joint registry
Recently Published Documents


TOTAL DOCUMENTS

190
(FIVE YEARS 91)

H-INDEX

24
(FIVE YEARS 6)

Author(s):  
Kevin Deere ◽  
Gulraj S. Matharu ◽  
Yoav Ben-Shlomo ◽  
J. Mark Wilkinson ◽  
Ashley W. Blom ◽  
...  

Aims A recent report from France suggested an association between the use of cobalt-chrome femoral heads in total hip arthroplasties (THAs) and an increased risk of dilated cardiomyopathy and heart failure. Cobalt-chrome is a commonly used material in orthopaedic implants. If the reported association is causal, the consequences would be significant given the millions of joint replacements and other orthopaedic procedures in which cobalt-chrome is used annually. We examined whether cobalt-chrome-containing THAs were associated with an increased risk of all-cause mortality, heart outcomes, cancer, and neurodegenerative disorders in a large national database. Methods Data from the National Joint Registry was linked to NHS English hospital inpatient episodes for 374,359 primary THAs with up to 14.5 years follow-up. We excluded any patients with bilateral THAs, knee replacements, indications other than osteoarthritis, aged under 55 years, and diagnosis of one or more outcome of interest before THA. Implants were grouped as either containing cobalt-chrome or not containing cobalt-chrome. The association between implant construct and the risk of all-cause mortality and incident heart failure, cancer, and neurodegenerative disorders was examined. Results There were 158,677 individuals (42.4%) with an implant containing cobalt-chrome. There were 47,963 deaths, 27,332 heart outcomes, 35,720 cancers, and 22,025 neurodegenerative disorders. There was no evidence of an association that patients with cobalt-chrome implants had higher rates of any of the outcomes. Conclusion Cobalt-chrome-containing THAs did not have an increased risk of all-cause mortality, or clinically meaningful heart outcomes, cancer or neurodegenerative disorders into the second decade post-implantation. Our findings will help reassure clinicians and the increasing number of patients receiving primary THA worldwide that the use of cobalt-chrome containing implants is not associated with significant adverse systemic effects.


2021 ◽  
Vol 16 (1) ◽  
Author(s):  
Talal Al-Jabri ◽  
Angela Brivio ◽  
Nicola Maffulli ◽  
David Barrett

Abstract Background Instability is one of the most common reasons for revision after a total knee replacement. It accounts for 17.4% of all single-stage revision procedures performed in the UK National Joint Registry. Through a careful patient evaluation, physical assessment and review of investigations one can identify the likely type of instability. Aims To critically examine the different types of instability, their presentation and evidence-based management options. Method A comprehensive literature search was conducted to identify articles relevant to the aetiology and management of instability in total knee replacements. Results Instability should be categorised as isolated or global and then, as flexion, mid-flexion, extension or recurvatum types. By identifying the aetiology of instability one can correctly restore balance and stability. Conclusion With careful judgement and meticulous surgical planning, instability can be addressed and revision surgery can provide patients with successful outcomes.


2021 ◽  
Vol 103-B (12) ◽  
pp. 1774-1782
Author(s):  
Hiren M. Divecha ◽  
Terence W. O'Neill ◽  
Mark Lunt ◽  
Tim N. Board

Aims The aim of this study was to determine if uncemented acetabular polyethylene (PE) liner geometry, and lip size, influenced the risk of revision for instability or loosening. Methods A total of 202,511 primary total hip arthroplasties (THAs) with uncemented acetabular components were identified from the National Joint Registry (NJR) dataset between 2003 and 2017. The effect of liner geometry on the risk of revision for instability or loosening was investigated using competing risk regression analyses adjusting for age, sex, American Society of Anesthesiologists grade, indication, side, institution type, surgeon grade, surgical approach, head size, and polyethylene crosslinking. Stratified analyses by surgical approach were performed, including pairwise comparisons of liner geometries. Results The distribution of liner geometries were neutral (39.4%; 79,822), 10° (34.5%; 69,894), 15° (21.6%; 43,722), offset reorientating (2.8%; 5705), offset neutral (0.9%; 1,767), and 20° (0.8%; 1,601). There were 690 (0.34%) revisions for instability. Compared to neutral liners, the adjusted subhazard ratios of revision for instability were: 10°, 0.64 (p < 0.001); 15°, 0.48 (p < 0.001); and offset reorientating, 1.6 (p = 0.010). No association was found with other geometries. 10° and 15° liners had a time-dependent lower risk of revision for instability within the first 1.2 years. In posterior approaches, 10° and 15° liners had a lower risk of revision for instability, with no significant difference between them. The protective effect of lipped over neutral liners was not observed in laterally approached THAs. There were 604 (0.3%) revisions for loosening, but no association between liner geometry and revision for loosening was found. Conclusion This registry-based study confirms a lower risk of revision for instability in posterior approach THAs with 10° or 15° lipped liners compared to neutral liners, but no significant difference between these lip sizes. A higher revision risk is seen with offset reorientating liners. The benefit of lipped geometries against revision for instability was not seen in laterally approached THAs. Liner geometry does not seem to influence the risk of revision for loosening. Cite this article: Bone Joint J 2021;103-B(12):1774–1782.


2021 ◽  
Author(s):  
Hasan R Mohammad ◽  
Andrew Judge ◽  
David W Murray

Background and purpose — Total knee replacement (TKR) can be implanted with or without bone cement. It is currently unknown how the functional outcomes compare. Therefore, we compared the patient-reported outcome measures (PROMS) of both fixation methods. Patients and methods — We performed a propensitymatched comparison of 14,404 TKRs (7,202 cemented and 7,202 cementless) enrolled in the National Joint Registry and the English National PROMs collection programme. Subgroup analyses were performed in different age groups (1) < 55 years; (2) 55–64 years; (3) 65–74 years; (4) ≥ 75 years. Results — The 6-month postoperative Oxford Knee Score (OKS) was significantly (p < 0.001) higher for cemented TKR (35, SD 9.7) than cementless TKR (34, SD 9.9). The OKS was also significantly higher for the cemented TKR in all age groups, except the 55–64-year group. A significantly higher proportion of cemented TKRs had an excellent OKS (≥ 41) compared with cementless (32% vs. 28%, p < 0.001) and a lower proportion of poor (< 27) scores (19% vs. 22%, p = 0.001). This was also observed for all age subgroups. Therewere no significant differences in EQ-5D points gained postoperatively between the groups respectively (0.31 vs. 0.30, p = 0.1). Interpretation — Cemented TKRs had a greater proportion of excellent OKS scores and lower proportion of poor scores both overall and across all age groups. However, the absolute differences are small and below the minimally clinically important difference, making both fixation types acceptable. Currently the vast majority of TKRs are cemented and the results from this study suggest that this isappropriate.


2021 ◽  
Vol 103-B (11) ◽  
pp. 1669-1677
Author(s):  
Hiren M. Divecha ◽  
Terence W. O'Neill ◽  
Mark Lunt ◽  
Tim N. Board

Aims To determine if primary cemented acetabular component geometry (long posterior wall (LPW), hooded, or offset reorientating) influences the risk of revision total hip arthroplasty (THA) for instability or loosening. Methods The National Joint Registry (NJR) dataset was analyzed for primary THAs performed between 2003 and 2017. A cohort of 224,874 cemented acetabular components were included. The effect of acetabular component geometry on the risk of revision for instability or for loosening was investigated using log-binomial regression adjusting for age, sex, American Society of Anesthesiologists grade, indication, side, institution type, operating surgeon grade, surgical approach, polyethylene crosslinking, and prosthetic head size. A competing risk survival analysis was performed with the competing risks being revision for other indications or death. Results The distribution of acetabular component geometries was: LPW 81.2%; hooded 18.7%; and offset reorientating 0.1%. There were 3,313 (1.5%) revision THAs performed, of which 815 (0.4%) were for instability and 838 (0.4%) were for loosening. Compared to the LPW group, the adjusted subhazard ratio of revision for instability in the hooded group was 2.31 (p < 0.001) and 4.12 (p = 0.047) in the offset reorientating group. Likewise, the subhazard ratio of revision for loosening was 2.65 (p < 0.001) in the hooded group and 13.61 (p < 0.001) in the offset reorientating group. A time-varying subhazard ratio of revision for instability (hooded vs LPW) was found, being greatest within the first three months. Conclusion This registry-based study confirms a significantly higher risk of revision after cemented THA for instability and for loosening when a hooded or offset reorientating acetabular component is used, compared to a LPW component. Further research is required to clarify if certain patients benefit from the use of hooded or offset reorientating components, but we recommend caution when using such components in routine clinical practice. Cite this article: Bone Joint J 2021;103-B(11):1669–1677.


2021 ◽  
Vol 2 (11) ◽  
pp. 921-925
Author(s):  
Afton K. Limberg ◽  
Cody C. Wyles ◽  
Michael J. Taunton ◽  
Arlen D. Hanssen ◽  
Mark W. Pagnano ◽  
...  

Aims Varus-valgus constrained (VVC) devices are typically used in revision settings, often with stems to mitigate the risk of aseptic loosening. However, in at least one system, the VVC insert is compatible with the primary posterior-stabilized (PS) femoral component, which may be an option in complex primary situations. We sought to determine the implant survivorship, radiological and clinical outcomes, and complications when this VVC insert was coupled with a PS femur without stems in complex primary total knee arthroplasties (TKAs). Methods Through our institution’s total joint registry, we identified 113 primary TKAs (103 patients) performed between 2007 and 2017 in which a VVC insert was coupled with a standard cemented PS femur without stems. Mean age was 68 years (SD 10), mean BMI was 32 kg/m2 (SD 7), and 59 patients (50%) were male. Mean follow-up was four years (2 to 10). Results The five-year survivorship free from aseptic loosening was 100%. The five-year survivorship free from any revision was 99%, with the only revision performed for infection. The five-year survivorship free from reoperation was 93%. The most common reoperation was treatment for infection (n = 4; 4%), followed by manipulation under anaesthesia (MUA; n = 2; 2%). Survivorship free from any complication at five years was 90%, with superficial wound infection as the most frequent (n = 4; 4%). At most recent follow-up, two TKAs had non-progressive radiolucent lines about both the tibial and femoral components. Knee Society Scores improved from 53 preoperatively to 88 at latest follow-up (p < 0.001). Conclusion For complex primary TKA in occasional situations, coupling a VVC insert with a standard PS femur without stems proved reliable and durable at five years. Longer-term follow-up is required before recommending this technique more broadly. Cite this article: Bone Jt Open 2021;2(11):921–925.


2021 ◽  
Vol 103-B (9) ◽  
pp. 1479-1487
Author(s):  
Edward T. Davis ◽  
Joseph Pagkalos ◽  
Branko Kopjar

Aims The aim of our study was to investigate the effect of asymmetric crosslinked polyethylene liner use on the risk of revision of cementless and hybrid total hip arthroplasties (THAs). Methods We undertook a registry study combining the National Joint Registry dataset with polyethylene manufacturing characteristics as supplied by the manufacturers. The primary endpoint was revision for any reason. We performed further analyses on other reasons including instability, aseptic loosening, wear, and liner dissociation. The primary analytic approach was Cox proportional hazard regression. Results A total of 213,146 THAs were included in the analysis. Overall, 2,997 revisions were recorded, 1,569 in THAs with a flat liner and 1,428 in THAs using an asymmetric liner. Flat liner THAs had a higher risk of revision for any reason than asymmetric liner THAs when implanted through a Hardinge/anterolateral approach (hazard ratio (HR) 1.169, 95% confidence interval (CI) 1.022 to 1.337) and through a posterior approach (HR 1.122, 95% CI 1.108 to 1.346). There was no increased risk of revision for aseptic loosening when asymmetric liners were used for any surgical approach. A separate analysis of the three most frequently used crosslinked polyethylene liners was in agreement with this finding. When analyzing THAs with flat liners only, THAs implanted through a Hardinge/anterolateral approach were associated with a reduced risk of revision for instability compared to posterior approach THAs (HR 0.561 (95% CI 0.446 to 0.706)). When analyzing THAs with an asymmetric liner, there was no significant difference in the risk of revision for instability between the two approaches (HR 0.838 (95% CI 0.633 to 1.110)). Conclusion For THAs implanted through the posterior approach, the use of asymmetric liners reduces the risk of revision for instability and revision for any reason. In THAs implanted through a Hardinge/anterolateral approach, the use of an asymmetric liner was associated with a reduced risk of revision. The effect on revision for instability was less pronounced than in the posterior approach. Cite this article: Bone Joint J 2021;103-B(9):1479–1487.