scholarly journals Near zero fluoroscopy in cryoballoon catheter ablation in patients with atrial fibrillation using the new KODEX-EPD imaging system

EP Europace ◽  
2021 ◽  
Vol 23 (Supplement_3) ◽  
Author(s):  
V Schillaci ◽  
G Stabile ◽  
A Arestia ◽  
G Shopova ◽  
F Solimene

Abstract Funding Acknowledgements Type of funding sources: None. Background. Fluoroscopy is commonly used in cryoballoon catheter ablation for pulmonary vein (PV) isolation procedures in patients with atrial fibrillation (AF). Objective The purpose of this study was to verify the feasibility of a new dielectric imaging system in reducing the radiation exposure during cryoballoon ablation in patients with AF. Methods We enrolled 26 consecutive patients with paroxysmal AF: 13 patients underwent the procedure under fluoroscopy guidance before the new system introduction, while 13 patient underwent the procedure under fluoroscopy and KODEX-EPD system guidance with its occlusion tool software. After transseptal access a detailed image reconstruction of left atrium and PVs was achieved with the Achieve octapolar circular mapping catheter and the cryoablation was performed with the Arctic Front Advance cryoballoon. Results Total time of the procedure was comparable between the two groups (90.15 ± 28.67 vs 80.77 ± 17.17 using KODEX-EPD, p = 0.34), while fluoroscopy time was significantly lower in the group using KODEX-EPD (16.92 ± 8.96 vs 5.54 ± 2.06, p < 0.01). Acute isolation was achieved in all PVs. No 30 days complication was observed. Conclusion This is the first study that demonstrates the feasibility of a reduce fluoroscopy workflow using the novel KODEX-EPD system in a cryoballoon procedure. Abstract Figure. Fluoroscopy time

2020 ◽  
Vol 41 (Supplement_2) ◽  
Author(s):  
V Schillaci ◽  
G Stabile ◽  
A Arestia ◽  
G Shopova ◽  
A Agresta ◽  
...  

Abstract Background Fluoroscopy is commonly used in cryoballoon catheter ablation for pulmonary vein (PV) isolation procedures in patients with atrial fibrillation (AF). Purpose The purpose of this study was to verify the feasibility of a new dielectric imaging system in reducing the radiation exposure during cryoballoon ablation in patients with AF. Methods We enrolled 40 consecutive patients with paroxysmal AF: 20 patients underwent the procedure under fluoroscopy guidance before the new system introduction, while 20 patient underwent the procedure under fluoroscopy and a new dielectric imaging system guidance with its occlusion tool software. After transseptal access a detailed image reconstruction of left atrium and PVs was achieved with an octapolar circular mapping catheter and the cryoablation was performed with a cryoballoon catheter. Results Total time of the procedure was comparable between the two groups (90.15±28.67 vs 80.77±17.17 using the dielectric imaging system, p=0.34), while fluoroscopy time was significantly lower in the group using the new dielectric imaging system (16.92±8.96 vs 5.54±2.06, p<0.01). Acute isolation was achieved in all PVs. No 30 days complication was observed. Conclusion This is the first study that demonstrates the feasibility of a reduce fluoroscopy workflow using the novel dielectric imaging system in a cryoballoon procedure. Funding Acknowledgement Type of funding source: None


2020 ◽  
Vol 41 (Supplement_2) ◽  
Author(s):  
F Solimene ◽  
F.M Cauti ◽  
G Stabile ◽  
P Rossi ◽  
V Schillaci ◽  
...  

Abstract Background Optimal pulmonary vein (PV) occlusion, checked with selective contrast injection, is mandatory to obtain an effective PV isolation with a cryoballoon. Purpose The purpose of this study was to verify the feasibility of a new dielectric sensing system in assessing PV occlusion during cryoballoon ablation in patients with atrial fibrillation (AF). Methods We enrolled 25 consecutive patients with paroxysmal or persistent AF. After transseptal access a detailed image reconstruction of left atrium and PVs was achieved with a decapolar circular mapping catheter and a novel dielectric imaging system. The degree of PV occlusion with the inflated cryoballoon catheter was verified by a new occlusion tool software of the dielectric imaging system and compared to the angiography with dye injection in each PV. Results A total of 114 PV cryoballoon occlusion were tested. The new occlusion tool software showed a 91.7% sensitivity and 81.5% specificity in assessing a complete PV occlusion verified with dye injection. The positive predictive value was 84.6% and the negative predictive value was 89.8%. Acute isolation was achieved in all PVs. No 30 days complications were observed. Conclusion This is the first study that demonstrates the feasibility of a new occlusion tool software, using the novel dielectric imaging system, in verifying the degree of PV occlusion during cryoballoon ablation. Funding Acknowledgement Type of funding source: None


Author(s):  
Antonio Bisignani ◽  
Luigi Pannone ◽  
Vincenzo Miraglia ◽  
Juan Sieira ◽  
Saverio Iacopino ◽  
...  

Introduction: Left atrial posterior wall isolation (LAPWI) is often performed in addition to pulmonary vein isolation (PVI) to improve outcome in patients with persistent atrial fibrillation (AF). In this setting, LAPWI + PVI using cryoballoon ablation had comparable results with radiofrequency ablation (RFA). The aim of the study is to evaluate the feasibility and safety of a new cryoballoon ablation system in PVI + LAPWI isolation, comparing it with the historical platform. Methods: The study was a prospective, non-randomized, single center study. Forty consecutive patients, undergoing PVI + LAPWI with the novel POLARx™, were compared to 40 consecutive patients who underwent the same procedure with the established Arctic Front Advance PRO™. Results: Acute isolation was achieved in all PVs in both groups and LAPWI was achieved in 38 patients (95%) in the POLARx group and in 36 patients (90%) in Arctic Front group. Procedural outcomes were similar between the two groups, except for lower temperatures during cryoapplications in the POLARx group, for both PVI and LAPWI. The complication rate was low and similar between groups. Conclusion: LAPWI+PVI with the novel POLARx™ Cryoballoon is feasible and safe; the results are comparable with the Arctic Front Advance PRO™ system.


2021 ◽  
Author(s):  
Vincenzo Schillaci ◽  
Giuseppe Stabile ◽  
Alberto Arestia ◽  
Gergana Shopova ◽  
Alessia Agresta ◽  
...  

2020 ◽  
Vol 41 (Supplement_2) ◽  
Author(s):  
K Ukita ◽  
A Kawamura ◽  
H Nakamura ◽  
K Yasumoto ◽  
M Tsuda ◽  
...  

Abstract Background Little has been reported on the outcome of contact force (CF)-guided radiofrequency catheter ablation (RFCA) and second generation cryoballoon ablation (CBA). Purpose The purpose of this study was to compare the outcome of CF-guided RFCA and second generation CBA for paroxysmal atrial fibrillation (PAF). Methods We enrolled the consecutive 364 patients with PAF who underwent initial ablation between September 2014 and July 2018 in our hospital. We compared the late recurrence of atrial tachyarrhythmia more than three months after ablation between RFCA group and CBA group. All RFCA procedures were performed using CF-sensing catheter and all CBA procedures were performed using second generation CB. Results There were significant differences in background characteristics: chronic kidney disease, serum brain natriuretic peptide level, and left ventricular ejection fraction. After propensity score matched analysis (Table), atrial tachyarrhythmia free survival was significantly higher in CBA group than in RFCA group (Figure). Conclusions Second generation CBA showed a significantly lower late recurrence rate compared to CF-guided RFCA. Kaplan-Meier Curve Funding Acknowledgement Type of funding source: None


2021 ◽  
Vol 42 (Supplement_1) ◽  
Author(s):  
R Schleberger ◽  
A Metzner ◽  
K H Kuck ◽  
D Andresen ◽  
S Willems ◽  
...  

Abstract Background Data on the optimal treatment strategy for antiarrhythmic drug therapy (AAD) after atrial fibrillation (AF) catheter ablation are inconsistent. While AAD potentially stabilizes sinus rhythm, it also increases the patients' treatment burden. Methods Patients from the prospective German Ablation Registry (n=3275) discharged with or without AAD after AF catheter ablation were compared regarding long-term success, cardiovascular events and patient reported outcome. Results In patients with paroxysmal AF (n=2138) recurrence and rehospitalization rates did not differ when discharged with (n=1051) or without (n=1087) AAD (recurrence: adjusted odds ratio (OR) 1.13, 95% confidence interval (CI) [0.95–1.35]; rehospitalization: OR 1.08, 95% CI [0.90–1.30]). The reablation rate was higher and reduced treatment satisfaction was reported more often in those discharged with AAD (reablation: OR 1.30, 95% CI [1.05–1.61]; reduced treatment satisfaction: OR 1.76, 95% CI [1.20–2.58]). Similar rates of recurrences, rehospitalisations, reablations and treatment satisfaction were found in patients with persistent AF (n=1137) discharged with (n=641) or without (n=496) AAD (recurrence: OR 1.22, 95% CI [0.95–1.56]; rehospitalization: OR 1.16, 95% CI [0.90–1.50]; reablation: OR 1.21, 95% CI [0.91–1.61]; treatment satisfaction: OR 1.24, 95% CI [0.74–2.08]). The incidence of cardiovascular events and mortality did not differ at follow-up in paroxysmal and persistent AF patients discharged with or without AAD. Conclusion The rates of recurrences, cardiovascular events and mortality did not differ between patients discharged with or without AAD after AF catheter ablation. However, AAD should be considered carefully in patients with paroxysmal AF, in whom it was associated with a higher reablation rate and reduced treatment satisfaction. FUNDunding Acknowledgement Type of funding sources: None.


2018 ◽  
Vol 2018 ◽  
pp. 1-10 ◽  
Author(s):  
Ali H. Hachem ◽  
Joseph E. Marine ◽  
Housam A. Tahboub ◽  
Sana Kamdar ◽  
Shaffi Kanjwal ◽  
...  

Background. Pulmonary vein isolation is commonly performed using radiofrequency energy with cryoablation gaining acceptance. We performed a meta-analysis of randomized controlled trials which compared radiofrequency versus cryoablation for patients with atrial fibrillation. Methods. A systematic search strategy identified both published and unpublished articles from inception to November 10, 2016, in multiple databases. The primary outcomes for this meta-analysis were long-term freedom from atrial fibrillation at 12-month follow-up and overall postoperative complication rates. For all included studies, the methodological quality was assessed through the Cochrane Collaboration’s tool for risk of bias. Results. A total of 247 articles were identified with eight being included in this review as they satisfied the prespecified inclusion criteria. Overall, there was no significant difference in freedom from atrial fibrillation at ≥12-month follow-up between those receiving cryoballoon and radiofrequency ablation, respectively (OR = 0.98, CI = 0.67–1.43, I2 = 56%, p=0.90). Additionally, the secondary outcomes of duration of ablation, fluoroscopy time, and ablation time failed to reach significance. Cryoballoon ablation had significantly greater odds of postoperative phrenic nerve injury at 12-month follow-up. Conclusions. Our meta-analysis suggests that cryoballoon ablation provides comparable benefits with regard to freedom from atrial fibrillation at medium-term follow-up, fluoroscopy time, ablation time, operative duration, and overall complication rate in comparison to radiofrequency ablation.


2019 ◽  
Vol 2 (51) ◽  
pp. 4-7
Author(s):  
Agnieszka Wojdyła-Hordyńska ◽  
Jakub Baran ◽  
Paweł Derejko

The first use of cryoablation in the treatment of arrhythmia has already been described over 40 years ago [1]. Since the introduction of cryoballoon in pulmonary veins isolation in atrial fibrillation treatment, the method has started to attract a lot of interest. Over 350,000 procedures around the word were carried out only by 2018 [2]. Recently, there have been several new publications on the results of second-generation cryoballoon ablation [2, 3, 4]. In view of technology changes, and to summarize years of experience in the treatment of atrial fibrillation, the first Cryousers conference was organized, and held in 2018 in Poland. During this meeting a survey was conducted, obtaining data on the practice of atrial fibrillation treatment in 38 Polish electrophysiological centers performing cryoablation of atrial fibrillation using both balloons, Arctic Front Advance, Medtronic Inc., Minneapolis MN, and radiofrequency point by point ablation. Around 3,745 cryoballoon procedures were performed in the surveyed centers during the year preceding the survey. The survey concerned practical issues related to the qualification and preparation of patients for the procedure, its course, and the results of pulmonary veins isolation in Poland.