Atypical cells in sysmex UN automated urine particle analyzer: a case report and pitfalls for future studies
AbstractObjectivesSysmex UN series fully automated urine analyzer reports a series of research parameters besides routine parameters including the “atypical cells” parameter. An automated instrument in clinical use capable of detecting neoplastic cells of the urinary tract will have paramount significance.Case presentationA 73 years old male patient with a recurrent high grade urothelial carcinoma admitted to our urology outpatient clinic due to hematuria.Urinalysis showed +3 hemoglobin, +3 leukocyte esterase, 200/HPF red blood cells, 300/HPF white blood cells. The instrument also reported atypical cells (7.6/µL or 1.3/HPF) under the heading of “research parameters.” Presence of atypical cells was confirmed by the manual microscopy. The patient has undergone transurethral resection of papillary lesions and the pathology report confirmed a recurrence. On follow-up, atypical cells fell to 0.1/µL after 40 days.ConclusionsThis case report presents a patient with atypical cells in urine, detected by a fully automated urine analyzer. The atypical cells presented on the screen of the analyzer were confirmed by the manual microscopy. This presentation may influence future studies pertaining to the subject.