scholarly journals Evaluation of a Rapid Immunochromatographic test kit to the Gold Standard Fluorescent Antibody Test for Diagnosis of Rabies in Animals in Bhutan

2020 ◽  
Author(s):  
Tenzin Tenzin ◽  
Kelzang Lhamo ◽  
Purna B Rai ◽  
Dawa Tshering ◽  
Pema Gyamtsho ◽  
...  

Abstract Background Rabies kills approximately 59,000 people in the world each year worldwide. Rapid and accurate diagnosis of rabies is important for instituting rapid containment measures and for advising the exposed people for postexposure treatment. The application of a rapid diagnostic tests in the field can greatly enhance disease surveillance and diagnostic activities, especially in resource poor settings. A total of 179 brain tissue samples collected from different rabies suspect animal species (113 dogs, 50 cattle, 10 cats, 3 goats, 2 horses, and 1 bear) were selected and tested using both rapid immunochromatographic kit and the reference standard fluorescent antibody test (FAT). We evaluated the test sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV) of a rapid antigen detection test kit produced by BioNote, Inc. (Hwaseong-si, Korea) to a FAT for its fit-for-purpose for confirmation of clinical cases of rabies for early response, prevalence of infection (rabies surveillance), freedom for infection and eradication of rabies. ResultsAmong 179 samples examined in this study, there was a concordance in results by the rapid test and FAT in 115 positive samples and 54 negative samples. Test results were discordant in 10 samples which were positive by FAT, but negative (false negative) by rapid kit. The rapid test kit showed a sensitivity of 92% (95% CI: 85.9 – 95.6) and specificity of 100% (95% CI: 93.4 – 100) using FAT as the reference standard. The positive and negative predictive values were found to be 100% (95% CI:96.7 – 100) and 84.4% (95% CI: 73.6 – 91.3), respectively. Overall there was 94.4% (95% CI: 90 – 96.9) test agreement between rapid test and FAT (Kappa value = 0.874) with a positive percent agreement and negative percent agreement of 92 and 100%, respectively.Conclusions Our finding demonstrated that the rapid test kit (BioNote) can be used for confirming clinical case of rabies in animals for making rapid decisions including early medical intervention in human exposure and controlling rabies outbreaks in resource poor settings. Keywords: rabies virus; diagnostic test; fluorescent antibody test; rapid anigen test; rapid immunochromatographic test; Bhutan

2020 ◽  
Author(s):  
Tenzin Tenzin ◽  
Kelzang Lhamo ◽  
Purna B Rai ◽  
Dawa Tshering ◽  
Pema Jamtsho ◽  
...  

Abstract Background: Rabies kills approximately 59,000 people in the world each year worldwide. Rapid and accurate diagnosis of rabies is important for instituting rapid containment measures and for advising the exposed people for postexposure treatment. The application of a rapid diagnostic tests in the field can greatly enhance disease surveillance and diagnostic activities, especially in resource poor settings. In this study, a total of 179 brain tissue samples collected from different rabies suspect animal species (113 dogs, 50 cattle, 10 cats, 3 goats, 2 horses, and 1 bear) were selected and tested using both rapid immunochromatographic kit and the reference standard fluorescent antibody test (FAT). We evaluated the sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV) of a rapid antigen detection test kit produced by BioNote, Inc. (Hwaseong-si, Korea) relative to a FAT for its fit-for-purpose for confirmation of clinical cases of rabies for early response and enhancing rabies surveillance. Results: Among 179 samples examined in this study, there was a concordance in results by the rapid test and FAT in 115 positive samples and 54 negative samples. Test results were discordant in 10 samples which were positive by FAT, but negative (false negative) by rapid kit. The rapid test kit showed a sensitivity of 92% (95% CI: 85.9 – 95.6) and specificity of 100% (95% CI: 93.4 – 100) using FAT as the reference standard. The positive and negative predictive values were found to be 100% (95% CI:96.7 – 100) and 84.4% (95% CI: 73.6 – 91.3), respectively. Overall, there was 94.4% (95% CI: 90 – 96.9) test agreement between rapid test and FAT (Kappa value = 0.874) with a positive percent agreement and negative percent agreement of 92 and 100%, respectively. Conclusions: Our finding demonstrated that the rapid test kit (BioNote) can be used for rabies surveillance and confirming clinical case of rabies in animals for making rapid decisions particularly controlling rabies outbreaks in resource poor settings.


2019 ◽  
Author(s):  
Tenzin Tenzin ◽  
Kelzang Lhamo ◽  
Purna B Rai ◽  
Dawa Tshering ◽  
Pema Gyamtsho ◽  
...  

Abstract Background: Rabies kills approximately 59,000 people in the world each year worldwide. Rapid and accurate diagnosis of rabies is important for instituting rapid containment measures and for advising the exposed people for postexposure treatment. The application of a rapid diagnostic tests in the field can greatly enhance disease surveillance activities, especially in resource poor settings.Methods: From 2012 to 2017, a total of 179 brain tissue samples collected from different animal species (113 dogs, 50 cattle, 10 cats, 3 goats, 2 horses, and 1 bear) suspected of having died due to rabies were selected and tested using the rapid immunochromatographic kit from BioNote© company and compared to the Gold Standard Fluorescent Antibody test (FAT) for diagnosis of rabies.Results: Among 179 samples examined in this study, there was concordance in results by the rapid test and FAT in 115 positive samples and 54 negative samples. Test result were discordant in 10 samples which were positive by FAT, but negative (false negative) by rapid kit. The rapid test kit showed a sensitivity of 92% (95% CI: 85.9 – 95.6) and specificity of 100% (95% CI: 93.4 – 100) using FAT as the gold standard. The positive and negative predicative values were found to be 100% (95% CI:96.7 – 100) and 84% (95% CI: 73.6 – 91.3), respectively. Overall there was 94.41% (95% CI: 90 – 96.9) test agreement (almost perfect agreement) between rapid test and FAT (Kappa value = 0.874).Conclusions: Our results demonstrate the potential value of the rapid test kit for countries with limited diagnostic resources, including Bhutan. The rapid kit’s inability to correctly detect 10 FAT-positive samples (10 out of 179 (5.6%) were false negatives) in our study could have been due to the low viral load in the samples (< 102.0LD50/0.03ml) which could not be detected by the rapid kit as compared with the FAT. The human factor related to the varying experiences of the technicians who performed the test in the field also may have influenced the test result. The rapid test kit is inexpensive, rapid and easy to use in the field or in laboratory setting without the need for special training and can support to enhance rabies surveillance in resource poor countries.


2013 ◽  
Vol 2013 ◽  
pp. 1-5
Author(s):  
O. O. Nwankiti ◽  
E. I. Ikeh ◽  
O. A. Arowolo ◽  
A. J. Nwankiti ◽  
M. O. Odugbo ◽  
...  

Scrapie, a disease of sheep and goats with a progressive course and fatal outcome, has not been identified in Nigeria. Anecdotal scrapie reports by livestock workers abound. Livestock diseases like scrapie form huddles in livestock economics of countries. For 8 months we surveyed for scrapie targeting emergency/casualty slaughter sheep and goats in Jos, Nigeria. We clinically examined 510 sheep and 608 goats of local breeds, aged from 12 months to 5 years. In total 31 (5.10%) goats and no sheep were clinically suspicious for scrapie. Caudal brainstem tissues of suspect animals collected postmortem were analyzed for the disease specific form of the prion protein, PrPSc, using Bio-Rad’s TeSeE ELISA rapid test kit. No sample was positive for scrapie. Fluorescent antibody test for rabies and H&E staining on samples were carried out for differential diagnosis. These showed no pathological lesions indicative for neurological disease. While our findings do not exclude the presence of scrapie in Jos, we demonstrate that targeted sampling of small ruminants for neuroinfectious disease is feasible in developing countries, pointing to the possibility of implementing such a monitoring scheme in Nigeria to prevent economic losses in small ruminant livestock as scrapie caveats from endemic countries have shown.


Author(s):  
Jesha Mundodan ◽  
Samina Hasnain ◽  
Hayat Khogali ◽  
Soha Shawqi Al Bayat ◽  
Dina Ali ◽  
...  

Background: In response to the growing coronavirus disease 2019 (COVID-19) pandemic and the shortage of laboratory based molecular testing capacity and reagents, multiple diagnostic test manufacturers have developed rapid and easy to use devices to facilitate testing outside laboratory settings. These kits are either based on detection of proteins from SARS-CoV-2 virus or detection of antigen or human antibodies generated in response to the infection. However, it is important to understand their performance characteristics and they must be validated in the local population setting.Design and Methods: The objective is to assess the validity of the rapid test for IgG and IgM immunoglobulins compared to the current gold standard reverse transcription polymerase chain reaction (RT-PCR) test. A total of 16951 asymptomatic individuals were tested by the Ministry of Public Health track-and-trace team using both rapid immunodiagnostic test and RT-PCR as part of screening across various random settings with potential risk of community interaction prior to gradual lifting of restrictions in Qatar.  Rapid test was considered to be posiive if both IgG and IgM are positive, while only IgG/IgM positive was considered as rapid test negative. Sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV) were calculated.Results: The sensitivity of rapid test kit was found to be 0.9%, whereas the specificity was found to be 97.8%. the PPV was found to be 0.3% whereas the NPV was found to be 99.4%.Conclusion: Based on the outcome and results of the study, it appears that the sensitivity and PPV of the rapid antibody test are low. As such, this test is not recommended for use to assist in taking clinic-based decisions or decisions related to quarantine/isolation.


2011 ◽  
Vol 23 (6) ◽  
pp. 1197-1201 ◽  
Author(s):  
Songsri Kasempimolporn ◽  
Wachiraporn Saengseesom ◽  
Samrerng Huadsakul ◽  
Supatsorn Boonchang ◽  
Visith Sitprija

An immunochromatographic test strip for Rabies virus was evaluated with dog saliva samples. The test was initially validated against 237 dogs of known infection status, and then evaluated in the field with 1,290 live dogs. By validation of paired saliva–brain specimens obtained from dogs at necropsy, the saliva strip test was 94.4% specific and 93.0% sensitive when compared to the gold standard fluorescent antibody test (FAT) on brain smears. The sensitivity and specificity of a nested polymerase chain reaction (nPCR) assay using saliva were 100% compared to the FAT results. The performance of strip test with field saliva samples from street dogs had a specificity of 98.7% in comparison to nPCR as the reference method. As the strip test kit can potentially be used outside the laboratory and be applicable as an on-site testing assay, it represents a powerful screening tool for epidemiological surveys and disease control. The test could be useful for the surveillance of rabies in dogs and, in particular, be used to monitor the success of rabies control programs.


2021 ◽  
Vol 15 (11) ◽  
pp. e0009859
Author(s):  
Hannah M. Cranford ◽  
A. Springer Browne ◽  
Karen LeCount ◽  
Tammy Anderson ◽  
Camila Hamond ◽  
...  

During 2019–2020, the Virgin Islands Department of Health investigated potential animal reservoirs of Leptospira spp., the bacteria that cause leptospirosis. In this cross-sectional study, we investigated Leptospira spp. exposure and carriage in the small Indian mongoose (Urva auropunctata, syn: Herpestes auropunctatus), an invasive animal species. This study was conducted across the three main islands of the U.S. Virgin Islands (USVI), which are St. Croix, St. Thomas, and St. John. We used the microscopic agglutination test (MAT), fluorescent antibody test (FAT), real-time polymerase chain reaction (lipl32 rt-PCR), and bacterial culture to evaluate serum and kidney specimens and compared the sensitivity, specificity, positive predictive value, and negative predictive value of these laboratory methods. Mongooses (n = 274) were live-trapped at 31 field sites in ten regions across USVI and humanely euthanized for Leptospira spp. testing. Bacterial isolates were sequenced and evaluated for species and phylogenetic analysis using the ppk gene. Anti-Leptospira spp. antibodies were detected in 34% (87/256) of mongooses. Reactions were observed with the following serogroups: Sejroe, Icterohaemorrhagiae, Pyrogenes, Mini, Cynopteri, Australis, Hebdomadis, Autumnalis, Mankarso, Pomona, and Ballum. Of the kidney specimens examined, 5.8% (16/270) were FAT-positive, 10% (27/274) were culture-positive, and 12.4% (34/274) were positive by rt-PCR. Of the Leptospira spp. isolated from mongooses, 25 were L. borgpetersenii, one was L. interrogans, and one was L. kirschneri. Positive predictive values of FAT and rt-PCR testing for predicting successful isolation of Leptospira by culture were 88% and 65%, respectively. The isolation and identification of Leptospira spp. in mongooses highlights the potential role of mongooses as a wildlife reservoir of leptospirosis; mongooses could be a source of Leptospira spp. infections for other wildlife, domestic animals, and humans.


Author(s):  
Ranti Permatasari ◽  
Aryati Aryati ◽  
Budi Arifah

Hepatitis C (HCV) infection could be spread by blood transfusion. Screening of HCV in donor blood could prevent HCV infection to the recipient. HCV antibody test using rapid test of multiple antibody detection by immunochromatography method is an easy and rapid test that could detect four HCV antibodies separately. The aim of this study was to evaluate the diagnostic value of antibody HCV using multiple antibody detection rapid test in diagnosing HCV infection. This was an analytical observational study with a cross sectional design. The samples consisted of 42 donors’ blood serum from the Surabaya Branch of the Indonesian Red Cross which underwent HCV infection test using ELISA method. The samples were then tested using PCR HCV RNA as the gold standard and antibody HCV multiple antibodydetection rapid test The diagnostic value of HCV antibody test using multiple antibody detection rapid test by immunochromatography method showed a diagnostic sensitivity of 100%, diagnostic specificity of 75%, positive predictive value of 66.7% and negative predictive value of 100%, a diagnostic efficiency of 83.3%, with a positive probability ratio of 4 times. The most often positive antibody pattern was four (4) positive antibodies (core protein, NS3, NS4 and NS5). Core protein (CP) and NS3 were the most often positive antibodies. Based on this study result, the HCV antibody test using multiple antibody detection rapid test by immunochromatography method has a good diagnostic value.


2020 ◽  
Vol 0 (0) ◽  
Author(s):  
Huseyin Agah Terzi ◽  
Ozlem Aydemir ◽  
Engin Karakece ◽  
Huseyin Hatipoglu ◽  
Mehmet Olmez ◽  
...  

AbstractObjectivesTo test the performance of the newly available rapid test for syphilis, we compared it with Treponema pallidum hemagglutination assay (TPHA). Additionally, we investigated the performance of rapid plasma reagin (RPR) and chemiluminescence microparticle immunoassays (CMIA) at our laboratory using TPHA as a gold standard.MethodsThe serum samples of 595 patients with the pre-diagnosis of syphilis were studied by four serological methods. The sensitivity, specificity, and predictive values of RPR, CMIA, and syphilis rapid test were assessed by utilizing TPHA as a gold standard for the diagnosis of syphilis.ResultsOf the patients, 6.2% (37/595) had positive RPR, 5.5% (33/595) had positive CMIA, 5.5% (33/595) had a positive rapid immunochromatographic method and 5% (30/595) had positive TPHA. When TPHA results were taken as the reference, the sensitivity of the rapid test for syphilis was 100%, the specificity was 99.5%, PPV was 90.9%, and NPV was 100.0%.ConclusionsIt was observed that the rapid test for syphilis used in the study was quite successful, its cost was appropriate, and the test was very fast and easy to apply. At the same time, the agreement between syphilis rapid test and TPHA was found to be excellent.


1988 ◽  
Vol 30 (6) ◽  
pp. 411-414 ◽  
Author(s):  
Beatriz J. Celeste ◽  
M. Carolina S. Guimarães ◽  
Edelma Maria Corrales L.

Performance indexes of the peroxidase antibody test were compared to that of the fluorescent antibody test. The peroxidase antibody test had a statistically higher sensitivity and negative predictive value and a higher efficiency than the fluorescent antibody test but its specificity and positive predictive value were within the 95% confidence limits for the values found for the fluorescent antibody test. Such differences did not change when Chagas' disease and visceral leishmaniasis sera were included in index calculations. Statistical analysis showed that the two tests have a substantial degree of agreement but the immunofluorescent test had a specificity index and a positive predictive value equal to 100.0% when Chagas' disease and visceral leishmaniasis sera were not included in the calculations of the performance index; in this instance, a positive test result equals a disclosure of the disease attribute due to the inexistence of false positive results. The enzyme/ protein ratio of the peroxidase conjugate, resulting in heavy or light-labeled conjugates may pose technical problems to its use in serology tests.


Sign in / Sign up

Export Citation Format

Share Document