analytical imprecision
Recently Published Documents


TOTAL DOCUMENTS

73
(FIVE YEARS 4)

H-INDEX

22
(FIVE YEARS 0)

Author(s):  
Anders Abildgaard ◽  
Cindy Søndersø Knudsen ◽  
Lise Nørkjær Bjerg ◽  
Sten Lund ◽  
Julie Støy

Abstract Objectives The glycated haemoglobin fraction A1c (HbA1c) is widely used in the management of diabetes mellitus, and the Siemens DCA Vantage™ point-of-care testing (POCT) instrument offers rapid HbA1c results even far from a clinical laboratory. However, the analytical performance has been questioned, and not much is known about effects of changing reagent lot, instrument and operator. We therefore compared the analytical performance of the DCA Vantage™ with established routine methods (Tosoh G8/G11 ion exchange HPLC) in a true clinical setting at two Danish hospitals. Methods We extracted all routine clinical HbA1c results incidentally drawn from the same patient within 48 h (n=960 pairs) and evaluated the effect of reagent lot, operator and instrument. We also performed a prospective method comparison in our diabetes out-patient clinic (n=97). Results The critical difference (CD) between two POCT results varied between 5.14 and 6.61 mmol/mol (0.47–0.55%), and the analytical imprecision of the DCA Vantage™ (CVA) was >3%. Significant effect of reagent lot and inter-instrument differences were found, whereas no effect of operator was seen. Conclusions The DCA Vantage™ HbA1c analysis does not fulfil the prevailing analytical performance specifications, but rigorous validation of new reagent lots and continuous recalibration of instruments may potentially improve the precision substantially. Our findings, therefore, clearly emphasise the necessity of a close collaboration between clinicians and laboratory professionals in the POCT field. Finally, POCT HbA1c results should always be interpreted together with other measures of glycaemic control to avoid inappropriate change of patient treatments due to measurement uncertainty.



Author(s):  
Simone Drummond ◽  
Ching-Tong Mark ◽  
Nadia Caruso ◽  
Lorna Clark ◽  
Pete Kavsak

There is much current interest in the use of low-normal high-sensitivity cardiac troponin (hsTn) concentrations, with or without minimal change, to rule out myocardial infarction (MI). Clifford-Mobley’s observations demonstrate that this a challenge even for platforms measuring hsTnT.1 Analytical imprecision may also affect algorithms that use hsTn change alone to rule in MI. For example, the imprecision observed with the Ortho hsTnI assay (Ortho Clinical Diagnostics, New Jersey, United States), using quality control or patient pools, has been found to exceed the European Society of Cardiology 0/1h algorithm criterion.2 The Ortho hsTnI assay has also been shown to yield high and non-reproducible results (i.e., outliers), in addition to the problems with imprecision.3 Outliers are often identified by repeat centrifugation and repeat testing. However, Ortho hsTnI results above the 99th percentile cutoffs may be discordant with respect to other cardiac troponin assays and the clinical diagnosis, even when imprecision from duplicate analysis is acceptable.4 As part of a stability study assessing Ortho hsTnI concentrations in both EDTA plasma and lithium heparin plasma over 24h, we have observed that this interference is random and not related to time on cells.



2021 ◽  
Vol 8 ◽  
Author(s):  
Emma H. Hooijberg ◽  
Karin Lourens ◽  
Leith C. R. Meyer

Pangolins are the world's most trafficked non-human mammals. A significant number of Temminck's pangolins (Smutsia temminckii) are presented for veterinary care and rehabilitation in southern Africa. Little is known about the physiology and normal health of this species, making diagnosis and medical management difficult. This study aimed to establish reference intervals (RIs) for hematology and plasma clinical chemistry in the Temminck's pangolin. RIs were generated according to international guidelines using samples from 27 healthy free-living (n = 18) and rehabilitated (n = 9) pangolins. Hematology was performed using the Abaxis VetScan HM5 analyzer with manual differentials; clinical chemistry was performed using heparin plasma on the Abaxis VetScan VS2 and Cobas Integra 400 Plus analyzers. Hematology RIs were: RBC 3.88–8.31 × 1012/L, HGB 73–150 g/L, HCT 26–51%, MCV 59–72 fL, MCH 15.6–21.4 pg, MCHC 257–325 g/L, RDW 14.3–19.1%, WBC 1.80–10.71 × 109/L. Vetscan VS2 clinical chemistry RIs were: albumin 27–41 g/L, ALP 26–100 U/L, ALT 25–307 U/L, amylase 267–826 U/L, bilirubin 4–10 μmol/L, calcium 2.1–2.2 mmol/L, globulin 21–55 g/L, glucose 3.8–10.0 mmol/L, phosphate 1.3–2.6 mmol/L, potassium 3.6–5.9 mmol/L, sodium 132–140 mmol/L total protein 52–84 g/L, and urea 5.3–11.4 mmol/L. RIs for creatinine were not calculated as analytical imprecision exceeded analytical performance goals. Cobas Integra clinical chemistry RIs were: albumin 22–33 g/L, ALP 20–104 U/L, ALT 17–291 U/L, amylase 466–1,533 U/L, bilirubin 1–14 μmol/L, calcium 2.0–2.4 mmol/L, creatinine <58 μmol/L, globulin 23–49 g/L, glucose 3.6–10.1 mmol/L, phosphate 1.0–2.2 mmol/L, potassium 3.1–5.8 mmol/L, sodium 137–150 mmol/L, total protein 47–72 g/L, and urea 6.0–12.5 mmol/L. There was significant bias between the two chemistry analyzers for several measurands. Differences were found for some analytes between free-living and rehabilitated animals, probably reflecting differences in nutrition and hydration. These are the first RIs generated for Temminck's pangolin. These results will allow veterinarians to better determine pangolin health status, formulate optimal treatment plans and increase patient survival rates in this endangered species.



Author(s):  
Johannes J. M. L. Hoffmann

AbstractBasophils (basophilic granulocytes) are the least abundant cells in blood. Nowadays, basophils are included in the complete blood count performed by hematology analyzers and therefore reported in practically all patients in whom hematologic investigations are requested. However, hematology analyzers are not reliable enough to report clinically useful results. This is due to a combination of very high analytical imprecision and poor specificity, because the chemical and physical methods used for basophil counting in hematology analyzers are ill-defined and thus basophils are not well recognized by the analyzers. As a result, false basophil counts are quite common. In view of increasing analytical performance demands, hematology laboratories should stop reporting basophil counts produced by hematology analyzers. Suggestions for alternative pathways are presented for those situations where basophils are of clinical relevance.



2020 ◽  
Author(s):  
Andrea Padoan ◽  
Chiara Cosma ◽  
Paolo Zaupa ◽  
Mario Plebani

BackgroundAbstractReliable SARS-CoV-2 serological assays are required for diagnosing infections, for the serosurveillance of past exposures and for assessing the response to future vaccines. In this study, the analytical and clinical performances of a chemiluminescent immunoassays for SARS-CoV-2 IgM and IgG detection (Mindray CL-1200i), targeting Nucleocapsid (N) and receptor binding domain (RBD) portion of the Spike protein, were evaluated.MethodsPrecision and linearity were evaluated using standardized procedures. A total of 157 leftover serum samples from 81 hospitalized confirmed COVID-19 patients (38 with moderate and 43 with severe disease) and 76 SARS-CoV-2 negative subjects (44 healthcare workers, 20 individuals with rheumatic disorders, 12 pregnant women) were included in the study. In an additional series of 44 SARS-CoV-2 positive, IgM and IgG time kinetics were also evaluated in a time-period of 38 days.ResultsPrecision was below or equal to 4% for both IgM and IgG, in all the studied levels, whilst a slightly significant deviation from linearity was observed for both assays in the range of values covering the manufacturer’s cut-off. Considering a time frame ≥ 12 days post symptom onset, sensitivity and specificity for IgM were 92.3% (95%CI:79.1%-98.4%) and 92.1% (95%CI:83.6%-97.0%). In the same time frame, sensitivity and specificity for IgG were 100% (95%CI:91.0%-100%) and 93.4% (95%CI:85.3%-97.8%). The assays agreement was 73.9% (Cohen’s kappa of 0.373). Time kinetics showed a substantial overlapping of IgM and IgG response, the latter values being elevated up to 38 days from symptoms onset.ConclusionsAnalytical imprecision is satisfactory as well as the linearity, particularly when taking into account the fact that both assays are claimed to be qualitative. Diagnostic sensitivity of IgG was excellent, especially considering specimens collected ≥12 days post symptom onset. Time kinetics suggest that IgM and IgG are detectable early in the course of infection, but the role of SARS-CoV-2 antibodies in clinical practice still requires further evaluations.



2020 ◽  
Vol 49 (1) ◽  
pp. 112-118 ◽  
Author(s):  
Fernando Tecles ◽  
Alberto Muñoz ◽  
José J Cerón ◽  
Kathleen Freeman


2019 ◽  
Vol 57 (12) ◽  
pp. 1999-2007 ◽  
Author(s):  
Eline A.E. van der Hagen ◽  
Christa M. Cobbaert ◽  
Ron Meijer ◽  
Marc H.M. Thelen

Abstract Background High-sensitivity cardiac troponin T/I (hs-cTnT/I) assays have improved analytical sensitivity for the detection of myocardial infarction (MI). To gain clinical specificity and sensitivity, interpretation of changes in cTn concentrations over time is crucial. The 2015 ESC NSTEMI guideline defines absolute delta values as additional rule-in and rule-out criteria for MI. A critical assumption for application of this rule is that total analytical imprecision within the delta period, including inter-instrument bias, is comparable to analytical imprecision in the validation studies. Methods Data from the Dutch External Quality Assessment Scheme (EQAS) were used to calculate inter-instrument bias and estimate imprecision for the measuring range where the proposed delta values are relevant: for Roche Elecsys hs-cTnT, 5–52 and 5–12 ng/L; for Abbott Architect hs-cTnI, 2–52 and 2–5 ng/L for rule-in and rule-out, respectively. Results For Elecsys, the median inter-instrument bias is 0.3 ng/L (n = 33 laboratories), resulting in reference change values (RCVs) of 3.0 and 1.7 ng/L, respectively, for rule-in and rule-out with imprecision as claimed by the manufacturer. With RCVs smaller than the guideline’s delta thresholds, 100% of the laboratories have adequate specifications. RCVs for rule-in/rule-out increased to 4.6 ng/L/2.5 ng/L, respectively, with individual imprecisions as estimated from EQA data, resulting in 64% and 82% of laboratories with adequate specifications. For Architect, 40% of instruments (n = 10) might falsely qualify the result as clinically relevant; hence, inter-instrument bias could not be determined. Conclusions We advise laboratories that use the fast 0/1-h algorithm to introduce stringent internal quality procedures at the relevant/low concentration level, especially when multiple analyzers are randomly used.



2019 ◽  
Vol 45 (03) ◽  
pp. 308-314 ◽  
Author(s):  
Giuseppe Lippi ◽  
Gian Salvagno ◽  
Matteo Gelati ◽  
Giovanni Poli ◽  
Davide Giavarina ◽  
...  

AbstractThis study aimed to provide a preliminary evaluation of the analytical performance of the new Roche cobas t 711 fully automated coagulation analyzer, which uses both liquid and lyophilized reagent cassettes. The analytical assessment included analysis of imprecision and linearity of prothrombin time (PT), activated partial thromboplastin time (APTT), and fibrinogen on cobas t 711 analyzer. Test results of 120 routine plasma samples were also compared with those obtained using two other coagulation analyzers (Instrumentation Laboratory ACL TOP 700 and Stago STA-R MAX). The accuracy, imprecision, and comparability of manual and automatic lyophilized material resuspension were also evaluated using 200 routine plasma samples. Overall, automatic resuspension was found to be more precise than, and equally accurate as, manual reconstitution, with coefficient of variations (CV%) three- to sixfold lower compared with manual reconstitution. The analytical imprecision was found to be excellent, as attested by total CV% of 0.7% for PT, 1.7 to 1.8% for APTT, and 1.9 to 3.2% for fibrinogen. Linearity was excellent over a clinically significant range of PT, APTT, and fibrinogen values, displaying correlation coefficients comprised between 0.994 and 0.999. Methods comparison studies revealed that results of PT, APTT, and fibrinogen on cobas t 711 are globally aligned with those obtained using identical plasma samples on IL ACL TOP 700 and Stago STA-R MAX, displaying correlation coefficients of 0.97 for PT, 0.81 and 0.88 for APTT, 0.90 and 0.94 for fibrinogen, respectively. In conclusion, the results of this preliminary evaluation demonstrate that PT, APTT, and fibrinogen on cobas t 711 coagulation analyzer displays excellent performance for routine use in clinical laboratories.



2018 ◽  
Vol 56 (5) ◽  
pp. 685-696 ◽  
Author(s):  
Denis Monneret ◽  
Martin Gellerstedt ◽  
Dominique Bonnefont-Rousselot

AbstractBackground:Detection of acute myocardial infarction (AMI) is mainly based on a rise of cardiac troponin with at least one value above the 99th percentile upper reference limit (99th URL). However, circulating high-sensitive cardiac troponin T (hs-cTnT) concentrations depend on age, sex and renal function. Using an analytical imprecision-based approach, we aimed to determine age- and sex-specific hs-cTnT 99th URLs for patients without chronic kidney disease (CKD).Methods:A 3.8-year retrospective analysis of a hospital laboratory database allowed the selection of adult patients with concomitant plasma hs-cTnT (<300 ng/L) and creatinine concentrations, both assayed twice within 72 h with at least 3 h between measurements. Absence of AMI was assumed when the variation between serial hs-cTnT values was below the adjusted-analytical change limit calculated according to the inverse polynomial regression of analytical imprecision. Specific URLs were determined using Clinical and Laboratory Standards Institute (CLSI) methods, and partitioning was tested using the proportion method, after adjustment for unequal prevalences.Results:After outlier removal (men: 8.7%; women: 6.6%), 1414 men and 1082 women with estimated glomerular filtration rate (eGFR) ≥60 mL/min/1.73 m2were assumed as non-AMI. Partitioning into age groups of 18–50, 51–70 and 71–98 years, the hs-cTnT 99th URLs adjusted on French prevalence were 18, 33, 66 and 16, 30, 84 ng/L for men and women, respectively. Age-partitioning was clearly required. However, sex-partitioning was not justified for subjects aged 18–50 and 51–70 years for whom a common hs-cTnT 99th URLs of about 17 and 31 ng/L could be used.Conclusions:Based on a laboratory approach, this study supports the need for age-specific hs-cTnT 99th URLs.



Sign in / Sign up

Export Citation Format

Share Document