A comprehensive biomechanical analysis of sacral alar iliac fixation: an in vitro human cadaveric model

2019 ◽  
Vol 30 (3) ◽  
pp. 367-375 ◽  
Author(s):  
Bryan W. Cunningham ◽  
Paul D. Sponseller ◽  
Ashley A. Murgatroyd ◽  
Jun Kikkawa ◽  
P. Justin Tortolani

OBJECTIVEThe objective of the current study was to quantify and compare the multidirectional flexibility properties of sacral alar iliac fixation with conventional methods of sacral and sacroiliac fixation by using nondestructive and destructive investigative methods.METHODSTwenty-one cadaveric lumbopelvic spines were randomized into 3 groups based on reconstruction conditions: 1) S1–2 sacral screws; 2) sacral alar iliac screws; and 3) S1–iliac screws tested under unilateral and bilateral fixation. Nondestructive multidirectional flexibility testing was performed using a 6-degree-of-freedom spine simulator with moments of ± 12.5 Nm. Flexion-extension fatigue loading was then performed for 10,000 cycles, and the multidirectional flexibility analysis was repeated. Final destructive testing included an anterior flexural load to construct failure. Quantification of the lumbosacral and sacroiliac joint range of motion was normalized to the intact spine (100%), and flexural failure loads were reported in Newton-meters.RESULTSNormalized value comparisons between the intact spine and the 3 reconstruction groups demonstrated significant reductions in segmental flexion-extension, lateral bending, and axial rotation motion at L4–5 and L5–S1 (p < 0.05). The S1–2 sacral reconstruction group demonstrated significantly greater flexion-extension motion at the sacroiliac junction than the intact and comparative reconstruction groups (p < 0.05), whereas the sacral alar iliac group demonstrated significantly less motion at the sacroiliac joint in axial rotation (p < 0.05). Absolute value comparisons demonstrated similar findings. Under destructive anterior flexural loading, the S1–2 sacral group failed at 105 ± 23 Nm, and the sacral alar iliac and S1–iliac groups failed at 119 ± 39 Nm and 120 ± 28 Nm, respectively (p > 0.05).CONCLUSIONSAlong with difficult anatomy and weak bone, the large lumbosacral loads with cantilever pullout forces in this region are primary reasons for construct failure. All reconstructions significantly reduced flexibility at the L5–S1 junctions, as expected. Conventional S1–2 sacral fixation significantly increased sacroiliac motion under all loading modalities and demonstrated significantly higher flexion-extension motion than all other groups, and sacral alar iliac fixation reduced motion in axial rotation at the sacroiliac joint. Based on comprehensive multidirectional flexibility testing, the sacral alar iliac fixation technique reduced segmental motion under some loading modalities compared to S1–iliac screws and offers potential advantages of lower instrumentation profile and ease of assembly compared to conventional sacroiliac instrumentation techniques.

2017 ◽  
Vol 27 (5) ◽  
pp. 570-577 ◽  
Author(s):  
Mohamed Macki ◽  
Rafael De la Garza-Ramos ◽  
Ashley A. Murgatroyd ◽  
Kenneth P. Mullinix ◽  
Xiaolei Sun ◽  
...  

OBJECTIVEAggressive sacral tumors often require en bloc resection and lumbopelvic reconstruction. Instrumentation failure and pseudarthrosis remain a clinical concern to be addressed. The objective in this study was to compare the biomechanical stability of 3 distinct techniques for sacral reconstruction in vitro.METHODSIn a human cadaveric model study, 8 intact human lumbopelvic specimens (L2–pelvis) were tested for flexion-extension range of motion (ROM), lateral bending, and axial rotation with a custom-designed 6-df spine simulator as well as axial compression stiffness with the MTS 858 Bionix Test System. Biomechanical testing followed this sequence: 1) intact spine; 2) sacrectomy (no testing); 3) Model 1 (L3–5 transpedicular instrumentation plus spinal rods anchored to iliac screws); 4) Model 2 (addition of transiliac rod); and 5) Model 3 (removal of transiliac rod; addition of 2 spinal rods and 2 S-2 screws). Range of motion was measured at L4–5, L5–S1/cross-link, L5–right ilium, and L5–left ilium.RESULTSFlexion-extension ROM of the intact specimen at L4–5 (6.34° ± 2.57°) was significantly greater than in Model 1 (1.54° ± 0.94°), Model 2 (1.51° ± 1.01°), and Model 3 (0.72° ± 0.62°) (p < 0.001). Flexion-extension at both the L5–right ilium (2.95° ± 1.27°) and the L5–left ilium (2.87° ± 1.40°) for Model 3 was significantly less than the other 3 cohorts at the same level (p = 0.005 and p = 0.012, respectively). Compared with the intact condition, all 3 reconstruction groups statistically significantly decreased lateral bending ROM at all measured points. Axial rotation ROM at L4–5 for Model 1 (2.01° ± 1.39°), Model 2 (2.00° ± 1.52°), and Model 3 (1.15° ± 0.80°) was significantly lower than the intact condition (5.02° ± 2.90°) (p < 0.001). Moreover, axial rotation for the intact condition and Model 3 at L5–right ilium (2.64° ± 1.36° and 2.93° ± 1.68°, respectively) and L5–left ilium (2.58° ± 1.43° and 2.93° ± 1.71°, respectively) was significantly lower than for Model 1 and Model 2 at L5–right ilium (5.14° ± 2.48° and 4.95° ± 2.45°, respectively) (p = 0.036) and L5–left ilium (5.19° ± 2.34° and 4.99° ± 2.31°) (p = 0.022). Last, results of the axial compression testing at all measured points were not statistically different among reconstructions.CONCLUSIONSThe addition of a transverse bar in Model 2 offered no biomechanical advantage. Although the implementation of 4 iliac screws and 4 rods conferred a definitive kinematic advantage in Model 3, that model was associated with significantly restricted lumbopelvic ROM.


2022 ◽  
pp. 1-9

OBJECTIVE The traditional anterior approach for multilevel severe cervical ossification of the posterior longitudinal ligament (OPLL) is demanding and risky. Recently, a novel surgical procedure—anterior controllable antedisplacement and fusion (ACAF)—was introduced by the authors to deal with these problems and achieve better clinical outcomes. However, to the authors’ knowledge, the immediate and long-term biomechanical stability obtained after this procedure has never been evaluated. Therefore, the authors compared the postoperative biomechanical stability of ACAF with those of more traditional approaches: anterior cervical discectomy and fusion (ACDF) and anterior cervical corpectomy and fusion (ACCF). METHODS To determine and assess pre- and postsurgical range of motion (ROM) (2 Nm torque) in flexion-extension, lateral bending, and axial rotation in the cervical spine, the authors collected cervical areas (C1–T1) from 18 cadaveric spines. The cyclic fatigue loading test was set up with a 3-Nm cycled load (2 Hz, 3000 cycles). All samples used in this study were randomly divided into three groups according to surgical procedures: ACDF, ACAF, and ACCF. The spines were tested under the following conditions: 1) intact state flexibility test; 2) postoperative model (ACDF, ACAF, ACCF) flexibility test; 3) cyclic loading (n = 3000); and 4) fatigue model flexibility test. RESULTS After operations were performed on the cadaveric spines, the segmental and total postoperative ROM values in all directions showed significant reductions for all groups. Then, the ROMs tended to increase during the fatigue test. No significant crossover effect was detected between evaluation time and operation method. Therefore, segmental and total ROM change trends were parallel among the three groups. However, the postoperative and fatigue ROMs in the ACCF group tended to be larger in all directions. No significant differences between these ROMs were detected in the ACDF and ACAF groups. CONCLUSIONS This in vitro biomechanical study demonstrated that the biomechanical stability levels for ACAF and ACDF were similar and were both significantly greater than that of ACCF. The clinical superiority of ACAF combined with our current results showed that this procedure is likely to be an acceptable alternative method for multilevel cervical OPLL treatment.


2014 ◽  
Vol 20 (2) ◽  
pp. 209-219 ◽  
Author(s):  
Sabrina A. Gonzalez-Blohm ◽  
James J. Doulgeris ◽  
Kamran Aghayev ◽  
William E. Lee ◽  
Andrey Volkov ◽  
...  

Object In this paper the authors evaluate through in vitro biomechanical testing the performance of an interspinous fusion device as a stand-alone device, after lumbar decompression surgery, and as supplemental fixation to expandable cages in a posterior lumbar interbody fusion (PLIF) construct. Methods Nine L3–4 human cadaveric spines were biomechanically tested under the following conditions: 1) intact/control; 2) L3–4 left hemilaminotomy with partial discectomy (injury); 3) interspinous spacer (ISS); 4) bilateral pedicle screw system (BPSS); 5) bilateral hemilaminectomy, discectomy, and expandable posterior interbody cages with ISS (PLIF-ISS); and 6) PLIF-BPSS. Each test consisted of 100 N of axial preload with ± 7.5 Nm of torque in flexion-extension, right/left lateral bending, and right/left axial rotation. Significant changes in range of motion (ROM), neutral zone stiffness (NZS), elastic zone stiffness (EZS), and energy loss (EL) were explored among conditions using nonparametric Friedman test and Wilcoxon signed-rank comparisons (p ≤ 0.05). Results The injury increased ROM in flexion (p = 0.01), left bending (p = 0.03), and right/left rotation (p < 0.01) and also decreased NZS in flexion (p = 0.01) and extension (p < 0.01). Both the ISS and BPSS reduced flexion-extension ROM and increased flexion-extension stiffness (NZS and EZS) with respect to the injury and intact conditions (p < 0.05), but the ISS condition provided greater resistance than BPSS in extension for ROM, NZS, and EZS (p < 0.01). The BPSS increased the rigidity (ROM, NZS, and EZS) of the intact model in lateral bending and axial rotation (p ≤ 0.01), except in EZS for left rotation (p = 0.23, Friedman test). The incorporation of posterior cages marginally increased (p = 0.05) the EZS of the BPSS construct in flexion but these interbody devices provided significant stability to the ISS construct in lateral bending and axial rotation for ROM (p = 0.02), in lateral bending for NZS (p = 0.02), and in flexion/axial rotation for EZS (p ≤ 0.03); however, both PLIF constructs demonstrated equivalent ROM and stiffness (p ≥ 0.16), except in lateral bending where the PLIF-BPSS was more stable (p = 0.02). In terms of EL, the injury increased EL in flexion-extension (p = 0.02), the ISS increased EL for lateral bending and axial rotation (p ≤ 0.03), and the BPSS decreased EL in lateral bending (p = 0.02), with respect to the intact condition. The PLIF-ISS decreased lateral bending EL with respect to the ISS condition (p = 0.02), but not enough to be smaller or, at least, equivalent, to that of the PLIF-BPSS construct (p = 0.02). Conclusions The ISS may be a suitable device to provide immediate flexion-extension balance after a unilateral laminotomy, but the BPSS provides greater immediate stability in lateral bending and axial rotation motions. Both PLIF constructs performed equivalently in flexion-extension and axial rotation, but the PLIF-BPSS construct is more resistant to lateral bending motions. Further biomechanical and clinical evidence is required to strongly support the recommendation of a stand-alone interspinous fusion device or as supplemental fixation to expandable posterior interbody cages.


Neurosurgery ◽  
2010 ◽  
Vol 66 (2) ◽  
pp. E404-E406 ◽  
Author(s):  
Neil R. Crawford ◽  
Şeref Doğan ◽  
K. Zafer Yüksel ◽  
Octavio Villasana-Ramos ◽  
Julio C. Soto-Barraza ◽  
...  

Abstract OBJECTIVE A standard top-loading lumbar pedicle screw-rod system is compared with a pedicle screw-plate system with smaller-diameter screws, more medial entry, and lower profile to assess the relative stability, strength, and resistance to fatigue of the 2 systems. METHODS Seven human cadaveric specimens were studied with each surgical construct. Nondestructive, nonconstraining pure moments were applied to specimens to induce flexion, extension, lateral bending, and axial rotation while recording L5–S1 motion optoelectronically. After initial tests, specimens were fatigued for 10 000 cycles and retested to assess early postoperative loosening. Specimens were then loaded to failure in hyperextension. RESULTS The standard screw-rod construct reduced range of motion to a mean of 20% of normal, whereas the screw-plate construct reduced range of motion to 13% of normal. Differences between systems were not significant in any loading mode (P &gt; 0.06). The 14% loosening of the screw-rod system with fatigue was not significantly different from the 10% loosening observed with the screw-plate system (P &gt; 0.15). Mean failure loads of 30 Nm for screw-rod and 37 Nm for screw-plate were also not significantly different (P = 0.38). CONCLUSION Posterior fixation at L5–S1 using the low-profile screw-plate system offers stability, resistance to fatigue, and resistance to failure equivalent to fixation using a standard cantilevered pedicle screw-rod system.


2017 ◽  
Vol 59 (3) ◽  
pp. 327-335 ◽  
Author(s):  
David Volkheimer ◽  
Fabio Galbusera ◽  
Christian Liebsch ◽  
Sabine Schlegel ◽  
Friederike Rohlmann ◽  
...  

Background Several in vitro studies investigated how degeneration affects spinal motion. However, no consensus has emerged from these studies. Purpose To investigate how degeneration grading systems influence the kinematic output of spinal specimens. Material and Methods Flexibility testing was performed with ten human T12-S1 specimens. Degeneration was graded using two different classifications, one based on X-ray and the other one on magnetic resonance imaging (MRI). Intersegmental rotation (expressed by range of motion [ROM] and neutral zone [NZ]) was determined in all principal motion directions. Further, shear translation was measured during flexion/extension motion. Results The X-ray grading system yielded systematically lesser degeneration. In flexion/extension, only small differences in ROM and NZ were found between moderately degenerated motion segments, with only NZ for the MRI grading reaching statistical significance. In axial rotation, a significant increase in NZ for moderately degenerated segments was found for both grading systems, whereas the difference in ROM was significant only for the MRI scheme. Generally, the relative increases were more pronounced for the MRI classification compared to the X-ray grading scheme. In lateral bending, only relatively small differences between the degeneration groups were found. When evaluating shear translations, a non-significant increase was found for moderately degenerated segments. Motion segment segments tended to regain stability as degeneration progressed without reaching the level of statistical significance. Conclusion We found a fair agreement between the grading schemes which, nonetheless, yielded similar degeneration-related effects on intersegmental kinematics. However, as the trends were more pronounced using the Pfirrmann classification, this grading scheme appears superior for degeneration assessment.


2019 ◽  
Vol 10 (8) ◽  
pp. 973-981
Author(s):  
Raymond J. Hah ◽  
Ram Alluri ◽  
Paul A. Anderson

Study Design: Biomechanics study. Objectives: To evaluate the biomechanical advantage of interfacet allograft spacers in an unstable single-level and 2-level anterior cervical discectomy and fusion (ACDF) pseudoarthrosis model. Methods: Nine single-level and 8 two-level ACDF constructs were tested. Range of motion in flexion-extension (FE), lateral bending (LB), and axial rotation (AR) at 1.5 N m were collected in 4 testing configurations: (1) intact spine, (2) ACDF with interbody graft and plate/screw, (3) ACDF with interbody graft and plate/loosened screws (loose condition), and (4) ACDF with interbody graft and plate/loosened screws supplemented with interfacet allograft spacers (rescue condition). Results: All fixation configurations resulted in statistically significant decreases in range of motion in all bending planes compared with the intact spine ( P < .05). One Level. Performing ACDF with interbody graft and plate on the intact spine reduced FE, LB, and AR 60.0%, 64.9%, and 72.9%, respectively. Loosening the ACDF screws decreased these reductions to 40.9%, 44.6%, and 52.1%. The addition of interfacet allograft spacers to the loose condition increased these reductions to 74.0%, 84.1%, and 82.1%. Two Level. Performing ACDF with interbody graft and plate on the intact spine reduced FE, LB, and AR 72.0%, 71.1%, and 71.2%, respectively. Loosening the ACDF screws decreased these reductions to 55.4%, 55.3%, and 51.3%. The addition of interfacet allograft spacers to the loose condition significantly increased these reductions to 82.6%, 91.2%, and 89.3% ( P < .05). Conclusions: Supplementation of a loose ACDF construct (pseudarthrosis model) with interfacet allograft spacers significantly increases stability and has potential applications in treating cervical pseudarthrosis.


2020 ◽  
Vol 10 (18) ◽  
pp. 6413
Author(s):  
Ji-Won Kwon ◽  
Hwan-Mo Lee ◽  
Tae-Hyun Park ◽  
Sung Jae Lee ◽  
Young-Woo Kwon ◽  
...  

The design and ratio of the cortico-cancellous composition of allograft spacers are associated with graft-related problems, including subsidence and allograft spacer failure. Methods: The study analyzed stress distribution and risk of subsidence according to three types (cortical only, cortical cancellous, cortical lateral walls with a cancellous center bone) and three lengths (11, 12, 14 mm) of allograft spacers under the condition of hybrid motion control, including flexion, extension, axial rotation, and lateral bending,. A detailed finite element model of a previously validated, three-dimensional, intact C3–7 segment, with C5–6 segmental fusion using allograft spacers without fixation, was used in the present study. Findings: Among the three types of cervical allograft spacers evaluated, cortical lateral walls with a cancellous center bone exhibited the highest stress on the cortical bone of spacers, as well as the endplate around the posterior margin of the spacers. The likelihood of allograft spacer failure was highest for 14 mm spacers composed of cortical lateral walls with a cancellous center bone upon flexion (PVMS, 270.0 MPa; 250.2%) and extension (PVMS: 371.40 MPa, 344.2%). The likelihood of allograft spacer subsidence was also highest for the same spacers upon flexion (PVMS, 4.58 MPa; 28.1%) and extension (PVMS: 12.71 MPa, 78.0%). Conclusion: Cervical spacers with a smaller cortical component and of longer length can be risk factors for allograft spacer failure and subsidence, especially in flexion and extension. However, further study of additional fixation methods, such as anterior plates/screws and posterior screws, in an actual clinical setting is necessary.


2009 ◽  
Vol 32 (2) ◽  
pp. 141-151 ◽  
Author(s):  
Pierre-Michel Dugailly ◽  
Stéphane Sobczak ◽  
Victor Sholukha ◽  
Serge Van Sint Jan ◽  
Patrick Salvia ◽  
...  

Neurosurgery ◽  
2001 ◽  
Vol 49 (6) ◽  
pp. 1399-1408 ◽  
Author(s):  
Andrzej Maciejczak ◽  
Michał Ciach ◽  
Maciej Radek ◽  
Andrzej Radek ◽  
Jan Awrejcewicz

ABSTRACT OBJECTIVE To determine whether the Cloward technique of cervical discectomy and fusion increases immediate postoperative stiffness of single cervical motion segment after application of interbody dowel bone graft. METHODS We measured and compared the stiffness of single-motion segments in cadaveric cervical spines before and immediately after interbody fusion with the Cloward technique. Changes in range of motion and stiffness of the C5–C6 segment were measured in a bending flexibility test (flexion, extension, lateral bending and axial rotation) before and after a Cloward procedure in 11 fresh-frozen human cadaveric specimens from the 4th through the 7th vertebrae. RESULTS The Cloward procedure produced a statistically significant increase in stiffness of the operated segment in flexion and lateral bending when compared with the intact spine. The less stiff the segment before the operation, the greater the increase in its postoperative flexural stiffness (statistically significant). The Cloward procedure produced nonuniform changes in rotational and extensional stiffness that increased in some specimens and decreased in others. CONCLUSION Our data demonstrate that Cloward interbody fusion increases immediate postoperative stiffness of an operated segment only in flexion and lateral bending in cadaveric specimens in an in vitro environment. Thus, Cloward fusion seems a relatively ineffective method for increasing the stiffness of a construct. This may add to discussion on the use of spinal instrumentation and postoperative management of patients after cervical discectomy, which varies from bracing in hard collars through immobilization in soft collars to no external orthosis.


2019 ◽  
Vol 9 (8) ◽  
pp. 826-833
Author(s):  
Ripul Panchal ◽  
Anup Gandhi ◽  
Chris Ferry ◽  
Sam Farmer ◽  
Jeremy Hansmann ◽  
...  

Study Design: In vitro biomechanical study. Objectives: The objective of this in vitro biomechanical range-of-motion (ROM) study was to evaluate spinal segmental stability following fixation with a novel anterior cervical discectomy and fusion (ACDF) device (“novel device”) that possesses integrated and modular no-profile, half-plate, and full-plate fixation capabilities. Methods: Human cadaveric (n = 18, C3-T1) specimens were divided into 3 groups (n = 6/group). Each group would receive one novel device iteration. Specimen terminal ends were potted. Each specimen was first tested in an intact state, followed by anterior discectomy (C5/C6) and iterative instrumentation. Testing order: (1) novel device (group 1, no-profile; group 2, half-plate; group 3, full-plate); (2) novel device (all groups) with lateral mass screws (LMS); (3) traditional ACDF plate + cage; (4) traditional ACDF plate + cage + LMS. A 2 N·m moment was applied in flexion/extension (FE), lateral bending (LB), and axial rotation (AR) via a kinematic testing machine. Segmental ROM was tracked and normalized to intact conditions. Comparative statistical analyses were performed. Results: Key findings: (1) the novel half- and full-plate constructs provided comparable reduction in FE and LB ROM to that of traditional plated ACDF ( P ≥ .05); (2) the novel full-plate construct significantly exceeded all other anterior-only constructs ( P ≤ .05) in AR ROM reduction; and (3) the novel half-plate construct significantly exceeded the no-profile construct in FE ( P < .05). Conclusions: The novel ACDF device may be a versatile alternative to traditional no-profile and independent plating techniques, as it provides comparable ROM reduction in all principle motion directions, across all device iterations.