scholarly journals RP-HPLC method development and validation for simultaneous estimation of efonidipine hydrochloride ethanolate and telmisartan in their synthetic mixture

Author(s):  
Grishma H Patel ◽  
Shreya D Adeshra ◽  
Dhananjay B Meshram

A Novel, selective, accurate and rapid Reversed Phase High Performance Liquid Chromatographic (RPHPLC) method for the analysis of Efonidipine Hydrochloride Ethanolate and Telmisartan in binary mixture has been developed and validated. The chromatographic system consisted of a Phenomenex Kinetex ® 5µ C18 Size: 150 * 4.6mm column and the separation was achieved by using ambient temperature with a mobile phase containing mobile Phase Acetonitrile:25mM Phosphate Buffer pH 4.9 (45:55). The samples were monitored at 253 nm for detection at a flow rate of 1.0 mL/min and the retention time was about 7.77 and 4.10 mins for Efonidipine Hydrochloride Ehanolate and Telmisartan respectively. The calibration curve was linear over the concentration range 5-30 and 10-60 ?g/mL for Efonidipine Hydrochloride Ehanolate and Telmisartan respectively. The proposed method is accurate in the range of 99.75% - 100.10% recovery and precise (%RSD of intraday variation and % RSD of inter day variation were found to be within the acceptance criteria). Therefore, this method can be used as a more convenient and efficient option for the analysis of Efonidipine Hydrochloride Ehanolate and Telmisartan in Quality control laboratory.

INDIAN DRUGS ◽  
2019 ◽  
Vol 56 (02) ◽  
pp. 39-46
Author(s):  
Babu D. China ◽  
◽  
C. Madhusudhana Chetty ◽  
Sk. Mastanamma

A simple, accurate, precise and robust reversed phase high performance liquid chromatographic (RP-HPLC) method was developed for the estimation of Ertugliflozin (ETZ) and metformin (MFN) in bulk and in tablet dosage form. The method was carried out by used Waters (5μm, C18 250 x 4.6 mm) column with mobile phase consists of 0.75 mM sodium dihydrogen orthophosphate buffer pH adjusted to 8.5, with NaOH, and acetonitrile in the ratio of 60:40 v/v, a flow rate of 1.5 mL/min and photodiode detection at 263 nm. The method was validated as per ICH guidelines with different parameters, the mean retention times of ertugliflozin and metformin were found to be 3.5& 2.0 min, respectively. The correlation coefficient values of calibration curves were found to be 0.999 for both ETZ and MFN, respectively. The LOD and LOQ for ertugliflozin and metformin were found to be 0.02-0.06 μg/mL and 17.5-58.3 μg/mL respectively.


INDIAN DRUGS ◽  
2015 ◽  
Vol 52 (08) ◽  
pp. 12-16
Author(s):  
S Vidyadhara ◽  
◽  
L. S Reddyvalam ◽  
T. Koduri ◽  
P. K. Borra ◽  
...  

A simple, accurate, precise high-performance liquid chromatographic (HPLC) method has been developed and validated for the simultaneous determination of abacavir sulphate (ABA) and lamivudine (LAM) in combined dosage form. Separation was performed on a C18 column [Agilent ODS UG 5 column, 250 mm x 4.5 mm], with methanol: water (50:50 V/V) isocratic elution using a flow rate of 1mL/min. Good sensitivity was observed with UV detection at 277 nm. After method development, the interference of other active compounds and excipients, repeatability and linearity, were investigated. Retention times of LAM and ABA were found to be 3.3 and 6.3 min, respectively. The method was validated over the range from 2.5-12.5 μg/mL for LAM and 5-25 μg/mL for ABA with correlation coefficients of 0.9997 and 0.9996, respectively. This method was shown to be accurate, robust, selective, linear, and repeatable and can be successfully employed in routine quality control for the simultaneous analysis of ABA and LAM in tablets.


2018 ◽  
Vol 8 (6-s) ◽  
pp. 78-82 ◽  
Author(s):  
Vishal Singh Solanki ◽  
RAM SINGH BISHNOI ◽  
Raviraj Baghel ◽  
Deepti Jain

A simple precise and economical reverse phase high performance liquid chromatographic method has been developed and validated for the simultaneous estimation of Cilnidipine (CDP), Atenolol (ATL) and Chlorthalidone (CTD).The chromatographic separation was achieved by using Hypersil- keystone C18 (4.6 x 250mm, 5μm) under isocratic conditions The mobile phase consisted of methanol and triple distilled water (80/20, v/v) having pH 7 with a flow rate of 1.0 mL/min. The eluents were monitored at 225 nm for simultaneous measurement.The selected chromatographic conditions were found to effectively separate CDP (Rt: 3.25 min), ATL (Rt: 5.366 min) and CTD (Rt: 9.025 min) having good resolution. The developed method was validated for linearity, accuracy, precision, LOD, LOQ, robustness and for system suitability parameters as per ICH guidelines. In this study, an excellent linearity was obtained with r2 = 0.999, r² = 0.999, r² = 1, for CDP, ATL and CTD respectively. The developed chromatographic method proved to be simple, precise, accurate, robust and reproducible Thus, this method would be employed for routine simultaneous quantification of CDP, ATL and CTD in bulk form or tablet dosage form.   Keywords: Cilnidipine, Atenolol and Chlorthalidone, RP-HPLC.


2017 ◽  
Vol 1 (2) ◽  
pp. 1-8
Author(s):  
Milena Cristina Ribeiro Souza Magalhães ◽  
Alisson Samuel Portes Caldeira ◽  
Hanna De Sousa Rocha Almeida ◽  
Sílvia Ligório Fialho ◽  
Armando Da Silva Cunha Junior

A reversed-phase high-performance liquid chromatographic (HPLC) method was developed and validated for the determination of encapsulation efficiency of zidovudine in nanoparticules. The method was carried out in isocratic mode using 0.040M sodium acetate: methanol: acetonitrile: glacial acetic acid (880:100:20:2) as mobile phase, a C8 column at 25ºC and UV detection at 240 nm. The method was linear (r2 ˃ 0.99) over the range of 25.0-150.0 μg/mL, precise (RSD ˂ 5%), accurate (recovery = 100.5%), robust and selective. The validated HPLC-UV method can be successfully applied to determine the rate of zidovudine in nanoparticules.


2019 ◽  
Vol 9 (01) ◽  
pp. 83-88
Author(s):  
Pinkal Patel ◽  
Nalini Patel ◽  
Kinjal Parmar

A simple, selective and rapid reversed phase High Performance Liquid Chromatographic (RP-HPLC) method has been developed and validated for the simultaneous analysis Gatifloxacin and flurbiprofen sodium in eye drops. The separation was carried out using a mobile phase consisting ACN: Buffer (pH 3.5) in the ratio of 55:45 v/v. The column used was Phenomenex luna ODS C18 (250mm X 4.6 mm i.d., 5 μm particle size) with flow rate of 1 ml / min using UV detection at 268 nm. The described method was linear over a concentration range of 2-12 μg/ml for both of Gatifloxacin and flurbiprofen sodium. The retention times of Gatifloxacin and flurbiprofen sodium were found to be 3.710 min. and 6.797 min respectively. Method was validated statistically and recovery studies were carried out. The proposed method has been applied successfully to the analysis of cited drugs either in pure form or in pharmaceutical formulations with good accuracy and precision. The method here in described can be employed for quality control and routine analysis of drugs in pharmaceutical formulations.


Author(s):  
Pooja Agarwal ◽  
Beena Kumari ◽  
Sangeeta Sangeeta ◽  
Manish Kumar

The combination of Domperidone and Lansoprazole is very useful in Gastro-esophageal disinfection (Dyspepsia). These methods provide means to separate the components characterize and quantify the components. An accurate, precise, specific and simple HPLC method was developed for simultaneous estimation of Domperidone and Lansoprazole. By this method retention time, linearity and accuracy data is respectively found for Domperidone and Lansoprazole. Mobile phase was prepared by mixing 51 volume of Acetonitrile and 49 volume of Ammonium Acetate (51:49 V/V) then 25mg each of Domperidone and Lansoprazole was dissolved in small volume of Acetonitrile: Ammonium Acetate (51:49 V/V) separately. Retention time was recorded 4.330 ± 0.003 minute and 5.820 ± 0.003 minute for Domperidone and Lansoprazole with 1.0 ml/min flow rate. The low value of % R.S.D indicates that this method is precise and accurate. Thus it can be concluded that the proposed method was good approach for obtaining reliable result.


INDIAN DRUGS ◽  
2018 ◽  
Vol 55 (11) ◽  
pp. 57-63
Author(s):  
V. Suryadevara ◽  
◽  
R. L Sasidhar ◽  
B. Venkateswara Rao ◽  
T. N. V. Ganesh Kumar ◽  
...  

A simple, accurate reverse phase high performance liquid chromatographic method has been developed for the simultaneous estimation of lamivudine, tenofovir disoproxil fumarate and efavirenz in bulk and pharmaceutical formulations. The analytical method development was carried on Agilent make HPLC instrument using RP - C18 column. The mobile phase employed for the estimation is phosphate Buffer pH 4.0 :acetonitrile adjusted to pH 4.0 with glacial acetic acid which was pumped at a flow rate of 1.0 mL min-1 in the ratio of 42:58 v/v. the eluents were monitored at 260 nm. Linearity was obtained in the concentration range of 20-100 μg/mL of lamivudine, tenofovir disoproxil fumarate and 100-500 μg/mL efavirenz. Degradation studies shows that all the three drugs were not degraded under acidic, alkaline, thermal and photolytic conditions.The method was statistically validated and RSD was found to be within limits. Due to its simplicity, rapidness, high precision and accuracy, the proposed HPLC method can be applied for determining lamivudine, tenofovir disoproxil fumarate and efavirenz in bulk and in pharmaceutical dosage form.


INDIAN DRUGS ◽  
2021 ◽  
Vol 58 (4) ◽  
pp. 78-81
Author(s):  
Rajesh Sharma ◽  
◽  
Mukesh C. Sharma ◽  
Gaurav Vijaywargiya

A simple, specific, accurate reversed phase high performance liquid chromatographic method was developed for the simultaneous estimation of clidinium bromide, chlordiazepoxide and dicyclomine hydrochloride. Chromatographic separation of the three drugs was performed on a Chromatopak C-18 column (25 cm x 4.6 i.d. x 5µm) as the stationary phase with a mobile phase composed of 0.1 % triethylamine in water pH adjusted by 5 % o-phosphoric acid and acetonitrile in the ratio 30:70 at a flow rate of 0.8mL/min, Detection was carried out at 210 nm. The retention times of clidinium bromide, chlordiazepoxide and, dicyclomine hydrochloride were found to be 3.9 min, 5.4 min, and 6.8 min, respectively. The proposed method was validated for linearity, accuracy, precision, LOD and LOQ.


2020 ◽  
Vol 13 (3) ◽  
pp. 206-216
Author(s):  
Madhusudan Bhoir ◽  
Nutan Rao

The objective of the study was to develop and validate a novel, stability indicating, simple, rapid, accurate, precise and isocratic reverse-phase high-performance liquid chromatographic (RP-HPLC) method for simultaneous estimation of benserazide HCl and levodopa in a marketed formulation. Chromatographic separation was achieved by using C18 Cosmosil 4.6 × 250 mm column with a mixture of phosphate buffer pH 2 and acetonitrile in proportion of 95:5 as mobile phase at a flow rate of 1.0 ml/min and column temperature 25°C. The detection was carried out at 210 nm using UV detector. The retention time for benserazide and levodopa was found to be 3.1 minutes and 6.6 minutes respectively and recoveries from tablet were between 98 and 102 %.


Pharmacia ◽  
2019 ◽  
Vol 66 (1) ◽  
pp. 1-5 ◽  
Author(s):  
Lily Peikova ◽  
Diana Tzankova ◽  
Alexandrina Dineva ◽  
Maya Georgieva ◽  
Alexander Zlatkov

A reversed phase enantio selective high performance liquid chromatographic method was developed for enantiomeric separation of 2-(1,3-dimethyl-2,6-dioxo-2,3-dihydro-1H-purine-7(6H)-yl)-N'-(3-fluorobenzylidene)-propanehydrazide isomers. The enantiomers of 2-(1,3-dimethyl-2,6-dioxo-2,3-dihydro-1H-purine-7(6H)-yl)-N'-(3-fluorobenzylidene)propanehydra-zide were resolved on a ACEÒEquivalenceTM С18 (250 × 4.6 mm, 5 μm) column using a mobile phase system containing methanol, water, phosphate buffer рН 7.4 (50:46:4 v/v/v). The resolution between enantiomers was found to be more than 2.0. The sample solution and mobile phase were found to be stable for at least 48 h. The final optimized method was successfully applied to separate the (R)- and the (S)-enantiomers of 2-(1,3-dimethyl-2,6-dioxo-2,3-dihydro-1H-purine-7(6H)-yl)-N'-(3-fluorobenzylidene)-propanehydrazide and was proven to be reproducible and accurate.


Sign in / Sign up

Export Citation Format

Share Document