scholarly journals Optical ammonia sensors based on fluorescent aza-BODIPY dyes— a flexible toolbox

2020 ◽  
Vol 412 (27) ◽  
pp. 7559-7567
Author(s):  
Maximilian Maierhofer ◽  
Veronika Rieger ◽  
Torsten Mayr

Abstract We present three types of optical ammonia sensors suitable for environmental, bioprocess, and reaction monitoring. A respective fluorescent BF2-chelated tetraarylazadipyrromethene dye (aza-BODIPYs) is physically entrapped in a polyurethane hydrogel (HydroMed D4) forming an emulsion system with vinyl-terminated polydimethylsiloxane (PDMS). The analyte-sensitive layer is covered by a hydrophobic membrane which excludes hydrophilic substances. Three different protection layers are tested, whereby the Teflon and the hydrophobic PES layers outperform a PDMS/TiO2 layer. Response times within their dynamic range of 15 s can be achieved, whereas the PDMS/TiO2-covered sensor requires at least 390 s. The three sensors entail the following concentration areas: first sensor 3 μg L−1–3 mg L−1 (LOD 0.23 μg L−1), second sensor 0.1–30 mg L−1 (LOD 28 μg L−1), and third sensor 3 mg L−1–1 g L−1 (LOD 0.51 mg L−1). Readout is performed with a commercially available phase fluorimeter combined with optical fibers. Dual-lifetime referencing (DLR) is used as referencing method and Egyptian blue acts as an inert reference material. No cross-sensitivity to pH changes can be detected.

Proceedings ◽  
2018 ◽  
Vol 2 (13) ◽  
pp. 1041 ◽  
Author(s):  
Maximilian Maierhofer ◽  
Sergey M. Borisov ◽  
Torsten Mayr

We present an optical ammonia sensor suitable for bioprocess monitoring. A fluorescent dye is physically entrapped in a polyurethane hydrogel forming an emulsion system with vinylterminated polydimethylsiloxane (PDMS). The sensing layer is covered by a hydrophobic porous membrane which excludes hydrophilic substances. Ammonia, diffuses through this barrier and PDMS to the protonated dye, whereby it deprotonates the dye and switches off its emission. Readout is performed with a miniaturized phase fluorimeter combined with optical fibers. Dual-lifetime referencing (DLR) acts as detection method and Egyptian blue as reference material.


2019 ◽  
Vol 42 (4) ◽  
pp. 917-923
Author(s):  
Jiss Paul ◽  
Jacob Philip

A modified inter-digital capacitive (IDC) transducer for relative humidity (RH) measurement is fabricated by coating a thick film of polyvinyl alcohol (PVA) on the structure. The effective capacitance of the sensor, measured at 1MHz, increases with RH at room temperature. The RH sensing characteristics of PVA coated IDC transducer are analysed in terms of its sensitivity, dynamic range, frequency response, repeatability and recovery time. It is found that the stability of such a PVA based RH transducer improves remarkably by diffusing nanoparticles of CoFe2O4 in the PVA layer. A standard deviation of ± 0.05 and ± 0.01 for a sensitive layer with pure PVA and PVA-CoFe2O4 combination respectively results in a measurement error of ± 0.005 and ± 0.003.


2013 ◽  
Vol 543 ◽  
pp. 385-388 ◽  
Author(s):  
Loukas Athanasekos ◽  
Nikos Aspiotis ◽  
Alexandros El Sachat ◽  
Stergios Pispas ◽  
Christos Riziotis

A novel approach on lysozyme sensing is proposed by employing Polymer Optical Fibers functionalized successfully with overlayers of block copolymer sensitive materials. The detection scheme is based on electrostatic interaction between lysozyme and the copolymer. Low concentration levels of lysozyme have been detected successfully with almost spontaneous response times.


1987 ◽  
Vol 33 (9) ◽  
pp. 1538-1542 ◽  
Author(s):  
W W Miller ◽  
M Yafuso ◽  
C F Yan ◽  
H K Hui ◽  
S Arick

Abstract A fluorescence-based fiber optic measurement system has been developed for monitoring pH, pCO2, and pO2 through a 20-gauge radial artery catheter without compromising capabilities for monitoring arterial pressure or for blood withdrawal. The measuring probe consists of three optical fibers to which the sensing chemistries are attached, and a thermocouple that measures temperature. The probe is designed to meet in-vivo biocompatibility requirements for a one-time use of up to 72 h. The components that are in contact with the patient's blood are nontoxic, nonhemolytic, nonthrombogenic, and sterilizable. Blood compatibility is enhanced by including covalently bound heparin. The in-vitro accuracy of the system has been tested against commercial blood-gas measurement instruments; comparison with tonometry and blood gas values gave r greater than or equal to 0.98 for all three sensors. The standard error for all sensors was within the College of American Pathologists' accuracy guidelines for measuring blood gas. Instrumental drift was minimal, indicating that system performance characteristics should not be the limiting factor in obtaining clinically useful information for up to 72 h. The response times of the sensors in animal and in-vitro studies were less than 2 min, suitable for monitoring physiological changes in blood gas values.


Blood ◽  
2007 ◽  
Vol 110 (11) ◽  
pp. 1013-1013 ◽  
Author(s):  
Susan Branford ◽  
Linda Fletcher ◽  
Nicolas C.P. Cross ◽  
Andreas Hochhaus ◽  
Martin C. Mueller ◽  
...  

Abstract BCR-ABL measurement by real-time quantitative PCR (RQ-PCR) has become an essential component for assessing treatment response for CML. A major molecular response (MMR) has prognostic significance and can be used to guide therapeutic decisions. However, the various methods are not standardized and the value representing MMR varies, which may lead to misinterpretation of molecular response. To align data, an international reporting scale (IS) was proposed where MMR is 0.10%. Conversion to the IS is achieved by applying laboratory (lab) specific conversion factors (CF). We aimed to calculate CF for diverse RQ-PCR methods by reference of patient BCR-ABL values to those generated in a reference lab with an established CF; validate the CF by subsequent patient sample exchange; examine the concordance of BCR-ABL values after IS conversion; determine if manufactured reference material is suitable for CF calculation. 34 labs from 13 countries (Australia/New Zealand 11, North/South America 9, Asia 8, Europe 6) sent 615 patient samples to the Adelaide reference lab to determine their specific CF. The RQ-PCR methods varied by the control gene (ABL 17, BCR 12, GUSβ 4, G6PDH 2, β2M 1, GAPDH 1; 3 labs used 2 controls therefore 37 methods), instrument, probe technology and standards. The CF for each method was calculated from the bias of patient BCR-ABL values between the originating lab and the reference lab, providing the bias was consistent across the dynamic range (Bland and Altman, Lancet,1986;1:307). CF were determined for 33 methods, 1 failed due to inconsistencies in the bias and 3 labs sent insufficient samples. CF were validated by sending subsequent sets of patient samples to the reference lab. The validation process is complete for 12 methods using 384 samples. The specific CF remained valid for each method. The mean bias between the reference and originating lab values was negligible after conversion. The limits of agreement indicated that 95% of values were within ±4.6-fold of the reference value. In contrast, prior to conversion 95% of values were within ±13-fold. Importantly after conversion the concordance in the range representing MMR was 87% (154/178 samples). In the future, conversion to the IS will be achieved using certified reference material, however this is currently not available. In order to mimic the patient bias CF calculation we prepared prototype reference material using BCR-ABL positive cells diluted to 4 levels using volunteer cells. The material was distributed to 29 labs and analysis completed for 24 methods. For 12 of the 24 the CF calculated using the reference material was consistent with the patient bias CF. This indicates that CF calculation is achievable using manufactured reference material but optimization is required before widespread distribution. In summary, alignment of BCR-ABL data generated from diverse methods is achievable using an international standardisation approach, and differences between laboratories are small enough to allow consistent interpretation of results and clinical decision-making.


2015 ◽  
Vol 23 (21) ◽  
pp. 27960 ◽  
Author(s):  
S. R. Sandoghchi ◽  
M. Petrovich ◽  
D. R. Gray ◽  
Y. Chen ◽  
N. V. Wheeler ◽  
...  

Sensors ◽  
2020 ◽  
Vol 20 (22) ◽  
pp. 6511
Author(s):  
Georgios Violakis ◽  
Tri Le-Quang ◽  
Sergey A. Shevchik ◽  
Kilian Wasmer

Acoustic Emission (AE) detection and, in particular, ultrasound detection are excellent tools for structural health monitoring or medical diagnosis. Despite the technological maturity of the well-received piezoelectric transducer, optical fiber AE detection sensors are attracting increasing attention due to their small size, and electromagnetic and chemical immunity as well as the broad frequency response of Fiber Bragg Grating (FBG) sensors in these fibers. Due to the merits of their small size, FBGs were inscribed in optical fibers with diameters of 50 and 80 μm in this work. The manufactured FBGs were used for the detection of reproducible acoustic waves using the edge filter detection method. The acquired acoustic signals were compared to the ones captured by a standard 125 μm-diameter optical fiber FBG. Result analysis was performed by utilizing fast Fourier and wavelet decompositions. Both analyses reveal a higher sensitivity and dynamic range for the 50 μm-diameter optical fiber, despite it being more prone to noise than the other two, due to non-standard splicing methods and mode field mismatch losses. Consequently, the use of smaller-diameter optical fibers for AE detection is favorable for both the sensor sensitivity as well as physical footprint.


1992 ◽  
Vol 46 (7) ◽  
pp. 1105-1112 ◽  
Author(s):  
William R. Moser ◽  
Joseph R. Berard ◽  
Peter J. Melling ◽  
Robert J. Burger

A new versatile spectroscopic technique for chemical reaction monitoring using mid-range infrared optical fibers has recently been developed. Chalcogenide glass optical fibers were used to direct infrared radiation from an FT-IR spectrometer through ZnSe Cylindrical Internal Reflectance (CIR) crystals embedded within laboratory scale reactors. The utility of this technique for studying chemical systems was demonstrated by monitoring various stoichiometric reactions at ambient conditions. A laboratory-scale glass reactor fabricated with the capability to mount a CIR crystal was used as the reaction vessel. The ability of this system to monitor high-pressure and/or high-temperature chemical reactions was also demonstrated by studying the cobalt catalyzed hydroformylation of olefins. A stainless steel CIR reactor, slightly modified to allow for connections with optical fibers, was used for experiments ranging from 50 to 90°C and under 750 to 800 psi synthesis gas (H2/CO mixture). In all cases sufficient signal strength at the detector and adequate penetration into the bulk reaction medium was achieved, resulting in infrared spectra of high quality and resolution. Spectral scans of the reaction in progress allowed the accurate determination of the concentration of reactants and products as a function of time.


Sign in / Sign up

Export Citation Format

Share Document