A Collaborative Study of a Proposed International Standard for Tissue Plasminogen Activator (t-PA)

1985 ◽  
Vol 53 (01) ◽  
pp. 134-136 ◽  
Author(s):  
P J Gaffney ◽  
A D Curtis

SummaryAn international collaborative study involving seven laboratories was undertaken to assess which of three lyophilised preparations might serve as an International Standard (I.S.) for tissue plasminogen activator (t-PA). Two of the preparations were isolates from human melanoma cell cultures while one was of pig heart origin. A clot lysis assay was used by all participants in the study.The data suggested that both preparations of human cell origin were comparable, in that their log dose-response lines were parallel, while that of the porcine preparation was not. Accelerated degradation studies indicated that one melanoma extract (denoted 83/517) was more stable than the other and it was decided to recommend preparation 83/517 as the standard for t-PA. The International Committee for Thrombosis and Haemostasis (Stockholm 1983) has recommended the use of this material as a standard and it has been established by the Expert Committee on Biological Standardization of the World Health Organization as the International, Standard for tissue plasminogen activator, with an assigned potency of 1000 International Units per ampoule.

1987 ◽  
Vol 58 (04) ◽  
pp. 1085-1087 ◽  
Author(s):  
P J Gaffney ◽  
A D Curtis

SummaryAn international collaborative study involving ten laboratories located in eight different countries was undertaken in order to replace the current International Standard (I.S.) for tissue plasminogen activator (t-PA). Two lyophilised candidate preparations of high purity were assessed in comparison with the current I.S. for t-PA using only a clot lysis assay. One preparation (coded 861670) was purified from a cultured melanoma cell supernatant and was about 98% single chain t-PA while the other preparation (coded 861624) was derived from Chinese hamster ovary (CHO) cells following DNA recombinant procedures and was 75% single chain t-PA.Both candidate preparations of t-PA compared in quite a satisfactory manner with the current I.S. from the viewpoint of the biometrics of parallel line bioassays and both preparations were quite stable for long periods at low temperatures and stable from up to 1 month at temperatures of 20° and 38° C. Both fultil the criteria to serve as a satisfactory Znd International Standard for t-PA. The Fibrinolysis Subcommittee of the International Committee for Thrombosis and Haemostasis recommended the melanoma source t-PA (861670) as the next I.S. in order to maintain continuity with the 1st I.S. which was also a melanomatype preparation. The data from the ten laboratories indicated that each ampoule of the new proposed standard contains 850 international units of t-PA activity by the clot lysis assay. It is planned to present the results of this study to the Expert Committee on Biological Standardization of the World Health Organization at its next meeting and to request that the preparation of t-PA, coded 861670, be established as the 2ndlnternational Standard for t-PA.


2002 ◽  
Vol 88 (08) ◽  
pp. 294-297 ◽  
Author(s):  
Dawn Sands ◽  
Colin Whitton ◽  
R. Merton ◽  
Colin Longstaff

SummaryAn international collaborative study was organised to replace the 2nd International Standard (IS) for tissue plasminogen activator (tPA). The 2nd IS for tPA (86/670) was used to calibrate the replacement Standard, which was selected from two candidate materials included in the collaborative study. Participants were provided with five sets of four samples (A, B, C, D) and asked to use sample A (2nd IS, 86/670, 850 IU/ml) to determine the activity of B (86/624, approximately 850 IU/ml), C and D (coded duplicates of the same material, 98/714 approximately 11000 IU/ml). A total of 14 laboratories returned results from Europe, USA, Japan and Australia, providing data from 60 independent assays. Four laboratories used a reference method based on a published monograph from the European Pharmacopoeia for Alteplase for Injection, 1998, and the remaining 10 used their own method. Fibrin was used as promoter of tPA activity by 12 out of the 14 laboratories, the remaining two used kits where fibrinogen fragments were the promoter. Data from this collaborative study and the previous study to establish the 2nd IS for tPA show that tPA from melanoma cells and recombinant tPA from CHO cells are both suitable materials as International Standards. It was agreed that sample C, D, recombinant tPA, 98/714, be established as the 3rd International Standard for tPA with a potency of 10000 IU per ampoule, calculated as the mean value from laboratories using fibrin as a promoter of tPA activity. The standard was established by WHO in November 2000.


Author(s):  
Susan J. Thorpe ◽  
Dawn Sands ◽  
Alan B. Heath ◽  
Malcolm S. Hamilton ◽  
Sheena Blackmore ◽  
...  

AbstractFolate measurements, particularly for whole blood, show wide inter-laboratory and inter-methodology variability. This variability appears to be due in part to the lack of internationally accepted reference materials. A whole blood haemolysate, lyophilised in ampoules and designated 95/528, was therefore evaluated by 15 laboratories in five countries for its suitability as an International Standard (IS) for whole blood folate. The preparation was assayed using a variety of microbiological and protein-binding methodologies against local standards and calibrators. A consensus folate content was assigned to 95/528. The inclusion of three whole blood samples in the study with widely differing folate levels demonstrated a considerable reduction in inter-laboratory variability when the folate content of the samples was determined relative to the proposed IS 95/528 rather than to laboratories' local standards and calibrators. Accelerated degradation studies indicated that the folate content of 95/528 is stable when stored at −20°C. On the basis of the results presented here, the World Health Organization Expert Committee on Biological Standardization established 95/528 as an IS for whole blood folate.


1989 ◽  
Vol 121 (1) ◽  
pp. 157-NP ◽  
Author(s):  
D. Schulster ◽  
R. E. Gaines Das ◽  
S. L. Jeffcoate

ABSTRACT Three ampouled preparations of purified human prolactin were assessed by 20 laboratories in eight countries for their suitability to serve as International Standards for the estimation of human prolactin in serum. Bioassays (pigeon crop sac assays and NB2 cell assays) were carried out in two laboratories, radioreceptor assays by one laboratory and radioimmunoassays by 17 laboratories. By physicochemical analysis the preparations appeared similar. Each preparation contained small amounts of contaminants and/or prolactin variants. No major differences among the three preparations were detected by immunoassay although, in one radioreceptor assay system, one of the preparations was found to differ from the other two. On the basis of all the available information, the Expert Committee on Biological Standardization of the World Health Organization (ECBS) in 1986 established the preparation in ampoules coded 83/562 as the Second International Standard for Prolactin and in October 1988 established the preparation in ampoules coded 84/500 as the Third International Standard for Prolactin. A value of 0·053 IU (53 mIU) prolactin activity/ampoule was assigned to both the Second and Third IS on the basis that this unitage would, insofar as possible, maintain continuity of the IU defined by the First International Reference Preparation of Prolactin, human, for Immunoassay (coded 75/504). Journal of Endocrinology (1989) 121, 157–166


Biomolecules ◽  
2021 ◽  
Vol 11 (11) ◽  
pp. 1610
Author(s):  
Haiyan Jia ◽  
Parvathy Harikumar ◽  
Eleanor Atkinson ◽  
Peter Rigsby ◽  
Meenu Wadhwa

Several Bevacizumab products are approved for clinical use, with many others in late-stage clinical development worldwide. To aid the harmonization of potency assessment across different Bevacizumab products, the first World Health Organization (WHO) International Standard (IS) for Bevacizumab has been developed. Two preparations of a Bevacizumab candidate and comparator were assessed for their ability to neutralize and bind vascular endothelial growth factor (VEGF) using different bioassays and binding assays in an international collaborative study. Relative potency estimates were similar across different assays for the comparator or the duplicate-coded candidate sample. Variability in relative potency estimates was reduced when the candidate standard was used for calculation compared with various in-house reference standards, enabling harmonization in bioactivity evaluations. The results demonstrated that the candidate standard is suitable to serve as an IS for Bevacizumab, with assigned unitages for VEGF neutralization and VEGF binding activity. This standard coded 18/210 was established by the WHO Expert Committee on Biological Standardization, which is intended to support the calibration of secondary standards for product development and lifecycle management. The availability of IS 18/210 will help facilitate the global harmonization of potency evaluation to ensure patient access to Bevacizumab products with consistent safety, quality and efficacy.


1990 ◽  
Vol 64 (03) ◽  
pp. 398-401 ◽  
Author(s):  
P J Gaffney ◽  
A B Heath

SummaryAn international collaborative study involving eleven laboratories located in eight countries was undertaken to establish an International Standard for high molecular weight urinary-type plasminogen activator (HMW/u-PA). The current International Reference Preparation (IRP code numbered 66146) for urinarytype plasminogen activator (u-PA) ot urokinase (see Nomenclature footnote) is a 66/34 molar ratio mixture of low molecular weight (LMW) - and high molecular weight (HMW) - u-PA’s and is considered unsuitable as a standard for homogeneous preparations of HMW/u-PA. The putative standard for HMWu-PA (code number, 87/594) was compared for potency in a clot lysis assay with the current IRP for u-PA (code numbet, 66146) and a lyophilised preparation of single chain urinary-type plasminogen activator (SCuPA), the latter being used in the assay without prior activation by plasmin to its active two chain form (TCuPA).Both the proposed standard for HMWu-PA (871594) and the SCuPA compared in a statistically satisfactory manner in parallel line bioassays with the current IRP for u-PA (66146), thus allowing potency estimates to be obtained for these two materials in relation to defined international units. Data from the eleven laboratories indicated that each ampoule of the proposed standard for HMWu-PA contained 4,300 i. u. of activity and was stable for over 1 year at 4° C. Most participants indicated that SCuPA expressed only a small amount of its activity without a prior activator step and this suggests that SCuPA assays need to be preceded by a plasmin activation stepThe Expert Committee on Biological Standardization of the World Health Organisation (Geneva, Oct 1989) established 871 594 as the International Standard for high molecular weight two chain urinary type plasminogen activator (HMW/TCuPA, of HMWu-PA), with an assigned unitage of 4,300 international units per ampoule.


1990 ◽  
Vol 64 (02) ◽  
pp. 267-269 ◽  
Author(s):  
A B Heath ◽  
P J Gaffney

SummaryAn International Standard for Streptokinase - Streptodomase (62/7) has been used to calibrate high purity clinical batches of SK since 1965. An international collaborative study, involving six laboratories, was undertaken to replace this standard with a high purity standard for SK. Two candidate preparations (88/826 and 88/824) were compared by a clot lysis assay with the current standard (62/7). Potencies of 671 i.u. and 461 i.u. were established for preparations A (88/826) and B (88/824), respectively.Either preparation appeared suitable to serve as a standard for SK. However, each ampoule of preparation A (88/826) contains a more appropriate amount of SK activity for potency testing, and is therefore preferred. Accelerated degradation tests indicate that preparation A (88/826) is very stable.The high purity streptokinase preparation, coded 88/826, has been established by the World Health Organisation as the 2nd International Standard for Streptokinase, with an assigned potency of 700 i.u. per ampoule.


1992 ◽  
Vol 67 (04) ◽  
pp. 424-427 ◽  
Author(s):  
P J Gaffney ◽  
A B Heath ◽  
J W Fenton II

SummarySince 1975 an International Standard for Thrombin of low purity has been used. While this standard was stable and of value for calibrating thrombins of unknown potency the need for a pure a-thrombin standard arose both for accurate calibration and for precise measurement of thrombin inhibitors, notably hirudin. An international collaborative study was undertaken to establish the potency and stability of an ampouled pure a-thrombin preparation. A potency of 97.5 international units (95% confidence limits 86.5-98.5) was established for the new a-thrombin standard (89/ 588) using a clotting-assay procedure. Stability data at various elevated temperatures indicated that the standard could be transported and stored with no significant loss of potency.Ampoules of lyophilised a-thrombin (coded 89/588) have been recommended as an International Standard for a-thrombin with an assigned potency of 100 international units per ampoule by the International Society for Thrombosis and Haemostasis (Thrombin and its Inhibitors Sub-Committee) in Barcelona, Spain in July 1990 while the Expert Committee on Biological Standardisation and Control of the World Health Organisation will consider its status at its next meeting in Geneva in 1991.


1983 ◽  
Vol 50 (03) ◽  
pp. 697-702 ◽  
Author(s):  
T W Barrowcliffe ◽  
A D Curtis ◽  
D P Thomas

SummaryAn international collaborative study was carried out to establish a replacement for the current (2nd) international standard for Factor VIII: C, concentrate. Twenty-six laboratories took part, of which 17 performed one-stage assays, three performed two-stage assays and six used both methods. The proposed new standard, an intermediate purity concentrate, was assayed against the current standard, against a high-purity concentrate and against an International Reference Plasma, coded 80/511, previously calibrated against fresh normal plasma.Assays of the proposed new standard against the current standard gave a mean potency of 3.89 iu/ampoule, with good agreement between laboratories and between one-stage and two- stage assays. There was also no difference between assay methods in the comparison of high-purity and intermediate purity concentrates. In the comparison of the proposed standard with the plasma reference preparation, the overall mean potency was 4.03 iu/ampoule, but there were substantial differences between laboratories, and the two-stage method gave significantly higher results than the one stage method. Of the technical variables in the one-stage method, only the activation time with one reagent appeared to have any influence on the results of this comparison of concentrate against plasma.Accelerated degradation studies showed that the proposed standard is very stable. With the agreement of the participants, the material, in ampoules coded 80/556, has been established by the World Health Organization as the 3rd International Standard for Factor VIII :C, Concentrate, with an assigned potency of 3.9 iu/ampoule.


1987 ◽  
Author(s):  
I Keber ◽  
K Potisk ◽  
D Keber ◽  
M Stegnar ◽  
N Vene

To determine the origin of tissue plasminogen activator (t-PA) release during physical activity, we studied the separate and combined effects of venous occlusion and acute physical activity on t-PA release in arm and leg. In 15 healthy volunteers 20 min venous occlusions of arm and leg were performed simultaneously before physical activity ( maximal stress testing on treadmill)(occlusion I), immediately after physical activity and 45 min later (occlusion II). Blood samples were drawn from unoccluded arm before occlusion and after physical activity, and from occluded arm and leg after occlusion. Fibrinolytic activity was measured by euglobulin clot lysis time (ECLT) and t-PA activity assay. The amount of released t-PA during different stimuli (fibrinolytic potential) was calculated as the difference between post- and prestimulation fibrinolytic activity. Before physical activity there was a great increase in fibrinolytic activity due to t-PA in the occluded arm but no increase in the occluded leg. Physical activity itself caused a similar increase of systemic fibrinolytic activity as arm occlusion locally. After physical activity arm occlusion evoked equally good response than before it. Fibrinolytic activity during leg occlusion behaved differently: there was an increase in t-PA activity in the occluded leg which persisted one hour after physical activity, when systemic fibrinolytic activity already fell to initial level.These results demonstrated that walking and running triggered t-PA release from the leg vessels. Since leg occlusion was not a stimulus for t-PA release, it served only as a method to demonstrate the effect of physical activity.


Sign in / Sign up

Export Citation Format

Share Document