keV from 0.1 to 0.7. A polymer phase plate made by additive printing achieved an increase in the Strehl ratio of a lens stack at 35 keV with NA of 0.18 × 10−3 from 0.15 to 0.89, demonstrating diffraction-limited nanofocusing at high X-ray energies.


2020 ◽  
Vol 40 (11) ◽  
pp. 2995-3004
Author(s):  
Jiahui Ma ◽  
Shanshan Wei ◽  
Xiaodan Jiang ◽  
Yilin Chou ◽  
Yuexin Wang ◽  
...  

Abstract Purpose To explore objective visual quality in dry eye diseases (DED) and the correlation between corneal nerves and objective visual quality. Methods Ninety-eight eyes of 49 patients with DED were included. Each patient was evaluated with the ocular surface disease index (OSDI), eyelid margin signs and meibomian gland assessments; corneal staining; tear film breakup time (TBUT); tear meniscus height (TMH); in vivo confocal microscopic (IVCM); objective visual quality including the objective scatter index (OSI), mean objective scattering index (mOSI), modulation transfer function (MTF) cutoff value and Strehl ratio. Results A significant correlation was found between the OSDI and mOSI (r = 0.422, p = 0.005), MTF cutoff value (r = − 0.355, p = 0.020), and Strehl ratio (r = − 0.446, p = 0.003). The OSI was significantly correlated with TBUTf (r = − 0.213, p = 0.042). The mOSI, MTF cutoff value, Strehl ratio were correlated with eyelid margin signs and meibomian assessments. Additionally, there was a statistically significant correlation between corneal nerve length and the mOSI (r = − 0.239, p = 0.037), OSI (r = − 0.294, p = 0.028), MTF cutoff value(r = 0.282, p = 0.012), and Strehl ratio (r = 0.299, p = 0.008). Conclusions Our study explored that objective visual quality was correlated with clinical symptoms and signs in DED patients. Furthermore, for the first time, our study explored the relationship between corneal nerves and objective visual quality and discovered that longer and wider corneal nerves were associated with better objective visual quality, which suggested that nerve changes may be a factor that related to poor visual quality in DED patients.


Sign in / Sign up

Export Citation Format

Share Document