Development of Electron Mirror Microscope

Author(s):  
T. Someya ◽  
M. Watanabe

Our electron mirror microscope has been improved by adopting magnetic lenses, which enable a wide range of magnification to be obtained. This improved version, known as the JEM-M1 (35kV acceleration) uses the straight design and shadow projection method.The construction of the newly designed instrument is illustrated in Fig. 1 and Fig. 2 shows a sectional view of its electron optical column. The electron gun assembly is located under the console table. Incident beam alignment is carried out by means of the deflection coil knob, the electron gun horizontal movement knob and the condenser lens knob located on the table. The zircaloy crystal located above the screen and on the optical axis enables the alignment procedure to be carried out easily and simply because it radiates light when impinged by the electron beam. The projector lens, having a short focal length, combined with the condenser lens, forms a small source for the shadow projection image on its back focal plane. The intermediate lens, acting on the reflected beam, produces image magnification by changing the excitation current. Magnification in the range x200 – x1500 is possible without any appreciable distortion.

Author(s):  
A. V. Crewe

A 100 kv transmission scanning microscope is now being constructed which should have a point resolution of 2.5 to 3 Å. The design of this microscope is similar to the design of our existing 30 kv 5 Å microscope, but there are several significant changes which are based upon some difficulties and sources of inflexibility of that microscope.A field emission electron gun of our usual design will be used as the source of electrons, the only difference being that the spacing between the anodes has been increased from 2 to 3 cm. The electron beam will then pass through a condenser lens which will produce a parallel beam of electrons. This parallel beam will then be focused onto the specimen by means of a short focal length lens (approximately 1 mm focal length). The reason for using a condenser lens to produce the parallel beam of electrons is that in the future a quadrupole-octupole correction system will be installed in this section of the microscope in order to attempt to correct the spherical aberrations of the objective lens and thereby improve its resolution.


2005 ◽  
Vol 872 ◽  
Author(s):  
W. Park ◽  
E. Schonbrun ◽  
M. Tinker ◽  
Q. Wu ◽  
J.-B. Lee

AbstractWe report a novel tunable nanophotonic device concept based on Mechanically Controlled Photonic Crystal (MCPC), which is comprised of a periodic array of high index dielectric material and a low index polymer. Tunability is achieved by applying mechanical force with nano-/micro-electron-mechanical system actuators. The mechanical stress induces changes in the periodicity of the photonic crystal, to which the photonic band structure is extremely sensitive. This consequently produces tunability much greater than that achievable by electro-optic materials such as liquid crystal. Our theoretical investigations revealed that we could achieve dynamic beam steering over a wide range of angles up to 75° with only 10% mechanical stretching. We also predicted tunable sub-wavelength imaging in which we could tune the frequency response and focal length of negative index PC lens. For experimental demonstration, we fabricated the PC structures on Si-on-insulator substrates. Optical characterizations clearly showed the anticipated negative refraction in which the incident beam was refracted back to the side it was incident. The experimental demonstration of negative refraction at optical frequencies in a Si-based photonic crystal structure is a significant step toward the next-generation nanophotonics.


Author(s):  
J. S. Wall ◽  
J. P. Langmore ◽  
H. Isaacson ◽  
A. V. Crewe

The scanning transmission electron microscope (STEM) constructed by the authors employs a field emission gun and a 1.15 mm focal length magnetic lens to produce a probe on the specimen. The aperture size is chosen to allow one wavelength of spherical aberration at the edge of the objective aperture. Under these conditions the profile of the focused spot is expected to be similar to an Airy intensity distribution with the first zero at the same point but with a peak intensity 80 per cent of that which would be obtained If the lens had no aberration. This condition is attained when the half angle that the incident beam subtends at the specimen, 𝛂 = (4𝛌/Cs)¼


Author(s):  
David A. Ansley

The coherence of the electron flux of a transmission electron microscope (TEM) limits the direct application of deconvolution techniques which have been used successfully on unmanned spacecraft programs. The theory assumes noncoherent illumination. Deconvolution of a TEM micrograph will, therefore, in general produce spurious detail rather than improved resolution.A primary goal of our research is to study the performance of several types of linear spatial filters as a function of specimen contrast, phase, and coherence. We have, therefore, developed a one-dimensional analysis and plotting program to simulate a wide 'range of operating conditions of the TEM, including adjustment of the:(1) Specimen amplitude, phase, and separation(2) Illumination wavelength, half-angle, and tilt(3) Objective lens focal length and aperture width(4) Spherical aberration, defocus, and chromatic aberration focus shift(5) Detector gamma, additive, and multiplicative noise constants(6) Type of spatial filter: linear cosine, linear sine, or deterministic


Author(s):  
L. F. Allard ◽  
E. Völkl ◽  
T. A. Nolan

The illumination system of the cold field emission (CFE) Hitachi HF-2000 TEM operates with a single condenser lens in normal imaging mode, and with a second condenser lens excited to give the ultra-fine 1 nm probe for microanalysis. The electron gun provides a guaranteed high brightness of better than 7×l08 A/cm2/sr, more than twice the guaranteed brightness of Schottky emission guns. There have been several articles in the recent literature (e.g. refs.) which claim that the geometry of this illumination system yields a total current which is so low that when the beam is spread at low magnifications (say 10 kX), the operator must “keep his eyes glued to the binoculars” in order to see the image. It is also claimed that this illuminating system produces an isoplanatic patch (the area over which image character does not vary significantly) at high magnification which is so small that the instrument is ineffective for recording high resolution images.


Author(s):  
David C Joy

The electron source is the most important component of the Scanning electron microscope (SEM) since it is this which will determine the overall performance of the machine. The gun performance can be described in terms of quantities such as its brightness, its source size, its energy spread, and its stability and, depending on the chosen application, any of these factors may be the most significant one. The task of the electron gun in an SEM is, in fact, particularly difficult because of the very wide range of operational parameters that may be required e.g a variation in probe size of from a few angstroms to a few microns, and a probe current which may go from less than a pico-amp to more than a microamp. This wide range of operating parameters makes the choice of the optimum source for scanning microscopy a difficult decision.Historically, the first step up from the sealed glass tube ‘cathode ray generator’ was the simple, diode, tungsten thermionic emitter.


Author(s):  
F. Hosokawa ◽  
Y. Kondo ◽  
T. Honda ◽  
Y. Ishida ◽  
M. Kersker

High-resolution transmission electron microscopy must attain utmost accuracy in the alignment of incident beam direction and in astigmatism correction, and that, in the shortest possible time. As a method to eliminate this troublesome work, an automatic alignment system using the Slow-Scan CCD camera has been introduced recently. In this method, diffractograms of amorphous images are calculated and analyzed to detect misalignment and astigmatism automatically. In the present study, we also examined diffractogram analysis using a personal computer and digitized TV images, and found that TV images provided enough quality for the on-line alignment procedure of high-resolution work in TEM. Fig. 1 shows a block diagram of our system. The averaged image is digitized by a TV board and is transported to a computer memory, then a diffractogram is calculated using an FFT board, and the feedback parameters which are determined by diffractogram analysis are sent to the microscope(JEM- 2010) through the RS232C interface. The on-line correction system has the following three modes.


2019 ◽  
Vol 9 (1) ◽  
Author(s):  
Sergio Jiménez-Gambín ◽  
Noé Jiménez ◽  
José M. Benlloch ◽  
Francisco Camarena

AbstractWe report zero-th and high-order acoustic Bessel beams with broad depth-of-field generated using acoustic holograms. While the transverse field distribution of Bessel beams generated using traditional passive methods is correctly described by a Bessel function, these methods present a common drawback: the axial distribution of the field is not constant, as required for ideal Bessel beams. In this work, we experimentally, numerically and theoretically report acoustic truncated Bessel beams of flat-intensity along their axis in the ultrasound regime using phase-only holograms. In particular, the beams present a uniform field distribution showing an elongated focal length of about 40 wavelengths, while the transverse width of the beam remains smaller than 0.7 wavelengths. The proposed acoustic holograms were compared with 3D-printed fraxicons, a blazed version of axicons. The performance of both phase-only holograms and fraxicons is studied and we found that both lenses produce Bessel beams in a wide range of frequencies. In addition, high-order Bessel beam were generated. We report first order Bessel beams that show a clear phase dislocation along their axis and a vortex with single topological charge. The proposed method may have potential applications in ultrasonic imaging, biomedical ultrasound and particle manipulation applications using passive lenses.


Crystals ◽  
2019 ◽  
Vol 9 (8) ◽  
pp. 431 ◽  
Author(s):  
Li-Lan Tian ◽  
Fan Chu ◽  
Hu Dou ◽  
Lei Li ◽  
Qiong-Hua Wang

An electrically tunable-focusing liquid crystal (LC) microlens array exhibiting a wide-range tunable focal length is proposed. The lower substrate has strip indium tin oxide (ITO) electrodes, the upper substrate has periodic ITO electrodes with a certain gap coated on the inner surface., and an LC microlens is generated between the two strip electrodes. For each LC microlens, the gap between the top planar electrodes is directly above the center of the microlens. Unlike the conventional LC lens, the individual LC microlens is not coated with ITO electrodes on the central part of its upper and lower substrates, which helps to maintain the LC’s horizontal orientation. In the voltage-off state, the focal length of the microlens array is infinity because of the homogeneous LC alignment. At a given operating voltage, an ideal gradient refractive index distribution is induced over the homogeneous LC layer, which leads to the focusing effect. The simulation result shows that the focal length of the LC microlens could be gradually drawn to 0.381 mm with a change of voltage.


Sign in / Sign up

Export Citation Format

Share Document