scholarly journals Effects of Simultaneous Liquid or Gel Sodium Hypochlorite Irrigation on the Cyclic Fatigue of Two Single-File Nickel-Titanium Instruments

2020 ◽  
Vol 10 (19) ◽  
pp. 6666
Author(s):  
Eugenio Pedullà ◽  
Giusy Rita Maria La Rosa ◽  
Marco Sesto Albani ◽  
Gaetano Isola ◽  
Taha Özyürek ◽  
...  

To evaluate the effect of simultaneous liquid or gel sodium hypochlorite (NaOCl) irrigation on cyclic fatigue of F6 SkyTaper (F6ST) and OneCurve (OC) single files, 180 new 25/0.06 F6ST and OC files were divided into 6 groups (n = 15) for each brand. Groups 1 and 4 included new instruments not exposed to NaOCl at 20 °C and 37 °C, respectively. Groups 2 and 5 included files activated with liquid NaOCl at 20 °C and 37 °C, respectively. Groups 3 and 6 consisted of instruments tested with NaOCl gel at 20 °C and 37 °C, respectively. Instruments were subjected to a fatigue test using a novel customized device. Data were expressed as time to fracture (TtF) and statistically analyzed (p < 0.05) after checking their normality through the Shapiro–Wilk test. Because they were normally distributed, 2-way analyses of variance (ANOVA) and the Tukey multiple comparison post-hoc test were used. Time to fracture of all tested instruments decreased at 37 °C (p < 0.05). At 20 °C, NaOCl improved TtF of F6ST and OC (p < 0.05). NaOCl liquid increased TtF of F6ST (p < 0.05) in comparison with gel, while there was no difference between the two formulations for OC. At 37 °C, both NaOCl formulations had no significant influence on TtF for F6ST, while they increased TtF of OC (p < 0.05). NaOCl improved the cyclic fatigue resistance of OC, independently of the temperature, while for F6ST the negative impact of higher temperature reduced the irrigant benefits.

2020 ◽  
Vol 2020 ◽  
pp. 1-5
Author(s):  
Hind F. Abuhulaibah ◽  
Ammar AbuMostafa

Background/purpose. To assess the resistance to cyclic fatigue (CF) displayed by two nickel-titanium (NiTi) files, One Curve (OC) and ProTaper Gold (PTG), after their immersion in 2.5% sodium hypochlorite (NaOCl) solution at body temperature. Materials and Methods. Forty-five files from each brand were randomly assigned to three groups (n = 15) and subjected to the following: no immersion (control), 1-minute immersion, and 5-minute immersion in 2.5% NaOCl at 37°C. CF for all the files was tested within a well-lubricated stainless-steel artificial canal in a water bath at 37°C simulating body temperature. The procedure was video recorded, and the number of cycles to fracture (NCF) was calculated by multiplying the time taken to fracture, with the number of rotations per second. The data were analyzed for normality, and thereafter, one-way ANOVA with multiple Bonferroni was used as the post hoc test. Results. The resistance to CF of OC files was significantly higher than PTG files in all groups. In groups immersed in NaOCl for one minute, OC files showed significant drop in the CF resistance; PTG files showed a drop as well but not significantly. Both files demonstrated a significant drop in resistance to CF after immersion in NaOCl for five minutes compared to the control group. Conclusion. OC files were significantly more resistant to CF compared to PTG in all groups. Immersion in 2.5% NaOCl for 5 minutes significantly reduced the resistance to CF for all the files.


Author(s):  
Giusy Rita Maria La Rosa ◽  
Carlotta Palermo ◽  
Sebastiano Ferlito ◽  
Gaetano Isola ◽  
Francesco Indelicato ◽  
...  

2019 ◽  
Vol 13 (1) ◽  
pp. 57-60
Author(s):  
Damla Özsu Kırıcı ◽  
Ertuğrul Karataş ◽  
Ahmet Demirhan Uygun ◽  
Ezgi Doğanay Yıldız ◽  
Kezban Meltem Çolak ◽  
...  

Background. The aim of the present study was to compare the cyclic fatigue resistance of novel nickel titanium rotary pathfinding instruments. Methods. Twenty instruments were selected for each file system. A simulated stainless steel root canal, with a 90° angle of curvature and a curvature radius of 3 mm, was used for cyclic fatigue test of the ProGlider (#16, progressive taper: 0.02‒ 0.085), PathGlider (#15, taper: .03), and One G (#14, taper: .03) instruments. Statistical analyses were performed with oneway ANOVA (P=0.05). Post hoc Tukey tests were used to determine any statistically significant differences between the groups. Results. The ProGlider instruments exhibited significantly more cyclic fatigue resistance than both PathGlider and One G instruments (P<0.001). One G instruments had significantly more resistance to fracture than PathGlider instruments (P<0.05). Conclusion. ProGlider instruments had better cyclic fatigue resistance than PathGlider and One G instruments.


2020 ◽  
Vol 10 (16) ◽  
pp. 5604 ◽  
Author(s):  
Gianluca Gambarini ◽  
Andrea Cicconetti ◽  
Dario Di Nardo ◽  
Gabriele Miccoli ◽  
Alessio Zanza ◽  
...  

Protaper Universal (PTU), Protaper Gold (PTG) (Maillefer, Ballaigues, CH), EdgeTaper (ET), and EdgeTaper Platinum (ETP) (Albuquerque, NM, USA) were tested for both torsional and flexural resistance. The aim of the present study was to evaluate the influence of proprietary heat treatment on the metallurgical properties of the aforementioned instruments. Four groups of 30 different instruments (size 20.07) were tested, then divided into two subgroups of 15 instruments—one for the cyclic fatigue test in a curved canal (90°—2 mm radius) at 300 rpm and 2.5 Ncm. The time to fracture (TtF) and fragment length (FL) were recorded. The other subgroup was subjected to the torsional test (300 rpm, 5.5 Ncm). The torque to fracture and TtF were recorded. All the instruments underwent a SEM analysis. The heat-treated instruments showed a significantly higher fatigue resistance than the non-heat-treated instruments (p < 0.05). No significant differences were found in the torsional resistance between the ET and PTU, and the ETP and PTG. However, when comparing all the groups, the heat-treated instruments showed less torsional resistance. The improvement from heat treatment was mainly found in the cyclic fatigue resistance.


Materials ◽  
2019 ◽  
Vol 12 (16) ◽  
pp. 2523 ◽  
Author(s):  
Gianluca Gambarini ◽  
Gabriele Miccoli ◽  
Marco Seracchiani ◽  
Tatyana Khrenova ◽  
Orlando Donfrancesco ◽  
...  

The aim of this study was to investigate the role of the flat-designed surface in improving the resistance to cyclic fatigue by comparing heat-treated F-One (Fanta Dental, Shanghai, China) nickel–titanium (NiTi) rotary instruments and similar prototypes, differing only by the absence of the flat side. The null hypothesis was that there were no differences between the two tested instruments in terms of cyclic fatigue lifespan. A total of 40 new NiTi instruments (20 F-One and 20 prototypes) were tested in the present study. The instruments were rotated with the same speed (500 rpm) and torque (2 N) using an endodontic motor (Elements Motor, Kerr, Orange, CA, USA) in the same stainless steel, artificial canal (90° angle of curvature and 5 mm radius). A Wilcoxon–Mann–Whitney test was performed to assess the differences in terms of time to fracture and the length of the fractured segment between the flat- and non-flat-sided instruments. Significance was set at p = 0.05. The differences in terms of time to fracture between non-flat and flat were statistically significant (p < 0.001). In addition, the differences in terms of fractured segment length were statistically significant (p = 0.034). The results of this study highlight the importance of flat-sided design in increasing the cyclic fatigue lifespan of NiTi rotary instruments.


Author(s):  
Zuha Ayad Jaber ◽  
Sattar Jabbar Abdul-Zahra Al-Hmedat ◽  
Sarmad Adel Hameed ◽  
Suhad Jabbar Hamed Al-Nasrawi ◽  
Abtesam Imhemed Aljdaimi ◽  
...  

2021 ◽  
Author(s):  
Hamed Karkehabadi ◽  
Abbas Farmani ◽  
Zahra Pakseresht ◽  
Faraz Sedaghat ◽  
Saber Yavari Niya

Abstract Objective: This study sought to assess the effect of cryogenic treatment on cyclic fatigue of Neoniti and Reciproc nickel-titanium (NiTi) rotary files. This in vitro, experimental study was performed on 48 Neoniti and Reciproc NiTi rotary files (#25, 6% taper) in two subgroups with and without cryogenic treatment. The files in cryogenic subgroups were stored in liquid nitrogen chamber at -196°C for 24 h. Next, they were placed at room temperature (25°C) to gradually warm up. Stainless steel (SS) blocks were used to standardize the degree of rotation of the files. The SS blocks simulated a root canal with the negative pattern of gutta-percha with 0.08 taper, #25 tip size, and 45° canal curvature. Each endodontic file was rotated in this canal until fracture. The time until fracture was recorded by a digital chronometer. Time to fracture was used to calculate the number of rotations before file fracture as the cyclic fatigue scale. Results: The mean cyclic fatigue resistance of both Neoniti and Reciproc rotary files in cryogenic subgroups was significantly higher than that in non-cryogenic control subgroups (P<0.05). Cryogenic treatment can significantly increase the cyclic fatigue resistance of Neoniti and Reciproc NiTi rotary files.


2018 ◽  
Vol 44 (10) ◽  
pp. 1563-1566 ◽  
Author(s):  
Hussam Alfawaz ◽  
Abdullah Alqedairi ◽  
Hala Alsharekh ◽  
Eman Almuzaini ◽  
Shahd Alzahrani ◽  
...  