scholarly journals Routine Analyses of Trace Elements in Geological Samples using Flow Injection and Low Pressure On-Line Liquid Chromatography Coupled to ICP-MS: A Study of Geochemical Reference Materials BR, DR-N, UB-N, AN-G and GH

2001 ◽  
Vol 25 (2-3) ◽  
pp. 187-198 ◽  
Author(s):  
Jean Carignan ◽  
Pascal Hild ◽  
Guy Mevelle ◽  
Jacques Morel ◽  
Delphine Yeghicheyan
2015 ◽  
Vol 93 (11) ◽  
pp. 1226-1231 ◽  
Author(s):  
R. Douglas Evans ◽  
Andrei Izmer ◽  
Karima Benkhedda ◽  
Andrew Toms ◽  
Angelo Fernando ◽  
...  

A measurement system capable of continuous on-line matrix removal, pre-concentration and analysis of 226Ra using pre-packed columns coupled to a flow injection system and an ICP-MS was developed. Full instrumental control of both the ICP-MS and the flow injection system provided automatic integration of the transient signals. The flow injection system was programmed to control column conditioning, sample loading, column rinsing, analyte elution and column cleaning operations employing appropriate solutions. The application of this system to the 226Ra analysis of an industrial liquid effluent was demonstrated. Using this particular instrument together with pre-concentration and matrix removal procedures, a limit of detection of 5.4 fg L−1 (2 mBq L−1) and a method detection limit of 16.2 fg L−1 (6 mBq L−1) were achieved for the measurement of 226Ra using a 25 mL sample volume. Total time for sample handling and analysis is approximately 10 minutes. The concentration of 226Ra in a discharged effluent sample was 0.73 pg L−1 (27 mBq L−1), which is in good agreement with the value of 0.81 pg L−1 (30 mBq L−1) measured using conventional alpha counting techniques.


2003 ◽  
Vol 19 (5) ◽  
pp. 791-793 ◽  
Author(s):  
Hideo HAYASHI ◽  
Takahiro HONDA ◽  
Taku IWATA ◽  
Tomokazu TANAKA ◽  
Masataka HIRAIDE

Sign in / Sign up

Export Citation Format

Share Document