scholarly journals Comparison of the accuracy and proximal shunt failure rate of freehand placement versus intraoperative guidance in parietooccipital ventricular catheter placement

2016 ◽  
Vol 41 (3) ◽  
pp. E10 ◽  
Author(s):  
Thomas J. Wilson ◽  
Kathleen E. McCoy ◽  
Wajd N. Al-Holou ◽  
Sergio L. Molina ◽  
Matthew D. Smyth ◽  
...  

OBJECTIVE The aim of this paper is to compare the accuracy of the freehand technique versus the use of intraoperative guidance (either ultrasound guidance or frameless stereotaxy) for placement of parietooccipital ventricular catheters and to determine factors associated with reduced proximal shunt failure. METHODS This retrospective cohort study included all patients from 2 institutions who underwent a ventricular cerebrospinal fluid (CSF) shunting procedure in which a new parietooccipital ventricular catheter was placed between January 2005 and December 2013. Data abstracted for each patient included age, sex, method of ventricular catheter placement, side of ventricular catheter placement, Evans ratio, and bifrontal ventricular span. Postoperative radiographic studies were reviewed for accuracy of ventricular catheter placement. Medical records were also reviewed for evidence of shunt failure requiring revision. Standard statistical methods were used for analysis. RESULTS A total of 257 patients were included in the study: 134 from the University of Michigan and 123 from Washington University in St. Louis. Accurate ventricular catheter placement was achieved in 81.2% of cases in which intraoperative guidance was used versus 67.3% when the freehand technique was used. Increasing age reduced the likelihood of accurate catheter placement (OR 0.983, 95% CI 0.971–0.995; p = 0.005), while the use of intraoperative guidance significantly increased the likelihood (OR 2.809, 95% CI 1.406–5.618; p = 0.016). During the study period, 108 patients (42.0%) experienced shunt failure, 79 patients (30.7%) had failure involving the proximal catheter, and 53 patients (20.6%) had distal failure (valve or distal catheter). Increasing age reduced the likelihood of being free from proximal shunt failure (OR 0.983, 95% CI 0.970–0.995; p = 0.008), while both the use of intraoperative guidance (OR 2.385, 95% CI 1.227–5.032; p = 0.011), and accurate ventricular catheter placement (OR 3.424, 95% CI 1.796–6.524; p = 0.009) increased the likelihood. CONCLUSIONS The use of intraoperative guidance during parietooccipital ventricular catheter placement as part of a CSF shunt system significantly increases the likelihood of accurate catheter placement and subsequently reduces the rate of proximal shunt failure.

2013 ◽  
Vol 119 (1) ◽  
pp. 66-70 ◽  
Author(s):  
Thomas J. Wilson ◽  
William R. Stetler ◽  
Wajd N. Al-Holou ◽  
Stephen E. Sullivan

Object The objective of this study was to compare the accuracy of 3 methods of ventricular catheter placement during CSF shunt operations: the freehand technique using surface anatomy, ultrasonic guidance, and stereotactic neuronavigation. Methods This retrospective cohort study included all patients from a single institution who underwent a ventricular CSF shunting procedure in which a new ventricular catheter was placed between January 2005 and March 2010. Data abstracted for each patient included age, sex, diagnosis, method of ventricular catheter placement, site and side of ventricular catheter placement, Evans ratio, and bifrontal ventricular span. Postoperative radiographic studies were reviewed for accuracy of ventricular catheter placement. Medical records were also reviewed for evidence of shunt failure requiring revision through December 2011. Statistical analysis was then performed comparing the 3 methods of ventricular catheter placement and to determine risk factors for inaccurate placement. Results There were 249 patients included in the study; 170 ventricular catheters were freehand passed, 51 were placed using stereotactic neuronavigation, and 28 were placed under intraoperative ultrasonic guidance. There was a statistically significant difference between freehand catheters and stereotactic-guided catheters (p < 0.001), as well as between freehand catheters and ultrasound-guided catheters (p < 0.001). The only risk factor for inaccurate placement identified in this study was use of the freehand technique. The use of stereotactic neuronavigation and ultrasonic guidance reduced proximal shunt failure rates (p < 0.05) in comparison with a freehand technique. Conclusions Stereotactic- and ultrasound-guided ventricular catheter placements are significantly more accurate than freehand placement, and the use of these intraoperative guidance techniques reduced proximal shunt failure in this study.


2007 ◽  
Vol 22 (4) ◽  
pp. 1-8 ◽  
Author(s):  
Kristen Upchurch ◽  
Murisiku Raifu ◽  
Marvin Bergsneider

Object Patients with symptomatic isolated fourth ventricle and multicompartmentalized hydrocephalus benefit from operative treatment, but the optimal surgical approach and technique have yet to be established. The authors report on their experience with the treatment of symptomatic adult patients by endoscope-assisted placement of a fourth ventricle shunt catheter via a frontal transventricular approach. Methods The authors describe a retrospective series of four patients treated for isolated fourth ventricle. The surgical technique is described in detail: use of a flexible endoscope with dual-port intraventricular access for direct visualization and for mechanical manipulation of a multiperforated panventricular catheter guided by frameless stereotaxy. The transventricular approach allowed optimal catheter placement within the fourth ventricle. The use of the flexible endoscope permitted the neurosurgeon to use the endoscope as a tool to guide the ventricular catheter tip within the third ventricle and through the cerebral aqueduct. Clinical outcomes demonstrated neurological and radiographically verified improvement in all patients. Conclusions The endoscope-assisted dual-port technique provides a solution to the technical difficulties of fourth ventricle shunt placement. The multiple advantages of this technique include a single ventricular catheter shunt system that equalizes ventricular pressures, a frontal location for the ventricular catheter that facilitates valve placement and programming, and ventricular catheter placement within the fourth ventricle that does not allow the catheter to impinge on the fourth ventricle floor and makes the catheter less prone to obstruction.


2019 ◽  
Vol 5 (1) ◽  
pp. 59-63
Author(s):  
DM Arman ◽  
Sheikh Muhammad Ekramullah ◽  
Sudipta Kumer Mukherjee ◽  
Samantha Afreen ◽  
Md Anwarul Hoque Faraji ◽  
...  

Background: Inaccurate placement of VP shunt catheter is related to shunt failure. Objective: The objective of this study was to determine the accuracy of ventricular catheter placement during ventriculoperitoneal shunt operations using the freehand technique. Methodology: This prospective observational study included all patients from a single institution who underwent a ventriculoperitoneal shunt procedure in which a new ventricular catheter was placed between September 2013 and August 2016 for a period of three (03) years. Data abstracted for each patient included age, sex, diagnosis, site and side of ventricular catheter placement. Postoperative CT scan images were reviewed for accuracy of ventricular catheter placement. Results: There were 140 patients included in the study; accuracy ventricular catheter tip placement were 55 (39.28%) using freehand technique. Conclusion: Mechanical malfunction and infection are the most significant problems associated with shunts for the treatment of hydrocephalus. Journal of National Institute of Neurosciences Bangladesh, 2019;5(1): 59-63


Neurosurgery ◽  
2015 ◽  
Vol 77 (3) ◽  
pp. 321-331 ◽  
Author(s):  
Cody L. Nesvick ◽  
Nickalus R. Khan ◽  
Gautam U. Mehta ◽  
Paul Klimo

Abstract BACKGROUND: Ventricular shunt placement for treating hydrocephalus is one of the most common neurosurgical procedures. The rate of shunt failure, however, has not appreciably changed with time. OBJECTIVE: To investigate whether intraoperative image guidance using ultrasound or stereotaxy contributes to accurate shunt catheter placement and survival. METHODS: We performed a systematic literature review using PubMed and MEDLINE databases for studies that use ultrasound and frameless stereotaxy for ventricular catheter placement for hydrocephalus. All articles assessed the accuracy of catheter tip placement and/or overall shunt survival, and the rate of accurate shunt catheter placement, the overall failure rate, and the average time to shunt failure were extracted for analysis. RESULTS: Although each modality (ultrasound/stereotaxy) did not increase catheter placement accuracy, a combined random-effects meta-analysis of 738 catheters (136 guided by ultrasound, 168 guided by frameless stereotaxy, and 434 freehand) demonstrated a weak benefit of image guidance (risk ratio: 1.19, 95% confidence interval: 1.02-1.39, P = .02), but this result was limited by considerable heterogeneity among studies (I2 = 86%, P &lt; .001 by Cochrane's Q test). A meta-analysis could not be performed for shunt survival due to heterogeneity in data reporting. CONCLUSION: Although image guidance offers a promising solution to lower the risk of inaccurate catheter placement, which could lead to lower premature failure of ventricular shunts, our review demonstrated that there is not yet a clear benefit of these technologies. Current literature is limited to case series and cohort studies, and significant between-study heterogeneity in methodology and reporting currently limits a higher order analysis.


2008 ◽  
Vol 1 (3) ◽  
pp. 237-239 ◽  
Author(s):  
Michael J. Ellis ◽  
Colin J. Kazina ◽  
Marc R. Del Bigio ◽  
Patrick J. McDonald

✓Shunt failure is commonly associated with infection or mechanical obstruction of the shunt system. The presence of eosinophilia in the cerebrospinal fluid (CSF) has been associated with CSF shunt failure and may be related to both latex and shunt allergies. The authors describe the case of a child with a latex allergy who presented with 10 episodes of shunt failure over a period of 93 months. Cerebrospinal fluid sampling demonstrated persistent eosinophilia (3–36%) and negative cultures. Pathological examination of the ventricular catheter on 3 occasions demonstrated mechanical obstruction by inflammatory debris consisting largely of eosinophils and multinucleated giant cells. On the suspicion that the child might have some uncharacterized allergy to the shunt hardware, shunt replacement was performed using an “extracted” shunt system. The child has remained free of shunt malfunction for > 2 years since the last surgery. Immune responses to unpolymerized silicone are discussed.


2010 ◽  
Vol 6 (3) ◽  
pp. 299-302 ◽  
Author(s):  
Anand I. Rughani ◽  
Bruce I. Tranmer ◽  
Jeffrey E. Florman ◽  
James T. Wilson

Accurate assessment of imaging studies in patients with ventriculoperitoneal shunts can be aided by empirical findings. The authors characterize an objective measurement easily performed on head CT scans with the goal of producing clear evidence of shunt fracture or disconnection in patients with a snap shunt–type system. The authors describe 2 cases of ventriculoperitoneal shunt failure involving a fracture and a disconnection associated with a snap-shunt assembly. In both cases the initial clinical symptoms were not convincing for shunt malfunction, and the interpretation of the CT finding failed to immediately identify the abnormality. As the clinical picture became more convincing for shunt malfunction, each patient subsequently underwent successful shunt revision. The authors reviewed the CT scans of 10 patients with an intact and functioning snap-shunt system to characterize the normal appearance of the snap-shunt connection. On CT scans the distance between the radiopaque portion of the ventricular catheter and the radiopaque portion of the reservoir dome measures an average of 4.72 mm (range 4.6–4.9 mm, 95% CI 4.63–4.81 mm). In the authors' patient with a fractured ventricular catheter, this interval measured 7.8 mm, and in the patient with a disconnection it measured 7.7 mm. In comparison with the range of normal values, a radiolucent interval significantly greater than 4.9 mm should promptly raise concern for a disconnected or fractured shunt in this system. This measurement may prove particularly useful when serial imaging is not readily available.


2019 ◽  
Vol 9 (1) ◽  
pp. 16-21
Author(s):  
DM Arman ◽  
Sheikh Muhammad Ekramullah ◽  
Sudipta Kumer Mukherjee ◽  
Joynul Islam ◽  
Mirza Hafizur Rashid ◽  
...  

Object: The objective of this study was to determine the relationship of the location of the ventricular catheter tip and function of the ventriculoperitoneal shunt. Methods: This prospective observational study included 140 patients from a single institution who underwent a ventriculoperitoneal shunt procedure in which a new ventricular catheter was placed between September, 2013 and September 2016. Data abstracted for each patient included age, sex, diagnosis, site and side of ventricular catheter placement. Postoperative CT scan images were reviewed for accuracy of ventricular catheter placement. Patients were followed up over 2 ½ to 4 ½ years.63 patients were available for follow up. We lost communication with rest of the patients. The relationship of the location of the ventricular catheter tip and function of the ventriculoperitoneal shunt was analyzed in 63 patients. Results: There were 140 patients included in the study; accuracy ventricular catheter tip placement were 55 (39.28%) using freehand technique. VP shunt functioned well in 43 (68.25%) of 63 patients. Among the 43 patients with well functioning shunt ,19 were in accurate group ,7 were in suboptimal group and 17 were in inaccurate group.26 patients (41.27%) had good outcome with normal development and normal IQ Conclusions: Mechanical malfunction and infection are the most significant problems associated with shunts for the treatment of hydrocephalus. Above all, a significant proportion of shunt failure was due to obstruction of the ventricular catheter, and accurate placement of the shunt catheter is highly important to reduce the incidence of shunt malfunction. Bang. J Neurosurgery 2019; 9(1): 16-21


2016 ◽  
Vol 27 ◽  
pp. 95-98 ◽  
Author(s):  
Kealeboga Josephine Jeremiah ◽  
Catherine Louise Cherry ◽  
Kai Rui Wan ◽  
Jennifer Ah Toy ◽  
Rory Wolfe ◽  
...  

2010 ◽  
Vol 5 (1) ◽  
pp. 4-16 ◽  
Author(s):  
Rolf W. Gruber ◽  
Bernd Roehrig

Object This 25-year follow-up study was performed on 120 children with hypertensive hydrocephalus to evaluate the influence of the early prophylactic implantation of the Integra antisiphon device (ASD, Integra Neurosciences Ltd.) on the rate of proximal shunt obstructions and the frequency of symptomatic slit ventricle syndrome (SVS). The adaptability of the ASD to growth, proper positioning of the ASD as a necessity for its successful performance, and the 3 phases of SVS development are discussed. Method Since 1978, the ASD has consistently been implanted either at the time of primary shunt insertion (66 neonates, mean follow-up 11 years) or during revisions of preexisting shunts (54 children, mean follow-up 11.8 years). The complication rate among the 54 children before ASD implantation (mean follow-up 8.3 years) was compared with that among all 120 patients once an ASD had been inserted. Shunt complications were documented as ventricular catheter, distal catheter, and infectious complications. Results The study revealed a significant long-term reduction in ventricular catheter obstructions and hospitalizations due to intermittent intracranial hypertension symptoms (symptomatic SVS) after both primary and secondary ASD implantation. Data in the study suggest that the high rate of ventricular catheter obstruction in pediatric shunt therapy is caused by hydrostatic suction induced by differential-pressure valve shunts during mobilization of the patient and that the development of a SVS can be traced back to this constant suction, which causes chronic CSF overdrainage and ventricular noncompliance. Recurrent ventricular catheter obstruction and SVS can be prevented by prophylactic supplementation of every shunt system with an ASD. Conclusions To inhibit chronic hydrostatic suction, to prevent overdrainage and proximal shunt obstruction, and to avoid SVS and thus improve the patient's quality of life, the prophylactic implantation of an ASD in every pediatric hydrocephalus shunt is recommended.


1982 ◽  
Vol 57 (4) ◽  
pp. 570-573 ◽  
Author(s):  
Hiroshi Yamada

✓ A method of regulating flow rate through cerebrospinal fluid (CSF) shunts with the addition of a flow-regulating device (FRD) is reported. The FRD consists of a small-caliber (0.4 mm) Teflon tube placed in the usual connector. This device has the advantage of converting the original shunt valves to valves for higher-resistance flow without replacing the entire distal catheter. An experimental study revealed that this device reduced the CSF flow rate in the shunt system by approximately 30%. The author has found it reliable in 32 patients, without causing CSF obstruction. The device is easily installed and removed.