scholarly journals Evaluation of Organ Dose and Image Quality Metrics of Pediatric CT Chest-Abdomen-Pelvis (CAP) Examination: An Anthropomorphic Phantom Study

2021 ◽  
Vol 11 (5) ◽  
pp. 2047
Author(s):  
Nor Azura Muhammad ◽  
Zunaide Kayun ◽  
Hasyma Abu Hassan ◽  
Jeannie Hsiu Ding Wong ◽  
Kwan Hoong Ng ◽  
...  

The aim of this study is to investigate the impact of CT acquisition parameter setting on organ dose and its influence on image quality metrics in pediatric phantom during CT examination. The study was performed on 64-slice multidetector CT scanner (MDCT) Siemens Definition AS (Siemens Sector Healthcare, Forchheim, Germany) using various CT CAP protocols (P1–P9). Tube potential for P1, P2, and P3 protocols were fixed at 100 kVp while P4, P5, and P6 were fixed at 80 kVp with used of various reference noise values. P7, P8, and P9 were the modification of P1 with changes on slice collimation, pitch factor, and tube current modulation (TCM), respectively. TLD-100 chips were inserted into the phantom slab number 7, 9, 10, 12, 13, and 14 to represent thyroid, lung, liver, stomach, gonads, and skin, respectively. The image quality metrics, signal to noise ratio (SNR) and contrast to noise ratio (CNR) values were obtained from the CT console. As a result, this study indicates a potential reduction in the absorbed dose up to 20% to 50% along with reducing tube voltage, tube current, and increasing the slice collimation. There is no significant difference (p > 0.05) observed between the protocols and image metrics.

2019 ◽  
Vol 9 (1) ◽  
Author(s):  
Miguel García García ◽  
Siegfried Wahl ◽  
Dibyendu Pusti ◽  
Pablo Artal ◽  
Arne Ohlendorf

AbstractTo evaluate the impact of multifocal contact lens wear on the image quality metrics across the visual field in the context of eye growth and myopia control. Two-dimensional cross-correlation coefficients were estimated by comparing a reference image against the computed retinal images for every location. Retinal images were simulated based on the measured optical aberrations of the naked eye and a set of multifocal contact lenses (centre-near and centre-distance designs), and images were spatially filtered to match the resolution limit at each eccentricity. Value maps showing the reduction in the quality of the image through each optical condition were obtained by subtracting the optical image quality from the theoretical physiological limits. Results indicate that multifocal contact lenses degrade the image quality independently from their optical design, though this result depends on the type of analysis conducted. Analysis of the image quality across the visual field should not be oversimplified to a single number but split into regional and groups because it provides more insightful information and can avoid misinterpretation of the results. The decay of the image quality caused by the multifocal contacts alone, cannot explain the translation of peripheral defocus towards protection on myopia progression, and a different explanation needs to be found.


Author(s):  
Nigar Salimova ◽  
Jan B. Hinrichs ◽  
Marcel Gutberlet ◽  
Bernhard C. Meyer ◽  
Frank K. Wacker ◽  
...  

Abstract Objectives To evaluate the impact of the reconstructed field-of-view (FOV) on image quality in computed-tomography angiography (CTA) of the lower extremities. Methods A total of 100 CTA examinations of the lower extremities were acquired on a 2 × 192-slice multidetector CT (MDCT) scanner. Three different datasets were reconstructed covering both legs (standard FOV size) as well as each leg separately (reduced FOV size). The subjective image quality was evaluated for the different vessel segments (femoral, popliteal, crural, pedal) by three readers using a semi-quantitative Likert scale. Additionally, objective image quality was assessed using an automated image quality metric on a per-slice basis. Results The subjective assessment of the image quality showed an almost perfect interrater agreement. The image quality of the small FOV datasets was rated significantly higher as compared to the large datasets for all patients and vessel segments (p < 0.05) with a tendency towards a higher effect in smaller vessels. The difference of the mean scores between the group with the large FOV and small FOV was 0.68 for the femoral level, 0.83 for the popliteal level, 1.12 for the crural level, and 1.08 for the pedal level. The objective image quality metric also demonstrated a significant improvement of image quality in the small FOV datasets. Conclusions Side-separated reconstruction of each leg in CTA of the lower extremities using a small reconstruction FOV significantly improves image quality as compared to a standard reconstruction with a large FOV covering both legs. Key Points • In CT angiography of the lower legs, the side-separated reconstruction of each leg using a small field-of-views improves image quality as compared to a standard reconstruction covering both legs. • The side-separated reconstruction can be readily implemented at every commercially available CT scanner. • There is no need for additional hardware or software and no additional burden to the patient.


1997 ◽  
Vol 36 (26) ◽  
pp. 6583 ◽  
Author(s):  
Robert T. Brigantic ◽  
Michael C. Roggemann ◽  
Kenneth W. Bauer ◽  
Byron M. Welsh

2021 ◽  
Vol 94 (1125) ◽  
pp. 20210430
Author(s):  
Puja Shahrouki ◽  
Kim-Lien Nguyen ◽  
John M. Moriarty ◽  
Adam N. Plotnik ◽  
Takegawa Yoshida ◽  
...  

Objectives: To assess the feasibility of a rapid, focused ferumoxytol-enhanced MR angiography (f-FEMRA) protocol in patients with claustrophobia. Methods: In this retrospective study, 13 patients with claustrophobia expressed reluctance to undergo conventional MR angiography, but agreed to a trial of up to 10 min in the scanner bore and underwent f-FEMRA. Thirteen matched control patients who underwent gadolinium-enhanced MR angiography (GEMRA) were identified for comparison of diagnostic image quality. For f-FEMRA, the time from localizer image acquisition to completion of the angiographic acquisition was measured. Two radiologists independently scored images on both f-FEMRA and GEMRA for arterial and venous image quality, motion artefact and diagnostic confidence using a 5-point scale, five being best. Signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) in the aorta and IVC were measured. The Wilcoxon rank-sum test, one-way ANOVA with Tukey correction and two-tailed t tests were utilized for statistical analyses. Results: All scans were diagnostic and assessed with high confidence (scores ≥ 4). Average scan time for f-FEMRA was 6.27 min (range 3.56 to 10.12 min), with no significant difference between f-FEMRA and GEMRA in diagnostic confidence (4.86 ± 0.24 vs 4.69 ± 0.25, p = 0.13), arterial image quality (4.62 ± 0.57 vs 4.65 ± 0.49, p = 0.78) and motion artefact score (4.58 ± 0.49 vs 4.58 ± 0.28, p > 0.99). f-FEMRA scored significantly better for venous image quality than GEMRA (4.62 ± 0.42 vs 4.19 ± 0.56, p = 0.04). CNR in the IVC was significantly higher for steady-state f-FEMRA than GEMRA regardless of the enhancement phase (p < 0.05). Conclusions: Comprehensive vascular MR imaging of the thorax, abdomen and pelvis can be completed in as little as 5 min within the magnet bore using f-FEMRA, facilitating acceptance by patients with claustrophobia and streamlining workflow. Advances in knowledge: A focused approach to vascular imaging with ferumoxytol can be performed in patients with claustrophobia, limiting time in the magnet bore to 10 min or less, while acquiring fully diagnostic images of the thorax, abdomen and pelvis.


2020 ◽  
Author(s):  
Anne Poulsen ◽  
Diane Jang ◽  
Mahmood Khan ◽  
Zaina Nabil Al-Mohtaseb ◽  
Michael Chen ◽  
...  

Purpose: To investigate the repeatability of a combined Dual-Scheimpflug placido disc corneal topographer (Zeimer Galilei G4) with respect to keratometric indices used to monitor progression of keratoconus (KCN). Methods: Patients with KCN were prospectively enrolled. For each eye lacking history of corneal surgery, 5 measurements were taken in succession. Eyes in which 3 or more measurements could be obtained (defined by the device's 4 image quality metrics) were included in the analysis. The repeatability limits (RL) and interclass correlation coefficients (ICC) were calculated for various parameters. Results: 32 eyes from 25 patients met all image quality metrics, and 54 eyes from 38 patients met at least 3/4 criteria (all except the placido image quality metric). RLs for key parameters when 4/4 or 3/4 image quality metrics were met included: 0.37 and 0.77 diopters (D) for steep simulated keratometry, 0.79 and 1.65 D for maximum keratometry, 13.80 and 13.88 degrees for astigmatism axis, 0.64 and 0.56 um for vertical coma magnitude, and 3.76 and 3.84 um for thinnest pachymetry, respectively. The ICCs for all parameters were excellent [above 0.87 except for spherical aberration (0.77)]. Conclusions: The dual-Scheimpflug placido disc corneal topographer is highly repeatable in quantifying parameters used in monitoring KCN. Excellent placido images are difficult to capture in eyes with KCN, but when available, increase the reliability of the measurements. The RLs may be especially helpful in detecting progression in mild KCN when interventions such as corneal cross-linking or intrastromal corneal ring segments are most beneficial.


2020 ◽  
Author(s):  
Katy Vecchiato ◽  
Alexia Egloff ◽  
Olivia Carney ◽  
Ata Siddiqui ◽  
Emer Hughes ◽  
...  

Background and Purpose: Head motion causes image degradation in brain MRI examinations, negatively impacting image quality, especially in pediatric populations. Here, we used a retrospective motion correction technique in children and assessed image quality improvement for 3D MRI acquisitions. Material and Methods: We prospectively acquired brain MRI at 3T using 3D sequences, T1-weighted MPRAGE, T2-weighted Turbo Spin Echo and FLAIR, in 32 unsedated children, including 7 with epilepsy (age range 2-18 years). We implemented a novel motion correction technique: Distributed and Incoherent Sample Orders for Reconstruction Deblurring using Encoding Redundancy (DISORDER). For each subject and modality, we obtained 3 reconstructions: as acquired (Aq), after DISORDER motion correction (Di), and Di with additional outlier rejection (DiOut). We analyzed 288 images quantitatively, measuring 2 objective no-reference image quality metrics: Gradient Entropy (GE) and MPRAGE White Matter Homogeneity (WM-H). As a qualitative metric, we presented blinded and randomized images to 2 expert neuroradiologists who scored them for clinical readability. Results: Both image quality metrics improved after motion correction for all modalities and improvement correlated with the amount of intrascan motion. Neuroradiologists also considered the motion corrected images as of higher quality (Wilcoxon z=-3.164 MPRAGE, z=-2.066 TSE, z=-2.645 FLAIR, for all p<0.05). Conclusions: Retrospective image motion correction with DISORDER increased image quality both from an objective and qualitative perspective. In 75% of sessions, at least one sequence was improved by this approach, indicating the benefit of this technique in un-sedated children for both clinical and research environments.


Author(s):  
Yang-Ting Hsu ◽  
Jo-Chi Jao

Radiologic technologists face various types of patients during multi-detector computed tomography (CT) examinations. In emergency departments, it is common to have patients who cannot follow instructions for the examinations. The asymmetric axial view of the head CT might affect the correctness of the clinician’s diagnosis. This study aimed to assess the impact of head positioning on the image quality of head CT using two phantoms. All scans were performed on a 16-slice CT scanner. In the control group, the tilted angle of the phantoms was 0[Formula: see text], and no multiplanar reconstruction (MPR) was performed. In the experimental groups, the tilted angles of the phantoms were 5[Formula: see text], 10[Formula: see text] and 15[Formula: see text], respectively, and MPR was performed afterwards. The results showed that if the head was tilted during the head CT examinations, image asymmetry and artifacts appeared without MPR. After MPR, one phantom showed that there were significant differences and the other phantom showed no significant differences quantitatively in image symmetry and artifacts between experimental groups and the control group, while both phantoms showed no significant differences qualitatively in image symmetry and artifacts between experimental groups and the control group. Although MPR can correct the image asymmetry and artifacts caused by tilted head positioning to some extent, it consumes time. Therefore, technologists should position the head as exactly as possible when performing head CT examinations.


Author(s):  
Gareth D Hastings ◽  
Raymond A Applegate ◽  
Alexander W Schill ◽  
Chuan Hu ◽  
Daniel R Coates ◽  
...  