Hip Instability Leading to Metallosis, Polyethylene Wear, and Failure of an Oxinium Femoral Head: A Case Report and Review of Literature

Author(s):  
Ahmed Nageeb Mahmoud ◽  
Timour F. El-Husseini ◽  
Wael S. Osman ◽  
Ali M. Maziad

AbstractOxinium is a zirconium metal alloy with an oxidized (ceramicized) surface that aims to mimic the superior wear properties of a ceramic femoral head while maintaining the high strength of metal-bearing surfaces. There are several reports that illustrate the wear and/or failure of Oxinium femoral heads following abnormal head contact with metal cup, with only a few studies reporting on the occurrence of metallosis. We present a case of metallosis and failure of Oxinium femoral head following two incidents of hip instability and one closed reduction. Successful revision was performed with cemented cup and metal femoral head, with 68 months of follow-up. We aimed to highlight the importance of achieving stable and well-oriented hip components when using Oxinium-bearing surfaces and to present a short literature review regarding the wear of oxidized zirconium-bearing surfaces in hip arthroplasty.

2012 ◽  
Vol 529-530 ◽  
pp. 279-284 ◽  
Author(s):  
Taishi Sato ◽  
Yasuharu Nakashima ◽  
Mio Akiyama ◽  
Takuaki Yamamoto ◽  
Taro Mawatari ◽  
...  

The purpose of this study was to examine the effects of ceramic femoral head material on the wear of annealed, crosslinked ultra-high molecular weight polyethylene (UHMWPE) (XLPE) in total hip arthroplasty compared to non-crosslinked conventional UHMWPE (CPE). XLPE was fabricated by crosslinking with 60 kGy irradiation and annealing. Femoral heads made from zirconia and alumina ceramics, and cobalt-chrome (CoCr) of 22 mm or 26 mm diameter were used. In this study, the femoral head penetration into the cup was measured digitally on radiographs of 70 hips with XLPE and 50 hips with CPE. The average follow-up periods were 6.1 and 12.7 years, respectively. The steady wear rate of XLPE was significantly lower than those of CPE (0.002 versus 0.08 mm/year, respectively). Zirconia displayed increased wear rates compared to alumina in CPE; however, there was no difference among head materials in XLPE (0.0028, 0.011 and 0.009 mm/year for zirconia, alumina and CoCr, respectively). Neither head size or implantation period impacted XLPE wear. In contrast to CPE, XLPE displayed low wear rates surpassing the effects of varying femoral head material, size, implantation period and patient demographics.


2019 ◽  
Vol 101-B (6_Supple_B) ◽  
pp. 57-61 ◽  
Author(s):  
B. P. Chalmers ◽  
D. G. Mangold ◽  
A. D. Hanssen ◽  
M. W. Pagnano ◽  
R. T. Trousdale ◽  
...  

Aims Modular dual-mobility constructs reduce the risk of dislocation after revision total hip arthroplasty (THA). However, questions about metal ions from the cobalt-chromium (CoCr) liner persist, and are particularly germane to patients being revised for adverse local tissue reactions (ALTR) to metal. We determined the early- to mid-term serum Co and Cr levels after modular dual-mobility components were used in revision and complex primary THAs, and specifically included patients revised for ALTR. Patients and Methods Serum Co and Cr levels were measured prospectively in 24 patients with a modular dual-mobility construct and a ceramic femoral head. Patients with CoCr heads or contralateral THAs with CoCr heads were excluded. The mean age was 63 years (35 to 83), with 13 patients (54%) being female. The mean follow-up was four years (2 to 7). Indications for modular dual-mobility were prosthetic joint infection treated with two-stage exchange and subsequent reimplantation (n = 8), ALTR revision (n = 7), complex primary THA (n = 7), recurrent instability (n = 1), and periprosthetic femoral fracture (n = 1). The mean preoperative Co and Cr in patients revised for an ALTR were 29.7 μg/l (2 to 146) and 21.5 μg/l (1 to 113), respectively. Results Mean Co and Cr levels were 0.30 μg/l and 0.76 μg/l, respectively, at the most recent follow-up. No patient had a Co level ≥ 1 μg/l. Only one patient had a Cr level ≥ 1 μg/l. That patient’s Cr level was 12 μg/l at 57 months after revision THA for ALTR (and decreased ten-fold from a preoperative Cr of 113 μg/l). Conclusion At a mean of four years, no patient with a modular dual-mobility construct and ceramic femoral head had elevated Co levels, including seven patients revised specifically for ALTR. While further studies are required, we support the selective use of a modular dual-mobility construct in revision and complex primary THAs for patients at high risk for instability. Cite this article: Bone Joint J 2019;101-B(6 Supple B):57–61.


2016 ◽  
Vol 10 (1) ◽  
pp. 779-784 ◽  
Author(s):  
Andreas Panagopoulos ◽  
Irini Tatani ◽  
Panagiotis Megas

Background: Although ceramic head fracture occurs infrequently today, in the event of a fracture, the resulting revision surgery can prove very challenging, since the ceramic particles lodge into the surrounding soft tissue and can cause rapid implant failure Case Presentation: A case of long term survivorship of a severed notched femoral stem after replacing the fractured femoral head with a cobalt-chromium one is reported in a 40-year old woman with hip dysplasia who underwent an uncomplicated total hip arthroplasty. The incident of ceramic femoral head fracture occurred 14 months postoperatively without reporting any significant trauma. Intraoperative findings at revision were a multifragmented femoral head and a damaged polyethylene insert along with diffuse metallosis and excessive wear of the cone of the stem. Both the stem and the acetabular component were stable. After removal of ceramic fragments, metallotic tissue excision and careful lavage of the joint, the inlay was replaced by a similar one and a cobalt-chromium femoral head was placed to the existing notched taper of the firmly incorporated stem. At the 13th year follow up examination, the patient had no pain, used no walking aids, and had normal activity with no signs of wearing or loosening in the plain x-rays. Conclusion: Despite current recommendations of using ceramic femoral heads in cases of fracture or to revise the severely damaged stems we were able to provide a long term survivorship up to 13 years postoperatively of a cobalt-chromium femoral head applied to a severe damaged stem.


2015 ◽  
Vol 2015 ◽  
pp. 1-3 ◽  
Author(s):  
H. Dortaj ◽  
A. Emamifar

Dislocation of the hip is a critical injury that results from high-energy trauma. This paper describes a case of posterior dislocation of the right hip in a 35-year-old woman with associated ipsilateral femoral head fracture. Initial treatment included reduction of the right hip through posterior approach and fixation of the femoral head fracture with three absorbable screws. After 15-month follow-up, a full range of motion has been achieved and there are no signs of avascular necrosis, hip instability, or limping. The authors describe their method of surgery.


2018 ◽  
Vol 100-B (12) ◽  
pp. 1592-1599 ◽  
Author(s):  
V. P. Galea ◽  
J. W. Connelly ◽  
B. Shareghi ◽  
J. Kärrholm ◽  
O. Sköldenberg ◽  
...  

Aims The primary aim of this study was to compare the wear properties of vitamin E-diffused, highly crosslinked polyethylene (VEPE) and one formulation of moderately crosslinked and mechanically annealed ultra-high molecular weight polyethylene (ModXLPE) in patients five years after primary total hip arthroplasty (THA). The secondary aim was to assess the clinical results of patients treated with VEPE by evaluating patient-reported outcome measures (PROMs), radiological evidence of fixation, and the incidence of mechanical failure. Patients and Methods A total of 208 patients (221 THAs) from four international centres were recruited into a prospective study involving radiostereometric analysis (RSA) and the assessment of clinical outcomes. A total of 193 hips (87%) were reviewed at the five-year follow-up. Of these, 136 (70%) received VEPE (vs ModXLPE) liners and 68 (35%) received ceramic (vs metal) femoral heads. PROMs and radiographs were collected preoperatively and at one, two, and five years postoperatively. In addition, RSA images were collected to measure PE wear postoperatively and at one, two, and five years after surgery. Results We observed similar bedding in one year postoperatively and wear two years postoperatively between the two types of liner. However, there was significantly more penetration of the femoral head in the ModXLPE cohort compared with the VEPE cohort five years postoperatively (p < 0.001). The only variables independently predictive of increased wear were ModXLPE (vs VEPE) liner type (β = 0.22, p = 0.010) and metal (vs ceramic) femoral head (β = 0.21, p = 0.013). There was no association between increased wear and the development of radiolucency (p = 0.866) or PROMs. No patient had evidence of osteolysis. Conclusion Five years postoperatively, patients with VEPE (vs ModXLPE) and ceramic (vs metal) femoral heads had decreased wear. The rates of wear for both liners were very low and have not led to any osteolysis or implant failure due to aseptic loosening.


Medicina ◽  
2021 ◽  
Vol 57 (11) ◽  
pp. 1188
Author(s):  
Maximilian G. Hanslmeier ◽  
Michael W. Maier ◽  
Manuel Feisst ◽  
Nicholas A. Beckmann

Background and Objectives: Femoral head and liner exchange is an established treatment for polyethylene wear but has had a more limited role in the treatment of other conditions including dislocation, because of concerns about an increased postoperative dislocation rate. Some authors have considered dislocation associated with polyethylene wear to be a contraindication for this procedure. Materials and Methods: Our retrospective review evaluated the outcome of head and liner exchange in a small consecutively operated heterogeneous cohort of 20 patients who presented with dislocation unrelated to trauma, component malposition or component loosening. Of this group, 12 had prior primary total hip arthroplasty, and 8 had prior revision total hip arthroplasty, and included 4 patients with prior revision for dislocation. Mean follow-up was 6 ± 3.5 years (range 1–145 months). Results: Kaplan–Meier analysis revealed a revision-free implant survival from any cause of 80% (confidence interval 95%:64.3–99.6%) at 5 years after head and liner exchange (index surgery). At final follow-up, 83.3% of patients (n = 10) with prior primary total hip arthroplasty and 62.5% of patients (n = 5) with prior revision total hip arthroplasty, had not required subsequent revision for any cause. None (0%) of the primary total hip arthroplasty group and 3 (38%) of the revision arthroplasty group had required revision for further dislocation. Of the eight revision arthroplasty patients, four had a prior revision for dislocation and three of these four patients required further revision for dislocation after index surgery. The fourth patient had no dislocation after index surgery. One additional patient who had prior revision surgery for femoral component fracture suffered dislocation after index surgery, but was successfully treated with closed reduction. Conclusions: In our study population, femoral head and liner exchange was an effective treatment option for patients with prior primary total hip arthroplasty and also for a highly select group of revision total arthroplasty patients with no prior history of dislocation. Femoral head and liner exchange does not appear to be a viable treatment option for patients who have had revision total arthroplasty after prior dislocations.


2020 ◽  
Vol 63 (2) ◽  
pp. 18-23
Author(s):  
Alejandro Jardón Gómez ◽  
Ana Cristina King ◽  
Carlos Pacheco Díaz

The clinical presentation of a proximal femoral fracture is completely different between young and adult patients. Unlike closed proximal femoral fractures, the incidence of exposed fractures is found in the young population between 15 and 30 years of age. Osteonecrosis of the femoral head is one the complications we can find in this type of fractures. Avascular necrosis (AVN or osteonecrosis) is defined as the interruption of blood supply to the femoral head due to trauma, infectionalcohol or steroid use, resulting in bone necrosis, joint collapse and osteoarthrosis. The treatment will depend on the clinical presentation, age of the patient and when the diagnosis is made. This is a case report of a 16-year-old patient with a gunshot wound on the hip. Surgical cleansing and closed reduction plus internal fixation with a nail in the center of the spine were performed. A 3-year clinical and radiographic follow up was made, observing the evolution of the fracture and the subsequent avascular necrosis that the patient presented. Key words: Proximal femoral fracture; hip; avascular necrosis (AVN, osteonecrosis); open fracture; osteoarthritis.


Sign in / Sign up

Export Citation Format

Share Document