scholarly journals Metacarpal shaft fixation: a biomechanical comparison of dorsal plating, lag screws, and headless compression screws

2021 ◽  
Vol 22 (1) ◽  
Author(s):  
Felix G. E. Dyrna ◽  
Daniel M. Avery ◽  
Ryu Yoshida ◽  
David Lam ◽  
Simon Oeckenpöhler ◽  
...  

Abstract Background Metacarpal shaft fractures are common and can be treated nonoperatively. Shortening, angulation, and rotational deformity are indications for surgical treatment. Various forms of treatment with advantages and disadvantages have been documented. The purpose of the study was to determine the stability of fracture fixation with intramedullary headless compression screws in two types of metacarpal shaft fractures and compare them to other common forms of rigid fixation: dorsal plating and lag screw fixation. It was hypothesized that headless compression screws would demonstrate a biomechanical stronger construct. Methods Five matched paired hands (age 60.9 ± 4.6 years), utilizing non-thumb metacarpals, were used for comparative fixation in two fracture types created by an osteotomy. In transverse diaphyseal fractures, fixation by headless compression screws (n = 7) and plating (n = 8) were compared. In long oblique diaphyseal fractures, headless compression screws (n = 8) were compared with plating (n = 8) and lag screws (n = 7). Testing was performed using an MTS frame producing an apex dorsal, three point bending force. Peak load to failure and stiffness were calculated from the load-displacement curve generated. Results For transverse fractures, headless compression screws had a significantly higher stiffness and peak load to failure, means 249.4 N/mm and 584.8 N, than plates, means 129.02 N/mm and 303.9 N (both p < 0.001). For long oblique fractures, stiffness and peak load to failure for headless compression screws were means 209 N/mm and 758.4 N, for plates 258.7 N/mm and 518.5 N, and for lag screws 172.18 N/mm and 234.11 N. There was significance in peak load to failure for headless compression screws vs plates (p = 0.023), headless compression screws vs lag screws (p < 0.001), and plates vs lag screws (p = 0.009). There was no significant difference in stiffness between groups. Conclusion Intramedullary fixation of diaphyseal metacarpal fractures with a headless compression screw provides excellent biomechanical stability. Coupled with lower risks for adverse effects, headless compression screws may be a preferable option for those requiring rapid return to sport or work. Level of evidence Basic Science Study, Biomechanics.

2021 ◽  
pp. 175319342110177
Author(s):  
John G. Galbraith ◽  
Lachlan S. Huntington ◽  
Paul Borbas ◽  
David C. Ackland ◽  
Stephen K. Tham ◽  
...  

We compared four methods of metacarpal shaft fixation: 2.2 mm intramedullary headless compression screw; 3.0 mm intramedullary headless compression screw; intramedullary K-wire fixation; and dorsal plate fixation. Transverse mid-diaphyseal fractures were created in 64 metacarpal sawbones and were assigned into four groups. Peak load to failure and stiffness were measured in cantilever bending and torsion. We found that dorsal plating had the highest peak load to failure. However, initial bending stiffness of the 3.0 mm intramedullary headless compression screw was higher than that of the dorsal plates. In torsion testing, dorsal plating had the highest peak torque, but there was no significant difference in torsional stiffness between the plate and intramedullary headless compression screw constructs. We concluded that intramedullary headless compression screw fixation is biomechanically superior to K-wires in cantilever bending and torsion; however, it is less stable than dorsal plating. In our study, the initial stability provided by K-wire fixation was sufficient to cope with expected loads in the early rehabilitation period, whereas dorsal plates and IHCS constructs provided stability far in excess of what is required.


Hand ◽  
2021 ◽  
pp. 155894472097411
Author(s):  
Luke T. Nicholson ◽  
Kristen M. Sochol ◽  
Ali Azad ◽  
Ram Kiran Alluri ◽  
J. Ryan Hill ◽  
...  

Background: Management of scaphoid nonunions with bone loss varies substantially. Commonly, internal fixation consists of a single headless compression screw. Recently, some authors have reported on the theoretical benefits of dual-screw fixation. We hypothesized that using 2 headless compression screws would impart improved stiffness over a single-screw construct. Methods: Using a cadaveric model, we compared biomechanical characteristics of a single tapered 3.5- to 3.6-mm headless compression screw with 2 tapered 2.5- to 2.8-mm headless compression screws in a scaphoid waist nonunion model. The primary outcome measurement was construct stiffness. Secondary outcome measurements included load at 1 and 2 mm of displacement, load to failure for each specimen, and qualitative assessment of mode of failure. Results: Stiffness during load to failure was not significantly different between single- and double-screw configurations ( P = .8). Load to failure demonstrated no statistically significant difference between single- and double-screw configurations. Using a qualitative assessment, the double-screw construct maintained rotational stability more than the single-screw construct ( P = .029). Conclusions: Single- and double-screw fixation constructs in a cadaveric scaphoid nonunion model demonstrate similar construct stiffness, load to failure, and load to 1- and 2-mm displacement. Modes of failure may differ between constructs and represent an area for further study. The theoretical benefit of dual-screw fixation should be weighed against the morphologic limitations to placing 2 screws in a scaphoid nonunion.


Author(s):  
Paul Borbas ◽  
Rafael Loucas ◽  
Marios Loucas ◽  
Maximilian Vetter ◽  
Simon Hofstede ◽  
...  

Abstract Introduction Coronal plane fractures of the distal humerus are relatively rare and can be challenging to treat due to their complexity and intra-articular nature. There is no gold standard for surgical management of these complex fractures. The purpose of this study was to compare the biomechanical stability and strength of two different internal fixation techniques for complex coronal plane fractures of the capitellum with posterior comminution. Materials and methods Fourteen fresh frozen, age- and gender-matched cadaveric elbows were 3D-navigated osteotomized simulating a Dubberley type IIB fracture. Specimens were randomized into one of two treatment groups and stabilized with an anterior antiglide plate with additional anteroposterior cannulated headless compression screws (group antiGP + HCS) or a posterolateral distal humerus locking plate with lateral extension (group PLP). Cyclic testing was performed with 75 N over 2000 cycles and ultimately until construct failure. Data were analyzed for displacement, construct stiffness, and ultimate load to failure. Results There was no significant difference in displacement during 2000 cycles (p = 0.291), stiffness (310 vs. 347 N/mm; p = 0.612) or ultimate load to failure (649 ± 351 vs. 887 ± 187 N; p = 0.140) between the two groups. Conclusions Posterolateral distal humerus locking plate achieves equal biomechanical fixation strength as an anterior antiglide plate with additional anteroposterior cannulated headless compression screws for fracture fixation of complex coronal plane fractures of the capitellum. These results support the use of a posterolateral distal humerus locking plate considering the clinical advantages of less invasive surgery and extraarticular metalware. Level of evidence Biomechanical study.


Hand ◽  
2016 ◽  
Vol 11 (4) ◽  
pp. 421-426 ◽  
Author(s):  
Eitan Melamed ◽  
Richard M. Hinds ◽  
Michael B. Gottschalk ◽  
Oran D. Kennedy ◽  
John T. Capo

Background: Recently, intramedullary headless screw (IMHS) has shown promise as an alternative to dorsal plate fixation of metacarpal fractures. The purpose of this study was to assess the biomechanical performance of IMHS versus plating. We hypothesized that IMHS fixation provides inferior stability to plating. Methods: Metacarpal fracture model with 3-mm of volar gapping in forty-four human cadaveric metacarpals was created. The specimens were divided into 5 groups: Group 1, 1.5-mm non-locking plate; Group 2, 1.5-mm locking plate; Group 3, 2.0-mm non-locking plate; Group 4, 2.0-mm locking plate; and Group 5, 2.4-mm short cannulated IMHS. A 4-point bending model was used to assess load-to failure (LTF) and stiffness. Results: Mean LTF was 364 ± 130 N for 1.5-mm non-locking plates, 218 ± 94 N for 1.5-mm locking plates, 421 ± 86 N for 2.0-mm non-locking plates, 351 ± 71 N for 2.0-mm locking plates, and 75 ± 20 N for IMHS. Mean stiffness was 91 ± 12 N/mm for 1.5-mm non-locking plates, 110 ± 77 N/mm for 1.5-mm locking plates, 94 ± 20 N/mm for 2.0-mm non-locking plates, 135 ± 16 N/mm for 2.0-mm locking plates, and 55 ± 15 N/mm for IMHS. IMHS demonstrated significantly lower LTF and stiffness than plates. Conclusions: IMHS fixation of unstable metacarpal shaft fractures offers less stability compared to plating when loaded in bending. The LTF and stiffness of IMHS versus plating of metacarpal shaft fractures has not been previously quantified. Our results reveal that IMHS fixation is less favorable biomechanically and should be carefully chosen in regards to fracture stability.


2018 ◽  
Vol 39 (7) ◽  
pp. 836-842 ◽  
Author(s):  
Fred T. Finney ◽  
Simon Lee ◽  
Jaron Scott ◽  
Todd A. Irwin ◽  
James R. Holmes ◽  
...  

Background: Lesser toe metatarsal-phalangeal (MTP) joint instability can be a major source of pain and dysfunction. Instability occurs when there is incompetence of the plantar plate and/or collateral ligaments. Newer operative treatments focus on performing anatomic repairs of the plantar plate. The goal of this study was to compare the biomechanical properties of 3 suture configurations that may be used for plantar plate repairs. Methods: Biomechanical analysis of 27 lesser toe plantar plates from fresh frozen human cadavers was completed. The plantar plate was detached from the proximal phalanx, and suture was placed in the distal plantar plate in a horizontal mattress, luggage-tag, or Mason-Allen suture configuration. Cyclic loading followed by load-to-failure testing was performed. Results: There was a significant difference in peak load-to-failure force between constructs (mattress: 115.53 ± 15.95 N; luggage-tag 102.42 ± 19.33 N; Mason-Allen: 89.96 ± 15.78 N; P = .015). Post hoc analysis demonstrated that the mattress configuration had significantly higher load-to-failure force compared with the Mason-Allen configuration ( P = .004). There were no significant differences between the mattress and the luggage-tag configurations or the luggage-tag and the Mason-Allen configurations. There were no differences in construct stiffness, axial displacement at the time of failure, or number of cycles required to produce 2 mm of displacement. Conclusion: The mattress configuration demonstrated better peak load-to-failure force compared with the Mason-Allen configuration but was not statistically different from the luggage-tag configuration. Although not significant, the mattress configuration trended toward higher load-to-failure force compared with the luggage-tag. Clinical Relevance: The horizontal mattress stitch may be the biomechanically superior configuration in plantar plate repairs.


Problems when calculating reinforced concrete structures based on the concrete deformation under compression diagram, which is presented both in Russian and foreign regulatory documents on the design of concrete and reinforced concrete structures are considered. The correctness of their compliance for all classes of concrete remains very approximate, especially a significant difference occurs when using Euronorm due to the different shape and sizes of the samples. At present, there are no methodical recommendations for determining the ultimate relative deformations of concrete under axial compression and the construction of curvilinear deformation diagrams, which leads to limited experimental data and, as a result, does not make it possible to enter more detailed ultimate strain values into domestic standards. The results of experimental studies to determine the ultimate relative deformations of concrete under compression for different classes of concrete, which allowed to make analytical dependences for the evaluation of the ultimate relative deformations and description of curvilinear deformation diagrams, are presented. The article discusses various options for using the deformation model to assess the stress-strain state of the structure, it is concluded that it is necessary to use not only the finite values of the ultimate deformations, but also their intermediate values. This requires reliable diagrams "s–e” for all classes of concrete. The difficulties of measuring deformations in concrete subjected to peak load, corresponding to the prismatic strength, as well as main cracks that appeared under conditions of long-term step loading are highlighted. Variants of more accurate measurements are proposed. Development and implementation of the new standard GOST "Concretes. Methods for determination of complete diagrams" on the basis of the developed method for obtaining complete diagrams of concrete deformation under compression for the evaluation of ultimate deformability of concrete under compression are necessary.


2019 ◽  
Author(s):  
Ripedah Nakana

BACKGROUND According to WHO, It has been reported that there has been a rise in road traffic accidents globally due to many factors. Consequently this has contributed to an increase in fracture of the femur in hospitals. The annual incidence of these Fractures has been estimated to be 10 per 100,000 persons and this incidence has been noticed to be higher among the youths decreasing after the age of 40 and then increasing in the elderly. These fractures are either managed conservatively or operatively. OBJECTIVE the aim of this study was to determine if there is a significant difference in knee joint functional outcome in patients with femur shaft fractures managed operatively and conservatively at Ndola Teaching Hospital at the orthopedics clinic and physiotherapy. This in future might help to place emphasis on government to sensitize the right measures needed to improve on the functional outcome of a knee joint in patients with femur fracture who are treated operatively or conservatively METHODS the method used for the study was a complete enumeration of all patients with femur fractures that were admitted at orthopedic ward, visited the orthopedic clinic and physiotherapy sessions at NTH between the period of July 2019 to September 2019. Data collected will be analyzed using a statistical data software called Statistical Package of Social Sciences (SPSS). The independent variable that will be considered in the study are age, DOI, DOA, femur shaft fractures, treatment and others that that will meet the selection Criteria. The dependent variable will be “Knee Joint Functional Outcome”. RESULTS Results did not show statistically significant difference in knee functional outcome for two methods (conservative or operative). In the current study it was noted that there was much reduction in swelling in the operative in comparison to the conservative group. We may therefore conclude that the operative management promises a better outcome especially when done on time. However it seemed the operative procedure was not preferred for any conservative approach however the study has shown an in depth feasibility study to affirm such findings in order to inform on policy or management paradigim shift. CONCLUSIONS There is need for further intervention in the management of femur fracture which complicates to poor knee joint function


2021 ◽  
Vol 9 (3) ◽  
pp. 232596712198928
Author(s):  
Heath P. Gould ◽  
Nicholas R. Delaney ◽  
Brent G. Parks ◽  
Roshan T. Melvani ◽  
Richard Y. Hinton

Background: Femoral-sided graft fixation in medial patellofemoral ligament (MPFL) reconstruction is commonly performed using an interference screw (IS). However, the IS method is associated with several clinical disadvantages that may be ameliorated by the use of suture anchors (SAs) for femoral fixation. Purpose: To compare the load to failure and stiffness of SAs versus an IS for the femoral fixation of a semitendinosus autograft in MPFL reconstruction. Study Design: Controlled laboratory study. Methods: Based on a priori power analysis, a total of 6 matched pairs of cadaveric knees were included. Specimens in each pair were randomly assigned to receive either SA or IS fixation. After an appropriate reconstruction procedure, the looped end of the MPFL graft was pulled laterally at a rate of 6 mm/s until construct failure. The best-fit slope of the load-displacement curve was then used to calculate the stiffness (N/mm) in a post hoc fashion. A paired t test was used to compare the mean load to failure and the mean stiffness between groups. Results: No significant difference in load to failure was observed between the IS and the SA fixation groups (294.0 ± 61.1 vs 250.0 ± 55.9; P = .352), although the mean stiffness was significantly higher in IS specimens (34.5 ± 9.6 vs 14.7 ± 1.2; P = .004). All IS reconstructions failed by graft pullout from the femoral tunnel, whereas 5 of the 6 SA reconstructions failed by anchor pullout. Conclusion: In this biomechanical study using a cadaveric model of MPFL reconstruction, SA femoral fixation was not significantly different from IS fixation in terms of load to failure. The mean load-to-failure values for both reconstruction techniques were greater than the literature-reported values for the native MPFL. Clinical Relevance: These results suggest that SAs are a biomechanically viable alternative for femoral-sided graft fixation in MPFL reconstruction.


2021 ◽  
Vol 10 (5) ◽  
pp. 995
Author(s):  
Marja Perhomaa ◽  
Tytti Pokka ◽  
Linda Korhonen ◽  
Antti Kyrö ◽  
Jaakko Niinimäki ◽  
...  

The preferred surgical fixation of forearm shaft fractures in children is Elastic Stable Intramedullary Nailing (ESIN). Due to known disadvantageous effects of metal implants, a new surgical method using biodegradable polylactide-co-glycolide (PLGA) intramedullary nails has been developed but its long-term outcomes are unclear. The aim of this study was to compare the long-term outcomes of Biodegradable Intramedullary Nailing (BIN) to ESIN and assess the biodegradation of the study implants via magnetic resonance imaging (MRI). The study population of the prospective, randomized trial consisted of paediatric patients whose forearm shaft fractures were treated with BIN (n = 19) or ESIN (n = 16). Forearm rotation at minimally four years’ follow-up was the main outcome. There was no clinically significant difference in the recovery of the patients treated with the BIN as compared to those treated with the ESIN. More than half of the implants (57.7%, n = 15/26) were completely degraded, and the rest were degraded almost completely. The PLGA intramedullary nails used in the treatment of forearm shaft fractures in this study resulted in good function and anatomy. No unexpected disadvantages were found in the degradation of the implants. However, two implant failures had occurred in three months postoperatively.


2021 ◽  
Vol 12 ◽  
pp. 215145932199274
Author(s):  
Hyojune Kim ◽  
Myung Jin Shin ◽  
Erica Kholinne ◽  
Janghyeon Seo ◽  
Duckwoo Ahn ◽  
...  

Purpose: This biomechanical study investigates the optimal number of proximal screws for stable fixation of a 2-part proximal humerus fracture model with a locking plate. Methods: Twenty-four proximal humerus fracture models were included in the study. An unstable 2-part fracture was created and fixed by a locking plate. Cyclic loading and load-to-failure tests were used for the following 4 groups based on the number of screws used: 4-screw, 6-screw, 7-screw, and 9-screw groups. Interfragmentary gaps were measured following cyclic loading and compared. Consequently, the load to failure, maximum displacement, stiffness, and mode of failure at failure point were compared. Results: The interfragmentary gaps for the 4-screw, 6-screw, 7-screw, and 9-screw groups were significantly reduced by 0.24 ± 0.09 mm, 0.08 ± 0.06 mm, 0.05 ± 0.01 mm, and 0.03 ± 0.01 mm following 1000 cyclic loading, respectively. The loads to failure were significantly different between the groups with the 7-screw group showing the highest load to failure. The stiffness of the 7-screw group was superior compared with the 6-screw, 9-screw, and 4-screw groups. The maximum displacement before failure showed a significant difference between the comparative groups with the 4-screw group having the lowest value. The 7-screw group had the least structural failure rate (33.3%). Conclusion: At least 7 screws would be optimal for proximal fragment fixation of proximal humerus fractures with medial comminution to minimize secondary varus collapse or fixation failure. Level of Evidence: Basic science study.