thoracic impedance
Recently Published Documents


TOTAL DOCUMENTS

203
(FIVE YEARS 25)

H-INDEX

22
(FIVE YEARS 2)

Circulation ◽  
2021 ◽  
Vol 144 (Suppl_2) ◽  
Author(s):  
Michelle Nassal ◽  
Xabier Jaureguibeitia ◽  
Elisabete Aramendi ◽  
Unai Irusta ◽  
Ashish R Panchal ◽  
...  

Introduction: Controlled ventilation is important in OHCA resuscitation, but there are few validated methods for accurate detection of ventilations. We sought to utilize changes in thoracic impedance (TI) to characterize resuscitation ventilations in the Pragmatic Airway Resuscitation Trial (PART). Methods: We analyzed CPR process files collected from adult OHCA enrolled in PART. We limited the analysis to cases with simultaneous capnography ventilation recordings at the Dallas-Ft Worth site. We identified ventilation waveforms in the thoracic impedance signal by applying automated signal processing with adaptive filtering techniques to remove overlying artifacts from chest compressions. We correlated detected ventilations with the end-tidal capnography signal. We determined the amplitudes (Ai, Ae) and durations (Di, De) of both insufflation and exhalation phases of the ventilation impedance signal (Figure 1). We compared differences between laryngeal tube (LT) and endotracheal intubation (ETI) airway management during mechanical or manual chest compressions using Mann-Whitney U-test. Results: We included 303 CPR process cases in the analysis; 209 manual (77 ETI, 132 LT), 94 mechanical (41 ETI, 53 LT). Ventilation Ai and Ae were higher for ETI than LT in both manual (ETI: Ai 0.71Ω, Ae 0.70Ω vs LT: Ai 0.46Ω Ae 0.45Ω, p<0.01 respectively) and mechanical chest compressions (ETI: Ai 1.22Ω, Ae 1.14Ω VS LT: Ai 0.74Ω, Ae 0.68Ω, p<0.01 respectively). Ventilations per minute, duration of TI amplitude insufflation and exhalation did not differ among groups. Conclusion: Compared with LT, ETI thoracic impedance ventilation insufflation and exhalation amplitude were higher while duration did not differ. TI may provide a novel approach to characterizing ventilation during OHCA.


Author(s):  
Michelle MJ Nassal ◽  
Xabier Jaureguibeitia ◽  
Elisabete Aramendi ◽  
Unai Irusta ◽  
Ashish Panchal ◽  
...  

2021 ◽  
Vol 70 ◽  
pp. 102941
Author(s):  
Hao Ge ◽  
Hui Qin ◽  
Shan Xue ◽  
Enkang Liu ◽  
Mingzhu Zhang ◽  
...  

Technologies ◽  
2021 ◽  
Vol 9 (3) ◽  
pp. 58
Author(s):  
Christos Dimas ◽  
Vassilis Alimisis ◽  
Ioannis Georgakopoulos ◽  
Nikolaos Voudoukis ◽  
Nikolaos Uzunoglu ◽  
...  

Electrical impedance tomography is a low-cost, safe, and high temporal resolution medical imaging modality which finds extensive application in real-time thoracic impedance imaging. Thoracic impedance changes can reveal important information about the physiological condition of patients’ lungs. In this way, electrical impedance tomography can be a valuable tool for monitoring patients. However, this technique is very sensitive to measurement noise or possible minor signal errors, coming from either the hardware, the electrodes, or even particular biological signals. Thus, the design of a good performance electrical impedance tomography hardware setup which properly interacts with the tissue examined is both an essential and a challenging concept. In this paper, we adopt an extensive simulation approach, which combines the system’s analogue and digital hardware, along with equivalent circuits of 3D finite element models that represent thoracic cavities. Each thoracic finite element model is created in MATLAB based on existing CT images, while the tissues’ conductivity and permittivity values for a selected frequency are acquired from a database using Python. The model is transferred to a multiport RLC network, embedded in the system’s hardware which is simulated at LT SPICE. The voltage output data are transferred to MATLAB where the electrical impedance tomography signal sampling and digital processing is also simulated. Finally, image reconstructions are performed in MATLAB, using the EIDORS library tool and considering the signal noise levels and different electrode and signal sampling configurations (ADC bits, sampling frequency, number of taps).


Heart Rhythm ◽  
2021 ◽  
Vol 18 (8) ◽  
pp. S448
Author(s):  
Stuart O. Schecter ◽  
Rita Jermyn

Medicina ◽  
2021 ◽  
Vol 57 (6) ◽  
pp. 618
Author(s):  
Seung-Young Roh ◽  
Jinhee Ahn ◽  
Kwang-No Lee ◽  
Yong-Soo Baek ◽  
Dong-Hyeok Kim ◽  
...  

Background and Objectives—Direct current cardioversion (DCCV) is a safe and useful treatment for atrial tachyarrhythmias. In the past, the energy delivered in DCCV was decided upon empirically, based only on the type of tachyarrhythmia. This conventional method does not consider individual factors and may lead to unnecessary electrical damage. Materials and Methods—We performed DCCV in patients with atrial tachyarrhythmias. The impedance and electrical current at the moment of shock were measured. The human thoracic impedance between both defibrillator patches and the electric current that was used were measured. Results—A total of 683 DCCVs were performed on 466 atrial tachyarrhythmia patients. The average impedance was 64 ± 11 Ω and the average successful current was 23 ± 6 mA. The magnitude of the electrical current that was successful depended upon the human impedance (linear regression, B = −0.266, p < 0.001) and the left atrial diameter (B = 0.092, p < 0.001). Impedance was directly proportional to body mass index (BMI) (B = 1.598, p < 0.001) and was higher in females than in males (77 ± 15 Ω vs. 63 ± 11 Ω, p < 0.001). Notably, the high-impedance (>70 Ω) group had a higher BMI (27 ± 4 kg/m2 vs. 25 ± 3 kg/m2, p < 0.001) and a higher proportion of females (37% vs. 9%, p < 0.001) than the low-impedance group (<70 Ω). However, thoracic impedance was not an independent predictor for successful DCCV. Conclusions—Human thoracic impedance was one of the factors that impacted the level of electrical current required for successful DCCV in patients with atrial arrhythmias. In the future, it will be helpful to consider individual predictors, such as BMI and gender, to minimize electrical damage during DCCV.


2021 ◽  
Author(s):  
K. Lim ◽  
C. Eastwood‐Sutherland ◽  
A. Marshall ◽  
T.J. Gale ◽  
P.A. Dargaville

2021 ◽  
Author(s):  
Oya Köksal ◽  
Erdem Haberal

Abstract Purpose Simultaneous monitoring of ECG and thoracic electrical bioimpedance (TEB) is important in evaluating cardiovascular performance. TEB is a non-invasive technique based on measuring the impedance value that changes in the chest area depending on the heartbeat. Within the framework of this study, it can be used in home monitoring and biotelemetry applications to measure thoracic electrical bioimpedance (TEB), ECG and ICG. Methods Within the scope of this study, a four-electrode TEB measurement system was designed and built using the Raspberry Pi single board computer and its original monitor, ESP32 and EVAL-ADAS1000SDZ evaluation board. With the designed system, ECG and thoracic impedance measurements at 50 kHz current frequency were taken as real-time over a single channel. Delta_Z and ICG signals were created from thoracic impedance values with the developed software.ResultsWhile the thoracic impedance value varies between 15-45 Ω, the 67 thoracic impedance value measured with the designed system is approximately 1000 times the 68 reference value. The impedance change in the thoracic region was measured with the designed 69 system between 0.1-0.2 Ω values, and the compatibility of these values with reference values was 70 determined. While the reference value of the dZ / dt signal is 0.8 - 3.5 Ω / s, this value is between 2.3 - 71 5.3 Ω / s in the measurements taken with the designed system.Conclusion The prototype is achieved in detecting small changes in the thoracic impedance signal. The prototype is cheap, portable, small-sized and medically safe, so it is suitable for home care services and clinics. In addition, the developed system can be adapted to wearable technology. In order to increase the success of the system, the impedances values added to the thoracic impedance value should be determined and a calibration procedure should be established.


2021 ◽  
Author(s):  
Yuan Lu ◽  
Karthik Murugiah ◽  
Paul W Jones ◽  
César Caraballo ◽  
Shiwani Mahajan ◽  
...  

ABSTRACTHospitalizations for acute cardiac conditions have markedly declined during the coronavirus disease 2019 (COVID-19) pandemic, yet the cause of this decline is not clear. Using remote monitoring data of 4,029 patients with implantable cardiac defibrillators (ICDs) living in New York City and Minneapolis/Saint Paul, we assessed changes in markers of cardiac status among these patients and compared thoracic impedance and arrhythmia burden in 2019 and 2020 from January through August. We found no change in several key disease decompensation markers among patients with implanted ICD devices during the first phase of COVID-19 pandemic, suggesting that the decrease in cardiovascular hospitalizations in this period is not reflective of a true population-level improvement in cardiovascular health.


Sign in / Sign up

Export Citation Format

Share Document