Comparison of Image Quality and Mean Absorbed Dose to the Breast for Two Mammographic Films

1993 ◽  
Vol 34 (4) ◽  
pp. 351-355 ◽  
Author(s):  
J. Persliden ◽  
V. Fransson ◽  
B. Vitak ◽  
G. Fagerberg

A film intended for mammography (Kodak SO 155 MRH-1) was tested and compared to one (Kodak SO 177 Ortho M) used earlier. Both films including the cassette could resolve 20 lp/mm. For a processing time of 90 s the new film gave the same overall image quality and irradiation dose to the breast as the old system. If, instead, the processing time was increased to 150 s, a 43% reduction in kerma could be attained with the new film. With a developing temperature of 36°C and 150 s processing time, the noise is clinically acceptable. A nonparametric test showed no significant difference between the 2 films on the 0.01 level. At an X-ray tube potential difference of 25 kV, the mean absorbed dose to a 4.5-cm-thick breast was reduced from 1.7 mGy with the old combination to 1.0 mGy with the new one. The measurements were made with a moving grid.

Author(s):  
A H Pakravan ◽  
S M R Aghamiri ◽  
T Bamdadian ◽  
M Gholami ◽  
M Moshfeghi

Background: Panoramic imaging is one of the most common imaging methods in dentistry. Regarding the side-effects of ionizing radiation, it is necessary to survey different aspects and details of panoramic imaging. In this study, we compared the absorbed x-ray dose around two panoramic x-ray units: PM 2002 CC Proline (Planmeca, Helsinki, Finland) and Cranex Tome (Soredex, Helsinki, Finland).Materials and Methods: In order to estimate x-ray dose, 15 thermoluminescet dosemeters (TLD-100) were placed in 3 semi-circles of 40cm, 80cm and 120cm radii. Around each unit, the number of TLDs in each semi-circle was 5 with equal intervals. The center of semicircles accords with the patient’s position. Each TLD was exposed 40 times. These dosemeters were read out with a Harshaw Model 4000 TLD Reader (USA). The calibration processing and the reading of dosemeters were performed by the Atomic Energy Organization of Iran.Results: The mean absorbed dose in three lines of PM 2002 CC Proline was 123.2±15.1, 118.0±11.0 and 108.0±9.1 µSv, (p=0.013). The results were 140.4±15.2, 120.2±10.4 and 111.6±11.2 µSv in Cranex Tome (p=0.208), which reveals no significant difference between two systems.Conclusion: There are no significant differences between the mean absorbed dose of surveyed models in panoramic imaging by two units (PM 2002 CC Proline and Cranex Tome). These results were less than occupational exposure recommended by ICRP, even at the highest calculated doses.


2010 ◽  
Vol 51 (3) ◽  
pp. 260-270 ◽  
Author(s):  
Peter Björkdahl ◽  
Ulf Nyman

Background: Concern has been raised regarding the mounting collective radiation doses from computed tomography (CT), increasing the risk of radiation-induced cancers in exposed populations. Purpose: To compare radiation dose and image quality in a chest phantom and in patients for the diagnosis of pulmonary embolism (PE) at 100 and 120 peak kilovoltage (kVp) using 16-multichannel detector computed tomography (MDCT). Material and Methods: A 20-ml syringe containing 12 mg I/ml was scanned in a chest phantom at 100/120 kVp and 25 milliampere seconds (mAs). Consecutive patients underwent 100 kVp ( n = 50) and 120 kVp ( n = 50) 16-MDCT using a “quality reference” effective mAs of 100, 300 mg I/kg, and a 12-s injection duration. Attenuation (CT number), image noise (1 standard deviation), and contrast-to-noise ratio (CNR; fresh clot = 70 HU) of the contrast medium syringe and pulmonary arteries were evaluated on 3-mm-thick slices. Subjective image quality was assessed. Computed tomography dose index (CTDIvol) and dose–length product (DLP) were presented by the CT software, and effective dose was estimated. Results: Mean values in the chest phantom and patients changed as follows when X-ray tube potential decreased from 120 to 100 kVp: attenuation +23% and +40%, noise +38% and +48%, CNR −6% and 0%, and CTDIvol −38% and −40%, respectively. Mean DLP and effective dose in the patients decreased by 42% and 45%, respectively. Subjective image quality was excellent or adequate in 49/48 patients at 100/120 kVp. No patient with a negative CT had any thromboembolism diagnosed during 3-month follow-up. Conclusion: By reducing X-ray tube potential from 120 to 100 kVp, while keeping all other scanning parameters unchanged, the radiation dose to the patient may be almost halved without deterioration of diagnostic quality, which may be of particular benefit in young individuals.


1997 ◽  
Vol 38 (6) ◽  
pp. 1010-1014 ◽  
Author(s):  
J. Persliden ◽  
P. Larsson ◽  
B. Norén ◽  
S. Wirell

Purpose: Image quality and the absorbed dose to the patient are issues of primary interest in the change-over from the conventional analogue technique to the digital technique in the examination of the colon by means of fluoroscopy. the aim of this study was to compare the incident radiation and to evaluate the image quality in two different X-ray equipment types, one digital and one analogue Material and Methods: A kerma-area product meter was used to measure the incident radiation to the patient. Both fluoroscopy and total-examination times were measured as was the number of images. an evaluation of image quality was made and statistically analysed Results and Conclusion: No significant difference in the irradiation dose was observed between the two techniques. the fluoroscopy time was significantly lower with the conventional technique but the total-examination time decreased by 18% with the digital technique. the total number of images taken was higher with the digital technique (25 images compared to 19) owing to the limited field of the image intensifier. Significantly more noise and less sharpness were observed with the digital system but there was no significant difference in contrast or image quality in the various anatomical structures. Although the change-over to the digital system produced a reduction in sharpness and an increase in noise, and no significant dose saving was measured, the digital system was faster to work with and could well be used for diagnostic purposes


1986 ◽  
Vol 27 (4) ◽  
pp. 467-479 ◽  
Author(s):  
B. Nielsen ◽  
G. Fagerberg

Some of the parameters determining image quality in mammography are analyzed: the effects of primary photon spectra, focal spot size and screen-film systems on spatial resolution are discussed as are scattered radiation, development temperature and absorbed dose. The parameters limiting spatial resolution and contrast are evaluated for the standard and magnification techniques. Methods of reducing scattered radiation to improve contrast are evaluated. Scatter to primary ratios for different scatter reducing methods are compared, using the physical quantity energy imparted. For the standard technique the spatial resolution has been found to be limited by the fluorescent screen. With magnification technique the focal spot is the weakest link for the spatial resolution. The contrast is mainly set by the amount of scatter using the standard technique considering the use of a low tube potential (∼25 kVp). Using the magnification technique the amount of scatter is so small, that the tube potential is the limiting factor. We have found the optimized standard mammographic technique to be achieved under the following conditions: 25 kVp, 0.3 to 0.6 mm focal spot, film-focus distance 500 mm, anti-scatter grid, developing temperature 36 to 38°C and 4 minutes total processing time with the screen-film system we have used. In magnification technique an air gap of at least 20 mm is desired. With an FFD of about 500 mm this will give a magnification ratio of 1.8 to 2.0 and a 0.1 mm × 0.1 mm focus spot is mandatory. With this technique, it is necessary to use a faster screen-film system than that used in standard mammography.


2020 ◽  
Vol 2020 ◽  
pp. 1-8
Author(s):  
Aobo Zhang ◽  
Qing Han ◽  
Bingpeng Chen ◽  
Chenyu Wang ◽  
Xue Zhao ◽  
...  

Background and purpose. A new method of three-dimensional (3D) reconstruction technology was used to take place of X-ray in measuring valgus correction angle (VCA) of both DDH patients and normal volunteers to improve precision. Two different ways to define VCA according to the various conditions of patients were compared and analyzed. Methods. Bilateral VCA of 50 DDH patients and 56 normal volunteers were measured by Mimics software in the 3D method and X-ray in 2D. Two VCA (the upper VCA and the lower VCA) were measured in both two methods. Every VCA was measured by observer A and observer B for twice separately. The statistical analyses of the differences were calculated among the measurements of the VCA. Results. The mean value of the upper VCA measured in 3D was 4.95°±0.76° in DDH group and 5.56°±0.62° in the normal group with significant difference (t=−6.457, p<0.01). The VCA of DDH group and normal group measured by 3D was larger than 2D, both the upper VCA and the lower VCA. The differences indicated statistically significant. The mean value of lower VCA was 0.60° smaller than the mean value of upper VCA in normal volunteers. The mean value of the lower VCA was 0.58° larger than the mean value of the upper VCA in DDH patients. Conclusions. Compared to X-ray, 3D reconstruction technology is more accurate without conventional limitations. The lower VCA of DDH patients should be regarded as the femoral intramedullary guide angle in TKA, especially for patients with femoral deformities.


2020 ◽  
Vol 14 (04) ◽  
pp. 651-656
Author(s):  
Hend Mahmoud Abdel-Azem ◽  
Ahmed Fawzy Abo Elezz ◽  
Rehab Khalil Safy

Abstract Objective This study aimed to investigate the efficacy of the Galla chinensis extract (GCE) on early dentin lesion remineralization. Materials and Methods Seventy-two dentin specimens were divided into three groups according to the treatment solution. In group S1, specimens were treated with GCE; in group S2, the specimens were treated with sodium fluoride (NaF); meanwhile, specimens of group S3 were treated with distilled water (DW). Each group was further subdivided into two subgroups according to the treatment time (1 minute and 5 minutes). An in vitro pH-cycling model for 12 days was done. Subsequently, surface microhardness (SMH) of the specimens, elemental analysis, and their micromorphological appearance were evaluated. Statistical Analysis Data were statistically analyzed. One-way analysis of variance was used to compare numerical (parametric) data between more than two separate groups followed by post hoc Tukey. Results There was no significant difference between the mean SMH of dentin between NaF and GCE groups. Regarding the time, the 5-minute treatment with NaF and DW groups recorded higher mean SMH value of dentin than the 1-minute treatment group. Meanwhile, for GCE groups, the 1-minute immersion recorded higher mean SMH value than the 5-minute immersion without any significant difference between them. The microhardness results were confirmed by environmental scanning electron microscope and energy dispersive X-ray analysis results. Conclusion GCE could be used as an effective alternative for dentin remineralization.


1977 ◽  
Vol 4 (1) ◽  
pp. 261-261
Author(s):  
R.J. Dickens ◽  
C. Moss ◽  
J.A. Dawe ◽  
B. Peterson

31 redshifts have been obtained for A1367, 34 for A262 and 61 for the Centaurus cluster (HMS 1247-4102). Full details of this work have been and will be presented in M.N.R.A.S. Both A1367 and A262 are spiral-rich in Oemler’s classification, while Centaurus is intermediate in type between spiral-rich and poor. In all 3 cases, the distributions of E and L galaxies are centrally concentrated, whilst the spirals are distributed in a more extensive and ragged fashion. The mean corrected redshifts and velocity dispersions for 2 morphological subsets are given in Table I. There is no significant difference in the mean velocities of the 2 subsets for each cluster. However, the velocity dispersions for the spirals are significantly greater than those for the E,L galaxies in both A1367 and A262. The differences in velocity dispersions of the 2 subsets for Centaurus are not statistically significant. Gott and Gunn have suggested that irregular clusters of the sort presently described have not undergone collapse. However, all 3 clusters show morphological separation and are x-ray sources. This is consistent with a collapsed core surrounded by a shell of infailing spirals.


Sign in / Sign up

Export Citation Format

Share Document