scholarly journals Performance and Operational Evaluation of the Access Bio CareStart Rapid Antigen Test in a High-throughput Drive-through Community Testing Site in Massachusetts

Author(s):  
Nira R. Pollock ◽  
Kristine Tran ◽  
Jesica R. Jacobs ◽  
Amber E. Cranston ◽  
Sita Smith ◽  
...  

AbstractBackgroundTo facilitate deployment of point-of-care testing for SARS-CoV-2, we evaluated the Access Bio CareStart COVID-19 Antigen test in a high-throughput, drive-through, free community testing site using anterior nasal (AN) swab RT-PCR for clinical testing.MethodsConsenting symptomatic and asymptomatic children (≤18 years) and adults received dual AN swabs. CareStart testing was performed with temperature/humidity monitoring. All tests had two independent reads to assess inter-operator agreement. Patients with positive CareStart results were called and instructed to isolate pending RT-PCR results. The paired RT-PCR result was the reference for sensitivity and specificity calculations.ResultsOf 1603 participants, 1245 adults and 253 children had paired RT-PCR/CareStart results and complete symptom data. 83% of adults and 87% of children were asymptomatic. CareStart sensitivity/specificity were 84.8% (95% confidence interval [CI] 71.1-93.7)/97.2% (92.0-99.4) and 85.7% (42.1-99.6)/89.5% (66.9-98.7) in adults and children, respectively, within 5 days of symptoms. Sensitivity/specificity were 50.0% (41.0-59.0)/99.1% (98.3-99.6) in asymptomatic adults and 51.4% (34.4-68.1)/97.8% (94.5-99.4) in asymptomatic children. Sensitivity in all 234 RT-PCR-positive people was 96.3% with cycle threshold (Ct) ≤25, 79.6% with Ct ≤30, and 61.4% with Ct ≤35. All 21 false positive CareStart tests had faint but normal bands. Inter-operator agreement was 99.5%. Operational challenges included identification of faint test bands and inconsistent swab elution volumes.ConclusionsCareStart had high sensitivity in people with Ct ≤25 and moderate sensitivity in symptomatic people overall. Specificity was unexpectedly lower in symptomatic versus asymptomatic people. Excellent inter-operator agreement was observed, but operational challenges indicate that operator training is warranted.

Author(s):  
Nira R Pollock ◽  
Kristine Tran ◽  
Jesica R Jacobs ◽  
Amber E Cranston ◽  
Sita Smith ◽  
...  

Abstract Background To facilitate deployment of point-of-care testing for SARS-CoV-2, we evaluated the Access Bio CareStart COVID-19 Antigen test in a high-throughput, drive-through, free community testing site using anterior nasal (AN) swab RT-PCR for clinical testing. Methods Consenting symptomatic and asymptomatic children (≤18 years) and adults received dual AN swabs. CareStart testing was performed with temperature/humidity monitoring. All tests had two independent reads to assess inter-operator agreement. Patients with positive CareStart results were called and instructed to isolate pending RT-PCR results. The paired RT-PCR result was the reference for sensitivity and specificity calculations. Results Of 1603 participants, 1245 adults and 253 children had paired RT-PCR/CareStart results and complete symptom data. 83% of adults and 87% of children were asymptomatic. CareStart sensitivity/specificity were 84.8% (95% confidence interval [CI] 71.1-93.7)/97.2% (92.0-99.4) and 85.7% (42.1-99.6)/89.5% (66.9-98.7) in adults and children, respectively, within 5 days of symptoms. Sensitivity/specificity were 50.0% (41.0-59.0)/99.1% (98.3-99.6) in asymptomatic adults and 51.4% (34.4-68.1)/97.8% (94.5-99.4) in asymptomatic children. Sensitivity in all 234 RT-PCR-positive people was 96.3% with cycle threshold (Ct) ≤25, 79.6% with Ct ≤30, and 61.4% with Ct ≤35. All 21 false positive CareStart tests had faint but normal bands. Inter-operator agreement was 99.5%. Operational challenges included identification of faint test bands and inconsistent swab elution volumes. Conclusions CareStart had high sensitivity in people with Ct ≤25 and moderate sensitivity in symptomatic people overall. Specificity was unexpectedly lower in symptomatic versus asymptomatic people. Excellent inter-operator agreement was observed, but operational challenges indicate that operator training is warranted.


2021 ◽  
Author(s):  
Nira R. Pollock ◽  
Jesica R. Jacobs ◽  
Kristine Tran ◽  
Amber Cranston ◽  
Sita Smith ◽  
...  

AbstractBackgroundRapid diagnostic tests (RDTs) for SARS-CoV-2 antigens (Ag) that can be performed at point-of-care (POC) can supplement molecular testing and help mitigate the COVID-19 pandemic. Deployment of an Ag RDT requires an understanding of its operational and performance characteristics under real-world conditions and in relevant subpopulations. We evaluated the Abbott BinaxNOW™ COVID-19 Ag Card in a high-throughput, drive-through, free community testing site in Massachusetts (MA) using anterior nasal (AN) swab RT-PCR for clinical testing.MethodsIndividuals presenting for molecular testing in two of seven lanes were offered the opportunity to also receive BinaxNOW testing. Dual AN swabs were collected from symptomatic and asymptomatic children (≤ 18 years) and adults. BinaxNOW testing was performed in a testing pod with temperature/humidity monitoring. One individual performed testing and official result reporting for each test, but most tests had a second independent reading to assess inter-operator agreement. Positive BinaxNOW results were scored as faint, medium, or strong. Positive BinaxNOW results were reported to patients by phone and they were instructed to isolate pending RT-PCR results. The paired RT-PCR result was the reference for sensitivity and specificity calculations.ResultsOf 2482 participants, 1380 adults and 928 children had paired RT-PCR/BinaxNOW results and complete symptom data. 974/1380 (71%) adults and 829/928 (89%) children were asymptomatic. BinaxNOW had 96.5% (95% confidence interval [CI] 90.0-99.3) sensitivity and 100% (98.6-100.0) specificity in adults within 7 days of symptoms, and 84.6% (65.1-95.6) sensitivity and 100% (94.5-100.0) specificity in children within 7 days of symptoms. Sensitivity and specificity in asymptomatic adults were 70.2% (56.6-81.6) and 99.6% (98.9-99.9), respectively, and in asymptomatic children were 65.4% (55.6-74.4) and 99.0% (98.0-99.6), respectively. By cycle threshold (Ct) value cutoff, sensitivity in all subgroups combined (n=292 RT-PCR-positive individuals) was 99.3% with Ct ≤25, 95.8% with ≤30, and 81.2% with ≤35. Twelve false positive BinaxNOW results (out of 2308 tests) were observed; in all twelve, the test bands were faint but otherwise normal, and were noted by both readers. One invalid BinaxNOW result was identified. Inter-operator agreement (positive versus negative BinaxNOW result) was 100% (n = 2230/2230 double reads). Each operator was able to process 20 RDTs per hour. In a separate set of 30 specimens (from individuals with symptoms ≤7 days) run at temperatures below the manufacturer’s recommended range (46-58.5°F), sensitivity was 66.7% and specificity 95.2%.ConclusionsBinaxNOW had very high specificity in both adults and children and very high sensitivity in newly symptomatic adults. Overall, 95.8% sensitivity was observed with Ct ≤ 30. These data support public health recommendations for use of the BinaxNOW test in adults with symptoms for ≤7 days without RT-PCR confirmation. Excellent inter-operator agreement indicates that an individual can perform and read the BinaxNOW test alone. A skilled laboratorian can perform and read 20 tests per hour. Careful attention to temperature is critical.


Author(s):  
Nira R. Pollock ◽  
Jesica R. Jacobs ◽  
Kristine Tran ◽  
Amber E. Cranston ◽  
Sita Smith ◽  
...  

Background: Rapid diagnostic tests (RDTs) for SARS-CoV-2 antigens (Ag) that can be performed at point-of-care (POC) can supplement molecular testing and help mitigate the COVID-19 pandemic. Deployment of an Ag RDT requires an understanding of its operational and performance characteristics under real-world conditions and in relevant subpopulations. We evaluated the Abbott BinaxNOW™ COVID-19 Ag Card in a high-throughput, drive-through, free community testing site in Massachusetts (MA) using anterior nasal (AN) swab RT-PCR for clinical testing. Methods: Individuals presenting for molecular testing in two of seven lanes were offered the opportunity to also receive BinaxNOW testing. Dual AN swabs were collected from symptomatic and asymptomatic children (≤ 18 years) and adults. BinaxNOW testing was performed in a testing pod with temperature/humidity monitoring. One individual performed testing and official result reporting for each test, but most tests had a second independent reading to assess inter-operator agreement. Positive BinaxNOW results were scored as faint, medium, or strong. Positive BinaxNOW results were reported to patients by phone and they were instructed to isolate pending RT-PCR results. The paired RT-PCR result was the reference for sensitivity and specificity calculations. Results: Of 2482 participants, 1380 adults and 928 children had paired RT-PCR/BinaxNOW results and complete symptom data. 974/1380 (71%) adults and 829/928 (89%) children were asymptomatic. BinaxNOW had 96.5% (95% confidence interval [CI] 90.0- 99.3) sensitivity and 100% (98.6-100.0) specificity in adults within 7 days of symptoms, and 84.6% (65.1-95.6) sensitivity and 100% (94.5-100.0) specificity in children within 7 days of symptoms. Sensitivity and specificity in asymptomatic adults were 70.2% (56.6-81.6) and 99.6% (98.9-99.9), respectively, and in asymptomatic children were 65.4% (55.6-74.4) and 99.0% (98.0-99.6), respectively. By cycle threshold (Ct) value cutoff, sensitivity in all subgroups combined (n=292 RT-PCR-positive individuals) was 99.3% with Ct ≤25, 95.8% with ≤30, and 81.2% with ≤35. Twelve false positive BinaxNOW results (out of 2308 tests) were observed; in all twelve, the test bands were faint but otherwise normal, and were noted by both readers. One invalid BinaxNOW result was identified. Inter-operator agreement (positive versus negative BinaxNOW result) was 100% (n = 2230/2230 double reads). Each operator was able to process 20 RDTs per hour. In a separate set of 30 specimens (from individuals with symptoms ≤7 days) run at temperatures below the manufacturer’s recommended range (46-58.5°F), sensitivity was 66.7% and specificity 95.2%. Conclusions: BinaxNOW had very high specificity in both adults and children and very high sensitivity in newly symptomatic adults. Overall, 95.8% sensitivity was observed with Ct ≤ 30. These data support public health recommendations for use of the BinaxNOW test in adults with symptoms for ≤7 days without RT-PCR confirmation. Excellent inter-operator agreement indicates that an individual can perform and read the BinaxNOW test alone. A skilled laboratorian can perform and read 20 tests per hour. Careful attention to temperature is critical.


2021 ◽  
Author(s):  
Laura Ford ◽  
Melissa J. Whaley ◽  
Melisa M. Shah ◽  
Phillip P. Salvatore ◽  
Hannah E. Segaloff ◽  
...  

Background: Performance characteristics of SARS-CoV-2 antigen tests among children are limited despite the need for point-of-care testing in school and childcare settings. We describe children seeking SARS-CoV-2 testing at a community site and compare antigen test performance to real-time reverse transcription-polymerase chain reaction (RT-PCR) and viral culture. Methods: Two anterior nasal specimens were self-collected for BinaxNOW antigen and RT-PCR testing, along with demographics, symptoms, and exposure information from individuals ≥5 years at a community testing site. Viral culture was attempted on residual antigen or RT-PCR positive specimens. Demographic and clinical characteristics, and the performance of SARS-CoV-2 antigen tests, were compared among children (<18 years) and adults. Results: About one in ten included specimens were from children (225/2110); 16.4% (37/225) were RT-PCR positive. Cycle threshold values were similar among RT-PCR positive specimens from children and adults (22.5 vs 21.3, p=0.46) and among specimens from symptomatic and asymptomatic children (22.5 vs 23.2, p=0.39). Sensitivity of antigen test compared to RT-PCR was 73.0% (27/37) among specimens from children and 80.8% (240/297) among specimens from adults; among specimens from children, specificity was 100% (188/188), positive and negative predictive value were 100% (27/27) and 94.9% (188/198) respectively. Virus was isolated from 51.4% (19/37) of RT-PCR positive pediatric specimens; all 19 had positive antigen test results. Conclusions : With lower sensitivity relative to RT-PCR, antigen tests may not diagnose all positive COVID-19 cases; however, antigen testing identified children with live SARS-CoV-2 virus.


2021 ◽  
Author(s):  
Sara Suliman ◽  
Wilfredo R. Matias ◽  
Isabel R Fulcher ◽  
Francisco J. Molano ◽  
Shannon Collins ◽  
...  

Background: Point-of-care antigen-detecting rapid diagnostic tests (RDTs) for Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2) represent a scalable tool for SARS-CoV-2 infections surveillance. Data on their performance in real-world community settings is paramount for their implementation. Method: We evaluated the accuracy of CareStartTM COVID-19 Antigen test (CareStart) in a testing site in Holyoke, Massachusetts. We compared CareStart to a SARS-CoV-2 reverse transcriptase quantitative polymerase chain reaction (RT-qPCR) reference, using anterior nasal swab samples. We calculated the sensitivity, specificity, and expected positive and negative predictive values at different SARS-CoV-2 prevalence estimates. Results: We performed 666 tests on 591 unique individuals. 573 (86%) were asymptomatic. There were 52 positive tests by RT-qPCR. The sensitivity of CareStart was 49.0% (95% Confidence Interval (CI): 34.8 - 63.4) and specificity was 99.5% (95% CI: 98.5 - 99.9). Among positive RT-qPCR tests, the median cycle threshold (Ct) was significantly lower in samples that tested positive on CareStart. Using a Ct ≤ 30 as a benchmark for positivity increased the sensitivity to 64.9% (95% CI: 47.5 - 79.8). Conclusions: CareStart has a high specificity and moderate sensitivity. The utility of RDTs, such as CareStart, in mass implementation should prioritize use cases in which a higher specificity is more important.


Author(s):  
Mar Masiá ◽  
Marta Fernández-González ◽  
Manuel Sánchez ◽  
Mar Carvajal ◽  
José Alberto García ◽  
...  

Abstract Background Performance of point-of-care tests in different clinical scenarios and on different samples remains undetermined. We comprehensively evaluated the performance of the nasopharyngeal Panbio COVID-19 antigen Rapid-Test-Device. Method Prospective study including consecutive patients attending three primary care centers (PCC) and an emergency department. The antigen test was performed at point-of-care in nasopharyngeal and nasal swabs, and in saliva. Positive and negative percent agreement (PPA, NPA) were calculated with the RT-PCR assay as reference standard. Results Of 913 patients included, 296 (32.3%) were asymptomatic and 690 (75.6%) came from the PCC. Nasopharyngeal swabs were collected from 913, nasal swabs from 659, and saliva from 611 patients. RT-PCR was positive in 196 (21.5%) nasopharyngeal samples (NPS). Overall PPA (95% CI) in NPS was 60.5% (53.3-67.4), and it was lower in nasal swabs (44.7%) and saliva (23.1%). Test performance in NPS was largely dependent on the cycle threshold (Ct) in RT-PCR, with PPA of 94% for Ct≤25 and 80% for Ct&lt;30. In symptomatic patients, the PPA was 95% for Ct≤25; 85% for Ct&lt;30, and 89% for the symptom triad of fever, cough and malaise. Performance was also dependent on age, with PPA of 100% in symptomatic patients &gt;50 years with Ct&lt;25. In asymptomatic patients, the PPA was 86% for Ct&lt;25. In all cases, NPA was 100%. Conclusion The nasopharyngeal Panbio COVID-19 antigen test performed at point-of-care has a good sensitivity in symptomatic patients with Ct&lt;30 and older age. The test was useful to identify asymptomatic patients with lower Ct values.


2022 ◽  
Author(s):  
Kristie J Sun ◽  
Mary Jane E Vaeth ◽  
Matthew L Robinson ◽  
Maryam Elhabashy ◽  
Ishaan Gupta ◽  
...  

SARS-CoV-2 continues to develop new, increasingly infectious variants, such as delta and omicron. Here, we evaluate the efficacy of the Abbott BinaxNOW Rapid Antigen Test against the gold standard of Reverse Transcription Polymerase Chain Reaction (RT-PCR) in 1054 pediatric participants presenting to a state-owned high-volume Coronavirus Disease 2019 (COVID-19) testing site. During the testing period, the delta variant was predominant. Prior to sample collection, symptomatic and exposure status was collected for all participants based on Centers for Disease Control (CDC) criteria. RT-PCR results demonstrated an overall prevalence rate of 5.2%. For all participants, the sensitivity of the rapid antigen tests was 92.7% (95% CI 82.4% - 98.0%) and specificity was 98.0% (95% CI 97.0%-98.8%). For symptomatic participants, the sensitivity was 92.3% (95% CI 74.9% - 99.1%), specificity was 96.6% (95% CI 93.6%- 98.4%), positive predictive value (PPV) was 72.7% (95% CI 54.5% - 86.7%) and negative predictive value (NPV) was 99.2% (95% CI 98.2% - 100%). Among asymptomatic participants, the sensitivity was 92.6% (95% CI 75.7% - 99.1%), specificity was 98.6% (95% CI 97.5% - 99.3%) the PPV was 71.4% (95% CI 53.7% - 85.4%) and the NPV was 99.7% (95% CI 99.0% - 100%). Our reported sensitivity and NPV are higher than other pediatric studies, but specificity and PPV are lower. Importance Children are especially impacted by the disease and its ability to disrupt educational opportunities. Although vaccinations have been approved for children 5 years and older, many children remain unvaccinated. Widespread testing may improve the ability for children to remain in in-person activities, minimizing absences from school and extracurriculars. Highly accurate rapid antigen tests may be vital to containing future COVID-19 waves while mitigating detrimental effects.


2021 ◽  
Vol 11 (1) ◽  
Author(s):  
Yuto Takeuchi ◽  
Yusaku Akashi ◽  
Daisuke Kato ◽  
Miwa Kuwahara ◽  
Shino Muramatsu ◽  
...  

AbstractThe clinical utility of antigen test using anterior nasal samples has not been well evaluated. We conducted a prospective study in a drive-through testing site located at a PCR center to evaluate the diagnostic performance of the antigen test QuickNavi-COVID19 Ag using anterior nasal samples and to compare the degrees of coughs or sneezes induction and the severity of pain between anterior nasal collection and nasopharyngeal collection. The study included a total of 862 participants, of which 91.6% were symptomatic. The median duration from symptom onset to sample collection was 2.0 days. Fifty-one participants tested positive for severe acute respiratory syndrome coronavirus 2 on reverse transcription PCR (RT-PCR) with nasopharyngeal samples, and all of them were symptomatic. In comparison to the findings of RT-PCR, the antigen test using anterior nasal samples showed 72.5% sensitivity (95% confidence interval [CI] 58.3–84.1%) and 100% specificity (95% CI 99.3–100%). Anterior nasal collection was associated with a significantly lower degree of coughs or sneezes induction and the severity of pain in comparison to nasopharyngeal collection (p < 0.001). The antigen test using anterior nasal samples showed moderate sensitivity in symptomatic patients who were at the early stages of the disease course but was less painful and induced fewer coughs or sneezes.


2020 ◽  
Author(s):  
Paul K. Drain ◽  
Madhavi Ampajwala ◽  
Christopher Chappel ◽  
Andre B. Gvozden ◽  
Melanie Hoppers ◽  
...  

AbstractBackgroundThe LumiraDx SARS-CoV-2 antigen test, which uses a high-sensitivity, microfluidic immunoassay to detect the nucleocapsid protein of SARS-CoV-2, was evaluated for diagnosing acute COVID-19 in adults and children across point-of-care settings.MethodsTwo paired anterior nasal swabs or two paired nasopharyngeal swabs were collected from each participant. Swabs were tested by the LumiraDx SARS-CoV-2 antigen test and compared with real-time PCR (rt-PCR; Roche cobas 6800 platform). Positive- and negative predictive values and likelihood ratios were calculated. Results stratified based on gender, age, duration of symptoms, and rt-PCR cycle threshold.ResultsOut of the 512 participants, aged 0-90 years, of this prospective validation study, 414 (81%) were symptomatic for COVID-19 and 123 (24%) swabs were positive for SARS-CoV-2 based on rt-PCR testing. Compared with rt-PCR, the 12-minute swab test had 97.6% sensitivity and 96.6% specificity within 12 days of symptom onset, representing the period of infectivity. All (100%) samples detected within 33 rt-PCR cycles were also identified using the antigen test. Results were consistent across age and gender. Despite being performed by minimally trained healthcare workers, the user error rate of the test system was 1%.ConclusionThe rapid high-sensitivity assay using nasopharyngeal or anterior nasal sampling may offer significant improvements for diagnosing acute SARS-CoV-2 infection in clinic- and community-based settings.SummaryA 12-minute nasal swab test detects 97.6% of COVID-19 infections, compared to gold standard real-time PCR testing, up to 12 days following symptom onset using a microfluidic immunoassay for SARS-CoV-2 nucleocapsid protein.


PLoS ONE ◽  
2021 ◽  
Vol 16 (4) ◽  
pp. e0249710
Author(s):  
Neeraj Sood ◽  
Rashmi Shetgiri ◽  
Anna Rodriguez ◽  
Dianna Jimenez ◽  
Sonia Treminino ◽  
...  

Background Rapid antigen tests hold much promise for use in the school environment. However, the performance of these tests in non-clinical settings and among one of the main target populations in schools—asymptomatic children—is unclear. To address this gap, we examined the positive and negative concordance between the BinaxNOW™ rapid SARS-CoV-2 antigen assay and an RT-PCR test among children at a community-based Covid-19 testing site. Methods We conducted rapid antigen (BinaxNOW™) and oral fluid RT-PCR (Curative Labs) tests on children presenting at a walk-up testing site in Los Angeles County from November 25, 2020 to December 9, 2020. Positive concordance was determined as the fraction of RT-PCR positive participants that were also antigen positive. Negative concordance was determined as the fraction of RT-PCR negative participants that were also antigen negative. Multivariate logistic regression models were used to examine the association between positive or negative concordance and participant age, race-ethnicity, sex at birth, symptoms and Ct values. Results 226 children tested positive on RT-PCR; 127 children or 56.2% (95% CI: 49.5% to 62.8%) of these also tested positive on the rapid antigen test. Positive concordance was higher among symptomatic children (64.4%; 95% CI: 53.4% to 74.4%) compared to asymptomatic children (51.1%; 95% CI: 42.5% to 59.7%). Positive concordance was negatively associated with Ct values and was 93.8% (95% CI: 69.8% to 99.8%) for children with Ct values less than or equal to 25. 548 children tested negative on RT-PCR; 539 or 98.4% (95% CI: 96.9% to 99.2%) of these also tested negative on the rapid antigen test. Negative concordance was higher among asymptomatic children. Conclusions Rapid antigen testing can successfully identify most COVID infections in children with viral load levels likely to be infectious. Serial rapid testing may help compensate for limited sensitivity in early infection.